Ann Thorac Surg 2005;80:962-968
© 2005 The Society of Thoracic Surgeons
a Department of Biomedical Engineering, Worcester Polytechnic Institute, University of Massachusetts Medical Center, Worcester, Massachusetts
b Department of Mechanical Engineering, Worcester Polytechnic Institute, University of Massachusetts Medical Center, Worcester, Massachusetts
c Department of Surgery, Division of Plastic Surgery, University of Massachusetts Medical Center, Worcester, Massachusetts
d Division of Cardiac Surgery, University of Massachusetts Medical Center, Worcester, Massachusetts
Accepted for publication March 21, 2005.
* Address reprint requests to Dr Billiar, Worcester Polytechnic Institute, 100 Institute Rd, Worcester, MA 01609 (Email: kbilliar{at}wpi.edu).
| Abstract |
|---|
|
|
|---|
METHODS: Lateral distraction tests were conducted on bisected polyurethane sternal models fixed with either a standard 7 wire configuration (n = 5) or a 3 plate configuration (n = 3). To assure controlled loading, the sternal models were attached to a computer-controlled test machine by a novel tethering system that distributes the total force (180N) equally to eight locations on the sternum. Stability was defined as the ability to restrict sternal separation at seven locations along the midline quantified using digital image analysis.
RESULTS: Our results indicate that rigid plate fixation significantly reduced lateral motion relative to wire fixation. The lower sternal region most noticeably benefited from plate fixation as the splaying observed for wire fixation was reduced.
CONCLUSIONS: Under these loading conditions, plating increased stability at the midline compared to wires; this increased stability may facilitate the recovery of high-risk patients undergoing cardiac operation. To enhance in vitro testing methods, future studies should incorporate additional in vivo loading conditions applied to the sternum. Alternate plating configurations should also be examined to further increase stability.
| Introduction |
|---|
|
|
|---|
A number of alternatives to stainless steel cerclage wires have been studied in recent years including stainless steel bands and side plates to reduce wire pull through [711] and rigid fixation plates [1214]. In craniofacial and orthopedic applications, where similar morbidity and complications including infection, delayed healing, and nonunion are observed with wire fixation, adoption of rigid plate fixation has been near universal [15, 16]. In fact, preliminary clinical studies indicate that rigid plate fixation improves sternal healing over wire fixation [4, 13, 14, 17]. However, cardiothoracic surgeons have been reluctant to adopt rigid plate fixation of the sternum because of concerns including drilling near the heart and bypass conduits, the added time and expense of plates, and the difficulty of reentry to the chest cavity in emergency situations [17]. A recent study by Song and colleagues [17] addressed these concerns by examining commercially available titanium plates specifically designed for sternal fixation (see discussion in Ref 17). In their study, plating reduced mediastinitis in high-risk patients. Since the number and placement of plates on the sternum was based on intuition rather than mechanical measurements, and no mechanical analysis of stability was performed, it is unclear if the plating configuration used in the study provided maximum stability with a minimal number of plates. For widespread acceptance of rigid plate fixation, systematic biomechanical studies must be conducted to determine simple and mechanically sound plating configurations for the sternum [21].
This paper describes the first in a series of systematic biomechanical studies analyzing the stability of rigid plate fixation for closure of median sternotomy. The overall hypothesis of this project is that improving the mechanical stability of sternal fixation through optimal plate usage will further enhance the rate of sternal healing and decrease the occurrence of complications associated with median sternotomy. To begin testing this hypothesis, this study describes a quantitative comparison of the mechanical stability of sternal models reapproximated using a standard wire configuration and a standard plate configuration using a computer-controlled in vitro test system.
| Material and Methods |
|---|
|
|
|---|
Specimen Preparation
Commercially available sternum-shaped polyurethane bone models with a density of 20 lb/ft3 and elastic modulus of 5.5 MPa (Sawbones, Pacific Research Laboratories, Vashon, WA) were bisected using standard midline sternotomy techniques and either plated using titanium sternal plates (SternaLock Plates, Walter Lorenz Surgical, Jacksonville, FL) or wired using stainless steel wires (Ethicon Inc, Somerville, NJ) by a surgeon at the University of Massachusetts Medical Center (author NF). Our plate configuration consisted of three X ("3X") plates spaced evenly down the sternum, as seen in Figure 1a. This configuration was considered the most basic arrangement of plates that would be used clinically and is not necessarily an optimal configuration of rigid metal plates. The wired samples were fixated using the standard 7S wire configuration (7 simple straight wires). Five sternal models were configured with wires and three sternal models were configured with plates. A greater number of wired samples were used as we expected relatively high variability in this group.
|
The sternal models were pulled laterally at 0.212 mm/s. In preliminary tests, forces exceeding 200N caused the cerclage wires to cut into, and in some cases fracture, the model sterna; thus, the maximum load was limited to 180N for this study. At increments of 60N the separation of the sternum was measured as described below.
Quantification of Stability of Fixation
Stability was quantified by measuring the amount of separation between the two halves of the sternum using digital image analysis. Small graphite chips, used as fiducial markers (
0.5 mm diameter, Fig 1b), were affixed in pairs on opposite sides of the sternal model halves at seven different locations using cyanoacrylate adhesive. A series of high resolution pictures was taken at each force increment using a digital camera (5.0 mega pixel Coolpix 5700; Nikon, Melville, NY). Marker locations at the manubrium (M), the proximal to distal midsternum (MS1, MS2, MS3, MS4, MS5), and the xiphoid (X) (Fig 1B) were determined using image analysis software (Scion Image; Scion Corporation, Fredrick, MD) by calculating their centroids after thresholding (Figs 1c and 1d). Scale markers were used to convert pixels to millimeters; the resolution of the digital images was 12 µm/pixel.
Based on previous studies [12], we also calculated the structural stiffness at each location for each sample (ie, slope of the force-displacement curve). However, we have not included these data since, due to the linearity of the force-displacement curves, this parameter does not provide any supplementary information in addition to the separation data presented herein.
Statistical Analysis
Overall differences in stability between the wire and plate configurations were determined using multivariate analysis of variance (MANOVA, SPSS 12.0; SPSS Inc, Chicago, IL) to compare the sternal separations at each location at the peak force (180N). A level of p less than 0.05 was considered significant. The MANOVA was necessary to compare the fixation types while accounting for the fact that the locations along the sternum are spatially related and thus are not independent. To determine differences at specific locations between the fixation configurations, post hoc analysis was performed using the Tukey HSD (honestly significantly different) test.
| Results |
|---|
|
|
|---|
|
|
|
|
400N) as the wired models (
200N). Additionally, the plated models fractured at the rib struts where they were attached to the tethering system and not the screws. In contrast, the wired models failed due to wire pull-through and not the attachment method. Since this study focused on subfailure behavior, all subsequent tests were conducted with a peak load of 180N for comparability between fixation methods. A more noticeable variation in stability was found along the length of the sternal models for the wire configuration than for the plate configuration. However, for both fixation methods the xiphoid region proved to be susceptible to large separations (Fig 3), and the separations in the M, MS5, MS4, MS3, and MS2 locations remained relatively low at all forces (Tables 1 and 2). The largest difference between fixation methods was in the lower sternal region. In the case of wires (Fig 3a), the lower sternal region (X and MS1 marker locations) experienced average sternal separations greater than 4 mm, whereas the remaining locations separated less than 1 mm. For plates, the separations at all regions were below 2 mm (Table 2). As a result of this disparity, the variation in mean separation along the length of the sternum of the plated sample data (0.08 to 0.70 mm) was lower than that of the wired group (-0.02 to 6.9 mm). The variability at each measurement location was also relatively high (eg, greater than 45% coefficient of variation in all locations for wire-fixed samples) despite the fact that polyurethane models were used and two additional wire-fixed samples were tested (Table 1). Interestingly, we observed negative separation values at the MS5 location for the wired models (Table 1) and the M and MS1 locations of the plated models (Table 2), possibly due to bending effects (see Comment).
| Comment |
|---|
|
|
|---|
Rigid plates provide significantly more stable fixation than wires, especially in the xiphoid region. The average separations in the lower sternal regions (X and MS1 locations) of wire-fixed sternal models are ten to fifteen times greater than the separations observed at the other five locations (Fig 2c) and much larger than observed in the plated samples (Fig 2d). Splaying at the xiphoid coincides with clinical observations and is a concern with the wiring technique. This phenomenon is most likely due to the lack of wire support directly in the area of the lower sternum. Dasika and colleagues [19] demonstrate in vitro that reinforcing the xiphoid region of sternal models with additional wires reduces these large separations; however in clinical practice, the lower sternum represents a difficult area of fixation as there is transition to a cartilaginous region with variable degrees of parasternal calcification. Although it may appear "unfair" to compare wired and plated configurations where the plate is situated further towards the xiphoid than the lowest wire, we chose to compare two standard fixation techniques, as each would be clinically performed.
In vivo, the sternum is loaded by a complex combination of tensile, shear, and compressive forces acting in multiple directions. Therefore, many loading paradigms including lateral [7, 9, 20], transverse, shear [21], and bending [12] have been utilized in in vitro studies to assess the mechanical stability of various sternal fixation methods. In cadaver studies, McGregor and colleagues [21] found that lateral distraction provides the most sensitive measure of the efficacy of fixation since bone-on-bone frictional forces do not have to be overcome in this direction. Based on this finding, they have developed efficient in vitro test methods to avoid the variability and expense inherent in cadaveric testing [19, 20].
Furthermore, in their cadaver studies, the authors [21] had the insight to measure the sternal separation at multiple locations. They observed substantial variation in the distraction along the length of the sternum, most notably large separations in the xiphoid region. These results were the first to demonstrate that the location where one measures sternal separation is critical in determining the effectiveness of a fixation method. This point is also clearly demonstrated by the fact that the difference in stability between the fixation methods in our study was not significantly different in the upper sternal region (M, MS5, MS4, and MS3 locations), whereas it was overwhelmingly different in the lower region (Fig 3). Previous studies comparing the stability of fixation between plates and wires [12] were limited because they only measured stability in one location. McGregor and colleagues [21] attribute the nonuniform separations to anatomic factors such as variable bone thickness and stabilizing structures (eg, clavicle); however, uneven distribution of loading along the sternum inherent in their method could also have contributed to the nonuniform separation. In their cadaver tests, the sternum was loaded at only three locations along its length (second, fourth, and sixth ribs) and, although the total force was known, the distribution of load between the three tethers was neither controlled nor quantified. To enable proper interpretation of the nonuniform separations along the sternum in our testing method, we applied a known distribution of forces along the length of the sternum. In contrast, the loads applied to sternal models in most in vitro systems are spread over multiple ribs by potting them in epoxy and/or clamping them [9, 12, 19, 20], yet the load distribution remains unknown.
In addition to the separation of the sternum being nonuniform, we observed negative sternal "separation" at the manubrium and the midsternum regions in multiple samples (Fig 3, Tables 1 and 2). This nonintuitive inward motion (only outward forces were applied) appears to be the result of flexure of the sternum that results from the wires and screws acting as pivots. It is important to note that although the negative separations were sufficiently large to be detected by our image analysis, bending was not visible to the naked eye and the deflections were relatively small (0.010.1mm) in comparison with the separations at the X and MS1 locations (> 3 mm).
To demonstrate the impact of the particular load distribution applied to the sternum during in vitro testing, our data from wire-fixed models at 60N, 120N, and 180N are graphed with analogous data from Trumble and colleagues [20] in Figure 4. In the previous study, the ribs are sandwiched between rigid plates (2 sets per side) and attached to the loading mechanism by 4 tethers (2 per set of plates). The "manubrium", "midsternum," and "xiphoid" locations correspond with MS1, MS3, and M locations defined herein, and the separations for 120N are linearly interpolated from the published data. The difference between the two data sets at the manubrium is minimal at all three loads (< 1 mm); however, the separations measured in the present study are considerably smaller at the midsternum and larger and at the xiphoid for all loading levels. These differences are most likely due to greater loading in the center of the sternum in the previous study relative to the loading of our models, although the true force distribution from the previous study cannot be calculated due to the linkage mechanisms. While these data demonstrate that the distribution of loading chosen for in vitro testing is critical, the actual in vivo forces at different points along the sternum are unknown, thus it is unclear which method provides more physiologically relevant data.
|
Clearly, the quasistatic loading methods described in this study are not sufficient to simulate demanding physiologic conditions such as chronic coughing secondary to emphysema or respiratory congestion, and vertical or transverse shear loading [9]. Furthermore, the homogenous nature of the polyurethane models does not simulate the bicortical structure of sternal bone or the cartilaginous xiphoid, making them unable to withstand large mechanical loads applied during coughing or predict the effects of wire and screw pull-through. Future studies should utilize multiple sternal models and test protocols including a more robust laminated polyurethane model exposed to multiple loading directions and repetitive cyclic loading [6, 9, 24].
This report is the first in a series of systematic studies examining the stability of rigid plate fixation for the closure of median sternotomy. The results of this study demonstrate that a simple 3 plate fixation configuration significantly reduces sternal separation in comparison to standard 7 wire fixation, particularly splaying at the xiphoid. It is important to note that the methods used in this study are meant to gauge the relative stability of each closure method, not to approximate actual clinical conditions. Such in vitro testing systems are highly desirable because they allow for uniform, inexpensive, rapid, and consistent mechanical comparisons of different closure methods that are not possible in vivo. We found that the local stability of the fixated sternum is dependent upon the placement of plates and wires, the loading distribution applied to the sternum, and the measurement locations. These findings highlight the need for more detailed analysis of the mechanics of sternal fixation. The in vitro testing system described in this study represents an important, yet incremental step toward developing a comprehensive biomechanical model system for effectively evaluating sternal closure techniques. Further studies are necessary for determining optimal plate configurations to reduce wound healing complications in a variety of patient populations.
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. W. Snyder, L. A. Graham, R. E. Byers, and W. L. Holman Primary sternal plating to prevent sternal wound complications after cardiac surgery: early experience and patterns of failure Interactive CardioVascular and Thoracic Surgery, November 1, 2009; 9(5): 763 - 766. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Z. Atkins, M. K. Wooten, J. Kistler, K. Hurley, G. C. Hughes, and W. G. Wolfe Does Negative Pressure Wound Therapy Have a Role in Preventing Poststernotomy Wound Complications? Surgical Innovation, June 1, 2009; 16(2): 140 - 146. [Abstract] [PDF] |
||||
![]() |
B. Voss, R. Bauernschmitt, G. Brockmann, M. Krane, A. Will, and R. Lange Complicated Sternal Dehiscence: Reconstruction With Plates, Cables, and Cannulated Screws Ann. Thorac. Surg., April 1, 2009; 87(4): 1304 - 1306. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schimmer, W. Reents, S. Berneder, P. Eigel, O. Sezer, H. Scheld, K. Sahraoui, B. Gansera, O. Deppert, A. Rubio, et al. Prevention of Sternal Dehiscence and Infection in High-Risk Patients: A Prospective Randomized Multicenter Trial Ann. Thorac. Surg., December 1, 2008; 86(6): 1897 - 1904. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schimmer, S.-P. Sommer, M. Bensch, T. Bohrer, I. Aleksic, and R. Leyh Sternal closure techniques and postoperative sternal wound complications in elderly patients. Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 132 - 138. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Voss, R. Bauernschmitt, A. Will, M. Krane, R. Kross, G. Brockmann, P. Libera, and R. Lange Sternal reconstruction with titanium plates in complicated sternal dehiscence. Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 139 - 145. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. C.H. John Modified closure technique for reducing sternal dehiscence; a clinical and in vitro assessment Eur. J. Cardiothorac. Surg., May 1, 2008; 33(5): 769 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Zeitani, A. Penta de Peppo, A. Bianco, F. Nanni, A. Scafuri, F. Bertoldo, A. Salvati, S. Nardella, and L. Chiariello Performance of a Novel Sternal Synthesis Device After Median and Faulty Sternotomy: Mechanical Test and Early Clinical Experience Ann. Thorac. Surg., January 1, 2008; 85(1): 287 - 293. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Raman, D. Straus, and D. H. Song Rigid Plate Fixation of the Sternum Ann. Thorac. Surg., September 1, 2007; 84(3): 1056 - 1058. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Zeitani, A. P. de Peppo, M. Moscarelli, L. G. Wolf, A. Scafuri, P. Nardi, F. Nanni, E. Di Marzio, P. De Vico, and L. Chiariello Influence of sternal size and inadvertent paramedian sternotomy on stability of the closure site: A clinical and mechanical study J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 38 - 42. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Losanoff Invited commentary Ann. Thorac. Surg., September 1, 2005; 80(3): 968 - 968. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |