Ann Thorac Surg 2004;77:1431-1433
© 2004 The Society of Thoracic Surgeons
Case report
Temporary assist device for postcardiotomy cardiac failure
Antonis A. Pitsis, MD, FETCSa,
Petros Dardas, MD, FESCa,
Nikolaos Mezilis, MD, FESCa,
Nikolaos Nikoloudakis, MDa*,
Gerasimos Filippatos, MD, FESCb,
Daniel Burkhoff, MD, PhDc
a Thessaloniki Heart Institute, St. Luke's Hospital, Thessaloniki, Greece
b Evangelismos General Hospital, Athens, Greece
c Columbia University, New York, New York, USA
Accepted for publication April 18, 2003.
* Address reprint requests to Dr Nikoloudakis, Thessaloniki Heart Institute, Ermou St 69, Thessaloniki, Greece
e-mail: nik_nik{at}freenet.de
 |
Abstract
|
|---|
Postcardiotomy left ventricular failure (LVF) complicating a cardiac surgical procedure is usually managed with intraaortic balloon pump (IABP) counterpulsation. We report two cases of postcardiotomy LVF unresponsive to inotropic support and IABP counterpulsation that were managed successfully with the use of the TandemHeart percutaneous ventricular assist device (pVAD) (CardiacAssist, Pittsburgh, PA). The TandemHeart pVAD appears to be safe and effective as a bridge to recovery in the treatment of postcardiotomy cardiac failure.
 |
Introduction
|
|---|
Postcardiotomy left ventricular failure (LVF) complicating a cardiac surgical procedure is usually managed with intraaortic balloon pump (IABP) counterpulsation [1]. However, weaning from cardiopulmonary bypass (CPB) is often difficult or impossible despite IABP and inotropic support [2, 3]. Under such conditions, temporary left ventricular assist devices have been used [25]. Devices used in the past for this indication generally require surgical implantation and explantation and specially trained personnel [6]. Here, we report our experience with two cases in which postcardiotomy LVF unresponsive to IABP and inotropic support were managed with the TandemHeart percutaneous ventricular assist device (pVAD) (CardiacAssist, Pittsburgh, PA). This system improves the hemodynamic condition of patients by reducing left atrial pressure while increasing mean arterial pressure and cardiac output [7]. These favorable effects are achieved by withdrawing oxygenated blood from the left atrium and returning it to the femoral artery with a simple extracorporeal circuit and centrifugal pump.
 |
Case reports
|
|---|
Patient 1
A 61-year-old man with unstable angina was admitted for redo coronary artery bypass grafting operation 10 years after a triple-vessel bypass grafting procedure. He had a history of claudication due to aortoiliac atherosclerotic disease. A preoperative angiogram revealed significant triple-vessel disease with blocked vein grafts to the circumflex and right coronary arteries and a 95% stenosis in a vein graft to the left anterior descending artery (LAD). On beating heart, without CPB (off-pump CAB [OPCAB]), the left internal thoracic artery (LITA) was used to bypass the LAD distally to the previous anastomosis. Also, a reversed saphenous vein graft was anastomosed between the posterior descending artery and the ascending aorta. Our OPCAB technique has been described previously [8]. After completing the anastomoses, ST segment elevation suddenly appeared in the anterior leads. At the same time, the flow through the LITA was measured (Medi-Stim Flowmeter, Medi-Stim AS, Oslo, Norway) and found to be decreased to 4 mL/min from 120 mL/min. This was thought to be the result of the embolization of the LAD from the old vein graft. Cardiopulmonary bypass was established, and the LITA-to-LAD anastomosis was reexplored. The LAD was found to be full of emboli. The heart was arrested using retrograde warm blood cardioplegia in an attempt to wash back the emboli from the distal LAD bed, and the anastomosis was redone. The flow through the LITA returned to 105 mL/min, and the ischemic electrocardiographic alterations disappeared. However, weaning from CPB was not possible despite full inotropic support (epinephrine, norepinephrine, and milrinone). Attempts to insert an IABP from either groin were unsuccessful because of the aortoiliac obliterative disease. An IABP was therefore inserted from the ascending aorta. However, the IABP failed to provide sufficient support to permit separation from CPB, and it was therefore removed. The TandemHeart pVAD was inserted using the right superior pulmonary vein for the insertion of a left atrial cannula (28 F) and the ascending aorta for the arterial cannula (17 F). The pVAD provided 6 L/min output at 6000 rpm, and CPB could be weaned while inotropic and pressor support were reduced. The sternum of the patient was left open, and the skin was closed. The patient was anticoagulated with intravenous heparin at an activated clotting time (ACT) of about 180 seconds. Left ventricular function recovered over the next 4 days, at which point the device was removed, and the patient's chest was closed without the need for any other support. The patient was discharged home on the 15th postoperative day. On follow-up 1 year later, the patient was in New York Heart Association class I and was without angina.
Patient 2
A 67-year-old man with class IV heart failure due to ischemic cardiomyopathy with large anteroseptal and inferior wall myocardial infarctions (left ventricular ejection fraction 15%) and grade 4+ mitral regurgitation was admitted for surgical treatment. While on CPB, the patient underwent a modified Dor procedure (beating heart), a mitral valve repair with an Alfieri stitch, and triple bypass grafting with the LITA to the LAD and saphenous vein grafts to the circumflex and right coronary arteries. Weaning from CPB was impossible despite maximum inotropic and IABP support. The IABP was removed, and a TandemHeart pVAD was inserted. A long 22-F cannula was inserted percutaneously through the right femoral vein and advanced to the left atrium from the right atrium through the foramen ovale. The right femoral artery was cannulated with a 17-F cannula (Seldinger technique) (Fig 1).
The TandemHeart pVAD provided 2.5 L/min flow at 6000 rpm, and the patient was weaned from CPB. After CPB, the cardiac index (CI) was found to be 2.5 L · min−1 · m−2. The ACT was kept at about 180 seconds during the support period. The patient was extubated on the 3rd postoperative day, and the device was removed on the 11th postoperative day, with hemostasis achieved by applying manual pressure in the groins. After the removal of the device, the CI was 2.6 L · min−1 · m−2. The patient was discharged on the 30th day but died from ventricular tachycardia 2 months later.
 |
Comment
|
|---|
Intraaortic balloon pumping was introduced more than 30 years ago and since then has been the most widely used means of providing mechanical support to a failing heart in a variety of settings. It has been preferentially used for postcardiotomy low-output syndrome, and it is considered to reduce mortality and morbidity in this setting [1]. When weaning from CPB is not possible despite inotropic and IABP support, a surgeon is left with no choice other than the use of ventricular assist devices (VADs) [25]. VADs can offer successful circulatory support, but they are quite expensive and require a second procedure to remove them and specially trained personnel to manage the patients.
The cost of the TandemHeart pVAD is a fraction of that of implantable VADs, although direct comparison is unfair because the latter can provide long-term support. When inserted percutaneously, it can be removed by withdrawing the catheters and applying manual pressure in the groin. This technique is quite similar to the one introduced by Clarence Dennis in the late 1940s and used clinically by Edmunds and colleagues [5], although a different type of pump has been used. The TandemHeart pVAD improves the hemodynamic condition of patients by reducing left atrial pressure while increasing mean arterial pressure and cardiac output. These favorable effects are achieved by withdrawing oxygenated blood from the left atrium and returning it to the femoral artery with a simple extracorporeal circuit and radial pump.
Another major advantage of the TandemHeart pVAD is its versatility. It can be inserted totally percutaneously under fluoroscopic or transesophageal echocardiographic guidance or partially percutaneously with the aid of a right atrial or pulmonary vein incision for the transseptal positioning of the left atrial cannula.
We should also mention that neither of the two patients presented here experienced any bleeding or infective complications or any mechanical device failure.
The TandemHeart pVAD has received the CE mark, which allows its clinical use in Europe. US Food and Drug Administration approval is pending.
In summary, the TandemHeart pVAD appears to be an effective, versatile, and economical VAD as a bridge to postcardiotomy recovery.
 |
References
|
|---|
- Mehlhorn U., Kroner A., de Vivie E.R. 30 years of clinical intra-aortic balloon pumping: facts and figures. Thorac Cardiovasc Surg 1999;47(Suppl 2):298-303.
- Hausmann H., Potapov E.V., Koster A., et al. Prognosis after the implantation of an intra-aortic balloon pump in cardiac surgery calculated with a new score. Circulation 2002;106(Suppl 1):I203-206.
- Goldstein D.J., Oz M.C. Mechanical support for postcardiotomy cardiogenic shock. Semin Thorac Cardiovasc Surg 2000;12:220-228.[Medline]
- Golding L.A., Crouch R.D., Stewart R.W., et al. Postcardiotomy centrifugal mechanical ventricular support. Ann Thorac Surg 1992;54:1059-1063.[Abstract]
- Edmunds L.H., Jr, Herrmann H.C., DiSesa V.J., Ratcliffe M.B., Bavaria J.E., McCarthy D.M. Left ventricular assist without thoracotomy: clinical experience with the Dennis method. Ann Thorac Surg 1994;57:880-885.[Abstract]
- Helman D.N., Morales D.L., Edwards N.M., et al. Left ventricular assist device bridge-to-transplant network improves survival after failed cardiotomy. Ann Thorac Surg 1999;68:1187-1194.[Abstract/Free Full Text]
- Thiele H., Lauer B., Hambrecht R., et al. Reversal of cardiogenic shock by percutaneous left atrial-to-femoral arterial bypass assistance. Circulation 2001;104:2917-2922.[Abstract/Free Full Text]
- Pitsis A.A., Angelini G.D. Off pump coronary bypass grafting of the circumflex artery. Eur J Cardiothorac Surg 1999;16:478-479.[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
M. Takagaki, C. Wurzer, R. Wade, R. Lee, S. C. Malaisrie, P. M. McCarthy, and E. C. McGee Jr
Successful Conversion of TandemHeart Left Ventricular Assist Device to Right Ventricular Assist Device After Implantation of a HeartMate XVE
Ann. Thorac. Surg.,
November 1, 2008;
86(5):
1677 - 1679.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. C. McGee Jr., P. M. McCarthy, and N. Moazami
Temporary Mechanical Circulatory Support
Card. Surg. Adult,
January 1, 2008;
3(2008):
507 - 534.
[Full Text]
|
 |
|

|
 |

|
 |
 
A. A. Pitsis, A. N. Visouli, D. Burkhoff, P. Dardas, N. Mezilis, G. Bougioukas, and G. Filippatos
Feasibility Study of a Temporary Percutaneous Left Ventricular Assist Device in Cardiac Surgery
Ann. Thorac. Surg.,
December 1, 2007;
84(6):
1993 - 1999.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Mehta and W. Pae
Conversion of the TandemHeart percutaneous ventricular assist device to long-term support
Interactive CardioVascular and Thoracic Surgery,
August 1, 2006;
5(4):
444 - 445.
[Abstract]
[Full Text]
[PDF]
|
 |
|