|
|
||||||||
a St. Lukes Hospital, Thessaloniki, Greece
b Columbia University, New York, New York
c Democritus University of Thrace, Alexandroupolis, Greece
d University of Athens, Athens, Greece
Accepted for publication June 15, 2007.
* Address correspondence to Dr Pitsis, St. Lukes Hospital, Panorama, Thessaloniki, 55236, Greece (Email: apitsis{at}otenet.gr).
| Abstract |
|---|
|
|
|---|
Methods: Between August 2001 and August 2004, 11 high-risk male patients who had undergone heart failure surgery or surgical revascularization were supported by the TandemHeart postcardiotomy. The major indication for pVAD insertion was failure to wean from cardiopulmonary bypass. Three different techniques were employed for cannulation: the closed percutaneous technique, the "open transeptal" technique with percutaneous cannulas insertion, and direct central cannulation.
Results: The mean duration of support was 88 hours. The mean pump flow was 3.09 L/min. The weaning rate was 72.72%. Survival to discharge and at 1 and 4 years was 54.54%, 45.45%, and 36.36%, respectively. The main complication was pericardial bleeding, noted mainly in patients receiving antiplatelet treatment preoperatively.
Conclusions: The TandemHeart appears to be safe for temporary support after cardiotomy. It is a versatile device allowing different techniques of insertion. Device application yielded high weaning rate and satisfactory early and long-term survival.
| Introduction |
|---|
|
|
|---|
Weaning from cardiopulmonary bypass (CPB) with use of IABP is well established, although evidence for the efficacy of this approach is hard to produce [2, 3]. Cardiac output increases by 10% to 20%, mainly owing to reduced afterload and increased coronary perfusion, but IABP does not directly pump blood and does not significantly redistribute blood flow [3, 4], suggesting that the degree of hemodynamic compromise that can be treated with this device is relatively limited.
Numerous devices, which generally require major surgery for insertion and removal, have been used with similar success for postcardiotomy MCS, none being inclusive of all the characteristics of an ideal device [4]. Devices for postcardiotomy support should allow easy implantation [4] and removal, and versatility of cannulation. Devices that allow peripheral cannulation include left ventricular assist devices (LVADs) with extracorporeal centrifugal pumps, intracorporeal miniaturized axial flow pumps, and the extracorporeal membrane oxygenation (ECMO) that provides cardiopulmonary support [4–6].
The TandemHeart percutaneous ventricular assist device (Tandem Heart pVAD; Cardiac Assist, Pittsburgh, Pennsylvania) allows percutaneous placement of inflow and outflow cannulas [7–9], with a technique resembling the one introduced by Dennis and used clinically by Edmunds and coworkers [10]. The TandemHeart includes a dual chambered, paracorporeal, continuous-flow centrifugal pump. The upper pump housing provides inflow and outflow ports. The lower housing provides communication with the controller and an integral infusion line to provide local cooling and anticoagulation [8]. The device is Conformité Européenne marked. It has been most extensively used for short-term left ventricular (LV) support during high-risk percutaneous coronary interventions and for treatment of peri-infarct cardiogenic shock [8, 9].
The goals of this feasibility study were to evaluate the safety and efficacy of the TandemHeart for postcardiotomy MCS.
| Patients and Methods |
|---|
|
|
|---|
Seven patients had chronic heart failure with severely dilated hearts due to ischemic cardiomyopathy or valvular heart disease. They mainly underwent heart failure surgery including surgical ventricular remodeling. Acute heart failure occurred as decompensation of preexisting end-stage chronic heart failure. Four patients underwent coronary artery bypass graft surgery for coronary artery disease and had acute de novo heart failure due to perioperative coronary syndromes.
Indications and Criteria for TandemHeart Insertion
The main indication for TandemHeart insertion was failure to wean from CPB (9 patients). Other indications were preoperative acute decompensation of end-stage chronic heart failure (1 patient) and postcardiotomy refractory cardiac arrest (1 patient).
High doses of at least two catecholamines including epinephrine were administered to all 11 patients before initiation of mechanical assistance. In 5 patients (all 4 patients with acute ischemia and 1 patient with chronic heart failure), the TandemHeart was inserted after IABP failed. In 6 patients with chronic heart failure, the TandemHeart was inserted without previous IABP. In these patients with chronically dilated and hypokinetic hearts, criteria for MCS were failure to wean from CPB on first attempt despite maximum inotropic support, and minimal intrinsic pumping activity and marked LV dilatation not only on CPB withdrawal but even with CPB flows less than 1.5 L/min.
The patients baseline characteristics and indications for MCS are shown in Table 1.
|
|
|
|
Pump Set-Up and Patient Management
The pump speed was initially set between 5,900 rpm and the maximum speed of 7,500 rpm aiming to a cardiac index greater than 2.2 L/min measured by thermodilution, a mean systolic arterial pressure greater than 70 mm Hg, and a mixed venous oxygen saturation greater than 70%.
After weaning from CPB, heparin was partially reversed with protamine. Continuous heparin infusion through the pump incorporated in the TandemHeart system was started intraoperatively or early postoperatively to maintain an activated clotting time of 180 s to 200 s.
In 3 of 4 patients with acute ischemia, IABP was maintained during support; in 1 patient with severe peripheral atherosclerosis, the balloon inserted through the ascending aorta was removed to facilitate insertion of the TandemHeart outflow cannula. In the 7 patients with decompensated chronic heart failure, IABP was not used during support, as the LV offloading provided by the device and pulsatility provided by residual LV function were adequate.
Progressively decreasing doses of inotropes were administered during support, and weaning was generally attempted after the first 24 hours. When the patients were hemodynamically stable, without excessive inotropic and vasoactive drugs (such as epinephrine or norepinephrine) and with good vital organ function, weaning was considered. Preload, afterload, cardiac rhythm, and ventricular function were optimized. Low to moderate doses of mirlinone, dobutamine, or dopamine were administered during the weaning procedure, which lasted between 2 and 12 hours. The pump flow was gradually decreased as tolerated, to a flow of 0.5 L/min (for short periods with activated clotting time > 220 s). If under these settings the cardiac index remained above 2.2 L/min, without ventricular dilatation and with adequate ventricular contraction, weaning was decided.
Weaning from the ventilator started during support in hemodynamically stable patients, and most patients were extubated shortly after device removal.
Device Removal
In the 7 patients with percutaneous inflow and outflow cannulas insertion, the device was removed by simple withdrawal of the cannulas, followed by manual compression to achieve hemostasis at the percutaneous insertion sites. When direct cannulation was employed the device was removed in theater.
Definitions and Measurements
Cardiac output during support is defined as the right ventricular output measured by thermodilution. Residual cardiac output is defined as the cardiac output during support minus the device flow measured by the device flowmeter. Residual cardiac output in the setting of partial support LVAD represents the LV output through the aortic valve, and it is attributed to residual intrinsic function of the assisted left ventricle. Residual cardiac output equals cardiac output (thermodilution) minus LVAD flow [11].
Hemodynamic measurements immediately before initiation of support are not available because most patients were on CPB (n = 9) and 1 patient was in cardiac arrest.
Hemodynamic and pump measurements obtained within the first 2 hours from initiation of support (or reinitiation of support in the patient with preoperative insertion) are defined as initial measurements. Measurements made at a mean of 24.7 ± 32.4 hours from initiation of support are defined as measurements during support. Comparisons of these measurements were performed with the paired t test.
| Results |
|---|
|
|
|---|
|
|
|
Causes of Death
All 3 patients who were not weaned from the device were taken to theater with evolving acute myocardial infarction and died of biventricular failure (Table 5). Their left ventricle never recovered; right-sided heart failure became evident (within 0.5 hour to 8 days after initiation of support) and was rapidly followed by death or multiple organ failure.
|
Adverse Events
The most common adverse event during support was pericardial bleeding with similar incidence in central and femoral cannulation (Table 6). Five patients (45.45%) required reexploration for bleeding complications. In 1 patient, bleeding was attributed to a small vessel, and in the remaining patients, to bleeding diathesis. Diffuse bleeding requiring reexploration was significantly higher among patients receiving aspirin or clopidogrel preoperatively (
2 test, p = 0.033).
|
Documented respiratory infection presented during support in 3 patients but its relationship to the device per se was not proven. All device cannula cultures after removal were negative.
No thromboembolic or cerebrovascular events or device failures were noted. Excluding the patient in whom multiple organ failure followed the right-sided heart failure, there were no hepatic or renal complications based on clinical and laboratory data. Creatinine remained below 1.4 mg/dL, liver enzymes remained below triple the higher normal value. The mean lactate dehydrogenase was 916 ± 336 U/L. The platelet count increased after the third day of support in all 5 patients supported longer than 3 days.
On transthoracic echocardiogram within 24 hours from device removal, among the 7 patients with transeptal inflow cannula insertion, no atrial septal defect was detected in 6 patients and a small, clinically insignificant atrial septal defect was detected in 1 patient. No atrial septal defect was detected on serial echocardiography.
Blood product transfusions are shown in Table 7.
|
| Comment |
|---|
|
|
|---|
TandemHeart is a versatile VAD allowing not only percutaneous insertion but also intraoperative "open transeptal" or direct cannulation. Although not purely percutaneous, the open transeptal technique involves percutaneous cannulas insertion and allows easy removal, avoiding surgical trauma.
We performed percutaneous cannulas insertion to avoid resternotomy for device removal, prolonged mechanical ventilation, increased risk of infection, and hemodynamic and respiratory derangement. We performed direct central cannulation in patients with severe peripheral vascular disease or when support was urgently required for resuscitation in a patient with open sternum. No interference of the inflow cannula with the left ventricle with either cannulation technique had advantages since it was used in patients with acute myocardial infarction or after LV surgical remodeling.
Although not maximal or directly comparable, in our patients, the TandemHeart flow with transthoracic cannulation was higher than that with percutaneous femoral cannulation. The flow with the transthoracically inserted TandemHeart (mean 4.02 ± 1.32 L/min, maximum 6.0 L/min) is comparable with that achieved by transthoracically inserted temporary VADs with other centrifugal or pulsatile pumps, microaxial flow pumps, and ECMO [4–6, 11, 13]. Although maximum flow of 4.5 L/min has been reported with the percutaneously inserted TandemHeart [14], in our patients, the maximum flow with percutaneous insertion was 3.1 L/min at a submaximal speed of 6,200 rpm (the mean flow being 2.55 ± 0.33). This flow is less than that achieved by ECMO and the Hemopump (Medtronic, Minneapolis, MN) inserted with open femoral cannulation and higher than that of the percutaneously inserted Hemopump [6, 13].
The incidence of pericardial bleeding is typical of that reported with commonly used extracorporeal centrifugal pumps and ECMO [4, 5], but it was higher that that reported for Impella (Cardiosystems AG, Aachen, Germany) [11], Hemopump [6], and temporary pulsatile VADs [4]. The incidence of bleeding was similar with percutaneous and direct cannulation, and it was higher among patients receiving antiplatelet treatment preoperatively. Delayed sternal closure could have been an option, but not without the disadvantages of restrainement to prevent cannulas dislodgment, sedation [5], prolonged mechanical ventilation, increased risk of infection, and reoperation for closure.
Thromboelastography could guide the strategy of closure and blood product administration. Heparin withdrawal in the early postoperative hours could minimize the need of reexploration.
The incidence of vascular complications is comparable with that reported with peripherally inserted ECMO [4, 15]. Anastomosis of a vascular graft to the common femoral artery to serve as a conduit for the arterial cannula could help in avoiding leg ischemia and cannula dislocation [4].
Despite advantages of easy device removal in percutaneous cannulation, owing to flow limitation and vascular complications, we would not recommend this approach in patients with no residual LV function, peripheral atherosclerosis, and small femoral arteries.
Evaluation of results is closely related to the indications for MCS. Although the recommendation for IABP before LVAD support in acute heart failure is characterized as class I, level of evidence C [2], Baskett and colleagues [3] consider that there is class IIa, level of evidence C, for postoperative IABP in coronary artery bypass graft surgery patients, as some authors suggest [13]; or apply in practice [4, 5, 11] earlier initiation of mechanical support when the shock is less prevalent without delay after failure of IABP. The high incidence of complications after VAD implantation is directly related to prolonged attempted weaning periods from CPB [5]. Early MCS is considered justifiable if the risk for the patients is not increased by the adverse events related to the device itself [6].
Two randomized studies comparing TandemHeart support to IABP in patients with cardiogenic shock showed improved hemodynamic parameters with the TandemHeart and similar survival benefit [9, 16].
Intra-aortic balloon pumping to treat low cardiac output rather than ischemia has been associated with poorer prognosis [3]. In all patients with acute ischemia and in 1 patient with decompensated chronic heart failure, IABP preceded TandemHeart support. All these patients would have probably died as they were not responsive to all means of conventional treatment, but TandemHeart support led to survival of 2 of 5 patients (40%). In 6 patients with decompensated chronic heart failure, the TandemHeart was inserted without previous IABP. These patients would have probably died without some kind of mechanical support as they were not responsive to maximal conventional medical treatment. Although we can not produce evidence of superiority of TandemHeart in comparison with IABP (and perhaps IABP or another LVAD would have led to the same or better results), it was TandemHeart assistance that led to survival of 4 of 6 patients (66.6%).
Weaning and survival to discharge with the TandemHeart was comparable with that reported with generally accepted centrifugal pumps, ECMO, pulsatile temporary VADs, and microaxial flow pumps in similar settings. Hospital discharge rates reported for patients receiving VADs postcardiotomy range between 20% and 60%, higher survival being observed when temporary VADs are used for bridging procedures [1, 4–6, 11, 13, 15, 17–19].
All patients who died on support had preoperative acute myocardial infarction and died of biventricular failure. The extent of myocardial injury has been underestimated in these patients who might have benefited by biventricular support or ECMO. Withdrawal of the inflow cannula to the right atrium and incorporation of an oxygenator to the circuit would allow switch to ECMO. Nevertheless, given that extensive myocardial damage is present in the majority of patients not surviving after postcardiotomy MCS, it has been considered unlikely that a specific VAD or mode of support could improve outcome [15]. In failure of cardiac recovery, early conversion to long term support could offer a life-saving alternative [15]. Mehta and Pae [14] recently reported conversion of the TandemHeart to long-term support.
The cause of death in patients weaned from the device was recurrence of chronic left-sided heart failure. In these patients, heart failure surgery combined with early initiation of support to allow recovery from the operative insult yielded good results (100% weaning, 71.4% survival). Longer support to allow myocardial remodeling [20] could improve survival.
In conclusion, versatility of cannulation, minimal invasiveness, and low cost are advantages of the TandemHeart. Rapid conversions to CPB with adequate flows and to long-term support have been recently shown. The adverse events during support were typical of those reported with other devices in similar settings. Device application yielded high cumulative weaning rate and satisfactory early and midterm survival. Percutaneous cannulation requires careful patient selection and biventricular failure necessitates urgent biventricular or ECMO support. We speculate that when poor outcome is anticipated, as in cases of extensive acute ischemic injury or persisting chronic heart failure, early implementation of a strategy to long-term support might improve outcomes.
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
W. T. Brinkman, J. E. Rosenthal, E. Eichhorn, T. M. Dewey, M. J. Magee, D. S. Savor, A. G. Riley, S. L. Prince, C. M. Worley, M. A. Herbert, et al. Role of a percutaneous ventricular assist device in decision making for a cardiac transplant program. Ann. Thorac. Surg., November 1, 2009; 88(5): 1462 - 1466. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. W. Miller Continuous Aortic Flow Augmentation: Not Enough MOMENTUM Circulation, September 16, 2008; 118(12): 1223 - 1224. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |