|
|
||||||||
Ann Thorac Surg 2007;83:542-548
© 2007 The Society of Thoracic Surgeons
a Heart and Diabetes Center North Rhine-Westphalia, Bad Oeynhausen
b Clinic for Heart Surgery, University of Bochum, Bochum
c Clinic for Heart Surgery, University of Rostock, Rostock
d Heart Center Dresden Ltd, Dresden, Germany
e Department of Cardiology, Rabin Medical Center, Petah Tiqva, Israel
f Clinic for Heart Surgery, University of Cologne, Cologne, Germany
Accepted for publication July 29, 2006.
* Address correspondence to Dr Fritzsche, Heart and Diabetes Center NRW, Georgstr 11, Bad Oeynhausen D-32545, Germany (Email: dfritzsche{at}hdz-nrw.de).
| Abstract |
|---|
|
|
|---|
METHODS: Between November 2003 and March 2006, ThromboCheck devices were handed out to 483 patients after mechanical heart valve replacement from six heart centers. In each case, the devices were calibrated for individual patients early postoperatively. Taking control measurements, the device calculated deviations from calibrated frequency and amplitude ranges using a fast Fourier transform and sent a warning message to the display if any relevant changes occurred. This message was taken as cause to determine morphologic correlations of the disturbance using imaging techniques, for example, echocardiography or fluoroscopy (observation period, 550.7 patient-years; median application period, 15.1 months).
RESULTS: Twenty-five true positive, 1 false positive, and no false negative measurements occurred. Eleven patients (44%) received thrombolytic therapy with streptokinase or recombinant tissue-type plasminogen activator, resulting in normalization of the digital frequency analysis. In 4 patients (16%), international normalized ratio levels were adjusted under temporary heparinization therapy. Normalization of the valve sounds were achieved within 3 to 7 days. Seven patients (28%) required reoperation despite lysis therapy. Sensitivity was 100% during the observation period; specificity was 99.8%.
CONCLUSIONS: Digital frequency analysis represents a reasonable addition to the monitoring of patients after artificial heart valve replacement to detect prosthesis-related complications early on.
| Introduction |
|---|
|
|
|---|
It is also known that in the presence of certain risk constellations, the statistic probability of a valvular dysfunction is even higher. These risks include mechanical valve replacement in the right heart (in particular, the tricuspid valve), dilated ventricles, dysrhythmias (especially atrial fibrillation), anomalies of the coagulation system, pregnancy, and so forth. Table 1 illustrates the risk of thromboembolic events after mechanical heart valve replacement using an overview of selected international publications on heart-valve-related complications in the last 10 years.
|
As a result of these observations and considerations, we decided to participate in the development and clinical testing of a novel early warning system for the detection of heart valve dysfunction. This multicentric and double-blind study shows the clinical results of a device enabling the early detection of prosthetic dysfunction by digital frequency analysis measurements.
| Patients and Methods |
|---|
|
|
|---|
The ThromboCheck device (CardioSignal GmbH, Hamburg, Germany) measures the frequency range of mechanical valve sounds digitally and fully automatically. A valve sound is composed of a number of frequencies with a certain amplitude, and each one is as unmistakably recognizable as a fingerprint, owing to the resonance characteristics of individual patients with particular types of valve. From the beats measured in 1 minute, the measurements outside the 95th percentile are excluded. From the other beats, the frequency spectrum is calculated by fast Fourier transformation. If an initial patient-typical signal is stored by the device, and if all sounds recorded later are compared with it using a highly developed software program, then even the slightest deviations are detected by the device as "suspicious." This test can be performed by patients themselves as often as they wish and takes no longer than 1 minute (Fig 1). Comprehensive experimental work (continuous observation of clot formation or lysis in thrombotic alloprostheses in animals) has shown that the acoustic phenomenon of a mechanical heart valve can be significantly altered even when all other imaging techniques fail to register anything suspicious [810]. That means that a prosthetic dysfunction in its initial stages (for example, due to thrombosis) can be detected before it is clinically, echocardiographically, or fluoroscopically recognizable (Fig 2AC).
|
|
As a result of the closely-knit controls, we expected to gain information regarding the frequency of valve dysfunction, as well as thromboembolic events. The early detection of dysfunction should enable suitable treatment strategies to be deduced. Biometric analysis should enable predictors for dysfunction or thromboembolic events to be recognized and future strategies for postoperative care to be deduced. In addition, acceptance of this method would be put to the test. In all patients, we evaluated an altered frequency range as early detection of a mechanical valve dysfunction and recommended medical clarification. In accordance with state-of-the-art recommendations, the diagnostic and therapeutic algorithm was applied as shown in Figure 3.
|
70 mm) or severely restricted left ventricular ejection fraction (30% or less), or both; patients desiring functional controls using ThromboCheck and who had given their written consent. Exclusion criteria were as follows: patients with expected or documented poor compliance; patients known to have a short life expectancy, induced by comorbidity; patients unable to understand or speak German or Hebrew; and patients abusing alcohol or drugs.
Between November 2003 and March 2006, a total of 483 ThromboCheck devices were handed out to patients from six heart centers. The observation period was 550.7 patient-years, with a median application period of 15.1 months (range, 1 to 24). In all, 332 patients with aortic valve replacements, 105 with mitral valve replacements, 41 with aortic valve replacement plus mitral valve replacement, 2 with aortic valve replacement plus pulmonary valve replacements, 2 with pulmonary valve replacement plus tricuspid valve replacements, and 1 patient with a mitral valve replacement plus tricuspid valve replacement were analyzed. These patients were all recipients of mechanical heart valves made by Medtronic, St. Jude, On-X, or Carbomedics. In each case, a device was calibrated for a patient at the early postoperative stage and the individual heart rate pattern stored on computer. Taking control measurements, the device calculated deviations in the frequency and amplitude ranges using a fast Fourier transformation and sent a warning message to the display if any relevant changes occurred. A warning message was taken as cause to determine the morphologic correlation of the disturbance using suitable imaging techniques, such as transthoracic echocardiography, transesophageal echocardiography, or fluoroscopy.
Patients were asked to complete questionnaires after 1, 6, and 12 months to submit precise information about clinical conspicuities that might have remained undetected. At the end of the investigation, a questionnaire was sent out to all patients asking them to evaluate the handling of the ThromboCheck device.
All results were multicentric, prospective, and double blind. ("Double-blind" refers to the following procedure, which was practiced on patients at the Heart and Diabetes Center NRW; 70% of all results were ascertained). Results were not displayed on the ThromboCheck device in the usual manner ["OK," "errorplease repeat," "2nd errorplease consult physician"], but were encrypted using a 10-figure binary code. The meaning of this code was known neither to the study center nor to the patients. The codes were protocolled by the patients and passed on continually to a person anonymous to both the study center and the patients. At this point the code was deciphered. If there were any deviations from the reference code, the study center was immediately informed and the patient in question contacted.)
| Results |
|---|
|
|
|---|
Eleven patients (44%) immediately received thrombolytic therapy with streptokinase or recombinant tissue-type plasminogen activator. That led to complete normalization in the digital frequency analysis of the sound pressure phenomenon of the artificial valve. In 4 patients (16%), INR levels were adjusted to within the therapeutic range by increasing oral anticoagulation therapy (and temporary heparinization). In this way, normalization of the valve sounds could be achieved within 3 to 7 days.
Nine patients (36%) required reoperation despite lysis therapy. In each case, the explanted valves displayed marked thrombotic formations on the occluder hinges and along the edges of the valves. In one case, tendinous fibers that had been left too long caused a mitral valve occluder to malfunction; in another case, a paravalvular leak in an artificial mitral valve was responsible for shifting the frequency focus.
One other patient displayed a small paravalvular leak, which could also be detected by digital frequency analysis. There had been no consequences to date, owing to a lack of clinical and paraclinical symptoms.
Table 2 summarizes all documented conspicuities from the digital frequency analysis, as well as the resulting therapeutic consequences. The patient with the suspected dysfunction that could not be verified using any of the imaging techniques (false positive) was very tightly controlled. Within a week, the valve sounds returned to normal spontaneously. No clinical events (thromboembolic, endocarditis, leak, and so forth) undetected by the ThromboCheck device occurred (0 false negative). Sensitivity was 100% during the observation period; specificity was 99.8%. No relevant hemorrhaging complications were reported.
|
| Comment |
|---|
|
|
|---|
Digital frequency analysis (the ThromboCheck method) represents a reasonable addition to the monitoring of patients with artificial heart valve replacements and the early detection of prosthesis-related complications, and should in the future receive particular attention for the discussed risk constellations or in conjunction with the high-risk situations of a large number of heart valve patients.
As the frequency spectrum representing the opening and closing sounds of a valve are measured at a certain time, valves in different positions can be differentiated chronologically in patients after double valve replacements (aortic and mitral valve replacement). In triple valve replacement, differentiation isat least so farimpossible. Even if a differentiation between valves is not yet possible and interferences in multiple valve replacement occur, changes in the frequency spectrum can be postulated and would suffice, as this device has to be seen as a very sensitive screening method.
These clinical findings and their practical relevance for a patients individual situation induced us to create a platform at the Institute for Applied Telemedicine at the Heart and Diabetes Center NRW in Bad Oeynhausen that would facilitate an online analysis of the functional integrity of mechanical valves by medical personnel at any time. This alternative is aimed at the early detection of thrombotic deposits, tissue ingrowth, and now also leakages, whereby the clarification of relevant or suspicious findings using suitable documentation and notification remains the responsibility of the cardiologist or center treating the patient.
There have been some crucial changes in the clinical follow-up of patients with mechanical heart valves. The fact that patients can now closely control their own coagulation and valve function means that the diagnostic window during which complications can arise unnoticed and trigger severe consequences (embolisms, hemorrhaging, reoperation) is effectively reduced to almost nothing. In our opinion, it is of vast importance that the ThromboCheck method is addressed in more detail and applied more broadly in the follow-up of patients with artificial heart valves, especially those with the risk constellations discussed above [11].
Against this background, it is our conviction and declared goal to undertake every effort to deepen the understanding of this method and to have it included in the guidelines of cardiological and cardiosurgical associations for the care of heart valve patients.[1220
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Schenk and D. Fritzsche Reply. Ann. Thorac. Surg., March 1, 2009; 87(3): 987 - 988. [Full Text] [PDF] |
||||
![]() |
A. Bagno, F. Anzil, R. Buselli, V. Tarzia, T. Bottio, G. Gerosa, and V. Pengo Is the Analysis Over the Time Domain or Over the Frequency Domain Significant for the Detection of Bileaflet Mechanical Heart Valve Dysfunction? Ann. Thorac. Surg., March 1, 2009; 87(3): 986 - 987. [Full Text] [PDF] |
||||
![]() |
B. K Redman When Is Patient Education Unethical? Nursing Ethics, November 1, 2008; 15(6): 813 - 820. [Abstract] [PDF] |
||||
![]() |
P. M Kilbridge and D. C Classen The Informatics Opportunities at the Intersection of Patient Safety and Clinical Informatics JAMIA, July 1, 2008; 15(4): 397 - 407. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Fritzsche Reply Ann. Thorac. Surg., March 1, 2008; 85(3): 1143 - 1143. [Full Text] [PDF] |
||||
![]() |
I. M. Hettich, W. B. Eichinger, and R. Lange A New Home Monitoring Device for the Detection of Mechanical Valve Dysfunction Ann. Thorac. Surg., March 1, 2008; 85(3): 1142 - 1143. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |