|
|
||||||||
Ann Thorac Surg 2002;74:2132-2137
© 2002 The Society of Thoracic Surgeons
a Department of Anesthesiology, Wake Forest University School of Medicine, Winston-Salem, North Carolina, USA
b Department of Department of Perfusion, Cleveland Clinic, Cleveland, Ohio, USA
Accepted for publication July 22, 2002.
* Address reprint requests to Dr Stump, Department of Anesthesiology, Wake Forest University School of Medicine, Medical Center Blvd, Winston-Salem, NC 27157-1009, USA.
e-mail: dstump{at}wfubmc
| Abstract |
|---|
|
|
|---|
METHODS: Five identical embolus detectors monitored the propagation of entrained air through a cardiopulmonary bypass (CPB) model. The ability of the CPB circuit to remove gaseous microemboli was studied with vacuum-assisted venous drainage and gravity siphon venous drainage using different pump speeds and rates of gaseous microemboli delivery.
RESULTS: Under all conditions entrained venous air resulted in the detection of gaseous microemboli in the perfusate after the arterial filter. In blood-primed circuits, increased flow rates and higher levels of vacuum-assisted venous drainage were independently associated with increased gaseous microemboli counts in the arterial line. Vacuum-assisted venous drainage at -40 mm Hg did not significantly increase gaseous microemboli activity when compared with gravity siphon venous drainage at 4 L/min flow rate.
CONCLUSIONS: Vacuum-assisted venous drainage at -40 mm Hg does not statistically reduce the ability of the CPB circuit to remove gaseous microemboli at lower pump rates. High levels of vacuum and increased pump flow rates should be avoided. Air should not be introduced into the venous line.
| Introduction |
|---|
|
|
|---|
The etiology of the above injuries is varied and probably multifactorial [5]. There is substantive evidence, however, particularly with regard to brain injury [6] implicating emboli as a cause of organ damage during CPB. Emboli may be gaseous, liquid, or particulate and they may originate in the circulation or be introduced into the circulation. Gaseous microemboli (GME) may either be entrained into the CPB circuit [7] or they may be generated by the circuit components such as the oxygenator [8], venous reservoir, or roller occlusive pump [9]. Circuit design and CPB technique are directed toward eliminating the generation of GME. The risk of inadvertent air embolism will always be present and the circuit components should be able to significantly reduce and preferably stop any GME from reaching the patient.
With the advent of minimally invasive cardiac surgery and the progression of congenital heart surgery there is a desire to reduce the diameter of the venous cannula to improve surgical access, decrease pump priming volumes, and enable cannulation of vessels remote from the heart. Unfortunately, decreasing cannula diameter increases resistance and reduces the adequacy of gravity siphon venous drainage (GSVD). Vacuum-assisted venous drainage (VAVD) is a method of active venous drainage dependent upon maintaining a negative pressure in the venous reservoir to ensure adequate venous drainage through small diameter cannula and tubing. Although the advantages of VAVD have been detailed, it has been suggested that the technique decreases the ability of the CPB circuit to remove GME thereby subjecting the patient to increased numbers of emboli [10].
The present study uses five identical embolus detectors to simultaneously monitor the propagation of venous entrained air through circuit components in an experimental model of adult CPB. Using reproducible volumes of entrained air we studied the influence of VAVD and GSVD on the ability of the CPB circuit to remove GME. In addition the influence of variations in pump flow and rate of GME delivery on GME behavior were studied.
| Material and methods |
|---|
|
|
|---|
|
All circuits were flushed with carbon dioxide for 5 minutes before priming and deairing according to clinical standards. The circuits were primed with fresh, anticoagulated bovine blood (Sierra Medical, Santa Fe Springs, CA) corrected to a hematocrit of 24% ± 1% using isotonic saline.
Air introduction
A Luer lock connector was cut into the venous line midway between the venous cannula and the venous reservoir to enable the introduction of air. A 60 ml quantity of air was introduced either rapidly as a bolus to replicate a sudden catastrophic event or as a steady stream to replicate a nonocclusive atrial purse string or leaking connector. The volume of air delivered was standardized by displacement of a 60 ml column of air by water either introduced rapidly from a pressurized (300 mm Hg) intravenous giving set or slowly by a Harvard Apparatus syringe pump. The rapid or bolus introduction lasted 12 seconds and the slower or stream method lasted 120 seconds.
Experimental conditions
All circuits were maintained at 37°C. The venous reservoir level was maintained at 800 ± 25 mL through a variably occlusive Hoffman Clamp placed on the venous line.
A perfusate pO2 of 200 ± 50 mm Hg was achieved using 95% air and 5% CO2, adjusting gas flow rates to match pump flow rates. The arterial line pressure was measured from the top of the arterial filter and maintained at 200 ± 10 mm Hg using a second Hoffman Clamp located on the arterial line proximal to the aortic cannula. For VAVD experiments the level of suction was set through the vacuum controller, and the corresponding negative pressure was recorded through a transducer located on the top of the venous reservoir.
Embolus detection and validation
Embolic activity was monitored simultaneously at five locations using five identical embolus detection and classification (EDAC) systems. The five transducers were attached to modified Bentley Oxysat Optical Transmission Cells (Baxter Healthcare) cut into the tubing in the venous line, after the venous reservoir, after the roller pump, postoxygenator, and postarterial filter. Preliminary studies demonstrated this method of transducer attachment to be associated with the least variation in counts arising from variations in probe orientation.
The EDAC system was developed jointly by Orincon Corporation and Embolus Incorporated, funded by National Institutes of Health and National Medical Technology Testbed. Each EDAC system consists of a 5 mHz "active sonar" transducer connected to a Pentium personal computer operating on a Windows-based system. Signal analysis and processing enables rejection of motion artifacts, radio frequency interference, and other such causes of false alarms. The system is able to count GME as small as 40 micron. Figure 2, taken during validation work, demonstrates synchronized acoustic signal collection and image capture of GME using a camera interfaced with the EDAC system.
|
Conduct of study
Three identical circuits were perfused with blood (hematocrit of 24% ± 1%). During each experiment the circuit was studied using GSVD, VAVD (-40 mm Hg), and VAVD (-65 mm Hg). Each type of venous drainage was studied at 4 L/min and 6 L/min pump flow rates. After achievement of steady-state and for all conditions, 60 mL air was introduced into the venous line either as a bolus or as a stream. Emboli were counted over the next 2 minutes. Each trial was repeated three times and between trials a steady-state was reestablished. The combination of circuit, method of venous drainage, flow rate, and air delivery was determined by a randomization schedule.
An additional three identical circuits were primed with deionized water and the experimental sequence was repeated. Deionized water was used as a protein deficient, clear perfusate to enable comparison of results with other investigators who have published their studies of the behavior of GME using non blood primes. The clear prime also enabled visualization and recording of the flow characteristics of GME with a video camera.
Statistical analysis
Repeated measures analysis of variance was used to test the effects of venous drainage, flow rate, and air delivery on the emboli count in the arterial line. Residual analyses determined that log(x + 1) transformation of the emboli counts was necessary to normalize the data. Geometric means with their 95% confidence limits are reported as appropriate after log-transformation. Bonferroni corrections were made for multiple comparisons on post-hoc comparisons only when the corresponding repeated measures main effects or their interactions were not statistically significant. All analyses were performed using SAS version 8 (SAS Institute, Cary, NC) at an
of 0.05.
| Results |
|---|
|
|
|---|
The emboli counts for the blood filled circuits are presented in Table 1. The rate of air delivery did not influence arterial line GME counts, but the main effects for suction and flow rate were significant. Increased flow rates resulted in significantly more emboli during GSVD (p= 0.022) and VAVD -65 mm Hg (p= 0.010), but not during VAVD -40 mm Hg (p= 0.144). Similarly, increased suction (VAVD -65 mm Hg) was associated with increased arterial GME counts compared with GSVD (p= 0.011). The effect of GSVD versus VAVD -40 mm Hg was not statistically significant at 4 L/min flow rate (p= 0.095), but it did become significant at 6 L/min (p= 0.044).
|
|
|
|
| Comment |
|---|
|
|
|---|
The observation that the GME handling characteristics of the oxygenator was reduced in the blood circuits may be explained by the fact that bubbles develop a platelet, lipoprotein capsule when exposed to blood [11, 12]. The formation of the capsule increases the residence time of the bubble [13] and may reduce the ability of CPB components to remove GME.
In a similar in vitro study, Willcox and associates [10] also detected GME after the arterial filter after the entrainment of air into the venous line when using VAVD (-60 mm Hg) and GSVD. They reported a direct relationship between the volume of air entrained and the number of GME detected postarterial filter. When introducing air at a fixed rate they were unable to demonstrate a significant difference in arterial line GME counts between GSVD and VAVD. In contrast when they introduced 50 mL air at an unrestricted rate there was a 10-fold increase in arterial line GME counts when using VAVD compared with GSVD. We introduced the same volume of air in a controlled manner at different rates, and this did not significantly influence arterial line GME counts under any condition. This suggests either the unrestricted introduction of a volume of air changes the physical property of the resulting GME or the pressure dynamics of the circuit or the difference may be accounted for by variations in study design. Willcox and associates [10] ventilated the oxygenator but the circuits were studied at room temperature as compared with the present study where the circuits were maintained at 37°C. Colder solutions contain a greater number of molecules and therefore become saturated sooner than warmer solutions [14]. In addition, they studied salvaged, reused clinical CPB circuits consisting of a Medtronic (Minneapolis, MN) venous reservoir and oxygenator in combination with a Baxter (Irvine, CA) arterial filter.
In a recent study LaPietra and associates [15] used a nonblood prime to study the effects of VAVD and kinetically assisted venous drainage on entrained venous air. In agreement with the present study and the findings of Willcox and associates they demonstrated GME after the arterial filter after the introduction of air into the CPB venous line. Undetermined volumes of air were introduced into the venous line and the group reported that VAVD at increasing levels of suction (-15 mm Hg, -45 mm Hg, and -75 mm Hg) was associated with increasing GME counts in the arterial line.
The association between intraoperative emboli and brain injury during cardiac surgery has been established [2, 16, 17] but our inability to identify the type of embolus has made it difficult to study the precise role of GME in postoperative brain injury. The optimal method for counting, sizing, and identifying different types of embolic material remains controversial [18, 19]. Different techniques and equipment produce markedly different embolic counts [20]. The most common technique used is ultrasound. Increasingly, the frequency, amplitude, intensity, and spectral array of the reflected signal are being analyzed to make assumptions about embolic numbers, size, and material; however, most systems have never been validated. Deal and associates [21] investigated the reliability of pulsed and continuous wave Doppler to count emboli in both in vitro and in vivo models. They noted that detectors from the same and different systems placed in series on a tube did not count the same number of emboli. Detectors are easily "tricked" by artifactual noise arising from tapping the tubing or from bubbles trapped in connectors or bifurcations. Similarly the angle of orientation, coating of acoustic gel, and application and reapplication of the transducer will influence the GME count. Changing ultrasound settings such as gain and power between recordings will result in changes in sensitivity and specificity making it difficult to compare GME counts. In the present study, we used five identically configured EDAC systems simultaneously with purposefully made detector holders to reduce these problems. The detectors were not disturbed until each experiment was complete and all settings were maintained throughout the study.
A consistent finding between VAVD and GSVD studies is the presence of GME in the arterial line after the introduction of air into venous line independent of the method of venous drainage. Clinically air is frequently introduced into the venous line due to nonocclusive atrial purse-strings or caval snares, during blood sampling, injecting drugs [7], and at the initiation of CPB. Traditionally it has been assumed that entrained venous air will be removed by the CPB circuit components. All studies to date have demonstrated this assumption to be incorrect.
The results of this study suggest that VAVD at -40 mm Hg does not statistically reduce the ability of the CPB circuit components to remove GME at lower pump flow rates. High levels of vacuum and increased pump flow rates should be avoided. Under all conditions and with all methods of venous drainage, entrained venous air results in the detection of GME after the arterial filter. Air in the venous line should be avoided and if present it should be dealt with promptly.
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. S. Murphy, E. A. Hessel II, and R. C. Groom Optimal Perfusion During Cardiopulmonary Bypass: An Evidence-Based Approach Anesth. Analg., May 1, 2009; 108(5): 1394 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Lynch and J. Riley Microemboli detection on extracorporeal bypass circuitsa Perfusion, January 1, 2008; 23(1): 23 - 32. [Abstract] [PDF] |
||||
![]() |
J. W. Hammon Extracorporeal Circulation: Organ Damage Card. Surg. Adult, January 1, 2008; 3(2008): 389 - 414. [Full Text] |
||||
![]() |
B. Kiaii, D. Bainbridge, and P. Fernandes Surgical, Anesthetic, Perfusion-Related Advances in Minimal Access Surgery Seminars in Cardiothoracic and Vascular Anesthesia, December 1, 2007; 11(4): 282 - 287. [Abstract] [PDF] |
||||
![]() |
N. Colangelo, L. Torracca, E. Lapenna, S. Moriggia, G. Crescenzi, and O. Alfieri Vacuum-assisted venous drainage in extrathoracic cardiopulmonary bypass management during minimally invasive cardiac surgery Perfusion, November 1, 2006; 21(6): 361 - 365. [Abstract] [PDF] |
||||
![]() |
D. Maselli, R. Pizio, and F. Musumeci Multifrequency transcranial Doppler for intraoperative automatic detection and characterisation of cerebral microemboli during port-access mitral valve surgery Interactive CardioVascular and Thoracic Surgery, February 1, 2006; 5(1): 32 - 35. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Barak and Y. Katz Microbubbles: Pathophysiology and Clinical Implications Chest, October 1, 2005; 128(4): 2918 - 2932. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kurusz and B. D Butler Bubbles and bypass: an update Perfusion, January 1, 2004; 19(1_suppl): S49 - S55. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |