|
|
||||||||
Ann Thorac Surg 2007;84:303-305
© 2007 The Society of Thoracic Surgeons
a Division of Pediatric Cardiology, University Childrens Hospital Zurich, Zurich, Switzerland
b Division of Congenital Cardiovascular Surgery, University Childrens Hospital Zurich, Zurich, Switzerland
c Department of Diagnostic Imaging, University Childrens Hospital Zurich, Zurich, Switzerland
Accepted for publication October 2, 2006.
* Address correspondence to Dr Bauersfeld, Division of Pediatric Cardiology, University Childrens Hospital, Steinwiesstrasse 75, CH-8032, Zurich, Switzerland (Email: urs.bauersfeld{at}kispi.unizh.ch).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
We describe a novel ICD system implantation technique with subpleural defibrillation coil electrodes, epicardial leads for pacing and sensing, and abdominal or intrathoracic device placement.
| Technique |
|---|
|
|
|---|
In 5 patients, epicardial electrodes and the coil lead were inserted through a muscle-sparing left lateral thoracotomy in the fourth intercostal space, as described in detail by our group [7]. The parietal pleura were dissected away from the thoracic wall to create a tunnel along the third intercostal space for insertion of the defibrillation lead. A Medtronic 6937-35 or 6937-52 coil electrode (Medtronic, Minneapolis, MN) was positioned in the prepared space and secured with a stitch. Bipolar steroid-eluting epicardial leads (CapSure Epi 4968-35 [Medtronic]) were sutured to the free lateral wall of the left ventricle and to the left atrial appendage in case of dual-chamber ICD systems.
Surgical access for lead positioning was different in patients requiring concomitant cardiac repair through a median sternotomy using cardiopulmonary bypass (CPB). The parietal pleura were also undermined along the third intercostal space, the coil electrode inserted in the space, and the suture secured. Epicardial electrodes were positioned on the left atrial appendage and the left ventricle during CPB with an unloaded heart. Likewise the extrapleural coil was inserted during CPB when the lung could be totally deflated.
The leads were brought in the upper abdomen through a separate incision in the first group of patients and through the sternotomy in the second group. They were connected to various ICD devices, which were placed in the first 5 patients in the left rectus muscle sheath. In an attempt to optimize the electrical field, the device was subsequently positioned intrathoracically in the diaphragm in a horizontal position underneath the right ventricle. To achieve this, the aponeurosis of the diaphragm was dissected away from the muscle. The dissection was pursued to permit the insertion of the device, often to the level of the inferior vena cava (Figs 1A, 1B). In patients with concomitant cardiac surgery, defibrillation thresholds of 20 Joules were accepted, and no further testing was undertaken at lower energy settings. Adequate defibrillation thresholds were obtained in 6 patients, even when testing was done immediately after CPB. A second epicardial coil electrode was used in a patient with hypertrophic obstructive cardiomyopathy to achieve a defibrillation threshold of 20 Joules. Testing in this patient was potentially hazardous immediately after surgery with CPB for resection of subaortic stenosis. Ventricular fibrillation was not inducible in 2 patients with hypertrophic cardiomyopathy.
|
During our observation period, three appropriate and successful ICD discharges were seen in 1 patient. One inappropriate shock was documented due to sinus tachycardia. One complication with dislodgement of a defibrillation electrode was detected 12 months after implantation on routine chest roentgenogram in 1 patient. Lead repositioning was combined with repositioning of the device intrathoracically, as described. A device recall prompted one device replacement.
| Comment |
|---|
|
|
|---|
We describe an implantation technique with subpleural placement of the defibrillation electrode. The device was placed abdominally or it was placed between the pericardium and diaphragm in a horizontal position underneath the heart to optimize the electrical field. The systems showed adequate defibrillation thresholds, even when testing was performed immediately after CPB surgery. Sensing and pacing thresholds of the epicardial atrial and ventricular leads remained stable with time, as reported in another study [7]. Re-testing in 2 patients demonstrated stable defibrillation thresholds
20 Joules. Compared with previous reported implantation techniques with pericardial or subcutaneous electrodes [3, 4, 8], a subpleural position of the defibrillation electrode guarantees a safe lead position with little or no lead stress through cardiac or lung movements. An intrathoracic horizontal position of the device underneath the heart assures an optimal electrical field and limits lead tension, compared with placement in the rectus muscle sheath. Moreover, the intrathoracic lead and device position ensure a safe ICD system position in active patients in case of physical impacts.
A main limitation includes the short follow-up period, revealing appropriate ICD shocks in only 1 patient, and one inappropriate shock. System revision due to one defibrillation lead dislodgement was necessary early in the learning period. So far we have had no need for device exchange with the intradiaphragmatic position, although no problems for device access would be expected.
In conclusion, the implantation of ICD systems with epicardial sensing and pacing leads, and a subpleural defibrillation electrode is feasible and safe. An intrathoracic position of the device optimizes the electrical field and results in a safe and protected device position. Further follow-up is required.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Roubertie, E. Le Bret, J. B. Thambo, and X. Roques Intra-diaphragmatic pacemaker implantation in very low weight premature neonate Interactive CardioVascular and Thoracic Surgery, October 1, 2009; 9(4): 743 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Juchem, M. Lang, K. Golczyk, M. Ulbrich, B. Reichart, and P. Lamm Successful use of transvenous coil electrodes as single element subcutaneous array leads Europace, March 1, 2009; 11(3): 391 - 394. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tomaske, R. Pretre, M. Rahn, and U. Bauersfeld Epicardial and pleural lead ICD systems in children and adolescents maintain functionality over 5 years Europace, October 1, 2008; 10(10): 1152 - 1156. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Pretre and U. Bauersfeld Minimally invasive implantation of a cardioverter in children J. Thorac. Cardiovasc. Surg., July 1, 2008; 136(1): 239 - 240. [Full Text] [PDF] |
||||
![]() |
M. Tomaske, B. Gerritse, L. Kretzers, R. Pretre, A. Dodge-Khatami, M. Rahn, and U. Bauersfeld A 12-Year Experience of Bipolar Steroid-Eluting Epicardial Pacing Leads in Children Ann. Thorac. Surg., May 1, 2008; 85(5): 1704 - 1711. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |