ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Ann Thorac Surg 2009;88:1838-1844. doi:10.1016/j.athoracsur.2009.07.007
© 2009 The Society of Thoracic Surgeons

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Feroze Mahmood
Joseph H. Gorman, III
Robert C. Gorman
Robert M. Hagberg
Kamal R. Khabbaz
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mahmood, F.
Right arrow Articles by Khabbaz, K. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mahmood, F.
Right arrow Articles by Khabbaz, K. R.
Related Collections
Right arrow Valve disease


Original Articles: Adult Cardiac

Three-Dimensional Echocardiographic Assessment of Changes in Mitral Valve Geometry After Valve Repair

Feroze Mahmood, MDa,*, Balachundhar Subramaniam, MDa, Joseph H. Gorman, III, MDd, Robert M. Levine, MDc, Robert C. Gorman, MDd, Andrew Maslow, MDa, Peter J. Panzica, MDb, Robert M. Hagberg, MDb, Swaminathan Karthik, MDb, Kamal R. Khabbaz, MDb

a Department of Anesthesia and Critical Care and Pain Management, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts
b Cardiovascular Institute, Beth Israel Deaconess Medical Center, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts
c Department of Medicine, Division of Cardiology, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts
d Gorman Cardiovascular Research Group, Glenoldan Research Laboratory, University of Pennsylvania, Glenoldan, Pennsylvania

Accepted for publication July 2, 2009.

* Address correspondence to Dr Mahmood, Vascular Anesthesia and Perioperative Echocardiography, Department of Anesthesia and Critical Care, CC-470, Deaconess 1 (West Campus), Beth Israel Deaconess Medical Center, Boston MA 02215 (Email: fmahmood{at}bidmc.harvard.edu).


    Abstract
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 References
 
Background: Application of annuloplasty rings during mitral valve (MV) repair has been shown to significantly change the mitral annular geometry. Until recently, a comprehensive two-dimensional echocardiographic evaluation of annular geometric changes was difficult owing to its nonplanar orientation. In this study, an analysis of the three-dimensional intraoperative transesophageal echocardiographic evaluation of the MV annulus is presented before and immediately after repair.

Methods: We performed three-dimensional geometric analysis on 75 patients undergoing MV repair during coronary artery bypass graft surgery for mitral regurgitation or myxomatous mitral valve disease. Geometric analysis of the MV was performed before and immediately after valve repair with full rings and annuloplasty bands. The acquired three-dimensional volumetric data were analyzed in the operating room. Specific measurements included annular diameter, leaflet lengths, the nonplanarity angle, and the circularity index. Before and after repair data were compared.

Results: Complete echocardiographic assessment of the MV was feasible in 69 of 75 patients (92%) within 2 to 3 minutes of acquisition. Placement of full rings resulted in an increase in the nonplanarity angle or a less saddle shape of the native mitral annulus (137 ± 14 versus 146 ± 14; p = 0.002. By contrast, the nonplanarity angle did not change significantly after placement of partial rings.

Conclusions: Mitral annular nonplanarity can be assessed in the operating room. Application of full annuloplasty rings resulted in the mitral annulus becoming more planar. Partial annuloplasty bands did not significantly change the nonplanarity angle. Neither of the two types of rings restored the native annular planarity.


    Introduction
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 References
 
The normal mitral valve (MV) annulus is a nonplanar, saddle-shaped structure with its highest point in the mid anterior annulus near the aortic valve and a second smaller peak near the mid portion of the posterior annulus [1–11]. The nonplanarity of the annulus is believed to contribute to leaflet stress reduction [12, 13]. Appreciation of the importance of this nonplanarity has led to the introduction of surgical repair techniques and prostheses that either maintain or restore the saddle-shape of mitral annulus [3, 6, 14]. Restoration of nonplanarity after repair can also potentially reduce leaflet stress and thus improve the durability of repair [10].

Owing to its saddle shape and difference in axes and levels of the various portions, the mitral annulus cannot be examined in a single two-dimensional ultrasound plane [7, 15–18]. The analysis of mitral annular nonplanarity has so far been performed using either video fluoroscopy to follow implanted annular markers in animals [4, 5, 19, 20] or magnetic resonance imaging [21, 22] or echocardiography in humans and animals [2, 6, 10, 23–26]. The annulus was then mathematically reconstructed off-line, thus making the information impractical for clinical use in the operating room. In recent years, the introduction of three-dimensional echocardiography has been instrumental in developing a better description and understanding of the shape of the mitral valve annulus [27]. Also, with the availability of rapid three-dimensional reconstruction, the geometric analysis of the MV is generally available with in a few minutes of acquisition [27].

In this study, we describe the feasibility and initial results of performing complex yet rapid analyses of MV annular shape routinely using intraoperative three-dimensional transesophageal echocardiography (TEE). Specifically, we compared the changes in the saddle shape of the native mitral annulus before and after repair with different annuloplasty devices by measurement of the nonplanarity angle (NPA). Our ability to appreciate the effects of annuloplasty rings and bands on the annular nonplanarity can potentially help the clinicians to select and choose the devices best suited for particular clinical situations.


    Material and Methods
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 References
 
After Institutional Review Board approval, a total of 75 patients undergoing MV surgery at our institution consented to participate in the study. Six patients were excluded from the study owing to incomplete or unsatisfactory data acquisition. Perioperative TEE examinations were carried out after induction of general anesthesia and endotracheal intubation. All studies were performed on a Sequoia C512 ultrasound system (Siemens, Mountainview, California) utilizing a V5M multiplane TEE transducer. The ultrasound system was also equipped with the TomTec (GmBH, Munich, Germany) "on-line perspective box," a commercially available computer capable of automatic triggered rotational acquisition of ultrasound images from 0 to 180 degrees in 3- to 5-degree rotations of the scan plane. All images were acquired before start of the surgical procedure during a brief period of apnea and then again after separation from cardiopulmonary bypass after MV repair.

Echocardiographic Imaging
After performing a comprehensive two-dimensional and color-flow Doppler examination [28], a three-dimensional acquisition of MV was performed. A mid-esophageal four-chamber view was obtained with the TEE, the region of interest (mitral valve) was identified, and a 5-degree rotational acquisition was started gated to the electrocardiographic R wave. The images were immediately rendered and reconstructed and analyzed utilizing the mitral valve assessment package software by TomTec (GmBH Munich, Germany) available on the perspective box. The mitral valve geometric reconstruction was performed according to the previously described methodology briefly outlined below [27].

We also carried out similar echocardiographic examination on 8 patients undergoing coronary artery bypass graft surgery with normal ejection fraction, trivial mitral regurgitation, and no other valvular pathology for assessment and calculation of "normal" geometric determinants for comparison.

Mitral Valve Three-Dimensional Reconstruction
Briefly, the mitral valve geometric reconstruction was performed in a series of sequential steps, as follows. Reference image set-up: identification of the aortic valve in the upper outer quadrant (Fig 1A). Image of interest: an end-systolic frame is identified. Landmark identification: anterior, posterior, anterolateral (AL), and posteromedial (PM) landmarks are identified on the orthogonal views from the reference image set-up, establishing the anteroposterior (AP) and mediolateral dimensions of the mitral annulus (Fig 1A, B). Line of sight adjustment: an en-face view of the mitral valve (maximal annular cross-sectional area). Mitral annulus marking: the mitral annulus is manually marked in a series of 10 rotated two-dimensional slices evenly placed between the already established AP and AL-PM dimensions of the mitral annulus (Fig 1C). Commissural identification (see Fig 1D, 1). Coaptation line identification: in a series of 10 rotated two-dimensional slices, the coaptation line is defined and manually marked (Fig 1D, Figs 2 and 3). Go Geometric analysis: in the last step, the geometric data based on the three-dimensional reconstruction are generated and analyzed.


Figure 1
View larger version (74K):
[in this window]
[in a new window]

 
Fig 1. (A) Anterior (ANT), posterior (POST), anterolateral (AL), and posteromedial (PM) landmark identification. (B) Annular markings displayed on the en-face view of the mitral valve. (C) Circumferential annular marking. (D) Identification of the commissures (1), identification of the coaptation line (2), and manually marked coaptation line (3).

 

Figure 2
View larger version (30K):
[in this window]
[in a new window]

 
Fig 2. Identification of the dimensions of the mitral valve automatically measured during geometric analysis. (AL-PM = anterolateral-posteromedial; AP = anteroposterior.)

 

Figure 3
View larger version (55K):
[in this window]
[in a new window]

 
Fig 3. Geometric measurements during mitral valve analysis. (A) Anteroposterior (AP) diameter is the distance between the anterior and the posterior landmark. (B) Anterolateral-posteromedial (AL-AP) diameter is the distance between the anterolateral and posteromedial landmarks. (C) Commissural diameter is the annular diameter at the point of identification of commissures. (D) Anterior annulus length. (E) Posterior annulus length. (F) Nonplanarity angle is the angle subtended between the anterior landmark (Ant. LM) and posterior landmark (Post. LM) at the commissural diameter.

 
Geometric Variables
The following geometric variables were measured and generated automatically by the mitral valve assessment package based on the above workflow (Figs 2, 3). Anteroposterior diameter is measured between the anterior and posterior landmarks identified in the third step of the workflow (Figs 2, 3A). The AL-PM diameter is measured between the AL and PM landmarks of the third step of the workflow (Figs 2, 3B). Commissural diameter is the diameter of the mitral annulus at the level of the two commissures (Figs 2, 3C). For annular circumference, see Figure 2. Anterior annulus length is the curve length between the endpoints of the commissural diameter through the anterior part of the annulus ring (Fig 3D). Posterior annulus length is the curve length between the endpoints of the commissural diameter through the posterior part of the annulus ring (Fig 3F). Nonplanarity angle is a three-dimensional measure of the nonplanarity of the mitral annulus (Figs 3F, 4); this is the angle subtended at the middle point of the commissural diameter, and the anterior and posterior landmarks, namely, the annular high points (Fig 4); the more nonplanar the annulus, the smaller the NPA. Circularity index is a derived ratio of AP diameter and AL-PM diameter and is automatically generated.


Figure 4
View larger version (67K):
[in this window]
[in a new window]

 
Fig 4. Measurement of nonplanarity angle by the mitral valve assessment package. (Ant = anterior; Post = posterior.)

 
Mitral Valve Geometric Assessment After Repair
The next three-dimensional data sets were also acquired as electrocardiography-triggered three-dimensional volume sets after separation from cardiopulmonary bypass and were analyzed in the same fashion. Before and after repair, geometric measurements were then compared. All geometric reconstruction and analyses were performed by an experienced echocardiographer (F.M.) after acquisition. The R-wave–gated three-dimensional acquisition of MV data sets is completed in 30 to 40 s for a patient with a heart rate of 70 beats per minute, and the reconstruction and geometric analysis and data in less than 2 minutes.

Statistical Methods
Only patients with complete data sets (before and after cardiopulmonary bypass) were included and analyzed in the study. The MV geometric data were checked for normality with skewness and kurtosis. These normally distributed data were analyzed by a paired t test for before and after repair comparisons and a Student t test for comparison of means between groups. Any p value less than 0.05 was considered significant. The software SPSS 15.0 for Windows (SPSS, Chicago, IL) was used for analyses.


    Results
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 References
 
Seventy-five consecutive patients scheduled to undergo mitral valve repair/replacement were enrolled in the study. We were able to acquire complete three-dimensional data sets (before and after repair) in 69 patients. In 6 patients, we were unable to acquire complete data sets owing to partial annulus drop. In total, there were 29 ischemic MV repairs performed for ischemic mitral regurgitation along with coronary artery bypass graft surgery and 40 for myxomatous MV disease. The ischemic MVs had a larger baseline NPA (138.06 degrees) as compared with the myxomatous MVs (132.88 degrees) and the normal patients (127 degrees), but this difference did not reach statistical significance (p = 0.07; Table 1). Before and after repair MV geometry data (circularity index and NPA) for all patients who had MV surgery with complete annuloplasty rings and partial bands for ischemic and myxomatous MV diseases is shown in Table 2. Patients whose MV repair included complete annuloplasty rings (all types) had significantly (p = 0.002) flatter native annuli after repair than before (NPA before and after repair was 137 ± 14 degrees and 146 ± 14 degrees, respectively; p = 0.002). Patients with MV repair using partial annuloplasty rings did not have any significant difference between before and after NPA (before repair 133 ± 16 degrees versus after repair 137 ± 16 degrees; p = 0.184).


View this table:
[in this window]
[in a new window]

 
Table 1 Baseline Characteristics of Nonplanarity Angle and Circularity Index for Normal, Myxomatous, and Ischemic Mitral Valves
 

View this table:
[in this window]
[in a new window]

 
Table 2 Comparison of Nonplanarity Angle and Circularity Index Between Full Rings (Ischemic/Myxomatous) and Partial Bands (Ischemic/Myxomatous)
 
The data were then analyzed based on the type of the annuloplasty device for ischemic and myxomatous MV disease. A total of 24 Carpentier-Edwards Physio-Rings (CE Rings [Edwards Lifesciences, Irwine, CA]) were used, 12 for ischemic and 12 for myxomatous mitral valves (Table 3). There were 38 Cosgrove-Edwards annuloplasty partial bands (Cosgrove bands [Edwards Lifesciences]), 13 for ischemic and 25 for myxomatous mitral valves (Table 4). Seven Medtronic (Minneapolis, MN) annuloplasty devices were used, 4 Collvin-Galloway full rings (CG-Future ring), all for ischemic mitral valves, and 3 Duran partial bands, all for myxomatous mitral valves (Table 5).


View this table:
[in this window]
[in a new window]

 
Table 3 Comparison of Before and After Repair Geometric Data for CE-Physio Rings (Full) for Ischemic and Myxomatous Mitral Valves
 

View this table:
[in this window]
[in a new window]

 
Table 4 Changes in Mitral Valve Geometry With Use of Cosgrove Partial Bands (n = 38) for Ischemic and Myxomatous Mitral Valves
 

View this table:
[in this window]
[in a new window]

 
Table 5 Changes in Mitral Valve Geometry With Use of Medtronics Duran Partial Bands and CG-Future Full Rings
 
The data were further subanalyzed based on the effects of specific annuloplasty devices on the valve geometry. The CE-Physio was the most frequently used full ring, for 34% of the cases. The use of this device resulted in a significant increase in the NPA angle, namely, the annulus became less saddle shaped whether it was used for ischemic (135.45 ± 12.45 versus 150.40 ± 15.00, p = 0.00) or myxomatous MV disease (136.87 ± 12.81 versus 147.71 ± 15.3, p = 0.043; Table 3). The use of CG-Future bands (55% of the cases), however, did not change the NPA significantly for ischemic MVs (n = 13 [139.20 ± 15.79 versus 137.90 ± 12.38, p = 0.176]) or myxomatous MVs (n = 25 [131.37 ± 16.79 versus 136.46 ± 17.44, p = 0.198; Table 4). A total of 7 Medtronic annuloplasty devices (11% of the cases) were used, 3 Duran partial bands for myxomatous MV disease and 4 CG-Future rings for ischemic MV disease (Table 5). A total of 24 patients underwent posterior leaflet resection during MV repair. Of these patients, 20 had partial bands and 4 had complete rings used for MV repair.

There was a significant reduction in the circularity index (anteroposterior diameter/anterolateral-posteromedial diameter) after MV repair with CE-Physio rings for ischemic MVs (0.96 ± 0.06 versus 0.87 ± 0.12, p = 0.030) and myxomatous MVs (0.96 ± 0.06 versus 0.82 ± 0.09, p = 0.043; Table 3). In patients with MV repair with partial CG-Future annuloplasty bands, there was no significant change in the circularity index after repair for either ischemic or myxomatous MVs (Table 4).


    Comment
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 References
 
Our study shows that the three-dimensional geometric alterations in the mitral annulus of the repaired valve structure may be determined by the nature of the implanted prosthetic device. Our results also indicate that complete annuloplasty rings increase the NPA, namely, make the mitral annulus more planar and less saddle shaped and decrease the circularity index. Furthermore, the results of this three-dimensional reconstruction were available within minutes of the acquisition of the data in the operating room.

The three-dimensional geometric reconstruction of mitral annulus in earlier studies has either been performed in animals by tracking surgically placed piezoelectric crystals or in humans with real-time three-dimensional transthoracic echocardiography or TEE [6, 8, 11, 12, 20, 21, 29]. The mitral annular nonplanarity in these studies is calculated as the ratio of annular heights (maximum and minimum) and commissural width (AHCWR), namely, the AHCWR measured relative to a plane of least-square fit through the three-dimensional annulus (Fig 5) [8, 9, 13, 23]. We used the NPA as a measure of the degree of the saddle shape of the annulus. The NPA has been used as for this purpose in other studies as well [30, 31]. Our observation of an unfavorable change in planarity of the native mitral annulus with application of complete annuloplasty rings during repair is consistent with a recent study assessing the annular planarity with a different methodology, the AHCWR [12, 29].


Figure 5
View larger version (66K):
[in this window]
[in a new window]

 
Fig 5. Comparison of the nonplanarity (NP) angle and the annular height commissural width ratio methods of assessment of the saddle shape of the mitral annulus. A: (AHMax = annular height maximum; AHMin = annular height minimum; Ant. = anterior annulus; MA plane = mitral annular plane of least square fit; Post. = posterior.) B: Assessment of mitral annular saddle shape by the nonplanarity angle.

 
Our methodology calculates the NPA as the angle subtended between the anterior and posterior annular landmarks (Figs 4, 5) and the middle point of the commissural diameter at end-systole (Fig 4) [30, 31]. A decrease in the NPA implies a more nonplanar annulus, with a corresponding increase in AHCWR (Fig 5). Similar findings were also recently reported in an animal study model in which authors reported almost a 50% reduction in the AHCWR with application of a full ring in mitral position [29].

Maintenance of annular nonplanarity has been shown to reduce leaflet stress in a finite model analysis of mitral leaflet structure [13]. Since the mitral annulus tends to become more flat during ischemic mitral regurgitation [32], repair techniques that restore its saddle shape can potentially prolong durability. Ring annuloplasty in conjunction with MV repair has been shown to prolong the durability of repair [33]. However, MV geometry (size and planarity) is significantly altered by the architecture and design of annuloplasty devices (complete versus partial and rigid versus flexible) [11, 29, 34]. A recent animal study has also demonstrated that the newly developed saddle shaped MV annuloplasty rings optimize leaflet stress and force distribution by increasing the nonplanarity of the native annulus [29]. Although there are no clinical data to support this claim, it is quite possible that annuloplasty ring selection may affect short- and long-term durability of MV repair by alteration of leaflet stress.

Further research would be needed to establish the significance and impact of these geometric alterations on leaflet stress and repair durability. The clinical significance of the availability of this information in the operating room and its impact on surgical decision making also remains to be determined. In the past, such three-dimensional geometric analysis was limited by cumbersome, time-consuming acquisition and off-line analysis. Our study describes that the three-dimensional acquisition and analysis can both be performed within minutes, with a software package tailored specifically to MV geometry [27]. Preoperative availability of information about the three-dimensional structure of the MV raises the possibility of the suitable selection of annuloplasty devices to achieve best results in individual patients.

Limitations of the Study
We have utilized a new methodology of assessment of planarity of the MV annulus, which would need further studies for validation. However, our results as well as prior results of application of NPA are consistent with studies utilizing the conventional methods of three-dimensional reconstruction. Our three-dimensional evaluation is based on R-wave–gated scan plane rotation, which can make the volumetric acquisition time consuming and difficult. However, the system can be adjusted to customize the R-R intervals to facilitate the acquisition, slightly prolonging the time to acquire the data sets.

In conclusion, the application of annuloplasty rings during MV repair surgery results in profound geometric changes in the architecture of the MV. The geometric changes can be measured before and after valve repair utilizing three-dimensional TEE in the operating room. Complete rings increase the NPA of the MV (namely, make the native mitral annulus less saddle shaped) as compared with the partial rings. Neither of the two types of rings restores the NPA to the normal range.


    References
 Top
 Abstract
 Introduction
 Material and Methods
 Results
 Comment
 References
 

  1. Fedak PW, McCarthy PM, Bonow RO. Evolving concepts and technologies in mitral valve repair Circulation 2008;117:963-974.[Free Full Text]
  2. Flachskampf FA, Chandra S, Gaddipatti A, et al. Analysis of shape and motion of the mitral annulus in subjects with and without cardiomyopathy by echocardiographic 3-dimensional reconstruction J Am Soc Echocardiogr 2000;13:277-287.[Medline]
  3. Gillinov AM, Cosgrove DM, Shiota T, et al. Cosgrove-Edwards annuloplasty system: midterm results Ann Thorac Surg 2000;69:717-721.[Abstract/Free Full Text]
  4. Glasson JR, Komeda M, Daughters GT, Bolger AF, Ingels NB, Miller DC. Loss of three-dimensional canine mitral annular systolic contraction with reduced left ventricular volumes Circulation 1996;94:II152-II158.[Medline]
  5. Glasson JR, Komeda M, Daughters GT, et al. Three-dimensional dynamics of the canine mitral annulus during ischemic mitral regurgitation Ann Thorac Surg 1996;62:1059-1068.[Abstract/Free Full Text]
  6. Gorman JH, Gorman RC, Jackson BM, Enomoto Y, St John-Sutton MG, Edmunds LH. Annuloplasty ring selection for chronic ischemic mitral regurgitation: lessons from the ovine model Ann Thorac Surg 2003;76:1556-1563.[Abstract/Free Full Text]
  7. Kaplan SR, Bashein G, Sheehan FH, et al. Three-dimensional echocardiographic assessment of annular shape changes in the normal and regurgitant mitral valve Am Heart J 2000;139:378-387.[Medline]
  8. Levine RA, Handschumacher, MD, Sanfilippo AJ, et al. Three-dimensional echocardiographic reconstruction of the mitral valve, with implications for the diagnosis of mitral valve prolapse Circulation 1989;80:589-598.[Abstract/Free Full Text]
  9. Ryan L, Jackson B, Parish L, et al. Quantification and localization of mitral valve tenting in ischemic mitral regurgitation using real-time three-dimensional echocardiography Eur J Cardiothorac Surg 2007;31839–4.
  10. Watanabe N, Ogasawara Y, Yamaura Y, et al. Mitral annulus flattens in ischemic mitral regurgitation: geometric differences between inferior and anterior myocardial infarction: a real-time 3-dimensional echocardiographic study Circulation 2005;112:I458-I462.[Medline]
  11. Yamaura Y, Yoshikawa J, Yoshida K, Hozumi T, Akasaka T, Okada Y. Three-dimensional analysis of configuration and dynamics in patients with an annuloplasty ring by multiplane transesophageal echocardiography: comparison between flexible and rigid annuloplasty rings J Heart Valve Dis 1995;4:618-622.[Medline]
  12. Ryan LP, Jackson BM, Enomoto Y, et al. Description of regional mitral annular nonplanarity in healthy human subjects: a novel methodology J Thorac Cardiovasc Surg 2007;134:644-648.[Abstract/Free Full Text]
  13. Salgo IS, Gorman JH, Gorman RC, et al. Effect of annular shape on leaflet curvature in reducing mitral leaflet stress Circulation 2002;106:711-717.[Abstract/Free Full Text]
  14. Liddicoat JR, Mac Neill BD, Gillinov AM, et al. Percutaneous mitral valve repair: a feasibility study in an ovine model of acute ischemic mitral regurgitation Catheter Cardiovasc Interv 2003;60:410-416.[Medline]
  15. Garwood S. Con: single-plane echocardiography does not provide an accurate and adequate examination of the native mitral valve J Cardiothorac Vasc Anesth 2002;16:515-520.[Medline]
  16. Pepi M, Barbier P, Doria E, Bortone F, Tamborini G. Intraoperative multiplane vs biplane transesophageal echocardiography for the assessment of cardiac surgery Chest 1996;109:305-311.[Abstract/Free Full Text]
  17. Pu M, Griffin BP, Vandervoort PM, Leung DY, Cosgrove DM, Thomas JD. Intraoperative validation of mitral inflow determination by transesophageal echocardiography: comparison of single-plane, biplane and thermodilution techniques J Am Coll Cardiol 1995;26:1047-1053.[Abstract]
  18. Stewart WJ, Griffin B, Thomas JD. Multiplane transesophageal echocardiographic evaluation of mitral valve disease Am J Card Imaging 1995;9:121-128.[Medline]
  19. Glasson JR, Komeda MK, Daughters GT, et al. Three-dimensional regional dynamics of the normal mitral anulus during left ventricular ejection J Thorac Cardiovasc Surg 1996;111:574-585.[Abstract/Free Full Text]
  20. Tsakiris AG, Von Bernuth G, Rastelli GC, Bourgeois MJ, Titus JL, Wood EH. Size and motion of the mitral valve annulus in anesthetized intact dogs J Appl Physiol 1971;30:611-618.[Free Full Text]
  21. Komoda T, Hetzer R, Uyama C, et al. Mitral annular function assessed by 3D imaging for mitral valve surgery J Heart Valve Dis 1994;3:483-490.[Medline]
  22. Young AA, Kramer CM, Ferrari VA, Axel L, Reichek N. Three-dimensional left ventricular deformation in hypertrophic cardiomyopathy Circulation 1994;90:854-867.[Abstract/Free Full Text]
  23. Levine RA, Triulzi MO, Harrigan P, Weyman AE. The relationship of mitral annular shape to the diagnosis of mitral valve prolapse Circulation 1987;75:756-767.[Abstract/Free Full Text]
  24. Ormiston JA, Shah PM, Tei C, Wong M. Size and motion of the mitral valve annulus in man. I. A two-dimensional echocardiographic method and findings in normal subjects Circulation 1981;64:113-120.[Abstract/Free Full Text]
  25. Ryan LP, Jackson BM, Parish LM, et al. Regional and global patterns of annular remodeling in ischemic mitral regurgitation Ann Thorac Surg 2007;84:553-559.[Abstract/Free Full Text]
  26. Tibayan FA, Rodriguez F, Langer F, et al. Annular remodeling in chronic ischemic mitral regurgitation: ring selection implications Ann Thorac Surg 2003;76:1549-1555.[Abstract/Free Full Text]
  27. Mahmood F, Karthik S, Subramaniam B, et al. Intraoperative application of geometric three-dimensional mitral valve assessment package: a feasibility study J Cardiothorac Vasc Anesth 2008;22:292-298.[Medline]
  28. Shanewise JS, Cheung AT, Aronson S, et al. ASE/SCA guidelines for performing a comprehensive intraoperative multiplane transesophageal echocardiography examination: recommendations of the American Society of Echocardiography council for intraoperative echocardiography and the Society of Cardiovascular Anesthesiologists task force for certification in perioperative transesophageal echocardiography Anesth Analg 1999;89:870-884.[Free Full Text]
  29. Ryan LP, Jackson BM, Hamamoto H, et al. The influence of annuloplasty ring geometry on mitral leaflet curvature Ann Thorac Surg 2008;86:749-760.[Abstract/Free Full Text]
  30. Kwan J, Qin JX, Popovic ZB, Agler DA, Thomas JD, Shiota T. Geometric changes of mitral annulus assessed by real-time 3-dimensional echocardiography: becoming enlarged and less nonplanar in the anteroposterior direction during systole in proportion to global left ventricular systolic function J Am Soc Echocardiogr 2004;17:1179-1184.[Medline]
  31. Kwan J, Yeom BW, Jones M, et al. Acute geometric changes of the mitral annulus after coronary occlusion: a real-time 3D echocardiographic study J Korean Med Sci 2006;21:217-223.[Medline]
  32. Kaji S, Nasu M, Yamamuro A, et al. Annular geometry in patients with chronic ischemic mitral regurgitation: three-dimensional magnetic resonance imaging study Circulation 2005;112:I409-I414.[Medline]
  33. Cohn LH, Couper GS, Aranki SF, Rizzo RJ, Kinchla NM, Collins JJ. Long-term results of mitral valve reconstruction for regurgitation of the myxomatous mitral valve J Thorac Cardiovasc Surg 1994;107:143-151.[Abstract/Free Full Text]
  34. van Rijk-Zwikker GL, Mast F, Schipperheyn JJ, Huysmans HA, Bruschke AV. Comparison of rigid and flexible rings for annuloplasty of the porcine mitral valve Circulation 1990;82:IV58-IV64.[Medline]



This article has been cited by other articles:


Home page
Ann. Thorac. Surg.Home page
M. Vergnat, B. M. Jackson, A. T. Cheung, S. J. Weiss, S. J. Ratcliffe, M. J. Gillespie, Y. J. Woo, J. E. Bavaria, M. A. Acker, R. C. Gorman, et al.
Saddle-Shape Annuloplasty Increases Mitral Leaflet Coaptation After Repair for Flail Posterior Leaflet
Ann. Thorac. Surg., September 1, 2011; 92(3): 797 - 803.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart J Cardiovasc ImagingHome page
E. G. Caiani, L. Fusini, F. Veronesi, G. Tamborini, F. Maffessanti, P. Gripari, C. Corsi, M. Naliato, M. Zanobini, F. Alamanni, et al.
Quantification of mitral annulus dynamic morphology in patients with mitral valve prolapse undergoing repair and annuloplasty during a 6-month follow-up
Eur Heart J Cardiovasc Imaging, May 1, 2011; 12(5): 375 - 383.
[Abstract] [Full Text] [PDF]


Home page
The EAE Textbook of EchocardiographyHome page
M. Monaghan and S. Adhya
Chapter 3 Three dimensional echocardiography
The EAE Textbook of Echocardiography, March 1, 2011; 1(1): med-9780199599639-chapter - med-9780199599639-chapter.
[Abstract] [Full Text]


Home page
Circ Cardiovasc ImagingHome page
D. B. Ennis, G. R. Rudd-Barnard, B. Li, C. G. Fonseca, A. A. Young, B. R. Cowan, and R. A. H. Stewart
Changes in Mitral Annular Geometry and Dynamics With {beta}-Blockade in Patients With Degenerative Mitral Valve Disease
Circ Cardiovasc Imaging, November 1, 2010; 3(6): 687 - 693.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
F. Mahmood, J. H. Gorman III, B. Subramaniam, R. C. Gorman, P. J. Panzica, R. C. Hagberg, A. B. Lerner, P. E. Hess, A. Maslow, and K. R. Khabbaz
Changes in Mitral Valve Annular Geometry After Repair: Saddle-Shaped Versus Flat Annuloplasty Rings
Ann. Thorac. Surg., October 1, 2010; 90(4): 1212 - 1220.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Feroze Mahmood
Joseph H. Gorman, III
Robert C. Gorman
Robert M. Hagberg
Kamal R. Khabbaz
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mahmood, F.
Right arrow Articles by Khabbaz, K. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mahmood, F.
Right arrow Articles by Khabbaz, K. R.
Related Collections
Right arrow Valve disease


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS