Ann Thorac Surg 2003;75:780-785
© 2003 The Society of Thoracic Surgeons
a Department of Cardiothoracic and Vascular Surgery, Deutsches Herzzentrum Berlin, Berlin, Germany
b Michael E. DeBakey Department of Surgery, Division of Transplantation and Assist Devices, Baylor College of Medicine, Houston, Texas, USA
* Address reprint requests to Dr Drews, Deutsches Herzzentrum Berlin, Augustenburger Platz 1, 13353 Berlin, Germany.
e-mail: dr.drews{at}berlin.de
Presented at the Thirty-eighth Annual Meeting of The Society of Thoracic Surgeons, Fort Lauderdale, FL, Jan 2830, 2002.
| Abstract |
|---|
|
|
|---|
METHODS: We report our 10-year experience with 108 patients on MCS, who were supported for more than 3 months. Group A consisted of 38 patients (25 Novacor, 13 Berlin Heart) who underwent assist implantation from 1996 to 2001. They had a mean support time of 454 days (range 100 to 1074 days) and spent a mean of 326 days (range 20 to 769 days) at home. Group B consisted of 70 patients (24 Novacor, 46 Berlin Heart) who underwent assist implantation between 1991 and 2000. They had a mean support time of 234 days (range 95 to 795 days) and were not discharged. The patients were monitored for complications, hospital readmissions, and causes of death including infections and thromboembolic and bleeding events during the MCS time.
RESULTS: Group A total mortality was 16% (6/38). Two patients died from cerebral embolism, one from cerebral hemorrhage, two from systemic infection, and one from multiorgan failure. Thirty-two patients (84%) required 95 readmissions to the hospital due to cerebral embolism (n = 9), bleeding (n = 1), wound infections (n = 23), coagulation disorder (n = 13) for heart transplantation (n = 5), and (n = 44). In group B the mortality was 43% (30/70) for noncardiac reasons and thus significant higher (p = 0.004,
2 test). Causes of death were cerebral embolism (n = 5), cerebral hemorrhage (n = 7), systemic infection (n = 14, significantly higher, p = 0.04,
2 test), and multiorgan failure (n = 4).
CONCLUSIONS: Our experience demonstrates that MCS can be used in outpatients without increased mortality and with an acceptable rate of readmissions (2.8/patient). It ensures the survival of the patient, enables recovery from multiorgan failure, and offers an acceptable quality of life.
| Introduction |
|---|
|
|
|---|
Since 1996 patients who recovered from secondary organ dysfunction have left our hospital to spend the time following assist device implantation at home. Thanks to the reliability of the wearable electrical support systems patients were able to go to locations as far as 1000 km (675 miles) away from the hospital.
In the present study the course of 38 patients who were discharged from the hospital with MCS is described. The results were compared with those of a group of 70 patients who could not be discharged for psychologic, social and familial, or nonmedical reasons.
At the time of implantation all patients presented with severe heart failure and evolving multiorgan failure. They were dependent on maximal inotropic support and refractory to medical therapy. The MCS was initiated as a lifesaving procedure in all patients. The goal was mostly to bridge the patient to heart transplantation. However, some patients were later weaned from support or presented with contraindications to transplantation and MCS became permanent support.
The goal of this study is to report our experience with patients discharged from the hospital while on mechanical support. The incidence of thromboembolism, bleeding, infection, and lethal complications during the time out of hospital is documented.
| Patients and methods |
|---|
|
|
|---|
|
Patients with two different devices are described [3]. The Novacor is an assist device for left ventricular support only (Fig 1), implanted into a pocket in the abdominal wall. It consists of a polyurethane sack that is compressed by two electrically powered magnet. The pump is connected by a transcutaneous cable with an external controller and two batteries allowing independence for a maximum of 4 hours. The blood is drained from the apex of the left ventricle and brought back to the ascending aorta. The Novacor has been implanted since 1993 at our institution and patients have been able to leave our hospital with it since 1996. This system is used in patients presenting with severe heart failure with beginning multiorgan failure refractory to medical therapy.
|
|
All patients received anticoagulation consisting of heparin, Coumadin (Du Pont Pharmaceuticals, Wilmington, DE), and platelet inhibitors [5]. During the early postoperative period i.v. heparin was administered; then the patient was switched to Coumadin. To inhibit platelet aggregation acetyl salicylic acid in low doses and dipyridamol in high doses were given. The platelet function was measured by thrombelastogram and platelet activation tests.
Before leaving hospital the patient and the patients family were intensively trained (Table 2). They were involved in the technical function of the assist device and the patient and one family member were trained in driver exchange in case of malfunction [6]. The ventricular assist device coordinator taught the patient and his caregivers sterile wound dressing and practiced this with them until the dressing had been changed three times without complications. The local family practitioner was instructed about the device and performed routine follow-up exams once a week (laboratory check: coagulation, blood count, kidney function, C-reactive protein) or when needed. If the patient had no family member who could be trained, the family practitioner would be trained in sterile wound dressing. Patients routinely returned to the hospital for follow-up examination at intervals of between 2 weeks and 3 months. In addition, a 24-hour phone support line for medical and technical problems was introduced.
|
| Results |
|---|
|
|
|---|
|
|
|
2 test.
|
2 test) at 20% although the time of support was half as long (234 days). Multiorgan failure occurred in 6% of patients and did not differ significantly from the incidence in group A. | Comment |
|---|
|
|
|---|
The 38 outpatients in group A spent a mean of 326 days at home, corresponding to 72% of their time on support. The number of readmissions per patient per year was 2.8. The most frequent cause was non-MCS related (46%), followed by wound infections (24%), cerebral bleeding (9%), and embolism (1%). Five percent of the readmissions were in order to prepare the patient for heart transplantation. The cumulative events per outpatient month were 0.0024 for cerebral bleeding, 0.022 for cerebral embolism, 0.057 for wound infection, 0.031 for coagulation disorder, and 0.11 for non-MCS related causes.
Richenbacher and colleagues [11] have recently reported results from 13 discharged patients on left ventricular assist devices (HeartMate; Thoratec Corp., Pleasanton, CA). On average, they spent 171 days (range 3 to 473 days) out of hospital. Eight of them required a total of 20 rehospitalizations. This corresponds to 3.28 readmissions per patient per year. The most frequent cause was in 40% device related infections, followed by 35% non-MCS related, 10% neurologic events, 10% device malfunction, and 5% epistaxis. Morales and coworkers [12] presented results of 44 patients on HeartMate discharged home for an average of 103 days. The cumulative event rate per outpatient month was 0.02 for bleeding, 0.0068 for thromboembolism, 0.053 for device infection, and 0.02 for major malfunction. Although their figures for bleeding are almost the similar to ours, their patients had a higher incidence of pocket infections and major device malfunctions. On the other hand the rate of cerebral embolism and of readmissions for non-MCS related causes was higher in our patient population.
Although no patient in group A died at home, the mortality rate was 16% during the support time. None of the discharged patients reported by Morales and colleagues [12] died. Frazier and coworkers [13] reported in a multicenter evaluation that in a group of 280 patient with an electrically vented HeartMate system the mortality rate was 29%. Most patients died of infection (40%), followed by bleeding (11%), neurologic dysfunction (5%), and thromboembolic events (6%). Although the rate of bleeding was equal to that observed by us, Frazier saw more infections and a lower rate of neurologic and thromboembolic events.
The mortality rate due to systemic infection was significantly lower in our outpatient group (5%) than for the patients who remained hospitalized (20%). The cause may be the number of opportunistic infections acquired in hospitals. Frazier [13] presented similar results with a morbidity of 40% due to infections.
The lifesaving effect of mechanical circulatory support in critically ill patients has been shown by several authors [1, 2, 6, 1217]. In the Rematch trial [17] advantages using VADs could be shown. However, experience with discharging patients on VADs from the hospital is limited especially for those implanted with biventricular paracorporeal assist devices. With regard to using such devices as an alternative to transplantation, the observations made in discharging these patients on the transplant waiting list are of special importance. In particular, the issue of quality of life for these patients needs to be addressed [18, 19]. Although others have reported a considerable incidence of device-related technical problems in patients on certain LVADs, it is noteworthy that in our patient population using Novacor and Berlin Heart assist devices, not a single case of device malfunction occurred during the extended observation time.
The comparison of 38 patients discharged from the hospital while on MCS with 70 patients continously hospitalized demonstrates that discharging patients on left and biventricular support devices does not increase the risk of complications in particular the number of systemic infections. The mortality rate was remarkably reduced in patients who spent extended periods of time at home. A dedicated team of nurses, technicians, and physicians contributed to the success of the discharge program.
However, 16% of the outpatients died before they could receive a heart transplant. The reason for this lies not only in the medical and technical problems but also in the mismatch of the number of potential organ recipients and the number of hearts available, which makes a waiting period, and hence the mechanical systems for outpatients necessary. Unfortunately this mismatch is likely to increase in the future, making it a matter of urgency to decrease the potential risks of MCS by improving both the systems and the postimplantation care of these patients.
| Acknowledgments |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
DR LOEBE: No. The TCI patients were managed only with aspirin. But actually our TCI experience in Berlin, as you know, is somewhat limited (n = 24). Since these patients did not spend more than 90 days on the device, they did not qualify for the present report of more than 3-month support experience.
DR FRAZIER: Has the long-term patient that you had over there (568 days) been transplanted? Is there anything unusual about that patients anticoagulation regimen?
DR LOEBE: No. This patient is actually ongoing. One of the reasons why we were able to generate such a long-term experience is that until very recently in the German organ allocation system a patient on assist device could not qualify for higher urgency, even with device-related problems. Therefore, these patients had to wait for this very long time with a mean waiting time of 1 year. I think it is another remarkable part of our experience that none of these patients experienced any device malfunctions despite the more than 1-year median support duration.
DR FRAZIER: Weve introduced the same problem here. Have you discharged the MicroMed patients?
DR LOEBE: MicroMed is expecting to be FDA approved for hospital discharge soon, but so far patients on MicroMed DeBakey LVADs cannot be discharged in the USA. In Berlin/Germany, however, we have been able to discharge patients on MicroMed pumps home and they did very well.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. Tschirkov, D. Nikolov, and V. Papantchev New technique for implantation of the inflow canula of Berlin Heart INCOR system. Eur. J. Cardiothorac. Surg., October 1, 2006; 30(4): 678 - 679. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Loisance Mechanical circulatory support systems--future opportunities Eur. J. Cardiothorac. Surg., December 1, 2004; 26(Suppl_1): S48 - S53. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Jurmann, Y. Weng, T. Drews, M. Pasic, E. Hennig, and R. Hetzer Permanent mechanical circulatory support in patients of advanced age Eur. J. Cardiothorac. Surg., April 1, 2004; 25(4): 610 - 618. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.-M. McMahon, C. van Doorn, M. Burch, P. Whitmore, S. Neligan, P. Rees, R. Radley-Smith, A. Goldman, K. Brown, G. Cohen, et al. Improved early outcome for end-stage dilated cardiomyopathy in children J. Thorac. Cardiovasc. Surg., January 1, 2004; 126(6): 1781 - 1787. [Abstract] [Full Text] [PDF] |
||||
![]() |
O.H. Frazier and R. M. Delgado Mechanical Circulatory Support for Advanced Heart Failure: Where Does It Stand in 2003? Circulation, December 23, 2003; 108(25): 3064 - 3068. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |