METHODS: Between April 2003 and May 2005, 14 patients with aortic stenosis underwent aortic valve bypass surgery at two institutions. All patients selected for aortic valve bypass surgery were deemed to be at very high risk for conventional aortic valve replacement. These patients represented 14 (5.8%) of all 243 patients undergoing isolated aortic valve surgery during the same time period. Mean Society of Thoracic Surgeons predicted risk for operative mortality (11%) was between the 90th and 95th percentile.
RESULTS: Twelve of 14 patients had previous cardiac surgery with patent bypass grafts. Average age was 78 years. Mean aortic valve area was 0.68 cm2. All operations were performed through a left thoracotomy on the beating heart (cross-clamp time, 0 minutes). Cardiopulmonary bypass was used for 6 patients (median cardiopulmonary bypass time, 15 minutes). There were 2 perioperative deaths. Median postoperative length of stay was 9 days. Two noncardiac late deaths occurred. Nine of 10 surviving patients are functional class I and are living independently. Early postoperative echocardiography confirms excellent aortic valve bypass function with preservation of ventricular ejection performance.
CONCLUSIONS: Treatment of high-risk aortic stenosis patients with aortic valve bypass surgery is promising. Avoidance of sternotomy and cardiopulmonary bypass supports broader application to moderate-risk patients with aortic stenosis and as a control arm for studies of novel interventional therapies.
| Introduction |
|---|
|
|
|---|
| Material and Methods |
|---|
|
|
|---|
Patients and Indications for Surgery
All patients selected for aortic valve bypass surgery were deemed to be at very high risk for conventional aortic valve replacement. An independent assessment of suitability for aortic valve bypass surgery was performed by at least two attending surgeons (J.S.G., B.P.G.) at the University of Maryland Medical Center and by the operating surgeon (J.W.B.) at Indiana University School of Medicine. Informed consent was obtained after consultation with the patient and family. This review was approved by the institutional review boards of the University of Maryland Medical Center and Indiana University School of Medicine.
Preoperative Evaluation
All patients underwent a standardized preoperative workup that included transesophageal echocardiography, cardiac catheterization, and multislice computed tomography. Patients with significant operable coronary disease, more than moderate aortic or mitral insufficiency, or severe calcification or mural disease of the descending thoracic aorta at the site of planned anastomosis were excluded.
Operative Technique
Patients were placed in the left lateral decubitus position. After double-lumen intubation, heparin (5,000 units) was administered intravenously, and a femoral vein accessed with a heparin-bonded Biomedicus cannula (Medtronic Inc, Minneapolis, MN). The aortic valve bypass conduit was constructed on the back table. In one configuration, a size 18 apical connector (Medtronic, Irvine, CA) was preclotted with thrombin and cryoprecipitate. It was anastomosed to a size 21 stentless porcine valve (Freestyle; Medtronic), which was in turn anastomosed to a size 20 Hemashield (Boston Scientific, Natick, MA) graft with a single 8-mm side branch, which was later used for arterial inflow from the heart-lung machine. The porcine coronary ostia were oversewn with pledgeted sutures. In the other configuration (previously described and illustrated [6]), the conduit consisted of a 20-mm diameter model 10 or 150 Medtronic valved conduit with a stentless porcine aortic valve inserted in its center. The conduit was in turn connected to an 18-mm diameter Medtronic apical connector. No preclotting of the graft was performed.
An anterolateral thoracotomy was performed over the apex of the heart. One author (J.S.G.) inserted a videoscope posteriorly through an exploratory incision to identify the location of the ventricular apex and guide placement of the thoracotomy incision. The inferior pulmonary ligament was mobilized, and the descending thoracic aorta exposed. The overall length of the conduit from the left ventricular apex to the anastomotic site on the descending aorta was determined, and the conduit cut to length. The distal anastomosis was constructed between the conduit and the aorta using running 4-0 monofilament suture or interrupted 2-0 polyester pledgeted sutures and a specially designed partial occluding clamp. The apex was exposed, and between 8 and 12 2-0 pledgeted monofilament sutures placed in a circumferential fashion about the location of the proposed ventriculotomy. During insertion of the apical connector, the heart was transiently paced at 200 beats per minute to minimize blood loss [7]. A stab wound was made in the apex, and a Foley catheter inserted and inflated with saline to a diameter that is 3 mm larger than the diameter of the circular coring knife (cork borer). The coring knife was used to create a clean apical hole, and the apical connector inserted and tied down. After deairing the graft with a 20G needle in the angled portion of the connector, blood flow was established from the apex to the descending thoracic aorta (Fig 1).
|
Follow-Up
Follow-up was performed by telephone interview or clinical examination, or both. Mean follow-up was 12 months (range, 2 to 39). All operative survivors underwent early transthoracic echocardiographic examination to evaluate left ventricular ejection performance and residual gradient across the native aortic valve.
| Results |
|---|
|
|
|---|
Perioperative Results
Cardiopulmonary bypass was used in 6 of 14 patients (Fig 2). Cross-clamp time was 0 minutes for all patients. No patient suffered a stroke, and no patient had permanent renal failure requiring hemodialysis. There was 1 perioperative death in a patient with advanced pulmonary fibrosis who could not be oxygenated in the operating room. Autopsy confirmed an intact and functional aortic valve bypass and severe parenchymal interstitial fibrosis.
|
|
| Comment |
|---|
|
|
|---|
Aortic valve bypass surgery is not a new operation. The concept of an apicoaortic valved conduit to bypass valvular aortic stenosis was conceived by Carrel in 1910 [8], performed experimentally by Sarnoff and colleagues [9] in 1955, and performed clinically by Templeton [6] in 1962. The technique did not gain popularity because of problems with hemolysis and emboli related to the conduit components and because of the success of direct aortic valve resection and replacement. In the early 1970s, a woven Dacron graft containing a porcine valve was introduced for right heart reconstruction. Brown and colleagues [6] used this valved conduit experimentally in the apicoaortic position and found that it functioned well to bypass the obstructed left ventricular outflow tract. Late animal data from Brown's group and early clinical experience by Brown and several other groups stimulated occasional clinical application of the apicoaortic conduit [6]. The published experience with the aortic valve bypass is summarized in Table 1.
|
There are several advantages of aortic valve bypass surgery compared with conventional aortic valve replacement. Cross-clamping of the ascending aorta is never required. Cardiopulmonary bypass is used infrequently, and when necessary, for a brief period of time. As we have gained experience with the technical aspects of aortic valve bypass surgery, cardiopulmonary bypass has been used less frequently (see Fig 2). The median duration of cardiopulmonary bypass in this series (15 minutes) is far less than that associated with conventional isolated aortic valve replacement (110 minutes) [2].
Conventional aortic valve replacement is associated with a significant risk of thromboembolism that is amortized over the life of the patient. Both mechanical and bioprosthetic aortic valves are associated with a risk of thromboembolism that is estimated to be between 0.5% and 2.0% per year [19]. A recent report of more than 3,000 patients having left heart valve operations and a cumulative follow-up of more than 20,000 patient-years demonstrated linearized embolic stroke rates of 1.3% ± 0.2% for aortic bioprostheses and 1.4% ± 0.2% for aortic mechanical replacement valves [20]. By 15 years after operation, approximately 20% of patients having aortic valve replacement had suffered a stroke. We believe that the basic configuration of the aortic valve bypass may be brain protective. Blood flow to the brain is through the native valve, while the conduit supplies the remainder of systemic blood flow. Thrombi that form on the prosthetic valve likely cannot reach the brain. The thromboembolic burden downstream from the valved conduit may be less than seen from a conventional prosthesis in the native position, because the sewing cuffs (and pledgets) are not exposed to the blood path. Careful scrutiny of the literature (Table 1) demonstrates a complete absence of late strokes among hospital survivors of aortic valve bypass surgery. We have continued to use aspirin alone after aortic valve bypass surgery, and have enjoyed complete freedom from perioperative and late stroke.
Paravalvular leak is present in as many as 6% of patients early after conventional aortic valve surgery [21]. Severe paravalular leak requiring reoperation occurs in 1.0% to 3.5% of patients after conventional aortic valve replacement [22], and is the second most common reason for reoperation after structural valve deterioration of a bioprosthetic valve. Paravalvular leak cannot occur after aortic valve bypass surgery.
Conduction system injury is a recognized complication of conventional aortic valve replacement that does not occur after aortic valve bypass surgery. The incidence of permanent pacemaker insertion after conventional aortic valve replacement is between 3% and 6% [23]. No patient in the current series required a new pacemaker during hospitalization.
Patient-prosthesis mismatch has been estimated to occur in as many as 50% of patients after conventional aortic valve replacement [24]. Patient-prosthesis mismatch has been associated with increased perioperative mortality, decreased postoperative exercise tolerance, and failure of left ventricular mass regression. The aortic valve bypass has two critical advantages over conventional aortic valve replacement in the sizing domain. Patient-prosthesis mismatch is completely avoided with the aortic valve bypass because the surgeon is not limited in the size valve and conduit that can be implanted. In our initial series, we have chosen a nominal size 21 stentless aortic valve coupled to a 20-mm conduit. Stentless valves are associated with a very low incidence of patient-prosthesis mismatch because they have a high performance index (ratio of internal orifice area to external orifice area). We believe that this size is more than adequate for all adult patients undergoing aortic valve bypass for aortic stenosis. The final left ventricular outflow area after aortic valve bypass is the sum of the area of the native stenotic valve (usually 0.5 to 1.0 cm2) and the area of the valve in the conduit. In comparison, the total effective orifice area after a conventional aortic valve replacement is limited to the internal orifice area of the prosthetic valve.
We observed excellent hemodynamic results after aortic valve bypass surgery in all patients. Left ventricular ejection performance was unchanged after operation. Measured gradients across the native valve were low (mean gradient, 8.8 ± 3.3 mm Hg) and similar to results after conventional aortic valve replacement [2527], suggesting excellent relief of left ventricular outflow obstruction. Measured gradients across the native valve are equivalent to gradients across the aortic valve bypass. Aortic valve bypass surgery affords durable relief of left ventricular outflow tract obstruction. Freeman and associates [28] recently reported a 24-year survivor of aortic valve bypass surgery, and we (J.W.B.) have a patient who is 25 years out from operation with a normally functioning conduit.
As our experience performing aortic valve bypass surgery has grown, we have learned the importance of certain technical aspects of the operation. The ventricular apex can be a friable structure, particularly in the elderly. The majority of the strength of the apex is within the epicardium. Sutures that are placed to the appropriate depth and that are tied in a manner that creates a hemostatic seal without tearing the epicardium are key to the construction of a hemostatic and reliable anastomosis. Our series is notable for 2 patients who required early reoperation for apical pseudoaneurysms. Control of bleeding from the friable apex and the development of pseudoaneurysms remain a significant concern and perhaps can be mitigated with improved technologynow under developmentfor inserting and securing the apical connector. We have routinely performed follow-up predismissal multislice computed tomography and have not seen evidence of additional apical anastomotic complications.
This study demonstrates that aortic valve bypass surgery is a promising alternative to conventional aortic valve replacement for the high-risk patient with aortic stenosis. Avoidance of sternotomy and cardiopulmonary bypass supports broader application to moderate-risk patients with aortic stenosis and as a control arm for studies of percutaneous interventional therapies.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Dimitrakakis, Z. Makhija, H. Luckraz, and P. A. O'Keefe Pseudo-aneurysm formation post apico-aortic conduit Interactive CardioVascular and Thoracic Surgery, August 1, 2009; 9(2): 377 - 378. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Balaras, K.S. Cha, B. P. Griffith, and J. S. Gammie Treatment of aortic stenosis with aortic valve bypass (apicoaortic conduit) surgery: an assessment using computational modeling. J. Thorac. Cardiovasc. Surg., March 1, 2009; 137(3): 680 - 687. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Matsushita, T. Kawase, E. Tsuda, and K. Kawazoe Apicoaortic conduit for the dilated phase of hypertrophic obstructive cardiomyopathy as an alternative to heart transplantation Interactive CardioVascular and Thoracic Surgery, February 1, 2009; 8(2): 232 - 234. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Gammie, L. S. Krowsoski, J. M. Brown, P. N. Odonkor, C. A. Young, M. J. Santos, J. S. Gottdiener, and B. P. Griffith Aortic Valve Bypass Surgery: Midterm Clinical Outcomes in a High-Risk Aortic Stenosis Population Circulation, September 30, 2008; 118(14): 1460 - 1466. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T.L. Chiam and C. E. Ruiz Percutaneous Transcatheter Aortic Valve Implantation: Assessing Results, Judging Outcomes, and Planning Trials: The Interventionalist Perspective J. Am. Coll. Cardiol. Intv., August 1, 2008; 1(4): 341 - 350. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. Rosengart, T. Feldman, M. A. Borger, T. A. Vassiliades Jr, A. M. Gillinov, K. J. Hoercher, A. Vahanian, R. O. Bonow, and W. O'Neill Percutaneous and Minimally Invasive Valve Procedures: A Scientific Statement From the American Heart Association Council on Cardiovascular Surgery and Anesthesia, Council on Clinical Cardiology, Functional Genomics and Translational Biology Interdisciplinary Working Group, and Quality of Care and Outcomes Research Interdisciplinary Working Group Circulation, April 1, 2008; 117(13): 1750 - 1767. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Doi, Y. Takahara, K. Mogi, and M. Hatakeyama Subepicardial aneurysm after apicoaortic bypass J. Thorac. Cardiovasc. Surg., July 1, 2007; 134(1): 237 - 238. [Full Text] [PDF] |
||||
![]() |
F. di Marco and G. Gerosa Percutaneous aortic valve replacement: Which patients are suitable for it? A quest for a controlled use J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 294 - 298. [Full Text] [PDF] |
||||
![]() |
P. McCarthy Invited commentary. Ann. Thorac. Surg., May 1, 2006; 81(5): 1610 - 1611. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |