|
|
||||||||
Ann Thorac Surg 2001;72:S999-S1003
© 2001 The Society of Thoracic Surgeons
a Department of Cardiovascular Surgery, University of Lausanne, Lausanne, Switzerland
b Department of Pathology, University of Lausanne, Lausanne, Switzerland
c Department of Cardiac Surgery, Lund University Hospital, Lund, Sweden
Address reprint requests to Dr Tozzi, Service de Chirurgie CardioVasculaireBH10, Centre Hospitalier Universitaire Vaudois, Rue du Bugnon, 46 1011 Lausanne, Switzerland
e-mail: tozzig{at}hotmail.com
Presented at the Seventh Annual Cardiothoracic Techniques and Technologies Meeting 2001, New Orleans, LA, Jan 2427, 2001.
| Abstract |
|---|
|
|
|---|
Methods. 10 sheep, 45 to 55 kg, underwent off-pump coronary artery bypass grafting (right internal mammary artery to left anterior descending artery). In 5 animals, the anastomosis was performed with a GraftConnector and in 5 animals with 7-0 running suture. Intraoperative fluoroscopy and a fluoroscopic control at 6 months were performed. After 6 months, the animals were sacrificed and the anastomoses were examined histologically.
Results. All animals survived at 6 months with 100% anastomosis patency rates in both groups. In the GraftConnector group, the anastomosis diameter at 6 months fluoroscopy was 118% of native left anterior descending artery versus 97% of the control group. Luminal anastomotic width at histology was 1.7 ± 0.2 mm in the device group versus 1.6 ± 0.1 mm in the control group. Mean intimal hyperplasia thickness was 0.21 ± 0.1 mm in the device group versus 0.01 mm in the control group.
Conclusions. The GraftConnector provides a consistent and reproducible coronary artery anastomosis and reduces technical demand and manual dexterity in coronary operations. Long-term results demonstrate that off-pump coronary artery bypass grafting performed with the GraftConnector had the same patency rate and luminal width as those performed with running suture.
| Introduction |
|---|
|
|
|---|
Through modern stent technology, Jomed International AB (Helsingborg, Sweden) has recently developed the GraftConnector (GC), which allows sutureless anastomoses between blood vessels. The GC reduces the time to perform an end-to-side anastomosis and facilitates a consistent and reproducible sutureless coronary artery anastomosis for minimally invasive and beating heart operations as already demonstrated in an animal model [2]. The aim of the present study was to evaluate histology and luminal width of end-to-side coronary arteries anastomoses performed in a sheep model with GC compared with standard running suture technique. These animals are part of a feasibility study and the present results concern sheep that were kept alive for 6 months after device implantation.
| Material and methods |
|---|
|
|
|---|
|
The left anterior descending artery (LAD) was then explored and a 4-0 polypropylene suture was placed around the proximal LAD for snaring it. In 5 animals, the end-to-side anastomosis between RIMA and LAD was performed with the device (GC group), whereas in another 5 animals the procedure was performed using the running suture (7-0 polypropylene) technique (control group). No stabilizer was used in both groups. The proximal LAD was ligated and the blood flow in the RIMA was measured by a transit time flow meter. The RIMA was cannulated and injected with contrast medium (Omnipaque, Nycomed) to acquire fluoroscopic images of the anastomosis.
Animals received ticlopidine 250 mg/day for 1 month and aspirin 100 mg/day for 6 months. At 1 and 6 months, under general anesthesia, the animals underwent fluoroscopy control through a direct cannulation of the RIMA at its origin from the subclavian artery behind the sternocleido muscle. All animals received human care in compliance with the European Convention on Animal Care, and our ethics committee approved the study. After 6 months all animals were sacrificed and anastomoses were carefully harvested.
Specimen preparation
Anastomoses carried out with the device were fixed in formaldehyde 10%, preserved in a special acrylic resin, and cut with a diamond-coated wire saw developed for cutting tissue samples with metallic implants in situ. Slice thickness varied between 60 and 150 µm. Anastomoses with the running suture technique were fixed in formaldehyde 10%, preserved in paraffin, and cut with a standard microtome. All specimens were stained with hematoxylin-eosin and elastine or toluidine blue. Specimens were cut according to the section plans showed in Figure 2. We prepared up to 15 slices for each specimen.
|
![]() |
|
![]() |
|
|
| Results |
|---|
|
|
|---|
|
|
| Comment |
|---|
|
|
|---|
Fluoroscopic and histologic data showed that all anastomoses examined were patent after 6 months. As fluoroscopic images demonstrated, the GC diameter was larger than native coronary diameter at the implantation (Table 1): anastomoses had a mean diameter of 136% of the distal LAD. After 6 months, fluoroscopic control demonstrated that mean minimal anastomotic diameter was 118% of the distal LAD. Blood flow in the RIMA after closure of the proximal LAD was 33 ± 3.8 mL/min in the GC group and 23.2 ± 3.6 mL/min in the control group (p value = NS). We can speculate that after 6 months, anastomotic flow is as good as the time of the device implantation because anastomotic diameter remained larger than distal LAD. Therefore, fluoroscopic evaluation of the anastomosis demonstrated that flow in the distal LAD was excellent in both groups after 6 months.
A review of the histologic results showed that the thickness of myointimal hyperplasia differed between groups: 0.21 ± 0.1 mm in the device group versus 0.01 mm in the control group (p < 0.001). Because myointimal hyperplasia is the first cause of anastomotic stenosis, we would presume it would cause a reduction of anastomotic diameter, but this was not the case. Mean anastomotic diameter was 1.7 mm in the GC group versus 1.6 mm in the control group (p value = NS). Mean anastomotic diameter in the GC group corresponded to mean LAD diameter of adult sheep. In other words, there was no angiographic or histopathological sign of stenosis in any groups even if myointimal hyperplasia was more important in the GC group, because anastomotic luminal width corresponds to the normal adult sheep LAD diameter.
Myointimal hyperplasia was less predominant in the tower but no statistical difference was noted in the different parts of the GC (tower and stent). Myointimal hyperplasia, although it did not affect anastomosis patency, was significantly higher in the GC group. Vessel injury, stent material, and stent geometry are factors known to correlate closely with late neointimal thickening in experimental models [3, 4]. The degree of hyperplasia has been related linearly to the degree of stent-induced vessel wall injury [3], but the specimens of the GC group had intact internal elastic lamina and we can reasonably assume that the GC did not provoke any significant vessel injury during deployment.
Animal studies [4] have demonstrated that stent design could cause late neointimal thickening because of the optimization of fluid flow at the bloodmetal interface, without affecting vessel patency. At deployment, the stent stretches the vessel, imposing a cross-sectional polygonal luminal shape that depends on the stent design. The lumen therefore initially assumes the geometric shape of a polygon, with the struts marking each vertex. Altered stent-imposed fluid dynamics may cause regions of turbulence or stagnation in the immediate vicinity of stent struts and provide a basis for the observation that restoring the lumen to a circular shape will optimize fluid flow characteristics at the bloodmetal interface. Figure 4 shows that stent struts alter vessel geometry; the restoration of luminal circularity by neointimal growth seems to be the physiologic response to optimize flow at the bloodmetal interface. The oversizing of the device at the implantation has also played a role in the degree of intimal proliferation. As fluoroscopy images demonstrated, all GC had a diameter larger than native LAD (Table 1) at the preoperative fluoroscopy and we can imagine that the difference between GC and native vessel diameter has been filled up with myointimal proliferation in order to optimize blood flow. Figure 5 illustrates the histologic reaction at the GC edges (transitional area) and elucidates this phenomenon: neointimal growth optimizes blood flow characteristics, reducing turbulence in the proximity of the GC edges.
The oversizing phenomenon may also explain the reduction of cross-sectional anastomotic area 35.6% of distal LAD in the device group versus 5% in the control group (Table 2). These values were calculated assuming that native vessel cross-sectional area corresponds to GC cross-sectional area as reported in Figure 4. This assumption, however, is valid only if the GC is oversized with respect to the native LAD diameter. The consequence of the GC oversizing is an overestimation of cross-sectional anastomotic area reduction. Therefore, the mean anastomotic cross-sectional area reduction of 35.6%, corresponding to a 15% reduction of anastomotic diameter, is overestimated.
No sign of artery wall degeneration has been found. Signs of mild chronic inflammatory reaction have been found in some specimens mostly in the proximity of PTFE layer. This tissue reaction, already described by other authors, seems to be related to the PTFE itself [5, 6]. There is no evidence in current literature that this tissue reaction has ever affected the anastomosis patency.
The segment of the IMA having been everted and tied onto a rigid structure (tower) was still viable even if it is not possible to recognize an intimal layer and the media has signs of chronic inflammatory reaction. Carrying out the histologic analysis, we also focused our attention on the rim of PTFE, which is between the end of the IMA and the endothelium of the coronary: this PTFE was covered with myointimal cells as were the other parts of the GC.
In conclusion, the GC provides a consistent and reproducible coronary artery anastomosis and reduces technical demand and manual dexterity in coronary operations. Long-term results demonstrate that OPCABG performed with GC had the same patency rate and luminal width as those performed with running suture. Myointimal hyperplasia was more important in GC group as consequence of vessel remodeling to optimize anastomotic blood flow. We think that the device oversizing at the implantation has played a major role in determining myointimal hyperplasia. More precise GC sizing could reduce myointimal hyperplasia even if, in our experience, size is not associated with stenosis. However, it remains to be determined whether this favorable outcome will also be present in humans.
| Acknowledgments |
|---|
|
|
|---|
| Footnotes |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Ferrari, P. Tozzi, and L. K. von Segesser The Vascular Join: a new sutureless anastomotic device to perform end-to-end anastomosis. Preliminary results in an animal model Interactive CardioVascular and Thoracic Surgery, February 1, 2007; 6(1): 5 - 8. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Tozzi, E. Borghi, E. Haesler, G. Siniscalchi, A. Motti, D. Hayoz, and L. K. von Segesser Progress in cardiovascular anastomoses: will the vascular join replace Carrel's technique? Eur. J. Cardiothorac. Surg., September 1, 2006; 30(3): 425 - 430. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. A. Khan, C. Torrens, D. E. Noakes, L. Poston, M. A. Hanson, L. R. Green, and S. K. Ohri Effects of pre-natal and early post-natal undernutrition on adult internal thoracic artery function Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 811 - 815. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, H. MacVaugh III, E. Bagaev, M. Maringka, S. Kirschner, J. Beilner, and A. Haverich Magnetic Vascular Port in Minimally Invasive Direct Coronary Artery Bypass Grafting Circulation, September 14, 2004; 110(11_suppl_1): II-55 - II-60. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Carrel, L. Englberger, D. Keller, S. Windecker, B. Meier, and F. Eckstein Clinical and angiographic results after mechanical connection for distal anastomosis in coronary surgery J. Thorac. Cardiovasc. Surg., June 1, 2004; 127(6): 1632 - 1640. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Carrel, F. S. Eckstein, L. Englberger, P. A. Berdat, and J. Schmidli Clinical experience with devices for facilitated anastomoses in coronary artery bypass surgery Ann. Thorac. Surg., March 1, 2004; 77(3): 1110 - 1120. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. van Andel, P. V. Pistecky, and C. Borst Mechanical properties of porcine and human arteries: implications for coronary anastomotic connectors Ann. Thorac. Surg., July 1, 2003; 76(1): 58 - 64. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Scheltes, C. J. van Andel, P. V. Pistecky, and C. Borst Coronary anastomotic devices: blood-exposed non-intimal surface and coronary wall stress J. Thorac. Cardiovasc. Surg., July 1, 2003; 126(1): 191 - 199. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Tozzi, A. F. Corno, and L. K. von Segesser Sutureless thoracic aorta to femoral artery bypass with robotic videoendoscopic approach: a fast track procedure Interactive CardioVascular and Thoracic Surgery, June 1, 2003; 2(2): 186 - 189. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Tozzi, A. F. Corno, and L. K. von Segesser Sutureless coronary anastomoses: revival of old concepts Eur. J. Cardiothorac. Surg., October 1, 2002; 22(4): 565 - 570. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |