ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Ani C. Anyanwu
David H. Adams
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Anyanwu, A. C.
Right arrow Articles by Adams, D. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Anyanwu, A. C.
Right arrow Articles by Adams, D. H.
Related Collections
Right arrow Mechanical Circulatory Assistance

Ann Thorac Surg 2007;84:1405-1407
© 2007 The Society of Thoracic Surgeons


How To Do It

Off-Pump Implant of the Jarvik 2000 Ventricular Assist Device Through Median Sternotomy

Ani C. Anyanwu, MD*, Gregory W. Fischer, MD, Irina Plotkina, PA-C, Sean Pinney, MD, David H. Adams, MD

Department of Cardiothoracic Surgery, Mount Sinai Medical Center, New York, NY

Accepted for publication February 2, 2007.

* Address correspondence to Dr Anyanwu, Department of Cardiothoracic Surgery, Mount Sinai Medical Center, 1190 Fifth Ave, Box 1028, New York, NY 10029 (Email: anelechi.anyanwu{at}mountsinai.org).


    Abstract
 Top
 Abstract
 Introduction
 Technique
 Operative Technique
 Comment
 References
 
Implantation of a left ventricular assist device through a median sternotomy usually requires cardiopulmonary bypass. However, the biological response to cardiopulmonary bypass is sometimes deleterious in end-stage heart failure patients, as it may compound pre-existing coagulopathy or multiorgan dysfunction. Therefore, there are potential advantages to avoiding cardiopulmonary bypass during left ventricular assist device placement. We describe a technique for off-pump implantation of the Jarvik 2000 left ventricular assist device (Jarvik Heart Inc, New York, NY) through a median sternotomy.


    Introduction
 Top
 Abstract
 Introduction
 Technique
 Operative Technique
 Comment
 References
 
As with other left ventricular assist devices (LVAD), the standard technique for insertion of the Jarvik 2000 LVAD (Jarvik Heart Inc, New York, NY) requires cardiopulmonary bypass (CPB) [1]. Because CPB has multiple systemic effects that compound the multiorgan dysfunction and coagulopathy that often pre-exists in LVAD patients, investigators have worked to reduce exposure to CPB during LVAD implantation. Notably, Frazier [2] and Frazier and colleagues [3] have developed techniques for off-pump implantation of the Jarvik 2000 through a left thoracotomy and subcostal incisions, respectively, and Piacentino and coworkers [4] have developed a technique for the off-pump implantation of the Heartmate XVE (Thoratec Corp, Pleasanton, CA). However, the off-pump Jarvik 2000 implantation is not undertaken through a sternotomy because of relative inaccessibility of the cardiac apex. We report a successful case of off-pump implantation of the Jarvik 2000 through a median sternotomy.


    Technique
 Top
 Abstract
 Introduction
 Technique
 Operative Technique
 Comment
 References
 
The patient was a 58-year-old man with nonischemic dilated cardiomyopathy who had been maintained on a continuous infusion of milrinone while awaiting cardiac transplantation. He was 185 cm tall, blood type O, and had not undergone prior cardiac surgery. He became increasingly decompensated, despite inotropic therapy, also with increasing serum creatinine (peak, 2.8 mg/dL). He was referred for placement of a LVAD as a bridge to cardiac transplantation. A Jarvik 2000 Flowmaker (Jarvik Heart Inc) was implanted as part of a Food and Drug Administration investigational trial. This trial was approved by our institutional review board and informed consent was obtained from the patient.


    Operative Technique
 Top
 Abstract
 Introduction
 Technique
 Operative Technique
 Comment
 References
 
General anesthesia was instituted. Milrinone infusion was maintained and supplemented by dobutamine. Transesophageal echocardiography confirmed a competent aortic valve and absence of flow across the intra-atrial septum. The left ventricle was markedly dilated with severe dysfunction (end-diastolic diameter, 97 mm). The right ventricle (RV) was also severely dilated and dysfunctional. Tricuspid regurgitation was moderate. A median sternotomy was performed and the pericardium was opened to reveal a gigantic heart. The patient was heparinized and was also cannulated in case of an emergent need for CPB. The left pleural cavity was widely opened. No pericardial retraction sutures were applied.

Using a chest tube to simulate the proposed path of the LVAD conduit, we measured the length from the left ventricle apex to the ascending aorta coursing through the left hemithorax. Routing through the left hemithorax allows a more natural lie, reducing the likelihood of kinking. The driveline was exteriorized through the right upper abdominal wall. The outflow graft was trimmed to the pre-identified length and placed inside a 20-mm ringed Gore-Tex vascular graft (W. L. Gore and Assoc, Flagstaff, AZ) trimmed to a similar length. This reinforced outer graft prevents kinking and offers protection during sternal re-entry. Care was taken to ensure there was no twisting of the outflow graft within the Gore-Tex tube (W. L. Gore & Assoc). A side-biting clamp was applied to the left side of the ascending aorta and a 15-mm arteriotomy was made. The graft was beveled and oriented such that the heel lay cephalad and slightly to the left. We ensured there was no rotation from the ventricular apex to insertion. The outflow graft was then anastomosed to the aorta using continuous 4-0 polypropelene. The graft was clamped and the side-biting clamp removed.

The patient was placed in a steep Trendelenberg position with slight rotation to the right. A suction device (Guidant XPOSE [Boston Scientific, Natick, MA]) was then deployed on the right ventricular apex, just lateral to the anterior descending artery, and was used to gradually retract the left ventricle apex into the surgical field, as described by Piacentino and colleagues [4] for off-pump insertion of the Heartmate XVE (Thoratec Corp). Arterial pressures were maintained with a vasopressin infusion and occasional catecholamine boluses. Overall, cardiac displacement was well tolerated. A circle was marked on the left ventricle apex to outline the circumference of the sewing ring. Ten interrupted plegeted 2-0 Ethibond (Ethicon Inc, Somerville, NJ) mattress sutures were placed through the left ventricle myocardium and subsequently through the sewing ring (Fig 1A). Irritation of the myocardium triggered brief runs of ventricular tachycardia, which were suppressed by administration of magnesium, lidocaine, and amiodarone. The sutures were tied to secure the sewing ring. The outflow clamp was released slightly to remove the air from the graft and the sewing ring was stabilized with four hemostats (Fig 1B). A cruciate incision was made in the apex, which was then immediately cored using the coring knife. As the surgeon removed the coring knife, the assistant simultaneously inserted the Jarvik pump into the ventricle. The pump was secured, the air was removed, and support was initiated (Fig 1C). The suction device was disengaged and the heart was released back into the left pleural cavity, heparin was reversed, and the chest was closed. The patient made an uncomplicated recovery and was discharged from the intensive care unit in 2 days, and from the hospital in 3 weeks. A postoperative computed tomographic scan showed a satisfactory position of the device (Fig 2).


Figure 1
View larger version (69K):
[in this window]
[in a new window]

 
Fig 1. Steps for off-pump Jarvik implantation (Jarvik Heart Inc, New York, NY). (A) After stabilization with suction device, sutures are passed through the myocardium and through the sewing cuff. (B) Sewing cuff has been tied down and device is filled with blood in preparation for apical coring. (C) Completed implantation.

 

Figure 2
View larger version (105K):
[in this window]
[in a new window]

 
Fig 2. Computed tomographic scan showing completed implantation. Routing the outflow graft through the left hemithorax reduces the likelihood of kinking within the pericardial cavity.

 

    Comment
 Top
 Abstract
 Introduction
 Technique
 Operative Technique
 Comment
 References
 
Avoidance of CPB during LVAD placement has several theoretical advantages including reduction in bleeding and less multiorgan injury. Although CPB can be avoided by using a thoracotomy or subcostal incision [2, 3], these approaches limit the surgeon’s ability to perform adjunctive procedures, and may have increased thromboembolism due to stasis in the ascending aorta [5]. Using a median sternotomy for off-pump device implantation offers the potential benefits of avoiding CPB, while maintaining surgical exposure and additionally allowing for the addition of adjunctive cardiac procedures if required. We have overcome the obstacles to off-pump LVAD placement through a sternotomy (ie, inaccessibility of apex and hemodynamic instability) by using a suction device and pharmacological management. Although routine off-pump insertion through a sternotomy is theoretically possible with other implantable devices [4], we believe the simplicity of the Jarvik implantation technique makes it particularly suited for this role. We have since used this technique successfully for implant of the Jarvik 2000 on a second patient. Further study is necessary to determine the applicability and benefits of this technique.


    References
 Top
 Abstract
 Introduction
 Technique
 Operative Technique
 Comment
 References
 

  1. Westaby S, Frazier OH, Pigott DW, Saito S, Jarvik RK. Implant technique for the Jarvik 2000 Heart Ann Thorac Surg 2002;73:1337-1340.[Abstract/Free Full Text]
  2. Frazier OH. Implantation of the Jarvik 2000 left ventricular assist device without the use of cardiopulmonary bypass Ann Thorac Surg 2003;75:1028-1030.[Abstract/Free Full Text]
  3. Frazier OH, Gregoric ID, Cohn WE. Initial experience with non-thoracic, extraperitoneal, off-pump insertion of the Jarvik 2000 Heart in patients with previous median sternotomy J Heart Lung Transplant 2006;25:499-503.[Medline]
  4. Piacentino 3rd V, Jones J, Fisher CA, et al. Off-pump technique for insertion of a HeartMate Vented Electric left ventricular assist device J Thorac Cardiovasc Surg 2004;127:262-264.[Free Full Text]
  5. Siegenthaler MP, Martin J, Gutwald R, et al. Anterior approach to implant the Jarvik 2000 with retroauricular power supply Ann Thorac Surg 2005;80:745-747.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
PerfusionHome page
K Griffith, E Jenkins, and F. Pagani
First American experience with the Terumo DuraHeartTM left ventricular assist system
Perfusion, March 1, 2009; 24(2): 83 - 89.
[Abstract] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Ani C. Anyanwu
David H. Adams
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Anyanwu, A. C.
Right arrow Articles by Adams, D. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Anyanwu, A. C.
Right arrow Articles by Adams, D. H.
Related Collections
Right arrow Mechanical Circulatory Assistance


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS