Ann Thorac Surg 2006;82:926-932
© 2006 The Society of Thoracic Surgeons
Department of Cardiothoracic Surgery, The University of Pittsburgh School of Medicine, Children's Hospital of Pittsburgh, Pittsburgh, Pennsylvania
Accepted for publication February 27, 2006.
* Address correspondence to Dr Sharma, Department of Cardiothoracic Surgery, 200 Lothrop St, Suite C-700 PUH, Pittsburgh, PA 15213. (Email: sharmams{at}upmc.edu).
| Abstract |
|---|
|
|
|---|
METHODS: A 15-year retrospective review was performed on all patients younger than 18 years of age undergoing insertion of a pulsatile VAD at our institution.
RESULTS: Eighteen patients underwent VAD placement during the study period. The mean age was 12 (range, 6 months to 18 years), with a mean body surface area of 1.48 (range, 0.25 to 2.3 m2). Diagnoses included dilated cardiomyopathy (n = 15), myocarditis (n = 2), and postcardiotomy ventricular failure (n = 1). Ten children underwent insertion of biventricular VADs, and 8 had implantation of left ventricular VADs. The mean support duration was 57 days (range, 2 to 175 days). Complications included bleeding requiring reoperation (n = 4), stroke (n = 5), and device-related infection (n = 2). Outcomes of VAD support were as follows: VAD explantation in 1 case, death while receiving mechanical support in 3 patients, and successful transplantation in 14 patients (77%). Survival at 6 months after orthotopic heart transplantation was 93% with 1-year and 5-year survival rates of 83%.
CONCLUSIONS: Currently available VADs are applicable for use as a bridge to orthotopic heart transplantation or in rare instances for myocardial recovery. The increasing accessibility of miniaturized devices allow for long-term support in smaller children and infants while awaiting orthotopic heart transplantation. Although the perioperative morbidity and mortality of VAD placement is not insignificant, survival for those who receive a heart transplant is excellent.
| Introduction |
|---|
|
|
|---|
The successful use of VADs as a bridge to transplant in adults has led to the introduction of such technology for children awaiting OHT [5]. In 1991, Warnecke and colleagues [6] reported the first successful bridge to transplantation using a pulsatile paracorporeal left ventricular VAD in an 8 year-old boy. Despite worldwide use with pulsatile pediatric VADs, centers in the United States have relied on adult-sized devices because of a lack of U.S. Food and Drug Administrationapproved miniaturized VADs [7].
We describe our single-center experience since 1990 with 18 children who were supported with pulsatile VADs as a bridge to OHT. Moreover, we include in this series our early experience with pediatric-sized VADs.
| Material and Methods |
|---|
|
|
|---|
Patients
The prospectively collected data from the Transplant Patient Management System were evaluated retrospectively in patients younger than 18 years of age who underwent insertion of a VAD at Children's Hospital of Pittsburgh, The University of Pittsburgh Medical Center from April 1990 through November 2005. Data points are summarized in tabular format and expressed as a mean value with range when appropriate. The median value is reported when statistically relevant.
Children and adolescents with intractable heart failure who received pulsatile paracorporeal VADs were included in this study. Patient demographics are summarized in Table 1. The study included 10 boys and 8 girls with a median age of 12 years (range, 6 months to 18 years) with 5 children younger than 8 years of age. Body surface area ranged from 0.25 to 2.3 m2 (mean, 1.48 ± 0.6 m2). The cause of heart failure was dilated cardiomyopathy in 15 patients: idiopathic (n = 13), peripartum (n = 2). One patient had a concomitant intracardiac fibroma. Two children suffered from acute myocarditis, and 1 child underwent repair of recurrent supravalvular aortic stenosis at another institution requiring postoperative ECMO. He was transferred to Children's Hospital of Pittsburgh, The University of Pittsburgh Medical Center, and subsequently had biventricular VAD placement.
|
|
25 kg, body surface area 1.0 m2) was the Thoratec paracorporeal pneumatic VAD (Thoratec Corp, Pleasanton, CA) as it has the flexibility of providing left, right, or biventricular support. Device selection based on age is depicted in Figure 1. This device consists of three components: (1) a blood pump, which has a 65-mL stroke volume and can deliver pulsatile flows of 1.3 to 7.1 L/min, (2) cannulas, which connect the blood pump to the heart, and (3) a drive console that powers the blood pump pneumatically. The VAD was placed in a paracorporeal position on the anterior abdominal wall. The left ventricular VAD cannulation was performed through the left ventricular apex (inflow) with return to the ascending aorta (outflow). We used left ventricular apical cannulation in all 12 patients. The right ventricular VAD was cannulated through the right atrium with return to the pulmonary artery. All Thoratec patients were run exclusively in volume mode.
|
Berlin heart excor ventricular assist device
Recently, 4 small children (body surface area, 0.25 to 1.0 m2) underwent placement of an appropriate sized (10 to 30 mL) blood pump (Berlin Heart AG Berlin, Germany) for left-ventricular support on a compassionate-use basis. The Berlin Heart consists of a paracorporeal pneumatically driven polyurethane blood pump with a multilayer flexible membrane that separates the blood and air chambers. Silicon cannulas connect the blood pump to the patient, and a triple-leaflet valve prevents blood reflux. All blood-contacting surfaces are heparin-coated (Carmeda, Inc, San Antonio, Texas). The pump is driven by a pulsatile electropneumatic system. The drive unit (IKUS 2000) has a triple operational control and pneumatic system, with synchronous and asynchronous operating modes. The synchronous mode was used in all cases.
Anticoagulation
Our anticoagulation protocol consisted of full systemic heparinization during insertion on cardiopulmonary bypass. Aprotinin was used at implant and removal. In adolescent patients, postoperative anticoagulation was started with Dextran 40% at 25 mL/h 6 hours after admission to the intensive care unit if bleeding was less than 100 mL/h. Subsequently, heparin was started when postoperative bleeding from the chest tubes was less than 50 mL/h for 3 consecutive hours. The goal for the partial thromboplastin time was 40 to 51 seconds for at least the first 72 hours or until the risk of bleeding from more aggressive anticoagulation was thought to be acceptable. Heparin was then increased to maintain a partial thromboplastin time of 50 to 70 seconds. Warfarin sodium (Coumadin; Bristol-Myers Squibb Company, Princeton, NJ) was introduced around postoperative day 10 to keep the international normalized ratio at 2.5 to 3.5. Heparin was discontinued after obtaining an international normalized ratio of at least 2.5.
The philosophy of anticoagulation was to maintain heparin until the patient demonstrated a low risk for bleeding complications and after there had been a period of stable gastrointestinal tract function and diet. This usually was found to occur around 10 to 14 days after implant. In our early experience, we did not add aspirin unless there was a specific concern about thromboembolism. Recently, we have altered our protocol to include use of aspirin. After discharge, the international normalized ratio was assessed at a minimum of 2 times per week for stable patients. Infants and small children were maintained on heparin alone with a target partial thromboplastin time range of 60 to 80 seconds followed by the addition of aspirin after 72 hours. Clopidogrel was added if there was visible fibrin or thrombus in the pump or thrombocythemia. We routinely monitored the efficacy of anticoagulation with thromboelastograms.
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Comment |
|---|
|
|
|---|
The decision as to which device to use for a particular situation in children is complex. Extracorporeal membrane oxygenation has certain advantages in the pediatric population in that it is relatively rapid, inexpensive, and readily available. Moreover, it allows the flexibility of peripheral and central cannulation and can provide total cardiopulmonary support. Our philosophy is to support children in the acute setting with ECMO as a bridge to recovery. Historically, we have also relied on ECMO for postcardiotomy support beyond 72 hours because of a lack of other options. Lengthy waiting times for donor organs, however, make extended support with ECMO and centrifugal VADs hazardous. Children on these devices are immobilized and become deconditioned. Extracorporeal membrane oxygenation and centrifugal pumps require high levels of anticoagulation, increasing the risk for hemorrhagic complications. Therefore, children with postcardiotomy heart failure who do not exhibit any recovery within 72 hours should be considered for conversion to a VAD.
The indications for pediatric VAD usage are evolving. Furthermore, there has been no fixed selection criteria such as those proposed for adults in whom indications and optimal timing of device implantation have been better defined. Our strategy for those children with significant biventricular failure caused by myocarditis or cardiomyopathy has been planned biventricular VAD implantation. Currently, however, we agree with Stiller and colleagues [15] that left ventricular VAD alone in this group may suffice. Left ventricular VAD therapy allows for maximal unloading of the left ventricle, which reduces the afterload of the right ventricle by up to 15 to 25 mm Hg. Combined with pharmacologic right heart support, the need for additional mechanical assistance is more limited. Studies in adult VAD recipients have shown that patients who do not exhibit signs of secondary organ malfunction have the greatest benefit from mechanical support, supporting the strategy of early intervention [16, 17]. Nevertheless, implantation of a VAD too early exposes the patient to unnecessary surgery and device-related morbidity. The importance of patient selection and timing of device implantation is illustrated in a study published by Stiller and colleagues [18] from the German Heart Center in Berlin. Of the 45 pediatric patients who received VADs during the period of 1990 to 2002, only 18 children underwent OHT, for a total survival rate of 51%. After careful analysis of their outcomes in this series, the authors concluded that most of these implantations occurred too late. Recently, the same group reported improved survival (67%) in postcardiotomy infants when compared with their historical control group [15]. The authors attribute this improvement in survival to earlier VAD implantation, exclusion of children with prolonged cardiopulmonary resuscitation, and refinements in device design, anticoagulation protocols, and intensive care unit management.
In the United States, lack of government approval for existing miniaturized VADs has necessitated the obligatory use of standard pulsatile devices in children. Although the Thoratec VAD has been used by several centers successfully in small children, specific concerns of using "oversized" devices in children have been documented [19]. These include (1) large stroke volumes (65 mL in volume mode) into a small aorta, leading to systolic hypertension and subsequent intracranial hemorrhage; (2) stasis in the device that can cause thromboembolic complications; and (3) obligatory use of multiple adult-size cannulas in a limited pericardial space. To compensate for these mechanical limitations, VADs in pediatric patients (
25 kg, body surface area, 1.0 m2) may be run using the fixed mode with higher rates (80 to 90 beats/min) and partial stroke volumes. This diminishes the effect of the large volumes in small children. Although this mode causes the VAD chamber to fill submaximally, physical inspection of the device at end systole can confirm complete emptying of the pneumatic chamber. The smallest patient to date supported with this device was a 7-year-old girl (body surface area, 0.77 m2) with biventricular failure from myocarditis. She was successfully bridged to OHT using a fixed-rate mode of 90 to 120 beats/min [20].
Studies examining pediatric bridge to transplant experiences report bridging success rates of 51% to 66% [18, 2123]. Our reported bridging success rate of 78% with no early deaths after OHT is favorable when compared with other centers using VADs for those children with heart failure as a result of cardiomyopathy or myocarditis. It is important to note that when VADs are used in the postcardiotomy setting or for complex congenital heart disease, survival on VAD is poor, ranging from 20% to 25% regardless of device [12, 23]. In an analysis of Thoratec VADsupported children, Reinhartz and colleagues [23] demonstrated that congenital heart disease and failure to wean off bypass were independent risk factors for death (p < 0.05).
When comparing adverse events in this study with the literature, bleeding requiring reoperation, infection, and device malfunction occurred with similar frequency to our series [18, 2123]. Neurologic events, defined as transient ischemic attack or stroke, varied from 11% to 45%. A subgroup analysis of our series reveals a stroke rate of 21% in the Thoratec group, which is less than a recent multicenter report in which the neurologic event rate was 35% [23]. Possible explanations for this include (1) a high percentage (24%) of children requiring left atrial or intraatrial groove cannulation, a demonstrated risk factor for stroke; (2) some centers using volume mode in smaller children, which expels a stroke volume of 65 mL into a relatively small aorta rather than a fixed-volume mode maintaining higher drive line pressures to ensure complete emptying of the pump; and (3) variations in anticoagulation protocols.
Both the Berlin Heart Excor and Medos HIA blood pumps are manufactured in multiple sizes, obviating large stroke volumes in small children. Device design includes the use of polyurethane valves instead of mechanical valves, which may be less thrombogenic, and transparent blood chambers and ports allowing visual control of filling and emptying and transillumination detection of thrombotic deposits. Nevertheless, neurologic event rates using these devices ranged from 11% to 45% in the literature with centers using a combination of left atrial and apical inflow cannulation and varied anticoagulation schemes [12, 21]. Pump exchanges owing to thrombus formation were not infrequent (28%) in the Berlin Heart group despite heparin coating of all blood-contacting surfaces [12]. Recent data may support the use of a modified anticoagulation protocol with heparin dose depending on a partial thromboplastin time of 60 to 80 seconds, antithrombin III substitution when less than 70%, and use of aspirin and dipyridamole depending on platelet number and function tests [15]. Of the 4 patients in our series supported with the Berlin Heart, 2 patients suffered a neurologic event. One patient had an underlying neoplasm and another patient suffered from a bloodborne infection that is known to enhance clotting potential.
Reports of 1-year and 5-year posttransplant survival in VAD-supported children vary from 62% to 88% and 47% to 72%, respectively [18, 2123]. Post-OHT survival after VAD support in this series was excellent, with 93% of children alive at 6 months. One-year and 5-year survival after OHT was 83% (Fig 4). Despite these favorable results, there exists concern regarding the effect of sensitization in VAD-supported patients because sensitized untreated left ventricular VAD recipients have been noted to have both a prolongation of waiting time to transplantation and an increased risk of acute rejection [24, 25]. The impact of elevated panel reactive antibody (PRA) in children awaiting OHT was addressed in a recent article that demonstrated that 30-day mortality was higher with elevated PRA versus those with PRA levels of less than 10% [26]. However, this difference did not reach statistical significance. In this report, 14% of patients showed evidence of sensitization after VAD support. These children all had negative donor cross-match results at OHT. Use of a VAD in our series does not appear to significantly increase antibody sensitization before transplantation and did not appear to have a negative impact on posttransplantation survival.
Improvements in patient selection, medical optimization, and surgical technique have improved outcomes for children suffering from heart failure. Nevertheless, a subset of these patients will have refractory myocardial dysfunction requiring mechanical circulatory support. Pulsatile paracorporeal VADs have been validated as an effective strategy to keep children alive while awaiting heart transplantation. The timing of implementing VAD support is critical. The success of our bridge to transplantation experience demonstrates the feasibility of VAD support in adolescents and older children. In addition, our limited use of miniaturized VADs to successfully bridge 4 small children to OHT underscores the need for continued development and implementation of this life-sustaining technology.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. Januszewska, E. Malec, J. Birnbaum, M. Loeff, R. Sodian, C. Schmitz, H. Netz, and B. Reichart Ventricular assist device as a bridge to heart transplantation in children Interactive CardioVascular and Thoracic Surgery, November 1, 2009; 9(5): 807 - 810. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. T. Mahle, G. Ianucci, R. N. Vincent, and K. R. Kanter Costs Associated With Ventricular Assist Device Use in Children Ann. Thorac. Surg., November 1, 2008; 86(5): 1592 - 1597. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Gandhi, C. B. Huddleston, D. T. Balzer, D. J. Epstein, T. A. Boschert, and C. E. Canter Biventricular Assist Devices as a Bridge to Heart Transplantation in Small Children Circulation, September 30, 2008; 118(14_suppl_1): S89 - S93. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. N. Ruygrok, D. S. Esmore, P. M. Alison, K. A. Finucane, S. P. McGuinness, A. D. McGeorge, J. Negri, K. Jones, and H. C. Gibbs Pediatric Experience With the VentrAssist LVAD Ann. Thorac. Surg., August 1, 2008; 86(2): 622 - 626. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Monagle, E. Chalmers, A. Chan, G. deVeber, F. Kirkham, P. Massicotte, and A. D. Michelson Antithrombotic Therapy in Neonates and Children: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines (8th Edition) Chest, June 1, 2008; 133(6_suppl): 887S - 968S. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. W. Duncan Invited commentary. Ann. Thorac. Surg., September 1, 2006; 82(3): 933 - 933. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||