Ann Thorac Surg 2002;74:1500-1505
© 2002 The Society of Thoracic Surgeons
a Columbia University, New York Presbyterian Hospital, New York, New York, USA
b Cardiovascular Institute, University of Dresden, Dresden, Germany
c AFx, Inc, Fremont, California, USA
Accepted for publication June 26, 2002.
* Address reprint requests to Dr Williams, Surgical Arrhythmia Program, Columbia University-College of Physicians and Surgeons, MHB 7-126, 177 Ft. Washington Blvd, New York, NY 10032, USA.
e-mail: mw365{at}columbia.edu
| Abstract |
|---|
|
|
|---|
METHODS: Specific absorption rate (SAR) and thermal profiles were determined after microwave energy ablation intended for clinical application. Lesion geometry measures obtained in vitro with tissue equivalent media were compared with values collected from animal myocardium.
RESULTS: Thermal profile analysis demonstrated lesion penetration depths of 2.5, 3, and 5 mm after 8, 16, and 24 seconds, respectively, with microwave power application of 40 W. Dosimetric studies performed using animal myocardium corresponded to the thermal profile analysis and showed that lesion depth was controlled by the output power and the ablation time.
CONCLUSIONS: Lesion geometry in tissue using microwave energy is similar to that predicted from in vitro analysis. The ablation depths and thermal profile of microwave ablation is favorable for performing atrial ablation, and this is corroborated by favorable early clinical results.
| Introduction |
|---|
|
|
|---|
Alternative approaches, using energy ablation in lieu of incisions, have evolved to further simplify surgical treatment of AF and avoid the drawbacks associated with the Maze procedure [69]. Several groups have used radiofrequency intraoperatively for treating AF during mitral valve repair or replacement [1013] and in patients with lone AF [11], whereas others have used microwave energy [14, 15] or cryotherapy [16, 17]. Energy ablation methods, which are easier and faster to perform than the "cut and sew" Maze procedure, have demonstrated promising success rates, and may pose less risk of peri- and posttherapy complications.
Despite recent use of newer energy modalities in the clinical setting, there has been very little "basic science" performed to understand the behavior and characteristics of clinical energy ablation. Lack of this basic knowledge may alter success or increase the morbidity with these procedures.
Microwave energy is only recently available in the US, and may offer several advantages. In general, these include the ability to produce long narrow lesions without gaps, high reproducibility in ablation depth and length, capacity to produce a flexible ablating probe, and capability for creating transmural lesions, with no surface sticking or charring. The purpose of this study was to characterize dynamics of lesion formation with microwave energy ablation developed for clinical use.
| Drs Williams and Bérubé disclose that they have a financial relationship with AFx, Inc.
|
| Material and methods |
|---|
|
|
|---|
|
) and the square of the electric field (E) emitted in the medium (SAR =
· E2/2). It is also proportional to the initial slope of the temperature (T) increase (SAR
T/
t).
The tissue-equivalent medium (liquid phantom) was made up of deionized water (67%) and glycerol (33%) [18]. For this type of analysis, the dielectric properties of this mixture have been shown to be similar to the myocardium at 2.45 GHz (
* = 57 - 31 · j) [18, 19]. During measurement of SAR intensity, the ablating probe was completely immersed into the liquid phantom mixture. A miniature SAR measurement system, consisting of three orthogonal dipoles and an electronic amplifier (Model 8021B/C and Amplifier 8010 [Narda, Hauppauge, NY]) was used to determine SAR patterns. SAR was measured in 2,000 locations on a 4 x 1-cm plane facing the ablating tip, where data were electronically collected with a regular data acquisition system.
Thermal profile measurements
Thermal profiles show propagation of the thermal front through tissue introduced by the ablation probe. Thermal profiles were used to examine consistency and depth of the temperature increase over time along the length of the microwave energy antenna. Thermal profile measures were performed using a tissue-equivalent medium (solid phantom) that had dielectric and thermal properties approximating those of cardiac tissue [20]. It consisted of water (75.48%), NaCl (1.05%), polyethylene powder (15.01%), and TX-151 (8.46%), a solidifying agent. During application of microwave energy (40 W from the generator), the ablating tip was positioned in direct contact with the solid phantom surface.
A fiberoptic thermometry system (Luxtron Corporation, Mountain View, CA) with four 0.5-mm diameter fiberoptic probes was used to record internal temperature of the solid phantom. The solid phantom was placed in a plastic container drilled with an array of 5 x 2-mm access holes to ensure accurate positioning of the fiberoptic probes. Temperature measurements were simultaneously recorded at all four locations in a series of successive trials to collect data from the entire array. Time for each ablation was set at 2 minutes. The solid phantom was allowed to cool for 5 minutes between successive trials.
Temperature values used to create thermal profiles were corrected from measured values to reflect the baseline phantom tissue temperature of 30°C. This adjustment was made in order to simulate thermal conditions of open heart surgery, where myocardial temperature ranges between 8°C and 20°C during cold cardioplegic arrest. Use of this model assumes that thermal conductivity of the solid phantom is constant within the temperature range applied for this study. The 50°C isotherm defined lesion size, as indicated by Haines and colleagues [21], who reported that the temperature at the margin between viable and nonviable tissue is approximately 48°C. The software Matlab (MathWorks, Natick, MA) was used to plot the thermal profiles.
In vivo analyses
Dosimetric study: power dependency
Results obtained from the solid phantom model were validated using freshly excised bovine hearts (n = 11). Two lesions were made on the left ventricle of each heart specimen. Each lesion was produced with the generator set at 40 or 60 W of power, applied for 25 seconds. After completing the ablation procedure, tissue specimens were stained with the vital dye 2,3,5-triphenyltetrazolium chloride (TTC) to identify the viable and nonviable tissue. Depth and width of ablation were measured at the midpoint of each lesion.
Dosimetric study: time dependency
Effect of ablation time on microwave energy penetration depth was also evaluated. Three ablations were performed on the left ventricle of each of four heart specimens. Output power from the generator was set at 40 W, with ablation time of either 10 or 40 seconds. Lesion geometry was measured as it was for the power dependency study. Results obtained from the power dependency portion of the study for microwave power application of 40 W for 25 seconds completed the performance curve analysis for time dependency.
| Results |
|---|
|
|
|---|
|
|
|
In vivo analyses
Dosimetric study: power dependency
With microwave application of 40 W over 25 seconds, average achieved penetration depth was 6.20 ± 0.92 mm and mean lesion width was 4.93 ± 0.98 mm. Mean penetration depth obtained with this energy application corresponded to results shown by thermal profile analysis (Fig 3), where after 25 seconds, temperature at 6 mm depth reached 50°C, which is sufficient for tissue ablation. Ablation width was significantly narrower than ablation depth (p = 0.003). With 60 W of power, average achieved penetration depth was 7.01 ± 0.83 mm and lesion width was 6.95 ± 1.49 mm. Depth and width did not differ statistically (p = 0.866). Penetration depth observed with application of 60 W power for 25 seconds was significantly deeper than depth achieved with 40 W for the same ablation time (p < 0.000). Lesion width also differed significantly between the two power settings (p < 0.000).
Dosimetric study: time dependency
Association between ablation time for a given power and lesion geometry is shown in Figure 5.
With power application of 40 W, average lesion depths were 2.05 ± 0.89 mm, 6.20 ± 0.92 mm, and 7.12 ± 1.12 mm for ablation times of 10, 25, and 40 seconds, respectively. Penetration depth differed significantly between ablation times (analysis of variance, p < 0.000); a post hoc test (Bonferroni) revealed that ablation depth differed significantly between ablation times of 10 and 25 seconds (p < 0.000) and 10 and 40 seconds (p < 0.000), but not between 25 and 40 seconds (p = 0.226). The association between ablation time and penetration depth was curvilinear, as indicated by thermal profile temperature elevation plots.
|
|
| Comment |
|---|
|
|
|---|
In vitro SAR analysis demonstrated good uniformity in distribution of the electromagnetic energy through the tissue and excellent penetration depth. Energy distribution was maximal near the center of the ablating element, indicating that depth of ablation was relatively deeper at the midpoint of a lesion. There was no indication of edge effect along the ablating tip, which could potentially produce overheating of the tissue surface and induce charring.
Findings from thermal profile analyses, using a solid phantom cardiac tissue model, corresponded with SAR results by showing continuous and even distribution of energy, with no areas of discontinuity over the length of the ablating probe. Temperature elevation plotted over time demonstrated smooth temperature increase in tissue cross sections, whereas temperature at the tissue surface remained below 100°C over the time required to produce a 6-mm-deep lesion. Because temperature remained below this critical value, charring and carbonization, as well as cavitation and wall disruption, should not occur. Indeed, neither of these problems occurred in any of the scenarios studied here. This is a critical finding because the ability to raise the tissue temperature to 50°C while maintaining one less than 100°C is paramount to both effective and safe hyperthermic ablation.
The results obtained on myocardial tissues demonstrated that ablation depth can be predictably controlled by modifying microwave energy power or ablation time. Ablation depth obtained after 25 seconds was significantly greater than depth achieved after 10 seconds; however, ablation depths achieved with 25 and 40 seconds did not differ significantly. This phenomenon can be explained by the fact that with progressively increased time of microwave energy application, a quasi-static state is reached where lesion growth essentially plateaus. The temperature elevation plot collected in vitro (Fig 4) corroborated observations in myocardial tissue. This was also an encouraging finding in that successful tissue ablation can be carried out in only 25 seconds, as opposed to the 2 minutes required with cryoablation.
Agreement between in vitro and in vivo studies provides validity for predicting microwave lesion formation in clinical use. Depth of ablation, assessed by measuring the zone of nonviable tissue in vivo, was the same as that shown by thermal profile analysis conducted in vitro for the same given microwave application. In vivo analyses demonstrated an average ablation depth of 2.05 ± 0.89 mm after 10 seconds and 6.20 ± 0.92 mm after 24 seconds, whereas in vitro analysis predicted an ablation depth of 2 to 3 mm after 8 seconds and 4 to 6 mm after 24 seconds of ablation. This reliable and consistent ablation depth plays an important role in clinical ablation, because as with other sources, there is no feedback mechanism to determine when a lesion has been successfully created. In the absence of a feedback mechanism, the reliable ablation depth demonstrated with microwave is important to assure safety and reproducibility in the clinical setting.
The microwave ablation probe used in this study was capable of producing narrow and deep lesions in the myocardium. The lesions were continuous with no gaps along the length of the antenna. Because of the metallic reflector built into the device, energy emission is unidirectional, which is important for ensuring that ablations can be performed safely. This study showed that penetration depth was predictably controlled by modifying microwave energy power and ablation time. At the suggested setting for this ablating probe (FLEX 2), which is 40 W of output power from the generator and 25 seconds of ablation time, geometry of an average lesion was approximately 6 mm deep and 5 mm wide. For the more commonly used FLEX 4, this power setting corresponds to 60 W for 45 seconds in the arrested heart. Because of good penetration of the microwave energy into the tissue, there was no evident tissue surface carbonization with ablation. Preliminary clinical experience has demonstrated that microwave energy using a system with this output design can be applied safely and successfully in cardiac tissue ablation [14].
Extrapolation between the models presented herein and clinical application requires addressing two potential limitations of this study. For the in vivo portion of the study, although our intention was to model atrial ablation, ventricular myocardial tissue was selected as the modus operandi in place of atrial tissue. Because the principal objective of the study was to characterize dosimetry of clinical application of microwave energy, it was essential to have adequate myocardial thickness in order to do so. Average atrial thickness in the normal human myocardium is 4 mm, significantly less than average achieved lesion depths reported here for selected power and time applications. It must be noted, however, that in patients with severe mitral valve disease (a large number of which could potentially receive cardiac ablation treatment), left atrial wall tissue can be fibrotic and significantly thicker (eg, 7 to 8 mm) than in persons with no disease. Because of the importance of ensuring lesion transmurality for efficacy of treatment, it was vital to define dosimetry over a clinically relevant range of energy applications; use of ventricular myocardial tissue allowed this type of study.
Second, our in vivo experimental model assumes that energy diffusion and temperature gradients are similar to those existing in the fully perfused in vivo clinical model. The results of this study are effectively only valid under arrested heart conditions and cannot be extended to beating heart surgery. In beating heart surgery, the circulating blood pool within the atria acts as a heat-sink, where significant thermal energy can be lost through convective heat transfer. McRury and colleagues showed that energy losses to micro-circulation, however, are not significant, and that difference between perfused and nonperfused tissue in terms of intramyocardial thermal gradients are minimal [23]. Their observations also suggest that the freshly excised myocardial model provides a good representation of the clinical arrested heart environment.
The results of these laboratory studies demonstrate the efficiency of microwave energy for atrial ablation. While no laboratory work can reliably predict clinical outcome, this study would suggest that microwave energy is a safe and effective energy source that can quickly produce transmural atrial lesions. Microwave energy ablation, in addition to the characterization of ideal patient populations and lesion sets, should play a key role as we progress to a more minimally invasive cure of AF. Initial work is promising, and further characterization of microwave energy ablation from the epicardium of the beating heart will allow for optimization of this energy source.
| Acknowledgments |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. T. McElderry, D. C. McGiffin, V. J. Plumb, K. Nanthakumar, A. E. Epstein, T. Yamada, and G. N. Kay Proarrhythmic Aspects of Atrial Fibrillation Surgery: Mechanisms of Postoperative Macroreentrant Tachycardias Circulation, January 15, 2008; 117(2): 155 - 162. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
J. R. Doty, D. B. Doty, K. W. Jones, J. H. Flores, M. Mensah, B. B. Reid, S. E. Clayson, G. Snow, E. Righter, and R. C. Millar Comparison of standard Maze III and radiofrequency Maze operations for treatment of atrial fibrillation J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 1037 - 1044. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Hashim, P. Tewari, and I. Birdi Wire perforation of Flex 10 microwave device during thoracoscopic atrial fibrillation ablation Interactive CardioVascular and Thoracic Surgery, December 1, 2006; 5(6): 744 - 745. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Lecoq, J. M. Gracia, C. Sureda, and A. Igual Subxyphoid approach for closed-chest atrial fibrillation surgery: the one hand operation Interactive CardioVascular and Thoracic Surgery, December 1, 2006; 5(6): 669 - 671. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Melby, A. Zierer, S. P. Kaiser, R. B. Schuessler, and R. J. Damiano Jr Epicardial microwave ablation on the beating heart for atrial fibrillation: The dependency of lesion depth on cardiac output. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 355 - 360. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Jahangiri, G. Weir, K. Mandal, I. Savelieva, and J. Camm Current strategies in the management of atrial fibrillation. Ann. Thorac. Surg., July 1, 2006; 82(1): 357 - 364. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Turek, L. R. Dibernardo, A. J. Lodge, S. S. Lin, R. D. Davis, C. A. Milano, and S. A. Simsir Histopathology and transmurality of acute microwave lesions on the beating human atrium Interactive CardioVascular and Thoracic Surgery, June 1, 2006; 5(3): 202 - 206. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. L. Gaynor, G. D. Byrd, M. D. Diodato, Y. Ishii, A. M. Lee, S. M. Prasad, J. Gopal, R. B. Schuessler, and R. J. Damiano Jr Microwave Ablation for Atrial Fibrillation: Dose-Response Curves in the Cardioplegia-Arrested and Beating Heart Ann. Thorac. Surg., January 1, 2006; 81(1): 72 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bisleri and C. Muneretto Innovative Monolateral Approach for Closed-Chest Atrial Fibrillation Surgery Ann. Thorac. Surg., November 1, 2005; 80(5): e22 - e25. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. C. Reade, J. O. Johnson, G. Bolotin, W. L. Freund Jr, N. L. Jenkins, C. E. Bower, S. Masroor, A. P. Kypson, L. W. Nifong, and W. R. Chitwood Jr Combining Robotic Mitral Valve Repair and Microwave Atrial Fibrillation Ablation: Techniques and Initial Results Ann. Thorac. Surg., February 1, 2005; 79(2): 480 - 484. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bisleri, A. Manzato, M. Argenziano, D. W. Vigilance, and C. Muneretto Thoracoscopic epicardial pulmonary vein ablation for lone paroxysmal atrial fibrillation Europace, January 1, 2005; 7(2): 145 - 148. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Wisser, C. Khazen, E. Deviatko, G. Stix, T. Binder, R. Seitelberger, H. Schmidinger, and E. Wolner Microwave and radiofrequency ablation yield similar success rates for treatment of chronic atrial fibrillation Eur. J. Cardiothorac. Surg., June 1, 2004; 25(6): 1011 - 1017. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. T Sie, W. P Beukema, A. Elvan, and A. R Ramdat Misier New strategies in the surgical treatment of atrial fibrillation Cardiovasc Res, June 1, 2003; 58(3): 501 - 509. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Manasse, P.G. Colombo, A. Barbone, P. Braidotti, G. Bulfamante, M. Roincalli, and R. Gallotti Clinical histopathology and ultrastructural analysis of myocardium following microwave energy ablation Eur. J. Cardiothorac. Surg., April 1, 2003; 23(4): 573 - 577. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||