|
|
||||||||
Ann Thorac Surg 2006;82:708-713
© 2006 The Society of Thoracic Surgeons
a Department of Cardiovascular Surgery, Christian-Albrechts-University of Kiel, School of Medicine, Kiel, Germany
b Department of Radiology, Christian-Albrechts-University of Kiel, School of Medicine, Kiel, Germany
Accepted for publication January 26, 2006.
* Address correspondence to Dr Lutter, Department of Cardiovascular Surgery, Christian-Albrechts-University of Kiel, School of Medicine, Arnold-Heller-Str. 7, Kiel, 24105 Germany (Email: lutter{at}kielheart.uni-kiel.de).
| Abstract |
|---|
|
|
|---|
DESCRIPTION: Bovine jugular xenografts were sutured into nitinol stents. Transfemoral implantation in the pulmonary position using a modified commercially available application device (with a 22-French outer diameter) was evaluated in 9 sheep.
EVALUATION: Two sheep died shortly after successful valved stent implantation due to internal venous hemorrhage. Another 1 sheep died 2.5 months after the procedure due to vegetations on the neovalve leading to subtotal stenosis. All other animals survived the 3-month study time (n = 6). An orthotopic pulmonary valved stent position was achieved in 4 animals and a supravalvular position in 1. During the deployment procedure, rhythm disturbances occurred in all animals, and mean arterial blood pressure dropped from 83.9 ± 26.0 mm Hg to 68.3 ± 22.3 mm Hg (p = 0.006) (n = 5). The peak-to-peak transvalvular gradient was 5.1 ± 4.0 mm Hg initially (n = 5), and 3.6 ± 1.6 mm Hg at follow-up (n = 5). Three-month angiographic and echocardiographic follow-up confirmed competent neovalves without paravalvular leakages.
CONCLUSIONS: After 3 months of implantation, percutaneously implanted memory nitinol valved stents demonstrated good function in the sheep.
| Introduction |
|---|
|
|
|---|
Our group demonstrated the feasibility of totally percutaneous transfemoral pulmonary valve implantation in an ovine model using self-expanding nitinol stents [6]. The continuously exerted radial force of self-expanding stents and their high flexibility assure a geometric adaptation to anatomical and tissue-property changes. Therefore the use of this kind of stent may be beneficial for the percutaneous treatment of patients with the previously mentioned anatomical anomalies.
The aim of this study was to evaluate the function of percutaneously implanted self-expanding stents carrying a biological valve into the pulmonary position of juvenile sheep during a 3-month period using angiographic and echocardiographic, hemodynamic, and macroscopic measurements.
| Technology |
|---|
|
|
|---|
| Technique |
|---|
|
|
|---|
The anesthesia and implanting procedure were carried out as previously described [6]. Seven French sheaths were introduced into the left femoral artery and vein for hemodynamic measurements. Through a 24-French sheath positioned in the right groin, the 22-French application device (Fig 1) was inserted and the valved stents were deployed directly over the native pulmonary valve under fluoroscopic control. The 24-French sheath was removed and after manual compression for 20 minutes, the skin in the right groin was closed with a single stitch. The animals were studied by analyzing various measurements described as follows.
|
Three months later the animals were reanalyzed (described as follows) and were sacrificed.
Angiography
An angiography unit (Multistar Top, Siemens, Erlangen, Germany) was used for fluoroscopic assessment of the position and function of the neovalves.
Echocardiography
Transthoracic echocardiography (TTE) was performed at the 3-month follow-up with the Ultrasound System Five (Vingmed Sound, Horten, Norway) using a multiplane 2.5-MHz transducer (n = 4). The echocardiograpic probe was applied at the right hemithorax between the fourth and sixth intercostal spaces. The measurements were recorded and stored on a magneto-optical disk and analyzed offline by an experienced investigator (RQ).
Hemodynamic Measurements
Arterial pressure, right ventricular pressure, left ventricular pressure, and pulmonary artery pressure (Micro-Tip Millar Catheter, Millar Instruments, Inc, Houston, TX) were recorded using Haemodyn-Software (Hugo Sachs Electronics, Hugstetten, Germany). The software calculates the maximum of the first derivative of the ventricular pressures (dp/dtmax) as an indicator for the contractility and its minimum (dp/dtmin) indicating the relaxation behavior.
Macroscopic Examination
Valved stents were grossly inspected and photographs were taken. Special attention was given to the retraction of the cusps or any deformed or indurated parts of the valve. The atrial and ventricular chambers and the pulmonary artery were exposed to look for catheter-induced damage and to look for penetration of stent struts.
Roentgenogram Assessment
Roentgenogram examination of the explanted valved stents (n = 4) was performed under mammography conditions to demonstrate and localize macroscopic calcification.
Statistics
Values are presented as mean ± standard deviation. Data were analyzed based on the Wilcoxon test to compare related data for non-normally distributed data using the SPSS 10.1 software (SPSS, Inc, Chicago, IL). The p values less than 0.05 were considered statistically significant.
| Clinical Experience |
|---|
|
|
|---|
The mean diameter of the pulmonary annuli was 20.0 ± 1.4 mm (range, 18.021.8 mm) as revealed by angiography. Valved stents with a maximal outer diameter from 20.6 to 23.7 mm were used.
Mean duration of the procedure from insertion of the application device through the sheath to deployment was 52.9 seconds (range, 9.0 to 124.5 seconds). The fluoroscopy time of the entire procedure ranged from 7.0 to 26.7 minutes (mean, 11.9 minutes).
Angiography
Angiography revealed no insufficiencies after 3 months and showed competent valved stents in all 6 surviving animals (Fig 2).
|
The mean systolic internal diameter of the valved nitinol stents was measured at 15.6 ± 1.1 mm, and the mean internal diastolic diameter was measured at 20.1 ± 0.9 mm (Fig 3).
|
The peak-to-peak gradient across the valved stents was 5.1 ± 4.0 mm Hg at 5 minutes after the implantation and 3.6 ± 1.6 mm Hg at the 3-month follow-up (n = 6) (Table 1).
|
No paravalvular defects were visible. The valved stents were pliable and the leaflets were thin without indurations. In one valve there was an evident lack of coaptation with one retracted leaflet. Slight fibrous overgrowth was seen at the inflow portions of two valved stents (Fig 4). Cardiac structures were unscathed. No macroscopic damage of the pulmonary artery was noted. In particular there was no penetration of stent struts.
|
|
| Comment |
|---|
|
|
|---|
This is one of the first reports of evaluation of percutaneously implanted self-expanding valved nitinol stents in the pulmonary position. So far, balloon-expandable valved stents are used in clinical practice [3]. The potential advantages of nitinol stents concerning the preservation of the valve and the deployment maneuver were previously discussed [6]. Interestingly in the present study, the echocardiographic examinations nicely demonstrated adaptation in the diameter of the valved stent to the radial motion of the RVOT during the cardiac cycle. The soft nitinol stent's ability to adapt to the surrounding structures could be beneficial for a physiologic blood flow in the proximal great arteries. However the radial force of this stent will not be high enough to overcome calcified valves. To reach a sufficient valve area, a preceding balloon dilation or resection procedure would be required [6, 9].
For the first time meticulous hemodynamic monitoring is reported for percutaneous heart valve replacement. The contractility of both right and left ventricle and their relaxation properties were unchanged at the end of the study. Thus the neovalves had no negative effect on the cardiac function. During the implantation, rhythm disturbances and a significant decrease of mean arterial pressure was observed without affecting the contractility as indicated by the left ventricular dP/dtmax.
The Millar catheter measurements revealed a low transvalvular pressure gradient shortly after the implantation as well as after 3 months. Macroscopy and roentgenogram assessment revealed no calcification of the leaflets, which would affect the valves function.
Limitations
The authors wish to address several limitations of this study. First, the large diameter of the application device has to be mentioned. Two animals died shortly after the percutaneous procedure due to hemorrhage from venous structures after injury by the large introducer sheath. This problem was not observed in our acute experiments, possibly because of the short follow-up period [6]. Only a further reduction of our catheter system's size will allow the clinical application in children without an inguinal cut down. Recently, Ruiz and coworkers [10] published a remarkable long-term study in pigs of catheter-placed low-profile biodegradable pulmonary valves made of small intestinal submucosa [10]. Their square stent-based valve required only an 8-French delivery system. Such a low-profile valve combined with the small flexible delivery system is ideal for use in pediatric cardiology.
Second, tissue deterioration and ensuing loss of function is a major problem with biological heart valve prostheses. In the present study, one of six valves was slightly affected. The 3-month observation period may be too short to provide valid information on the durability of the biological valve. For that reason the development of catheter-based techniques for the implantation of mechanical heart valves that are not affected by this problem should not be overlooked. An interesting feasibility study of percutaneous aortic disc valve prosthesis was reported by Sochman and coworkers [11]. Further studies in this field should be undertaken.
Third, bovine jugular veins have a limited maximal diameter of approximately 24 mm. Therefore they are not suitable for an enlarged RVOT or pulmonary trunk. Boudjemline and colleagues [5] are working intensively on this problem. In an ovine study they successfully tested a two-stage surgical and percutaneous procedure to overcome the previously mentioned problem [5]. Others chose a transthoracic beating heart approach to introduce a self-expanding valved stent of large size in the pulmonary position of pigs [4]. In contrast, our transfemorally inserted cylindrical valved stent can not provide a solution on this matter.
Nonetheless, percutaneously implanted memory nitinol valved stents are evaluated during a 3-month period in an ovine model in the present study showing a good structural and functional outcome.
| Disclosures and Freedom of Investigation |
|---|
|
|
|---|
| Disclaimer |
|---|
|
|
|---|
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G.-J. Zong, Y. Bai, H.-B. Jiang, W.-P. Li, H. Wu, X.-X. Zhao, and Y.-W. Qin Use of a novel valve stent for transcatheter pulmonary valve replacement: An animal study. J. Thorac. Cardiovasc. Surg., June 1, 2009; 137(6): 1363 - 1369. [Abstract] [Full Text] [PDF] |
||||
![]() |
G.-W. Meng, J.-Y. Zhou, Y. Tang, Z.-K. Ye, Y. Zhang, G.-M. Liu, and S.-S. Hu Off-Pump Pulmonary Valve Implantation of a Valved Stent With an Anchoring Mechanism. Ann. Thorac. Surg., February 1, 2009; 87(2): 597 - 601. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lutter and J. Cremer Invited commentary. Ann. Thorac. Surg., February 1, 2009; 87(2): 601 - 602. [Full Text] [PDF] |
||||
![]() |
M. J. Antunes Off-pump aortic valve replacement with catheter-mounted valved stents.: Is the future already here? Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
T. A. Vassiliades Jr Invited commentary Ann. Thorac. Surg., August 1, 2006; 82(2): 713 - 714. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |