|
|
||||||||
Ann Thorac Surg 2004;77:143-149
© 2004 The Society of Thoracic Surgeons
a Department of Thoracic and Cardiovascular Surgery, University Hospital, Muenster, Germany
b Heart Failure Center, Columbia University, New York, New York, USA
c Department of Cardiology and Angiology, University Hospital, Muenster, Germany
Accepted for publication June 25, 2003.
* Address reprint requests to Dr Klotz, Department of Thoracic and Cardiovascular Surgery, University Hospital Muenster, Albert-Schweitzer-Str 33, 48129 Muenster, Germany.
e-mail: stefan.klotz{at}thgms.uni-muenster.de
| Abstract |
|---|
|
|
|---|
METHODS: In 31 patients undergoing long-term LVAD implantation (nonpulsatile = 10, pulsatile = 21) preoperative and postoperative echocardiographic and hemodynamic assessment with right heart catheterization had been obtained.
RESULTS: All patients had similar echocardiographic, hemodynamic, and clinical heart failure characteristics at baseline. The degree of LV pressure unloading was the same in both device types, caused by similar reduction of mean pulmonary pressure (18.6 ± 5.1 versus 18.3 ± 7.5 mm Hg) and pulmonary capillary wedge pressure (8.9 ± 4.4 versus 8.0 ± 7.0 mm Hg). Left ventricular volume unloading was pronounced with a pulsatile device owing to a statistically significant higher pump output (5.1 ± 1.0 L/min) in comparison with nonpulsatile LVADs (3.6 ± 0.9 L/min, p < 0.001). Echocardiographic-determined end-systolic indicators confirm this augmentation in pulsatile LVADs. Etiology or the time interval of hemodynamic reassessment had no impact in left ventricular pressure unloading, but LV volume unloading decreased between day 60 and 120 in patients with nonpulsatile LVADs. The preoperative and postoperative transplant mortality was comparable in both groups.
CONCLUSIONS: Left ventricular pressure unloading is similar in patients with nonpulsatile as compared with pulsatile implantable long-term assist devices. Left ventricular volume unloading is pronounced in pulsatile LVADs.
| Introduction |
|---|
|
|
|---|
Improvement of cardiac function with pulsatile LVADs is well described [37]. Pulsatile LVADs provide profound LV volume and pressure unloading while simultaneously restoring adequate systemic blood flow in end-stage congestive heart failure patients [6]. However, comparison of LV pressure and volume unloading between nonpulsatile axial flow pumps and pulsatile pumps has not been investigated. While early studies with nonpulsatile blood flow in calves [8] suggest no adverse impact on nonphysiological blood flow pattern the debate regarding the effect of nonpulsatile flow on organ function and neurohormonal activation remains controversial [9, 10].
To address the question of left ventricular pressure and volume unloading we compared preoperative to postoperative echocardiographic and hemodynamic measurements in patients with implantable pulsatile and nonpulsatile long-term LVADs.
| Patients and methods |
|---|
|
|
|---|
Pulsatile LVAD implantations were performed from July 1994 to January 2000 and implantations of nonpulsatile LVADs began in February 2000. The LVAD allocation was based on an identical clinical algorithm [12] and the clinical characteristics were identical in both groups at baseline, despite the two varying time frames.
Devices
MicroMed DeBakey VAD.
The MicroMed DeBakey VAD is an electromagnetically actuated, implantable titanium axial flow pump that connects to the left ventricular apex and ascending aorta [13]. Because of the continuous flow properties of the axial flow pump no valves are needed in the system (Fig 1).
The axial flow motor is small and contains rotary blades that spin at 7,500 to 12,500 rpm and can pump approximately 5 to 6 L/min against 100 mm Hg pressure [14]. The DeBakey VAD was first implanted in our department in February 2000 [15].
|
Novacor LVAS.
The Novacor LVAS differs significantly from the TCI HeartMate in its method of pump actuation and use of a smooth blood-containing surface. During pump systole two opposing pusher plates compress a seamless polyurethane blood sac, causing ejection of blood. The inflow and outflow connections are similar to those described for the TCI HeartMate except for two 21-mm bioprosthetic valves to achieve unidirectional blood flow [16].
Hemodynamic protocol
Preoperative and postoperative hemodynamic measurements were done under the same conditions by using a Swan-Ganz catheter (Baxter Healthcare, Irvine, CA) through the left or right jugular vein, inserted with local anesthesia. Cardiac output was measured by thermodilution using rapid 10 mL bolus injections of cold saline. The averages of five measurements were used. Mean arterial pressure was measured noninvasively with the Dinamap XL (Johnson & Johnson Medical, Arlington, TX) or by invasive methods if noninvasive measurements were not possible in patients with nonpulsatile devices.
Transpulmonary gradient (TPG), pulmonary vascular resistance (PVR) and systemic vascular resistance (SVR) were calculated using the following formulas:
![]() |
![]() |
![]() |
During the postoperative assessment every patient was either in the normal ward or in our LVAD outpatient program. No patient was on inotropic agents, intraaortic balloon pump, or ventilation. All patients were on standard congestive heart failure medication titrated upward as clinically appropriate. Furthermore all patients were fully rehabilitated on their LVAD and able to perform a bicycle exercise protocol up to 50 W.
The time interval between preoperative hemodynamics and LVAD implantation was 29.6 ± 35.8 days (range, 0 to 120). The time of the postoperative hemodynamic measurements was 111.5 ± 44.9 days (range, 46 to 220) after LVAD implantation.
Echocardiographic measurements
In addition to the hemodynamic protocol we used two-dimensional echocardiography data to determine LV end-diastolic and end-systolic dimension, fractional shortening, end-diastolic and end-systolic volume, and flow pattern across the aortic valve (during LVAD support). Left ventricular ejection fraction was determined by radionuclide ventriculography or angiography. All preoperative measurements were done 6.1 ± 7.7 days before LVAD implantation; the postoperative measurements were performed during the time of the hemodynamic protocol (±15 days).
Statistical analysis
A paired Student's t test was used to compare groups. All data are expressed as mean ± SD. A p-value of less than 0.05 was considered statistically significant. For all pair-wise comparisons the Wilcoxon signed-rank test was employed to assess significance. The follow-up period ended on July 31, 2002.
| Results |
|---|
|
|
|---|
|
Hemodynamics
Hemodynamic baseline values (preoperative values) for the nonpulsatile versus the pulsatile group are shown in Table 2.
Mean pulmonary artery pressure and pulmonary capillary wedge pressure were elevated in both the nonpulsatile and pulsatile groups (mean pulmonary artery pressure, 40.2 ± 11.0 versus 33.7 ± 8.0 mm Hg, p = not significant; pulmonary capillary wedge pressure, 30.6 ± 9.5 versus 23.1 ± 7.6 mm Hg, p = 0.03) as was mean right atrial pressure (6.8 ± 5.9 versus 7.8 ± 8.1 mm Hg, p = not significant). The relatively high preoperative values of cardiac output were intensified by inotropic support or intraaortic balloon pump.
|
|
|
|
|
|
In the pulsatile group LVAD duration was 189 ± 71 days (range, 53 to 309; p = not significant versus nonpulsatile group). One patient died of multiorgan failure during LVAD support on postoperative day 53, another patient on postoperative day 116 owing to intracerebral bleeding, and a third patient on postoperative day 244 of unknown reason (the patient was found dead at home). In all other patients cardiac transplantation was performed. Four patients died after transplantation from multi organ failure, acute rejection, and acute right ventricular failure (n = 2).
| Comment |
|---|
|
|
|---|
Hemodynamic improvement with pulsatile assist devices has been studied extensively. The Berlin group [4] observed that optimal improvement in left ventricular size and function occurred within a few weeks (89 ± 79 days) after LVAD implantation. In another study from Columbia University the maximal benefits of LVAD support for reverse remodeling were between 80 and 120 days [21]. These investigators also noted that pulmonary capillary wedge pressure and mean arterial pressure decreased significantly after 30 days during LVAD support [5]. But there is a lack of comparison between pulsatile and nonpulsatile LVADs. In two studies concerning biochemical markers for brain injury or activation of the inflammatory system no differences were found in the early phases after LVAD surgery [9, 10].
Our postoperative assessment was made 112 ± 45 days after LVAD implantation and confirmed similar LV improvement as previously reported. Despite differing LVAD implantation time points in our hospital owing to availability and clinical trials both groups were similar in clinical heart failure status and hemodynamic and echocardiographic measurements. With LVAD support pulmonary vascular resistance, mean pulmonary pressure, and capillary wedge pressure reduction were highly significant in comparison with the preoperative values in failing hearts. The degree of reduction was independent of the type of device, etiology, and time of reassessment. Systemic vascular resistance and mean arterial pressure showed no improvement with a nonpulsatile LVAD, which is most likely due to the lack of pulsatile flow properties. While cardiac output increases with pulsatile and nonpulsatile LVAD support, we found a significant difference in LVAD output (p < 0.001). Pulsatile LVADs generate nearly complete LV volume unloading. In contrast nonpulsatile devices produce only a partial unloading of the left ventricle. These findings with pulsatile LVADs were supported by a pronounced decrease in end-systolic dimension and volume in comparison with nonpulsatile support. Furthermore we found a time-dependent decrease of LVVU between days 60 and 120 in these devices.
The impact of complete or partial LV volume unloading on cardiac recovery is not clear. However morbidity and mortality rates with a nonpulsatile device are similar to those of patients with a pulsatile device. Hemodynamic and echocardiographic data suggest that the completely unloaded ventricle could contract easier (without any pressure gradient) whereas the heart connected to a nonpulsatile device utilizes a part of its own contractility to create an antegrade flow. While partial unloading of the left ventricle seems to be sufficient enough for LVAD bridge to transplantation, the impact of a bridge to recovery with the goal to wean the patient from the device is not clear and further studies are warranted.
Study limitations
The small number of patients enrolled in this study limits our ability to make conclusions with respect to LV unloading or the etiology of LVAD support. Background medical management after LVAD implantation may have varied during the time period. In addition all reported values were determined under full device support (automode in both pulsatile devices, fixed speed at 10,000 rpm in the nonpulsatile device). Hence changes in hemodynamic and echocardiographic measurements with reduced LVAD flow were not determined and are warranted.
In conclusion nonpulsatile devices show left ventricular pressure unloading and pretransplant and posttransplant outcomes comparable with those of pulsatile devices. Nonpulsatile LVADs generate only partial LV volume unloading whereas pulsatile LVADs produce complete volume unloading. Further studies on the impact of weaning from the device depending on the device type are warranted.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. John, F. Kamdar, P. Eckman, M. Colvin-Adams, A. Boyle, S. Shumway, L. Joyce, and K. Liao Lessons Learned From Experience With Over 100 Consecutive HeartMate II Left Ventricular Assist Devices Ann. Thorac. Surg., November 1, 2011; 92(5): 1593 - 1600. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. H. Healy, N. O. Mason, M. E. Hammond, B. B. Reid, S. E. Clayson, S. G. Drakos, A. G. Kfoury, A. N. Patel, D. A. Bull, D. Budge, et al. Allograft Rejection in Patients Supported With Continuous-Flow Left Ventricular Assist Devices Ann. Thorac. Surg., November 1, 2011; 92(5): 1601 - 1607. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Krabatsch, M. Schweiger, M. Dandel, A. Stepanenko, T. Drews, E. Potapov, M. Pasic, Y.-G. Weng, M. Huebler, and R. Hetzer Is Bridge to Recovery More Likely With Pulsatile Left Ventricular Assist Devices Than With Nonpulsatile-Flow Systems? Ann. Thorac. Surg., May 1, 2011; 91(5): 1335 - 1340. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. John, K. Liao, F. Kamdar, P. Eckman, A. Boyle, and M. Colvin-Adams Effects on pre- and posttransplant pulmonary hemodynamics in patients with continuous-flow left ventricular assist devices J. Thorac. Cardiovasc. Surg., August 1, 2010; 140(2): 447 - 452. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. R. Rega, I. Vantichelen, H. Bollen, V. Leunens, G. Derjung, F. Kirchhof, E. Verbeken, and B. P. Meyns Pediatric heart support with a newly developed catheter based pulsatile 12F rotary blood pump: an animal study Eur J Cardiothorac Surg, December 1, 2008; 34(6): 1173 - 1178. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. John, F. Kamdar, K. Liao, M. Colvin-Adams, A. Boyle, and L. Joyce Improved Survival and Decreasing Incidence of Adverse Events With the HeartMate II Left Ventricular Assist Device as Bridge-to-Transplant Therapy Ann. Thorac. Surg., October 1, 2008; 86(4): 1227 - 1235. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Haft, W. Armstrong, D. B. Dyke, K. D. Aaronson, T. M. Koelling, D. J. Farrar, and F. D. Pagani Hemodynamic and Exercise Performance With Pulsatile and Continuous-Flow Left Ventricular Assist Devices Circulation, September 11, 2007; 116(11_suppl): I-8 - I-15. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. R. Schulman, T. P. Martens, P. J. Christos, M. J. Russo, G. M. Comas, F. H. Cheema, T. M. Naseem, R. Wang, K. A. Idrissi, S. H. Bailey, et al. Comparisons of infection complications between continuous flow and pulsatile flow left ventricular assist devices J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 841 - 842. [Full Text] [PDF] |
||||
![]() |
A. J. Clegg, D. A. Scott, E. Loveman, J. L. Colquitt, P. Royle, and J. Bryant Clinical and cost-effectiveness of left ventricular assist devices as a bridge to heart transplantation for people with end-stage heart failure: a systematic review and economic evaluation Eur. Heart J., December 2, 2006; 27(24): 2929 - 2938. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Klotz, J. Stypmann, H. Welp, C. Schmid, G. Drees, A. Rukosujew, and H. H. Scheld Does Continuous Flow Left Ventricular Assist Device Technology Have a Positive Impact on Outcome Pretransplant and Posttransplant? Ann. Thorac. Surg., November 1, 2006; 82(5): 1774 - 1778. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. C. Deng The challenges of generating evidence to guide mechanical circulatory support-based management of advanced heart failure Eur. Heart J., May 2, 2005; 26(10): 953 - 955. [Full Text] [PDF] |
||||
![]() |
S. Klotz and C. Schmid Arterial Pressure and Pump Flow Rate During Long-Term Pulsatile and Nonpulsatile Cardiac Support: Reply Ann. Thorac. Surg., March 1, 2005; 79(3): 1094 - 1094. [Full Text] [PDF] |
||||
![]() |
A. Undar and J. L. Myers Arterial Pressure and Pump Flow Rate During Long-Term Pulsatile and Nonpulsatile Cardiac Support Ann. Thorac. Surg., March 1, 2005; 79(3): 1093 - 1094. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |