|
|
||||||||
Ann Thorac Surg 2006;81:72-76
© 2006 The Society of Thoracic Surgeons
Division of Cardiothoracic Surgery, Department of Surgery, Washington University School of Medicine, St. Louis, Missouri
Accepted for publication June 10, 2005.
* Address correspondence to Dr Damiano, Division of Cardiothoracic Surgery, Washington University School of Medicine, 660 S Euclid Ave, Box 8234, St. Louis, MO 63110 (Email: damianor{at}wustl.edu).
| Abstract |
|---|
|
|
|---|
METHODS: Twelve domestic pigs (40 to 45 kg) were subjected to microwave ablation in either the arrested (n = 6) or beating heart (n = 6). The cardioplegia-arrested heart was maintained at 10° to 15°C while six atrial endocardial and seven right ventricular epicardial lesions were created in each animal. On the beating heart, six right atrial and seven ventricular epicardial lesions were created. Ablations were performed for 15, 30, 45, 60, 90, 120, and 150 seconds (65 W, 2.45 GHz). The tissue was stained with 2,3,5-triphenyl-tetrazolium chloride, and sectioned at 5-mm intervals. Lesion depth and width were determined from digital micrographs.
RESULTS: Mean atrial wall thickness was 2.8 mm (range, 1 to 8 mm). In the arrested heart, 94% of atrial lesions were transmural at 45 seconds and 100% were transmural at 90 seconds. In the beating heart, only 20% of atrial lesions were transmural despite prolonged ablation times (90 seconds). Ventricular lesion width and depth increased with duration of application, and were similar on the arrested and beating hearts.
CONCLUSIONS: Microwave ablation produces linear doseresponse curves. Transmural lesions can be reliably produced on the arrested heart, but not consistently on the beating heart.
| Introduction |
|---|
|
|
|---|
There are a number of obstacles that must be overcome for these new procedures to be as successful as the traditional cut and sew Cox-Maze procedure. The most important requirement is that the device must be able to produce transmural lesions on the arrested and beating heart safely without producing collateral damage to vital structures. This requires the development of doseresponse curves on living tissue that would allow surgeons to deliver appropriate energy based on the tissue depth, temperature, and the clinical situation. The purpose of this study was to investigate the FLEX 10 microwave device (Guidant, Inc, Santa Clara, CA) and to establish dosimetry in a surgically relevant model.
Microwave energy generates heat within tissue by inducing oscillation of dielectric molecules such as water [10]. The kinetic energy produced in these molecules by the microwave field is imparted to the tissue as heat. This form of heat production is known as dielectric heating. The depth of the microwave lesions is dependent on the power, the duration of application, and the type of microwave antennae [11]. The advantage of microwave over radiofrequency energy is that it produces a deeper lesion with even penetration and less surface heating [12]. Tissue temperatures exceed 50°C but remain below 100°C, hence reducing the risk of tissue charring and possible thromboembolism [13, 14].
In this study, the FLEX 10 device was used to produce endocardial atrial and epicardial right ventricular lesions on the cardioplegia- arrested heart and epicardial atrial and right ventricular lesions on the beating heart. The FLEX 10 device consists of a 26-mm antenna, which is encased in a flexible whiplike 53.5-cm polytetrafluoroethylene (Teflon) sheath (Fig 1). The antenna can be adjusted within the sheath to produce ablation lines 2 cm in length per application and a maximum ablatable segment length of 20 cm (10 x 2 cm segments) if the device is held in one position. The device principally was introduced for epicardial application on the beating heart. The manufacturer's recommended setting is 65 W for 90 seconds at a frequency of 2.45 GHz for the beating heart and 45 seconds for the arrested heart [14].
|
| Material and Methods |
|---|
|
|
|---|
In 6 other animals, epicardial atrial and ventricular lesions were created on the beating normothermic heart at 65 W. Atrial lesions were performed at 15, 30, 45, 60, 90, and 120 seconds whereas right ventricular lesions were performed at 15, 30, 45, 60, 90, 120, and 150 seconds.
Histologic Assessment
At the end of the procedure, 1% 2,3,5-triphenyl-tetrazolium chloride was perfused into the left ventricle of the beating heart and into the aortic root of the arrested heart before the animals were sacrificed. The hearts were then removed en bloc. The myocardial lesions were examined for evidence of charring and tissue disruption. The hearts were then placed in 1% 2,3,5-triphenyl-tetrazolium chloride solution and incubated at room temperature for 45 minutes. Each microwave lesion was sectioned four times, 5.0 mm apart, perpendicular to the long axis of the ablation line. Each section was then digitally photographed along with a 5-mm caliper to allow for calibration (Fig 2). Lesion width and depth and atrial tissue thickness were analyzed using commercial software (Adobe Photoshop, San Jose, CA). The lesion depth and width were measured from the unstained area to the pink halo region surrounding each lesion. The accuracy of this technique was ± 0.03 mm. A total of 13 lesions from each animal were examined.
|
| Results |
|---|
|
|
|---|
Atrial Ablation on the Arrested Heart
Ninety-four percent of ablated atrial sections created on the arrested heart were transmural at 45 seconds (n = 16) and 100% of ablated atrial sections were transmural at 90 seconds (n = 14; Fig 3). In some instances, the ablation lines produced on the atrium were less than 4 cm in length even though two continuous 2-cm segments of the device were activated. All sections of every tissue sample were transmural after 90 and 120 seconds of ablation for the arrested heart.
|
|
|
|
| Comment |
|---|
|
|
|---|
This study demonstrated that the FLEX 10 device was capable of producing transmural atrial lesions on the cardioplegia-arrested heart. The lesions were well defined and easily visible. All atrial lesions created on the arrested heart were transmural with ablation duration of 90 seconds. The lesions that did not achieve transmurality were over the trabeculated areas of the right atrium, particularly the crista terminalis, which can achieve thicknesses up to 8 mm. However, our doseresponse curves suggested that more prolonged ablation will be required in patients if atrial thickness exceeds 4 mm, as may be the case in the human diseased atria. In this study in a porcine model, the mean atrial thickness was less than 3 mm.
On the beating heart, it was difficult to produce transmural lesions consistently. Only 20% of ablated atrial sections were transmural even with prolonged ablation times. Interestingly, the mean lesion depth did not increase significantly with time. At 45 seconds, mean lesion depth was 2.3 ± 0.8 mm and at 120 seconds mean lesion depth was 2.6 ± 1.0 mm. The circulating blood pool with its cooling effect on the endocardium is likely responsible for this limitation on lesion depth. Epicardial fat also may have restricted the depth of some lesions. These findings should cause surgeons to cautiously use this technology in this setting because of its inability to create transmural lesions.
Other investigators have documented similar difficulty in creating transmural lesions on the beating heart with both microwave and other unipolar energy sources, particularly radiofrequency energy. Van Brakel and colleagues [21] experienced limited success with the FLEX 10 device in creating transmural lesions on the beating heart. In their study, 33% ± 19% of the lesions created on the beating heart acutely were transmural. However, the percentage of transmurality increased in the chronic setting to 66% ± 14% [21]. A similar shortcoming was demonstrated with unipolar radiofrequency ablation on the beating and arrested heart in an ovine model. Epicardial fat and the circulating blood pool were believed to have an important negative effect on the production of transmural lesions [22]. Our laboratory found unipolar radiofrequency to be incapable of creating transmural lesions in an ovine beating heart model [23]. In humans, only 8% of lesions created on the beating heart with unipolar radiofrequency were transmural [24].
There has been controversy in the literature as to whether the creation of transmural lesions is necessary to surgically cure atrial fibrillation. However, it is not arguable that lesions must create conduction block to be effective as a treatment for atrial fibrillation [25]. Ablation lines are designed to either electrically isolate the triggers in the pulmonary veins or block the macro-reentrant circuits responsible for permanent atrial fibrillation. Unfortunately, the only guarantee of complete and permanent conduction block is a transmural lesion. Thus, our data would suggest that microwave ablation should be used with caution on the beating heart unless pacing can be used to document conduction block. During pulmonary vein isolation, pacing to establish conduction block can be used as an indicator of adequate lesion depth. However, there is no such indicator for transmurality for other lesions used in the Cox-Maze or other procedures for atrial fibrillation.
The doseresponse curves yielded clinically relevant information about the biophysics of microwave ablation on living tissue. The importance of recognizing the dosimetry of this device cannot be overemphasized if the surgeon is to avoid injury to collateral structures. Interestingly, there was no significant difference in lesion depth between the two groups, and the dosimetry curves for both the beating and the arrested heart were similar. These data suggest that in thick ventricular tissue there is little effect of the circulating blood volume on lesion depth. The lesion depth increased with duration of ablation and exceeded a depth of 4 mm after 90 seconds of ablation in both the cardioplegia-arrested and beating heart. Although normal atrial thickness is less than 5 mm, the myocardium may be thicker in the region of the crista terminalis and in diseased atria [22]. However, the FLEX 10 device was capable of producing lesions in excess of 6 mm at extended periods of ablation in both the cardioplegia-arrested heart and the beating heart. It would appear that the ability of microwave energy to penetrate the ventricular tissue was unaffected by the tissue temperature. In contradistinction, it was observed that the lesions created on the right ventricle were significantly wider in the beating heart when compared with the arrested heart. This difference in lesion width may have been related to the lower baseline temperature of the arrested heart. In the arrested heart, the myocardial temperature was maintained at 10° to 15°C as compared with 37°C for beating heart procedures.
Study Limitations
The main limitation of this study was that it was performed in an acute model. It is possible that a chronic model would have yielded slightly different results, in that there may be lesion extension with time [21]. However, 2,3,5-triphenyl-tetrazolium chloride staining has been documented to be effective, reliable, and sensitive in determining the extent of necrosis [26]. Our laboratory has used this previously, and has found that acute results with 2,3,5-triphenyl-tetrazolium chloride staining correlated well with chronic histologic examination [27, 28].
The second limitation of this study was that it examined the effect of microwave ablation on normal porcine hearts. It is possible that results would differ under pathologic states. Usually, atrial thickness is increased in patients with valvular heart disease. This would exacerbate any difficulties seen with transmurality. Another unavoidable shortcoming of this study was that doseresponse curves were determined on right ventricular tissue. Unfortunately, atrial tissue is too thin to allow for accurate determination of dosimetry. However, there is no reason to expect that there would be significant biophysical differences between atrial and ventricular microwave ablation. Finally, the method used for determining lesion depth and width, although accurate, can be prone to observer error. A caliper set at 5 mm was always photographed along with the tissue sample, and this was used as a reference of measurement in an attempt to reduce error.
Conclusions
Doseresponse curves for the FLEX 10 device were established in the cardioplegia-arrested heart and on the beating heart. In myocardial tissue unaffected by the cooling effects of the circulating blood volume, the device was capable of producing lesions in excess of 4 mm in depth both on the cardioplegia-arrested and on the beating heart. However, the device failed to create transmural atrial lesions consistently on the beating heart. It is recommended that surgeons should avoid using this device for beating heart pulmonary vein isolation or at least incorporate pacing after the procedure to confirm conduction block.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. El Oumeiri, A. J. Poncelet, and G. El Khoury Why is freedom from atrial fibrillation still lower with endoscopic pulmonary vein isolation than with the Cox maze III procedure? J. Thorac. Cardiovasc. Surg., April 1, 2009; 137(4): 1036 - 1036. [Full Text] [PDF] |
||||
![]() |
J. Wang, X. Meng, H. Li, Y. Cui, J. Han, and C. Xu Prospective randomized comparison of left atrial and biatrial radiofrequency ablation in the treatment of atrial fibrillation Eur. J. Cardiothorac. Surg., January 1, 2009; 35(1): 116 - 122. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Vicol, D. Kellerer, P. Petrakopoulou, I. Kaczmarek, P. Lamm, and B. Reichart Long-term results after ablation for long-standing atrial fibrillation concomitant to surgery for organic heart disease: Is microwave energy reliable? J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1156 - 1159. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-i. Sakamoto, R. K. Voeller, S. J. Melby, S. C. Lall, N.-l. Chang, R. B. Schuessler, and R. J. Damiano Jr. Surgical ablation for atrial fibrillation: The efficacy of a novel bipolar pen device in the cardioplegically arrested and beating heart. J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1295 - 1301. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi, S. Nascimbene, A. Galanti, A. Fumero, E. Dorigo, V. Zerbi, M. Cioni, and O. Alfieri Complete left atrial ablation with bipolar radiofrequency Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 590 - 595. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov Choice of Surgical Lesion Set: Answers From the Data Ann. Thorac. Surg., November 1, 2007; 84(5): 1786 - 1792. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Balasubramanian, T. Theologou, and I. Birdi Microwave surgical ablation for atrial fibrillation during off-pump coronary artery surgery using total arterial-Y-grafts: an early experience Interactive CardioVascular and Thoracic Surgery, August 1, 2007; 6(4): 447 - 450. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Sagbas, B. Akpinar, I. Sanisoglu, B. Caynak, B. Tamtekin, K. Oral, and B. Onan Video-Assisted Bilateral Epicardial Pulmonary Vein Isolation for the Treatment of Lone Atrial Fibrillation Ann. Thorac. Surg., May 1, 2007; 83(5): 1724 - 1730. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Lall, S. J. Melby, R. K. Voeller, A. Zierer, M. S. Bailey, T. J. Guthrie, M. R. Moon, N. Moazami, J. S. Lawton, and R. J. Damiano Jr The effect of ablation technology on surgical outcomes after the Cox-maze procedure: A propensity analysis J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 389 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Melby, A. Zierer, S. P. Kaiser, R. B. Schuessler, and R. J. Damiano Jr Epicardial microwave ablation on the beating heart for atrial fibrillation: The dependency of lesion depth on cardiac output. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 355 - 360. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. G. Smedira and A. M. Gillinov Invited commentary Ann. Thorac. Surg., January 1, 2006; 81(1): 76 - 77. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |