| Abstract |
|---|
|
|
|---|
Methods: Supravalvular transcatheter valves were created using a Dacron covered stainless steel stent at the base and trileaflet pericardial leaflets in an open stent above the bioprosthesis. The transcatheter valves were implanted within 19-, 21-, and 23-mm Carpentier-Edwards Perimount bioprostheses with simulated degeneration using BioGlue to achieve a mean pressure gradient of 50 mm Hg. Hemodynamics of valve-in-valve implantation were studied in a pulse duplicator.
Results: Supravalvular transcatheter valves successfully relieved bioprosthetic stenosis. Acceptable hemodynamics were achieved with a significant reduction in mean pressure gradient of 54.0 ± 3.5 to 9.2 ± 6.3 mm Hg in 23-mm bioprostheses (p < 0.001), from 49.3 ± 3.1 to 14.4 ± 4.7 mmHg (p < 0.001) in 21 mm, and from 53.9 ± 3.8 to 28.3 ± 9.8 mm Hg (p = 0.013) in 19-mm bioprostheses. Effective orifice area after valve-in-valve implantation increased significantly and was comparable to rereplacement with the same size bioprosthesis.
Conclusions: Valve-in-valve implantation was performed using a novel supravalvular transcatheter valve, which successfully relieved bioprosthetic stenosis. The hemodynamics were comparable with standard surgical valve replacement. Further studies are required to assess device safety and efficacy in patients.
| Introduction |
|---|
|
|
|---|
We hypothesized that a possible solution to achieve better hemodynamics of TAVs within bioprostheses is to develop a supravalvular TAV where the valve within the TAV stent is situated above the bioprosthesis (Fig 1). As a result, the TAV is not constrained by the bioprosthesis and a larger valve can be deployed above the bioprosthesis to enhance hemodynamics. The objective of this study was to determine the feasibility and efficacy of supravalvular TAVs in degenerated bioprostheses.
|
| Material and Methods |
|---|
|
|
|---|
|
A customized cylindrical stainless steel stent (W.L. Gore and Associates, Flagstaff, AZ), 30-mm height, was dilated to an external diameter of 23 mm. The leaflets and Dacron sheet were secured within the stent. The 3 commissures were anchored at the top of the stent with 4-0 polypropylene interrupted stitches. The Dacron cylinder was attached to the stent at each metal intersection using 5-0 polypropylene interrupted stitches (Fig 2).
Degenerated Bioprosthetic Valves
Acquiring explanted degenerated bioprostheses from patients would be unpredictable with respect to pressure gradients as well as bioprosthetic valve sizes, and it would be difficult to achieve the quantity of each size sufficient for statistical analyses at one institution. Therefore, a reproducible model to simulate degeneration of normal bioprostheses was developed to investigate VIV hemodynamics. The model provided consistent transvalvular pressure gradients and reflected the in vivo pathology of the calcified valve.
To simulate calcification, the most frequent mode of failure in pericardial bioprostheses, BioGlue (CryoLife, Inc, Kennesaw, GA) was applied to the leaflets of normal bioprostheses to stiffen the leaflets and imitate calcification [6]. An additional sheet of pericardium was needed to maintain BioGlue adherence and prevent dislodgement during balloon predilatation of the bioprosthesis before TAV implantation. Although various epoxy and commercial glues were used, BioGlue was most effective at simulating degeneration.
Three pieces of Edwards Bovine Pericardial Patch were cut in half-circles with the same diameter as that of the leaflet's free edge. Each piece of pericardium was sutured to 1 of the 3 leaflets of the bioprosthesis on the aortic side by a polypropylene 6-0 running suture along the circular part of the piece, from the bottom of the bioprosthesis sinus to the commissure. No sutures were made on the free edge of the leaflet bioprosthesis in order to create a pocket for the BioGlue. Approximately 2 to 3 mL of BioGlue was then injected into the pockets to reach the desirable pressure gradient (Fig 3). A mean bioprosthetic gradient of 50 mm Hg was set as the goal based on echocardiographic data of degenerated aortic bioprostheses [7, 8]. The bioprosthetic degeneration model reproduced the hemodynamics of a patient with severe bioprosthetic aortic stenosis.
|
Hemodynamic Measurements
Valve hemodynamics were evaluated with three variables: effective orifice area, mean pressure gradient, and regurgitant volume. Pressure was measured in the left atrium, left ventricle, left ventricular outflow tract, and ascending aorta with strain gauge pressure transducers (Cobe Laboratories Inc, Lakewood, CO). An electromagnetic flowmeter (Carolina Medical Electronics Inc, Oldtown, NC) was used to measure the aortic valve flow rate and the regurgitant volume. Effective orifice area within the TAV was then calculated using the Gorlin equation. Regurgitant fraction was also calculated, defined as the aortic retrograde flow divided by systolic ejection flow. Two-dimensional echocardiography (ACUSON Sequoia C256, Siemens Medical Solutions USA, Inc, Malvern, PA) was used to identify leakage and leakage location and to assess valve opening and closing processes.
Data Acquisition and Analyses
To study the efficacy of VIV implantation, the following experiments were conducted in the pulse duplicator: First, supravalvular TAVs were tested alone in the pulse duplicator to determine the efficacy of the TAV before implantation. Data acquisition was run over 10 consecutive cardiac cycles, and transvalvular pressure gradient, regurgitant volume, and effective orifice area were determined. Then, a normal Carpentier-Edwards PERIMOUNT aortic bioprostheses was tested to obtain a hemodynamic baseline. Subsequently, the same sized degenerated Carpentier-Edwards PERIMOUNT aortic bioprostheses was tested in the pulse duplicator. Finally after balloon predilation of the degenerated bioprosthesis, the supravalvular TAV was implanted within the bioprosthesis. Measurements were performed for three bioprosthetic valve sizes (19, 21, and 23 mm; n = 3 of each bioprosthetic size). Hemodynamic measurements were compared using a one-way analysis of variance (ANOVA). Reported values are quoted as mean ± standard deviation and statistical analyses were performed using MATLAB 7.0 software (Natick, MA).
| Results |
|---|
|
|
|---|
|
|
However, regurgitant fraction also increased significantly in all bioprosthetic valves after 23-mm supravalvular TAV implantation (Fig 4C). Regurgitant fraction of supravalvular TAVs at baseline was determined in the pulse duplicator and found to be 12.58% ± 0.77%. Two-dimensional echocardiography assessment of the valve showed that the leakage was mainly paravalvular. After VIV implantation, regurgitant fraction increased significantly, from 4.34% ± 1.36% to 17.09% ± 1.65% (p < 0.001) in 19-mm bioprosthesis, from 4.82% ± 1.61% to 16.86% ± 5.35% in 21-mm bioprostheses (p = 0.020), and from 5.87 ± 1.81% to 20.58 ± 3.51% (p <0.001) in 23-mm bioprostheses.
VIV hemodynamics for each bioprosthetic size were compared with hemodynamics of normal Carpentier-Edwards PERIMOUNT bioprostheses to compare VIV therapy with surgical rereplacement with equivalently sized bioprostheses. Supravalvular TAV reduced mean pressure gradient to 9.2 ± 6.3 mm Hg in the 23-mm degenerated bioprostheses, which was not significantly different than the mean pressure gradient of 5.7 ± 0.9 mm Hg (p = 0.422) for normal 23-mm PERIMOUNT bioprostheses. Similarly, VIV mean transvalvular gradient of 14.4 ± 4.7 mm Hg in 21-mm bioprostheses was not significantly different than the 11.8 ± 1.9 mm Hg (p = 0.306) of normal 21-mm PERIMOUNT bioprostheses, Even in 19-mm bioprostheses, although the obtained VIV mean pressure gradient of 28.3 ± 9.8 mm Hg was higher than that of the normal 19-mm PERIMOUNT bioprostheses (16.2 ± 2.2 mm Hg), the difference was not statistically significant (p = 0.056).
| Comment |
|---|
|
|
|---|
Hemodynamic performance of 23-mm supravalvular TAVs was studied within Carpentier-Edwards pericardial bioprostheses. The supravalvular TAV successfully relieved the bioprosthetic stenosis in all three bioprosthetic sizes. The obtained VIV transvalvular gradients were comparable with standard surgical valve replacement of equivalent size. For a 19-mm degenerated bioprosthesis, however, a lower pressure gradient can be expected from surgical aortic valve replacement with a 19-mm valve. The corresponding difference in effective orifice area of 1.00 cm2 for VIV in a 19-mm bioprostheses vs 1.28 cm2 for rereplacement with a 19-mm bioprosthesis may or may not have significant clinical effect depending on the patient's body surface area and comorbidities that affect the risk/benefit ratio of open and minimally invasive procedures.
VIV implantation in these circumstances could still benefit nonsurgical and very high-risk patients. Careful clinical judgment regarding the surgical risks of reoperative valve replacement and the severity of dysfunction of the degenerated bioprosthesis should be taken into account before decisions are made between an operation and VIV implantation in these instances. Overall, our supravalvular TAVs demonstrated good hemodynamics within small-sized degenerated bioprostheses and were effective in relieving bioprosthetic stenosis.
A potential complication not encountered with surgical rereplacement with a bioprosthesis was the increase in regurgitation with VIV implantation. Regurgitant volume increased significantly after implantation of 23-mm supravalvular TAVs in all three bioprostheses. The leak was mainly paravalvular, which was assessed using 2-dimensional echocardiography. Intravalvular leak, which was caused by incomplete coaptation of leaflets, was also observed in the 19-mm bioprosthesis. The top portion of the supravalvular TAV could not be fully expanded in the 19-mm bioprosthesis because the lower portion was tightly constrained by the small size bioprosthesis. Therefore, the TAV leaflets were distorted because of an excess of pericardial tissue relative to the stent orifice area (Fig 5). A more flexible stent design using a nitinol stent in the supravalvular portion may facilitate the expansion of the supravalvular TAV and improve hemodynamics. Paravalvular leak has similarly been a problem for clinical implantation of Edwards SAPIEN TAV within native stenosed aortic valves [14, 15].
|
The VIV prostheses were stable and no TAV migration was observed. Placing a larger valve above the bioprosthesis helped to avoid any TAV displacement, especially during the diastolic phase where the dominant drag force was applied to TAV [16, 17]. Furthermore, because the lower portion of the stent was to hold the TAV within the bioprosthesis and the valve was located above the bioprosthesis, no TAV leaflet distortion or traumatic injury to the leaflets by calcification was observed within the 21- and 23-mm bioprostheses. Zegdi and colleagues [18] described such leaflet distortion during TAV implantation within native aortic valves.
The next step for determining the effect of the supravalvular TAV on bioprosthetic stenosis is to document the effect of this design on coronary flows after implantation. It is unknown whether movement of the valve above the bioprosthesis improves hemodynamics but adversely affects coronary flows. The height of the supravalvular TAV in relation to the height of the aortic sinuses is crucial for successful implantation of the 23-mm supravalvular TAV. Ideally, the supravalvular portion remains within the sinus; however, if it extends into the ascending aorta, the diameter of the ascending aorta above the sinuses is also a critical variable. The supravalvular TAV completely within the aortic sinus or smaller in diameter than the ascending aorta if extending to the ascending aorta should not obstruct coronary flow because the open stent would allow blood flow through the stent during valve closure.
In some patients, the coronaries reside close to the sowing cuff of the bioprosthesis or near the stent posts. Because the bioprosthetic cuff and stent posts constrain the TAV expansion in this region, we hypothesize that the covered Dacron stent would not obstruct the coronaries unless the degenerated bioprosthetic leaflets themselves had sufficient bulky calcification to obstruct low lying coronaries. The open valve stent of the TAV above the bioprosthesis expanded to full 23-mm size would allow flow into the sinus and coronaries during diastole while maximizing orifice area, particularly in small sized bioprostheses.
The effect of this design in altering the amount of coronary flow during diastole and pattern of flow should be studied not only experimentally but computationally. Precise echocardiographic measurement is required to determine the size of the ascending aorta and sinuses before TAV implantation. Future in vitro pulse duplicator and in vivo animal studies are required to address the effect of supravalvular TAV on coronary flow before safety trials in humans.
The primary limitation of the study is that the bioprosthetic valves used in our experiments were not degenerated and we used BioGlue to simulate calcification. TAV implantation within degenerated bioprosthesis could be more complicated due to leaflet calcification, stent deformation, or pannus. Although we cannot directly extrapolate our results to clinical practice, our in vitro study suggests that the 23-mm supravalvular TAV provides acceptable hemodynamics in 19-, 21-, and 23-mm degenerated bioprostheses. The regurgitant volume for VIV was higher than for a bioprosthesis, but the pressure gradient and effective orifice area were comparable with surgical valve replacement. The paravalvular leakage may have arisen from various sources, such as the suture line between the leaflets and Dacron sheet of the TAV itself, gaps between the tip of bioprosthetic leaflet and the base of the supravalvular TAV leaflet, or gaps between the bioprosthesis and TAV stent. Nevertheless, lower regurgitant volume is expected using blood instead of the working fluid, which does not have any coagulation properties.
In conclusion, we have developed a supravalvular TAV designed to treat bioprosthetic degeneration. This novel TAV consists of two components: a Dacron-covered stainless steel stent that sits within the bioprosthesis and a valve situated above the bioprosthesis. Successful VIV implantation was performed using this supravalvular TAV for treatment of bioprosthetic stenosis. The 23-mm supravalvular TAV successfully relieved bioprosthetic stenosis, and the obtained transvalvular pressure gradient was comparable with standard surgical valve replacement of equivalent size. Further studies are mandatory to examine the effect of the supravalvular TAV on coronary flows as well as device safety and efficacy in patients.
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
Related Article
This article has been cited by other articles:
![]() |
A. N. Azadani and E. E. Tseng Transcatheter Heart Valves for Failing Bioprostheses: State-of-the-Art Review of Valve-in-Valve Implantation Circ Cardiovasc Interv, December 1, 2011; 4(6): 621 - 628. [Full Text] [PDF] |
||||
![]() |
R. Gurvitch, A. Cheung, J. Ye, D. A. Wood, A. B. Willson, S. Toggweiler, R. Binder, and J. G. Webb Transcatheter Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves. J. Am. Coll. Cardiol., November 15, 2011; 58(21): 2196 - 2209. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Azadani, N. Jaussaud, L. Ge, S. Chitsaz, T. A. M. Chuter, and E. E. Tseng Valve-in-Valve Hemodynamics of 20-mm Transcatheter Aortic Valves in Small Bioprostheses Ann. Thorac. Surg., August 1, 2011; 92(2): 548 - 555. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Azadani, N. Jaussaud, P. B. Matthews, L. Ge, T. A. M. Chuter, and E. E. Tseng Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses Interact CardioVasc Thorac Surg, July 1, 2010; 11(1): 70 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Walther and C. Walther Invited Commentary Ann. Thorac. Surg., December 1, 2009; 88(6): 1870 - 1870. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |