Ann Thorac Surg 2002;74:S1390-S1394
© 2002 The Society of Thoracic Surgeons
a St. Elisabeth Heart Center and Institute of Physiology and Biomedical Engineering, Norwegian University of Science and Technology, Trondheim, Norway
* Address reprint requests to Dr Haaverstad, St. Elisabeth Heart Center, University Hospital of Trondheim, Trondheim N-7018 Norway.
e-mail: rune.haaverstad{at}medisin.ntnu.no
Presented at the Eighth Annual Cardiothoracic Techniques and Technologies Meeting 2002, Miami Beach, FL, Jan 2326, 2002.
| Abstract |
|---|
|
|
|---|
METHODS: Twenty-four LIMA-to-LAD anastomoses were evaluated with real-time epicardial ultrasound imaging using an ultrasound transducer positioned between the paddles of the stabilizer during off-pump procedures. The length of the anastomosis (DA), diameters of LIMA (DM), LAD at the toe of the anastomosis (D1), and 5 mm distally to the anastomosis (D2) were measured, and the ratios between these variables were calculated. The flow velocity through the anastomoses was visualized by color Doppler coding, and flow was assessed with transit-time flowmetry.
RESULTS: The epicardial color Doppler ultrasound allowed accurate assessment of the anastomoses. Twenty-three (96%) of the primary anastomoses were confirmed as patent. Mean ratios of D1/D2, DA/D2, and DM/D2 were 0.89 ± 0.13, 3.01 ± 1.04 and 1.32 ± 0.32, respectively. One anastomosis had a stenosis more than 50% detected by color Doppler ultrasound. After surgical revision, transit-time flow increased from 22 to 40 mL/min.
CONCLUSIONS: Intraoperative color Doppler ultrasound allowed adequate imaging for quality assessment of LIMA-to-LAD anastomoses performed on the beating heart. One anastomosis was revised due to a technical error detected by epicardial color Doppler imaging. Epicardial ultrasound scanning is a valuable tool for intraoperative assessment of LIMA-to-LAD anastomoses during off-pump coronary surgery.
| Introduction |
|---|
|
|
|---|
Angiography has until now been considered as the "golden standard" for assessment of anastomotic quality and several authors have recommended the routine use of early postoperative angiography [3]. Yet angiography is a time-consuming procedure not available in most operating rooms. The only widely recognized method of assessing graft function intraoperatively is transit-time flowmetry [4]. However, this method is of limited value because the observed flow is influenced by several hemodynamic variables such as blood pressure and vascular resistance.
Over the last 15 years several echocardiographic devices for visualization of coronary arteries and graft anastomoses have been tested in experimental models and intraoperatively in humans [57]. Unfortunately none of these devices have gained wide acceptance mainly because of their many technical limitations. We have previously reported our initial experience with a new linear array 10-MHz transducer specially designed for epicardial vascular ultrasound scanning during on-pump CABG [8], demonstrating its feasibility to detect and classify coronary artery stenoses. The severity score correlated significantly with the grading obtained from preoperative angiograms. Furthermore, ultrasound imaging allowed evaluation of the morphology of the anastomoses and a semiquantitative assessment of the flow velocity at the anastomotic site [8].
Based on our previous positive experience with on-pump CABG, we decided to test the device during OPCAB procedures. The present study was designed to establish a valid method of intraoperative quality-assessment of left internal mammary artery (LIMA) to left anterior descending artery (LAD) anastomoses with epicardial color Doppler ultrasound during off-pump surgery.
| Material and methods |
|---|
|
|
|---|
Surgical technique
All operations were performed by the same surgeon (R.H.). Median sternotomy was used in all cases. After full heparinization (3 mg/kg), the LIMA was harvested with its pedicle and a diluted solution of papaverine was injected intraluminally. The activated clotting time was not allowed to drift below 270 seconds. The LAD was identified and snared with a 4-0 pledgeted polypropylene suture (Prolene; Ethicon, Somerville, NJ) proximally to the incision. After 3 to 5 minutes of ischemic preconditioning, the snare was released and an epicardial stabilizer was used to immobilize the target site chosen for grafting. After the incision of LAD, an intracoronary shunt (CardioThoracic Systems, Inc, Cupertino, CA) was positioned into the vessel lumen and the coronary anastomosis was performed using a continuous 7-0 or 8-0 (Prolene) suture. The LIMA pedicle was secured with two epicardial stitches using 7-0 polypropylene on both sides. After assessment of the graft, heparin was reversed with protamine. The stabilizer was then removed.
Epicardial ultrasound scanning
After completing the LIMA-to-LAD anastomosis, with the stabilizer still in place, epicardial color Doppler scanning of the anastomosis was performed by means of a linear array 10 MHz GE Vingmed transducer (foot-print 27.3 x 9.6 mm) connected to a GE Vingmed System FiVe echocardiography unit (GE Vingmed, Horten, Norway) [8]. With sterile gel as conduction medium, the sterilized transducer was applied directly onto the epicardium between the paddles of the stabilizer (Figure 1).
Real-time ultrasound images and storage of data for later analysis were obtained within approximately 10 minutes in each patient. The ultrasound images were stored as digital data for later analysis using the Echo PAC software (GE Vingmed).
|
|
Statistical analysis
Data normally distributed were described as arithmetic mean and standard deviation (SD) was the measure of variability. Skewness test was used to test the distribution of the data. Skewed data are presented as median and range. A two-sampled Students t test was used to evaluate differences in normally distributed continuous data. Skewed distributed continuous data and categorical data were analyzed with the nonparametric Mann-Whitney test. Linear regression was used to evaluate the correlation between LAD flow rate and the D1/D2 ratio. A p value of less than 0.05 was considered statistically significant. Statistical analyses were performed with the NCSS program (Number Cruncher Statistical System, Kaysville, UT).
| Results |
|---|
|
|
|---|
Twenty-three (96%) anastomoses were fully patent according to color Doppler imaging (Figure 3), with median transit-time flow rate 29 mL/min and a range from 7 to 72 mL/min. Measurements of D1, D2, DA, and DM, as well as the ratios D1/D2, DA/D2, and DM/D2 are presented in Table 1. No statistically significant difference was observed between D1 and D2 in the entire group. All other dimensions and ratios were significantly different from each other (Table 1). No significant correlation was observed between LAD flow and the D1/D2 ratio (r2 = 0.05, p = 0.1). Nine anastomoses (38%) with D1/D2 less than 0.9 (mean 0.79 ± 0.14) had a minor intraluminal echo-dense protrusion observed at the toe.
|
|
| Comment |
|---|
|
|
|---|
A patent LIMA-to-LAD graft is the single most important determinant of long-term and event-free survival and imperfect graft anastomoses have a poor long-term patency after CABG [9, 10]. Epicardial color Doppler scanning allows an accurate assessment of the LIMA-to-LAD anastomoses and provides three important pieces of information: visualization of the anastomosis and its components (LAD and LIMA); measurement of the length of the anastomosis proper as well as the diameters of the LAD downstream and the LIMA; and color Doppler assessment of blood flow velocity at the anastomotic site. The necessary measurements were obtained easily without risk of any complications, are not time consuming, and provide important information about the geometry and the flow pattern of the anastomosis.
Minor intraluminal protrusions at the toe were observed in nine of the anastomoses with a D1/D2 ratio of less than 0.9. Incongruence or a slight bulging at the edges of the vessel walls where the graft and the LAD are joined together may explain these findings. However, no major changes in blood velocity were observed at these sites. Anastomoses may develop temporary edema, hematoma, and spasm, and some angiographic follow-up studies have suggested that some FitzGibbon B lesions seen with immediate or early postoperative angiography have spontaneously resolved at later coronary angiography [9, 11]. Transit-time flowmetry has been widely used to evaluate the quality of distal anastomoses after CABG. However, the lowest acceptable flow value in the LIMA graft has not been not defined and graft revision has been advised when flow ranges from 5 to 20 mL/min [4, 12]. This wide range indicates that the mean graft flow alone is not a reliable measurement for graft assessment, unless the stenosis is more than 75% to 90% [13, 14]. The only anastomosis with a marked stenosis observed in the present study was detected by epicardial imaging in a patient with a LIMA transit-time flow rate of 22 mL/min. During surgical revision, an intimal flap at the toe of the anastomosis was detected. After revision LIMA flow increased to 40 mL/min. Intimal flap is a well-known technical failure of the anastomosis, along with thrombosis, stenosis at the heel, coronary dissection, LIMA dissection, and vessel stenosis distal to the graft [4, 15, 16]. The epicardial color Doppler scanning technique used in the present study may be considered the only method able to identify all these problems, and may accordingly offer an important tool for detecting anastomotic errors leading to early graft failures.
Concern has been raised regarding the use of temporary snares and shunts because of the potential risk of endothelium alteration [17]. We were not able to detect any LAD lesion caused by snares or intraluminal shunts with intraoperative epicardial color Doppler scanning. However, our patient cohort will undergo angiographic follow-up to evaluate whether stenosis at these sites develop over time.
Previous experimental and clinical studies have demonstrated that the assessment of the technical adequacy of bypass grafts may be enhanced by the use of epicardial echocardiography [57, 18, 19]. The main reason why epicardial echocardiography has not gained more popularity is the large size of the ultrasonic probe, allowing only the visualization of LAD, diagonals, and proximal right coronary artery [57, 18, 19]. Coronary branches located laterally or at the back of the heart could be scanned only during CPB because the dislocation of the heart caused hemodynamic instability. When performing OPCAB through a median sternotomy, the heart is exposed with pericardial stitches and modern stabilizers, allowing epicardial scanning of coronary vessels also on the inferior and lateral walls. In the present study we emphasized quality assessment of LIMA-to-LAD anastomosis. However, the design of the transducer allows epicardial imaging of all areas, and we plan to use it for assessing all distal anastomoses with the stabilizer still in place in future studies.
The early-generation transducers were able to provide only tissue images. In this study the color Doppler modality was obtained as a semiquantitative measure. Although color Doppler imaging may be given as an absolute velocity with the device used in our study, this measurement is still not reliable intraoperatively in a clinical setting because of difficulties in adjusting for the angle created between the transducer and the blood flow within the vessels. The latter problem points out that the technology of intraoperative color Doppler imaging in off-pump coronary surgery still needs further development, and the prognostic value of graft patency has yet to be evaluated by follow-up angiography. The introduction of this new transducer for epicardial imaging of small vessels with color Doppler features may enhance the widespread use of epicardial scanning in coronary artery surgery.
| Conclusions |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Shimamoto, A. Marui, T. Nishina, and M. Komeda To-and-fro pattern: an indication of flow competition or a sign of incompetent coronary anastomosis? Interactive CardioVascular and Thoracic Surgery, January 1, 2009; 8(1): 155 - 156. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. S Ibrahim, L. Lovstakken, I. Kirkeby-Garstad, H. Torp, H. Vik-Mo, and R. Haaverstad Effect of the Cardiac Cycle on the Coronary Anastomosis Assessed by Ultrasound Asian Cardiovasc Thorac Ann, April 1, 2007; 15(2): 86 - 90. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Di Giammarco, M. Pano, S. Cirmeni, P. Pelini, G. Vitolla, and M. Di Mauro Predictive value of intraoperative transit-time flow measurement for short-term graft patency in coronary surgery. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 468 - 474. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, P. F. A. Bakker, R. Meijer, C. Borst, and P. F. Grundeman Ultrasound Mini-Transducer with Malleable Handle for Coronary Artery Surgery Ann. Thorac. Surg., January 1, 2006; 81(1): 322 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, W. J.L. Suyker, P. T.W. Suyker, C. W.J. Verlaan, R. Meijer, C. Borst, and P. F. Grundeman Quality assessment of distal S2AS connector anastomosis by 13MHz epicardial ultrasound Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 833 - 837. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Jones, T. Athanasiou, P. P. Tekkis, V. Malinovski, S. Purkayastha, A. Haq, J. Kokotsakis, and A. Darzi Does Doppler echography have a diagnostic role in patency assessment of internal thoracic artery grafts? Eur. J. Cardiothorac. Surg., November 1, 2005; 28(5): 692 - 700. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, T. C. Dessing, R. Meijer, P. F.A. Bakker, C. Borst, and P. F. Grundeman Robot-assisted 13 MHz epicardial ultrasound for endoscopic quality assessment of coronary anastomoses Interactive CardioVascular and Thoracic Surgery, December 1, 2004; 3(4): 616 - 620. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. K. Hol, P. S. Lingaas, R. Lundblad, K. A. Rein, K. Vatne, H.-J. Smith, S. Nitter-Hauge, and E. Fosse Intraoperative angiography leads to graft revision in coronary artery bypass surgery Ann. Thorac. Surg., August 1, 2004; 78(2): 502 - 505. [Abstract] [Full Text] [PDF] |
||||
![]() |