Ann Thorac Surg 2009;88:1864-1869. doi:10.1016/j.athoracsur.2009.08.004
© 2009 The Society of Thoracic Surgeons

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Elaine E. Tseng
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Azadani, A. N.
Right arrow Articles by Tseng, E. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Azadani, A. N.
Right arrow Articles by Tseng, E. E.
Related Collections
Right arrow Valve disease
Right arrowRelated Article


Original Articles: Adult Cardiac

Valve-in-Valve Implantation Using a Novel Supravalvular Transcatheter Aortic Valve: Proof of Concept

Ali N. Azadani, PhD, Nicolas Jaussaud, MD, Peter B. Matthews, BS, Liang Ge, PhD, T. Sloane Guy, MD, Timothy A.M. Chuter, MD, Elaine E. Tseng, MD*

Department of Surgery, University of California at San Francisco Medical Center and San Francisco Veterans Affairs Medical Center, San Francisco, California

Accepted for publication August 4, 2009.

* Address correspondence to Dr Tseng, Division of Cardiothoracic Surgery, UCSF Medical Center, 500 Parnassus Ave, Ste 405W, Box 0118, San Francisco, CA 94143-0118 (Email: elaine.tseng{at}ucsfmedctr.org).


    Abstract
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 
Background: Transcatheter valve implantation within degenerated bioprostheses is a potentially promising treatment for high-risk surgical patients. Clinical experience is limited; however, we have shown in vitro that currently available transcatheter aortic valve sizes did not provide acceptable hemodynamics in small bioprostheses. The objective of this study was to develop a new transcatheter valve that would provide good hemodynamics within degenerated bioprostheses.

Methods: Supravalvular transcatheter valves were created using a Dacron covered stainless steel stent at the base and trileaflet pericardial leaflets in an open stent above the bioprosthesis. The transcatheter valves were implanted within 19-, 21-, and 23-mm Carpentier-Edwards Perimount bioprostheses with simulated degeneration using BioGlue to achieve a mean pressure gradient of 50 mm Hg. Hemodynamics of valve-in-valve implantation were studied in a pulse duplicator.

Results: Supravalvular transcatheter valves successfully relieved bioprosthetic stenosis. Acceptable hemodynamics were achieved with a significant reduction in mean pressure gradient of 54.0 ± 3.5 to 9.2 ± 6.3 mm Hg in 23-mm bioprostheses (p < 0.001), from 49.3 ± 3.1 to 14.4 ± 4.7 mmHg (p < 0.001) in 21 mm, and from 53.9 ± 3.8 to 28.3 ± 9.8 mm Hg (p = 0.013) in 19-mm bioprostheses. Effective orifice area after valve-in-valve implantation increased significantly and was comparable to rereplacement with the same size bioprosthesis.

Conclusions: Valve-in-valve implantation was performed using a novel supravalvular transcatheter valve, which successfully relieved bioprosthetic stenosis. The hemodynamics were comparable with standard surgical valve replacement. Further studies are required to assess device safety and efficacy in patients.


    Introduction
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 
The feasibility of transcatheter aortic valve (TAV) implantation within degenerated bioprostheses has recently been demonstrated [1–4]. Although valve-in-valve (VIV) results are encouraging, clinical experience remains limited [1–3]. We have previously shown hemodynamic complications from implantation of currently available TAVs within small-size normal bioprostheses in in vitro experiments [5]. Successful VIV implantation was achieved only when the TAV matched the bioprosthetic size. The oversized TAV was constrained by the rigid bioprosthesis, resulting in TAV stenosis. The TAV was anchored inside the bioprosthesis and could not be dilated beyond the bioprosthetic annulus. Thus, one concern of using TAVs within small bioprostheses is that the effective orifice area may be reduced significantly, particularly when compared with surgical valve replacement of equal size.

We hypothesized that a possible solution to achieve better hemodynamics of TAVs within bioprostheses is to develop a supravalvular TAV where the valve within the TAV stent is situated above the bioprosthesis (Fig 1). As a result, the TAV is not constrained by the bioprosthesis and a larger valve can be deployed above the bioprosthesis to enhance hemodynamics. The objective of this study was to determine the feasibility and efficacy of supravalvular TAVs in degenerated bioprostheses.


Figure 1
View larger version (124K):
[in this window]
[in a new window]

 
Fig 1. Drawing shows a supravalvular transcatheter aortic valve within a degenerated bioprosthesis.

 

    Material and Methods
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 
Supravalvular TAVs
The custom designed supravalvular TAV consists of a stainless steel stent covered with Dacron (DuPont, Wilmington, DE) that sits within the bioprosthesis and a valve within an open stent situated above the bioprosthetic posts (Fig 2). To study the efficacy of VIV implantation, six 23-mm supravalvular TAVs were made for implantation within 19-, 21-, and 23-mm Carpentier-Edwards Perimount Magna (Edwards Lifesciences, Irvine, CA) bioprostheses (n = 3 of each bioprosthetic size). Each supravalvular TAV was reusable and functioned well after removal, recrimping, retesting hemodynamics, and reimplantation in a smaller bioprosthetic size.


Figure 2
View larger version (128K):
[in this window]
[in a new window]

 
Fig 2. Supravalvular transcatheter aortic valves are shown in (A) top and (B) side views.

 
Three trapezoidal-shaped leaflets (22.5- and 23.5-mm parallel sides, and 15-mm height) were cut from a flat piece of bovine pericardium (Edwards Bovine Pericardial Patch, Edwards Lifesciences). The lateral sides of the 3 leaflets were sutured together to create commissures. The base of the leaflets was sutured to a Dacron sheet (69 x 23 mm) at a height of 15 mm using 6-0 polypropylene running suture. Thus, 15 mm of the base of the Dacron sheet was below the valve.

A customized cylindrical stainless steel stent (W.L. Gore and Associates, Flagstaff, AZ), 30-mm height, was dilated to an external diameter of 23 mm. The leaflets and Dacron sheet were secured within the stent. The 3 commissures were anchored at the top of the stent with 4-0 polypropylene interrupted stitches. The Dacron cylinder was attached to the stent at each metal intersection using 5-0 polypropylene interrupted stitches (Fig 2).

Degenerated Bioprosthetic Valves
Acquiring explanted degenerated bioprostheses from patients would be unpredictable with respect to pressure gradients as well as bioprosthetic valve sizes, and it would be difficult to achieve the quantity of each size sufficient for statistical analyses at one institution. Therefore, a reproducible model to simulate degeneration of normal bioprostheses was developed to investigate VIV hemodynamics. The model provided consistent transvalvular pressure gradients and reflected the in vivo pathology of the calcified valve.

To simulate calcification, the most frequent mode of failure in pericardial bioprostheses, BioGlue (CryoLife, Inc, Kennesaw, GA) was applied to the leaflets of normal bioprostheses to stiffen the leaflets and imitate calcification [6]. An additional sheet of pericardium was needed to maintain BioGlue adherence and prevent dislodgement during balloon predilatation of the bioprosthesis before TAV implantation. Although various epoxy and commercial glues were used, BioGlue was most effective at simulating degeneration.

Three pieces of Edwards Bovine Pericardial Patch were cut in half-circles with the same diameter as that of the leaflet's free edge. Each piece of pericardium was sutured to 1 of the 3 leaflets of the bioprosthesis on the aortic side by a polypropylene 6-0 running suture along the circular part of the piece, from the bottom of the bioprosthesis sinus to the commissure. No sutures were made on the free edge of the leaflet bioprosthesis in order to create a pocket for the BioGlue. Approximately 2 to 3 mL of BioGlue was then injected into the pockets to reach the desirable pressure gradient (Fig 3). A mean bioprosthetic gradient of 50 mm Hg was set as the goal based on echocardiographic data of degenerated aortic bioprostheses [7, 8]. The bioprosthetic degeneration model reproduced the hemodynamics of a patient with severe bioprosthetic aortic stenosis.


Figure 3
View larger version (202K):
[in this window]
[in a new window]

 
Fig 3. A 21-mm degenerated bioprosthesis in which BioGlue was used to simulate calcification.

 
Pulse Duplicator System
An in vitro study provides a consistent and well-controlled environment to examine VIV hemodynamics. The valves were tested at room temperature in a custom-built pulse duplicator system developed for TAV implantation (Vivitro Systems Inc, Victoria, BC, Canada). This system is a pulsatile-flow model of the left side of the heart and systemic circulation. Heart rate, blood pressure, and cardiac output were used as control variables for the waveform generator controlling a servo pump. Each stroke of the pump's piston changed the pressure surrounding a compliant silicone left ventricle, causing ejection through the aortic valve. A recirculating fluid of 36% by volume glycerin solution in normal saline solution was used as a blood analogue fluid, which mimicked blood viscosity at 37°C when tested at room temperature. The physiologic circulation was simulated through viscoelastic ventricular contraction, blood-simulating fluid, and control of local compliance and peripheral resistance [9–11]. Pulse duplicator input variables were used to match International Organization for Standardization (ISO) 5840 and Food and Drug Administration standards for testing heart valves: heart rate of 70 beats/min, 35% systolic duration of cycle period, mean atrial and aortic pressures of 10 and 100 mm Hg, and cardiac output 5 L/min [12, 13]. These hemodynamic variables were maintained constant throughout the study.

Hemodynamic Measurements
Valve hemodynamics were evaluated with three variables: effective orifice area, mean pressure gradient, and regurgitant volume. Pressure was measured in the left atrium, left ventricle, left ventricular outflow tract, and ascending aorta with strain gauge pressure transducers (Cobe Laboratories Inc, Lakewood, CO). An electromagnetic flowmeter (Carolina Medical Electronics Inc, Oldtown, NC) was used to measure the aortic valve flow rate and the regurgitant volume. Effective orifice area within the TAV was then calculated using the Gorlin equation. Regurgitant fraction was also calculated, defined as the aortic retrograde flow divided by systolic ejection flow. Two-dimensional echocardiography (ACUSON Sequoia C256, Siemens Medical Solutions USA, Inc, Malvern, PA) was used to identify leakage and leakage location and to assess valve opening and closing processes.

Data Acquisition and Analyses
To study the efficacy of VIV implantation, the following experiments were conducted in the pulse duplicator: First, supravalvular TAVs were tested alone in the pulse duplicator to determine the efficacy of the TAV before implantation. Data acquisition was run over 10 consecutive cardiac cycles, and transvalvular pressure gradient, regurgitant volume, and effective orifice area were determined. Then, a normal Carpentier-Edwards PERIMOUNT aortic bioprostheses was tested to obtain a hemodynamic baseline. Subsequently, the same sized degenerated Carpentier-Edwards PERIMOUNT aortic bioprostheses was tested in the pulse duplicator. Finally after balloon predilation of the degenerated bioprosthesis, the supravalvular TAV was implanted within the bioprosthesis. Measurements were performed for three bioprosthetic valve sizes (19, 21, and 23 mm; n = 3 of each bioprosthetic size). Hemodynamic measurements were compared using a one-way analysis of variance (ANOVA). Reported values are quoted as mean ± standard deviation and statistical analyses were performed using MATLAB 7.0 software (Natick, MA).


    Results
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 
The 23-mm supravalvular TAVs made in the laboratory had a mean pressure gradient of 5.2 ± 3.7 mm Hg and an effective orifice area of 2.5 ± 0.7 cm2 when tested alone in the pulse duplicator. Normal Carpentier-Edwards PERIMOUNT bioprostheses were tested in the pulse duplicator to compare VIV hemodynamics with standard surgical aortic valve replacement. A summary of hemodynamic results from the normal Carpentier-Edwards PERIMOUNT bioprostheses is presented in Table 1. Bioprostheses with simulated degeneration achieved the desired gradient, mean pressure gradient was 53.9 ± 3.8 mmHg in 19 mm, 49.3 ± 3.1 mm Hg in 21 mm, and 54.0 ± 3.5 mm Hg in 23 mm valves (Fig 4A). The 23-mm supravalvular TAV reduced mean pressure gradient significantly for all degenerated bioprostheses (Fig 4A). In the 23-mm degenerated bioprostheses, mean pressure gradient decreased significantly from 54.0 ± 3.5 to 9.2 ± 6.3 mm Hg (p < 0.001). Furthermore, a significant reduction in the mean pressure gradient was observed after 23-mm supravalvular TAV implantation in the 21-mm (49.3 ± 3.1 to 14.4 ± 4.7 mm-Hg, p < 0.001) and in the 19-mm degenerated bioprostheses (53.9 ± 3.8 to 28.3 ± 9.8 mm-Hg, p = 0.013).


View this table:
[in this window]
[in a new window]

 
Table 1 Hemodynamic Characteristics of Normal Carpentier-Edwards PERIMOUNT Bioprosthesis
 

Figure 4
View larger version (31K):
[in this window]
[in a new window]

 
Fig 4. (A) Mean transvalvular pressure gradient of bioprostheses before and after transcatheter aortic valve (TAV) implantation (*p < 0.02). (B) Effective orifice area of bioprostheses before and after TAV implantation (*p < 0.025. (C) Regurgitant volume of bioprostheses before and after TAV implantation (*p ≤ 0.02). Mean data are presented with the standard deviation (error bar).

 
The 23-mm supravalvular TAV significantly increased the effective orifice area of all degenerated bioprostheses (Fig 4B). Effective orifice area increased from 0.70 ± 0.02 to 1.87 ± 0.55 (p = 0.021) in the 23-mm degenerated bioprostheses, from 0.74 ± 0.02 to 1.39 ± 0.22 (p =0.007) in 21-mm, and from 0.70 ± 0.02 to 1.00 ± 0.14 (p = 0.024) in 19-mm degenerated bioprostheses.

However, regurgitant fraction also increased significantly in all bioprosthetic valves after 23-mm supravalvular TAV implantation (Fig 4C). Regurgitant fraction of supravalvular TAVs at baseline was determined in the pulse duplicator and found to be 12.58% ± 0.77%. Two-dimensional echocardiography assessment of the valve showed that the leakage was mainly paravalvular. After VIV implantation, regurgitant fraction increased significantly, from 4.34% ± 1.36% to 17.09% ± 1.65% (p < 0.001) in 19-mm bioprosthesis, from 4.82% ± 1.61% to 16.86% ± 5.35% in 21-mm bioprostheses (p = 0.020), and from 5.87 ± 1.81% to 20.58 ± 3.51% (p <0.001) in 23-mm bioprostheses.

VIV hemodynamics for each bioprosthetic size were compared with hemodynamics of normal Carpentier-Edwards PERIMOUNT bioprostheses to compare VIV therapy with surgical rereplacement with equivalently sized bioprostheses. Supravalvular TAV reduced mean pressure gradient to 9.2 ± 6.3 mm Hg in the 23-mm degenerated bioprostheses, which was not significantly different than the mean pressure gradient of 5.7 ± 0.9 mm Hg (p = 0.422) for normal 23-mm PERIMOUNT bioprostheses. Similarly, VIV mean transvalvular gradient of 14.4 ± 4.7 mm Hg in 21-mm bioprostheses was not significantly different than the 11.8 ± 1.9 mm Hg (p = 0.306) of normal 21-mm PERIMOUNT bioprostheses, Even in 19-mm bioprostheses, although the obtained VIV mean pressure gradient of 28.3 ± 9.8 mm Hg was higher than that of the normal 19-mm PERIMOUNT bioprostheses (16.2 ± 2.2 mm Hg), the difference was not statistically significant (p = 0.056).


    Comment
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 
This study describes a supravalvular TAV designed for treatment of bioprosthetic degeneration. Although current transcatheter aortic valves have demonstrated encouraging results for a 21-mm PERIMOUNT and 23-mm Mitralflow bioprosthesis, published clinical experience remains limited [1–3]. We have previously shown that the current Edwards SAPIEN intravalvular TAV design resulted in unacceptable hemodynamics within small normal bioprostheses in in vitro studies, particularly the 19-mm size [5]. The fundamental principle of our supravalvular design is to move the valve leaflets above the bioprosthesis to avoid constraining the size of the valve by the bioprosthetic annulus and stent posts. A Dacron-covered stainless stent within the degenerated bioprosthesis would yield the maximum circular orifice area achievable within the bioprosthesis and minimize paravalvular leakage, while the valve leaflets within the open stent above would allow flow into the coronary orifices. In this study, we examined the hemodynamics of our supravalvular TAV within pericardial bioprostheses degenerated using BioGlue in a pulse duplicator setup.

Hemodynamic performance of 23-mm supravalvular TAVs was studied within Carpentier-Edwards pericardial bioprostheses. The supravalvular TAV successfully relieved the bioprosthetic stenosis in all three bioprosthetic sizes. The obtained VIV transvalvular gradients were comparable with standard surgical valve replacement of equivalent size. For a 19-mm degenerated bioprosthesis, however, a lower pressure gradient can be expected from surgical aortic valve replacement with a 19-mm valve. The corresponding difference in effective orifice area of 1.00 cm2 for VIV in a 19-mm bioprostheses vs 1.28 cm2 for rereplacement with a 19-mm bioprosthesis may or may not have significant clinical effect depending on the patient's body surface area and comorbidities that affect the risk/benefit ratio of open and minimally invasive procedures.

VIV implantation in these circumstances could still benefit nonsurgical and very high-risk patients. Careful clinical judgment regarding the surgical risks of reoperative valve replacement and the severity of dysfunction of the degenerated bioprosthesis should be taken into account before decisions are made between an operation and VIV implantation in these instances. Overall, our supravalvular TAVs demonstrated good hemodynamics within small-sized degenerated bioprostheses and were effective in relieving bioprosthetic stenosis.

A potential complication not encountered with surgical rereplacement with a bioprosthesis was the increase in regurgitation with VIV implantation. Regurgitant volume increased significantly after implantation of 23-mm supravalvular TAVs in all three bioprostheses. The leak was mainly paravalvular, which was assessed using 2-dimensional echocardiography. Intravalvular leak, which was caused by incomplete coaptation of leaflets, was also observed in the 19-mm bioprosthesis. The top portion of the supravalvular TAV could not be fully expanded in the 19-mm bioprosthesis because the lower portion was tightly constrained by the small size bioprosthesis. Therefore, the TAV leaflets were distorted because of an excess of pericardial tissue relative to the stent orifice area (Fig 5). A more flexible stent design using a nitinol stent in the supravalvular portion may facilitate the expansion of the supravalvular TAV and improve hemodynamics. Paravalvular leak has similarly been a problem for clinical implantation of Edwards SAPIEN TAV within native stenosed aortic valves [14, 15].


Figure 5
View larger version (77K):
[in this window]
[in a new window]

 
Fig 5. (A) A 23-mm supravalvular transcatheter aortic valve (TAV) implanted within a 21-mm bioprosthesis. (B) A 23-mm supravalvular TAV implanted within a 19-mm bioprosthesis.

 
The risks of reoperative valve procedures may have to be weighed against an imperfect result with VIV, where hemodynamic stenosis is relieved with some degree of aortic regurgitation. Owing to the irregularity of the landing zone in both bioprosthetic and native aortic stenosis, complete coaptation of the TAV stent to the bioprosthesis or native annulus would not be expected and some degree of paravalvular leakage would be expected.

The VIV prostheses were stable and no TAV migration was observed. Placing a larger valve above the bioprosthesis helped to avoid any TAV displacement, especially during the diastolic phase where the dominant drag force was applied to TAV [16, 17]. Furthermore, because the lower portion of the stent was to hold the TAV within the bioprosthesis and the valve was located above the bioprosthesis, no TAV leaflet distortion or traumatic injury to the leaflets by calcification was observed within the 21- and 23-mm bioprostheses. Zegdi and colleagues [18] described such leaflet distortion during TAV implantation within native aortic valves.

The next step for determining the effect of the supravalvular TAV on bioprosthetic stenosis is to document the effect of this design on coronary flows after implantation. It is unknown whether movement of the valve above the bioprosthesis improves hemodynamics but adversely affects coronary flows. The height of the supravalvular TAV in relation to the height of the aortic sinuses is crucial for successful implantation of the 23-mm supravalvular TAV. Ideally, the supravalvular portion remains within the sinus; however, if it extends into the ascending aorta, the diameter of the ascending aorta above the sinuses is also a critical variable. The supravalvular TAV completely within the aortic sinus or smaller in diameter than the ascending aorta if extending to the ascending aorta should not obstruct coronary flow because the open stent would allow blood flow through the stent during valve closure.

In some patients, the coronaries reside close to the sowing cuff of the bioprosthesis or near the stent posts. Because the bioprosthetic cuff and stent posts constrain the TAV expansion in this region, we hypothesize that the covered Dacron stent would not obstruct the coronaries unless the degenerated bioprosthetic leaflets themselves had sufficient bulky calcification to obstruct low lying coronaries. The open valve stent of the TAV above the bioprosthesis expanded to full 23-mm size would allow flow into the sinus and coronaries during diastole while maximizing orifice area, particularly in small sized bioprostheses.

The effect of this design in altering the amount of coronary flow during diastole and pattern of flow should be studied not only experimentally but computationally. Precise echocardiographic measurement is required to determine the size of the ascending aorta and sinuses before TAV implantation. Future in vitro pulse duplicator and in vivo animal studies are required to address the effect of supravalvular TAV on coronary flow before safety trials in humans.

The primary limitation of the study is that the bioprosthetic valves used in our experiments were not degenerated and we used BioGlue to simulate calcification. TAV implantation within degenerated bioprosthesis could be more complicated due to leaflet calcification, stent deformation, or pannus. Although we cannot directly extrapolate our results to clinical practice, our in vitro study suggests that the 23-mm supravalvular TAV provides acceptable hemodynamics in 19-, 21-, and 23-mm degenerated bioprostheses. The regurgitant volume for VIV was higher than for a bioprosthesis, but the pressure gradient and effective orifice area were comparable with surgical valve replacement. The paravalvular leakage may have arisen from various sources, such as the suture line between the leaflets and Dacron sheet of the TAV itself, gaps between the tip of bioprosthetic leaflet and the base of the supravalvular TAV leaflet, or gaps between the bioprosthesis and TAV stent. Nevertheless, lower regurgitant volume is expected using blood instead of the working fluid, which does not have any coagulation properties.

In conclusion, we have developed a supravalvular TAV designed to treat bioprosthetic degeneration. This novel TAV consists of two components: a Dacron-covered stainless steel stent that sits within the bioprosthesis and a valve situated above the bioprosthesis. Successful VIV implantation was performed using this supravalvular TAV for treatment of bioprosthetic stenosis. The 23-mm supravalvular TAV successfully relieved bioprosthetic stenosis, and the obtained transvalvular pressure gradient was comparable with standard surgical valve replacement of equivalent size. Further studies are mandatory to examine the effect of the supravalvular TAV on coronary flows as well as device safety and efficacy in patients.


    Acknowledgments
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 
This work was supported by American Heart Association (AHA) postdoctoral fellowship, AHA Beginning Grant-in-Aid 0565148Y, the Northern California Institute for Research and Education, and Fédération Française de Cardiologie/Société Française de Cardiologie, Société Française de Chirurugie Thoracique et Cardio-Vasculaire. We thank Christine Frolich for providing us with normal Carpentier-Edwards PERIMOUNT bioprostheses and John Hiebert for providing us with BioGlue.


    References
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 Acknowledgments
 References
 

  1. Walther T, Falk V, Dewey T, et al. Valve-in-a-valve concept for transcatheter minimally invasive repeat xenograft implantation J Am Coll Cardiol 2007;50:56-60.[Abstract/Free Full Text]
  2. Walther T, Kempfert J, Borger MA, et al. Human minimally invasive off-pump valve-in-a-valve implantation Ann Thorac Surg 2008;85:1072-1073.[Abstract/Free Full Text]
  3. Webb JG. Transcatheter valve in valve implants for failed prosthetic valves Catheter Cardiovasc Interv 2007;70:765-766.[Medline]
  4. Wenaweser P, Buellesfeld L, Gerckens U, Grube E. Percutaneous aortic valve replacement for severe aortic regurgitation in degenerated bioprosthesis: the first valve in valve procedure using the Corevalve Revalving system Catheter Cardiovasc Interv 2007;70:760-764.[Medline]
  5. Azadani AN, Jaussaud N, Matthews PB, et al. Aortic valve-in-valve implantation: impact of transcatheter- bioprosthesis size mismatch J. Heart Valve Dis 2009;18:367-373.[Medline]
  6. Roselli EE, Smedira NG, Blackstone EH. Failure modes of the Carpentier-Edwards pericardial bioprosthesis in the aortic position J Heart Valve Dis 2006;15:421-427discussion 427–8.[Medline]
  7. Myken PS, Berggren HE, Larsson S, Roberts B, Wallentin I, Caidahl K. Long-term Doppler echocardiographic results of aortic or mitral valve replacement with Biocor porcine bioprosthesis J Thorac Cardiovasc Surg 1998;116:599-608.[Abstract/Free Full Text]
  8. Vogt PR, Brunner-LaRocca H, Sidler P, et al. Reoperative surgery for degenerated aortic bioprostheses: predictors for emergency surgery and reoperative mortality Eur J Cardiothorac Surg 2000;17:134-139.[Abstract/Free Full Text]
  9. Baldwin JT, Campbell A, Luck C, Ogilvie W, Sauter J. Fluid dynamics of the CarboMedics kinetic bileaflet prosthetic heart valve Eur J Cardiothorac Surg 1997;11:287-292.[Abstract/Free Full Text]
  10. De Paulis R, Schmitz C, Scaffa R, Nardi P, Chiariello L, Reul H. In vitro evaluation of aortic valve prosthesis in a novel valved conduit with pseudosinuses of Valsalva J Thorac Cardiovasc Surg 2005;130:1016-1021.[Abstract/Free Full Text]
  11. Gerosa G, Tarzia V, Rizzoli G, Bottio T. Small aortic annulus: the hydrodynamic performances of 5 commercially available tissue valves J Thorac Cardiovasc Surg 2006;131:1058-1064.[Abstract/Free Full Text]
  12. Food and Drug Administration Draft Replacement Heart Valve Guidance 1994.
  13. International Organization for Standardization Cardiovascular implants–cardiac valve prostheses ISO 5840 2005:1-83.
  14. Cribier A, Eltchaninoff H, Tron C, et al. Treatment of calcific aortic stenosis with the percutaneous heart valve: mid-term follow-up from the initial feasibility studies: the French experience J Am Coll Cardiol 2006;47:1214-1223.[Abstract/Free Full Text]
  15. Webb JG, Pasupati S, Humphries K, et al. Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis Circulation 2007;116:755-763.[Abstract/Free Full Text]
  16. Dwyer HA, Matthews PB, Azadani AN, Ge L, Guy TS, Tseng EE. Migration forces of transcatheter aortic valves in patients with noncalcific aortic insufficiency. J Thorac Cardiovasc Surg [in press].
  17. Dwyer HA, Matthews PB, Azadani AN, et al. Hemodynamics of transcatheter aortic valve degeneration. Interactive Cardiovasc Thorac Surg [in press].
  18. Zegdi R, Ciobotaru V, Noghin M, et al. Is it reasonable to treat all calcified stenotic aortic valves with a valved stent?. Results from a human anatomic study in adults. J Am Coll Cardiol 2008;51:579-584.[Abstract/Free Full Text]

Related Article

Invited Commentary
Thomas Walther and Claudia Walther
Ann. Thorac. Surg. 2009 88: 1870. [Extract] [Full Text] [PDF]



This article has been cited by other articles:


Home page
Circ Cardiovasc IntervHome page
A. N. Azadani and E. E. Tseng
Transcatheter Heart Valves for Failing Bioprostheses: State-of-the-Art Review of Valve-in-Valve Implantation
Circ Cardiovasc Interv, December 1, 2011; 4(6): 621 - 628.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. Gurvitch, A. Cheung, J. Ye, D. A. Wood, A. B. Willson, S. Toggweiler, R. Binder, and J. G. Webb
Transcatheter Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves.
J. Am. Coll. Cardiol., November 15, 2011; 58(21): 2196 - 2209.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
A. N. Azadani, N. Jaussaud, L. Ge, S. Chitsaz, T. A. M. Chuter, and E. E. Tseng
Valve-in-Valve Hemodynamics of 20-mm Transcatheter Aortic Valves in Small Bioprostheses
Ann. Thorac. Surg., August 1, 2011; 92(2): 548 - 555.
[Abstract] [Full Text] [PDF]


Home page
Interact CardioVasc Thorac SurgHome page
A. N. Azadani, N. Jaussaud, P. B. Matthews, L. Ge, T. A. M. Chuter, and E. E. Tseng
Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses
Interact CardioVasc Thorac Surg, July 1, 2010; 11(1): 70 - 77.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
T. Walther and C. Walther
Invited Commentary
Ann. Thorac. Surg., December 1, 2009; 88(6): 1870 - 1870.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Elaine E. Tseng
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Azadani, A. N.
Right arrow Articles by Tseng, E. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Azadani, A. N.
Right arrow Articles by Tseng, E. E.
Related Collections
Right arrow Valve disease
Right arrowRelated Article


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS