|
|
||||||||
a Department of Thoracic and Cardiovascular Surgery, Aarhus University Hospital, Skejby, Denmark
b Department of Cardiology, Aarhus University Hospital, Skejby, Denmark
c Department of Anesthesiology, Aarhus University Hospital, Skejby, Denmark
Accepted for publication November 10, 2008.
* Address correspondence to Dr Thuesen, Department of Cardiology, Aarhus University Hospital, Skejby, Aarhus N, 8200, Denmark (Email: leif.thuesen{at}ki.au.dk).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
An 82-year-old woman underwent aortic valve replacement (21-mm Mitroflow) in 2002 because of severe aortic valve stenosis. Two years later, she was treated with a DDD pacemaker because of sick sinus syndrome. The patient also had intermittent claudication because of stenoses of the iliac and femoral arteries. Since 2007, the patient experienced a rapidly deteriorating condition with shortness of breath, chest pain, and overt heart failure.
Significant stenoses of the right and the circumflex coronary arteries were treated with drug-eluting stents, but with no clinical improvement. The aortic bioprosthesis was found to be severely stenosed, with a valve area of 0.4 cm2 and a peak gradient of 100 mm Hg according to transthoracic (TTE) and transesophageal echocardiography (TEE). Left ventricular function was normal. The systolic pulmonary pressure was 65 mm Hg according to TTE. Renal function was moderately impaired (serum creatinine, 140 µmol/L).
The patient was considered for conventional aortic valve replacement, but with a logistic European System for Cardiac Operative Risk Evaluation (EuroSCORE) of 53%, the perioperative risk was considered unacceptably high. Instead, we suggested dilatation of the stenotic prosthesis and transapical insertion of an Edwards SAPIEN THV. We were aware—and the patient was informed—that this would be an experimental and nonrecommended use of the device. In advance, we had compared a Mitroflow 21-mm valve with a 23-mm Edwards SAPIEN THV, and found it possible to cover the Mitroflow cusps if the SAPIEN THV was implanted about 3 mm below the ring of the Mitroflow valve. We anticipated that this position would secure stability and optimal functional result of the stent valve implantation.
The procedure was performed with extensive hemodynamic monitoring under general anesthesia but without artificial cardiopulmonary support. Through a 5-cm-long thoracotomy in the fifth left intercostal space and after insertion of myocardial pacing electrodes, the left ventricle was punctured at the apex. A soft, 0.35-inch J-tip wire was inserted through the stenosed aortic valve prosthesis. A right coronary catheter was used to change to an Amplatz Extra Stiff 0.35-inch wire (Cook Medical, Bloomington, IN). A 14F introducer sheath was inserted over the guidewire, with the tip of the sheath positioned in the middle of the left ventricle.
During rapid cardiac pacing (220 beats/min), the aortic valve bioprosthesis was dilated with a 5-cm-long, 23-mm dilatation balloon catheter (Fig 1). We changed to a 26F sheath, and the 23-mm Edwards SAPIEN THV was implanted during rapid pacing (220 beats/min) within the degenerated Mitroflow valve. Fluoroscopy was used to position the inflow part of the stent valve just below the ring of the Mitroflow valve as planned (Figs 2 and 3).
|
|
|
|
| Comment |
|---|
|
|
|---|
Before implanting the Edwards SAPIEN THV, we discussed several potential problems:
We decided to position the inferior edge of the stent valve immediately below the radiopaque ring of the Mitroflow valve. We anticipated that a position too low might result in insufficient dilatation and opening of the agglutinated leaflets of the Mitroflow valve. Conversely, a position just above the Mitroflow valve ring might be too high and result in obstruction of the coronary arteries or peripheral dislodgement of the Edwards SAPIEN THV. The apical approach allowed an excellent alignment between the Edwards SAPIEN THV Ascendra delivery system and the aortic root. The choice of position was apparently a correct decision, because the implanted Edwards SAPIEN THV remained in a stable position and covered the degenerated leaflets of the Mitroflow valve.
The concern of underexpansion was handled using a 23-mm dilatation balloon for predilatation. During the balloon predilatation, we suspected and noticed that the Mitroflow valve could only be partially dilated. We found, however, that the waist of the dilatation balloon was sufficiently small to proceed with the procedure (Fig 1). During and after the implantation of the Edwards SAPIEN THV, a moderate waist of the stent was observed. A smaller waist may be seen after Edwards SAPIEN THV treatment of genuine aortic valve stenoses. The moderate waist of the stent valve did not result in failure of the valve leaflet function. However, the waist probably resulted in the Doppler pressure gradient of 40 mm Hg across the Edwards SAPIEN THV and a valve area of only 1 cm2, features less favorable than after implantation of a stent valve in a native aortic valve stenosis.
At present, the Edwards SAPIEN THV Valve is commercially available in a 23- and a 26-mm version. When used in stenotic genuine aortic valves, the stent valve is usually oversized by 2 to 3 mm to secure stability and minimize paravalvular leakage. It is possible that oversizing should not be done in the valve-in-valve situation, and that a 21-mm device (not commercially available) might have performed better in the present case. Nevertheless, the treatment resulted in marked hemodynamic and clinical improvement compared with the pretreatment situation
As in previous reports, the present case may give rise to some optimism concerning stent valve treatment of degenerated bioprosthetic heart valves. However, these procedures are still highly experimental, and extended follow-up are needed to evaluate the potential of the treatment.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. H. Ong, R. Bauernschmitt, G. Schuler, and R. Mueller Short- and mid-term safety and effectiveness of transcatheter aortic valve implantation in a failing surgical aortic bioprosthesis Eur J Cardiothorac Surg, August 1, 2012; 42(2): 268 - 276. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Seiffert, L. Conradi, S. Baldus, M. Knap, J. Schirmer, O. Franzen, D. Koschyk, T. Meinertz, H. Reichenspurner, and H. Treede Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses J. Thorac. Cardiovasc. Surg., March 1, 2012; 143(3): 617 - 624. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Descoutures, D. Himbert, C. Radu, B. Iung, C. Cueff, D. Messika-Zeitoun, G. Ducrocq, E. Brochet, P. Nataf, and A. Vahanian Transarterial Medtronic CoreValve System Implantation for Degenerated Surgically Implanted Aortic Prostheses Circ Cardiovasc Interv, October 1, 2011; 4(5): 488 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. H. M. Nielsen, H. Egeblad, K.-E. Klaaborg, V. E. Hjortdal, and L. Thuesen Transatrial Stent-Valve Implantation in a Stenotic Tricuspid Valve Bioprosthesis Ann. Thorac. Surg., May 1, 2011; 91(5): e74 - e76. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Silva, J.-H. Stripling, L. Hansen, and F.-C. Riess Aortic Valve Replacement After Transapical Valve-in-Valve Implantation Ann. Thorac. Surg., January 1, 2011; 91(1): e5 - e7. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Azadani, N. Jaussaud, P. B. Matthews, L. Ge, T. A. M. Chuter, and E. E. Tseng Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses Interact CardioVasc Thorac Surg, July 1, 2010; 11(1): 70 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Ferrari, C. Marcucci, C. Sulzer, and L. K. von Segesser Which available transapical transcatheter valve fits into degenerated aortic bioprostheses? Interact CardioVasc Thorac Surg, July 1, 2010; 11(1): 83 - 85. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Kempfert, A. Van Linden, A. Linke, M. A. Borger, A. Rastan, C. Mukherjee, J. Ender, G. Schuler, F. W. Mohr, and T. Walther Transapical Off-Pump Valve-in-Valve Implantation in Patients With Degenerated Aortic Xenografts Ann. Thorac. Surg., June 1, 2010; 89(6): 1934 - 1941. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. C. Maroto, J. E. Rodriguez, J. Cobiella, and P. Marcos Transapical off-pump aortic valve-in-a-valve implantation in two elderly patients with a degenerated porcine bioprosthesis Eur J Cardiothorac Surg, March 1, 2010; 37(3): 738 - 740. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |