|
|
||||||||
Ann Thorac Surg 2001;71:1369-1371
© 2001 The Society of Thoracic Surgeons
a Cardiopulmonary Research Science Technology Institute, Dallas, Texas, USA
Accepted for publication May 18, 2000.
Address reprint requests to Dr Mack, COR Specialty Associates of North Texas, P.A., 7777 Forest Ln, Suite A-323, Dallas, TX 75230
e-mail: mmack{at}crsti.org
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Two types of AVD have been described. Kinetic-assisted venous drainage (KAVD) is performed by placement of a centrifugal pump in the venous line [1]. Alternatively vacuum-assisted venous drainage (VAVD) has been used as another enabling perfusion modality for smaller cannula systems [2, 3]. Vacuum-assisted AVD requires a vacuum source with a pressure-regulating device, a negative pressure measuring device, and a sealed venous reservoir. Currently, equipment manufacturers offer sealed (open) venous reservoirs with and without pressure release valves, pressure-regulating kits, and preassembled lines [4, 5]. Issues of excessive negative pressure applied to the sealed venous reservoir causing damage to the cellular elements of the blood as well as over-pressurization of the venous reservoir with delivery of retrograde venous air embolus to the patient are paramount.
Generally entrainment of venous air during cardiopulmonary bypass using gravity venous drainage is not uncommon and is usually believed to be controllable and insignificant because of the de-airing capabilities of the conventional cardiopulmonary bypass circuit. We describe in this case report a complication of VAVD of significant air entrainment during minimal access repair of an atrial septal defect because of inadvertent positive pressurization of the venous circuit during vacuum-assisted venous drainage.
A 33-year-old woman was referred for elective repair of a secundum-type atrial septal defect. A minimal access surgical approach was planned using a right inframammary incision. Cardiopulmonary bypass was instituted using access through a right groin incision. A Heartport (Heartport Inc, Redwood City, CA) endoarterial cannula (21F) was placed into the femoral artery and a Medtronic-DLP Carpentier (Medtronic Inc, Grand Rapids, MI) two-stage venous cannula (24/29F) was placed through the right femoral vein and positioned in the right atrium and superior vena cava.
The cardiopulmonary bypass circuit consisted of a Sarns Turbo 440 (Terumo/Sarns Inc, Ann Arbor, MI) oxygenator with a hard-shell venous reservoir, a Sarns Delphin centrifugal arterial pump, a 40-µm arterial line filter, standard suction, and vent lines. A DLP 60000 pressure monitor (Medtronic) was used to measure venous line pressure, with the alarm set to trigger at negative 100 mm Hg. A standard vacuum wall source with regulator was used in combination with a new device consisting of a disposable needle valve with the ability to precisely control negative pressure during recirculation at 1 mm Hg increments. The valve came preassembled with a Y line open to atmosphere for reservoir venting before institution of VAVD, a vapor trap tubing and a Roberts clamp. The ports on the venous reservoir were sealed with caps on the vacuum line with the needle valve assembly connected to the vent port assuring that the Y line was open to the atmosphere. After assembly of the circuit, cannulation and adequate heparinization, the suction lines were activated 2 to 3 minutes before the anticipated institution of cardiopulmonary bypass. The clamp on the venous line was removed and immediately depriming retrograde toward the patient occurred. The line was immediately clamped. A quick examination of the circuit confirmed that the Y line of the vacuum regulating device was open to the atmosphere. Air was noted high in the venous cannula and the confirmation of positive pressurization of venous reservoir was confirmed when a Luer lock cap was removed resulting in audible positive pressure release. It then became evident that the venous line had deprimed toward the patient. An immediate examination of the heart by transesophageal echocardiography revealed air in the right atrium, left atrium, and left ventricle. The clinical assessment, therefore, was that the positively pressurized venous line caused an air embolus into the right atrium with enough positive pressure that it superceded the pressure of the left to right shunt across the atrial septal defect resulting in a paradoxic embolus to left side of the heart. The needle valve assembly was expeditiously removed and replaced with a Y tubing vented to the atmosphere.
The surgical options of managing air embolism at that point were entertained including deep Trendelenberg positioning, retrograde cerebral venous rescue, or institution of cardiopulmonary bypass with aortic cross-clamping and venting. It was decided that minimal amounts of air had crossed the septal defect to the left side of the heart and that surgical correction should proceed as planned. The defect was repaired with an aortic cross-clamp time of 38 minutes.
The patient was extubated in the operating room; however, immediate neurologic examination revealed a left hemiparesis and mental obtundation and a left visual field of homonymous hemianopsia. The patient was immediately taken to a hyperbaric oxygen chamber for decompression therapy. Over a period of 5 hours at a pressure of 3 atm the patient had immediate recovery of all neurologic function including the left hemiparesis and homonymous hemianopsia. Diffusion-weighted magnetic resonance imaging performed immediately on completion of decompression as well as on the fourth postoperative day were normal. The patient was discharged on the fourth postoperative day with a normal clinical neurologic examination.
| Comment |
|---|
|
|
|---|
|
|
We still use VAVD for select procedures; however, we have adopted the following changes. First, venous inlet pressure is monitored using a DLP 60000 (Medtronic), the alarm is set to trigger at nominal positive pressure (5 to 10 mm Hg) instead of a negative pressure. Second, we continue to monitor negative inlet pressure and maintain at less than -80 mm Hg. We also advocate measuring the hard shell venous reservoir/integrated cardiotomy pressure with an independent device, set to alarm at a minimal positive pressure. Cardiopulmonary bypass is initiated with the vacuum source unattached to the reservoir and applied once gravity drainage has begun. This will insure a vented reservoir in the event that suctions were activated before bypass. The needle valve is place between the Y and the vacuum source rather than between the Y and the venous reservoir (Fig 1). We believe that it is important for all perfusionists and surgeons to be aware of this real and potential hazard of VAVD and appropriate measures be taken to prevent this complication.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. W. Hammon Extracorporeal Circulation: Perfusion System Card. Surg. Adult, January 1, 2008; 3(2008): 350 - 370. [Full Text] |
||||
![]() |
A. F. Corno Systemic venous drainage: can we help Newton? Eur J Cardiothorac Surg, June 1, 2007; 31(6): 1044 - 1051. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. M. Eckmann and V. N. Lomivorotov Microvascular gas embolization clearance following perfluorocarbon administration J Appl Physiol, March 1, 2003; 94(3): 860 - 868. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Shen, C. Giacomuzzi, and R. M. Ungerleider Current strategies for optimizing the use of cardiopulmonary bypass in neonates and infants Ann. Thorac. Surg., February 1, 2003; 75(2): S729 - S734. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. A. Hessel II and L. H. Edmunds Jr. Extracorporeal Circulation: Perfusion Systems Card. Surg. Adult, January 1, 2003; 2(2003): 317 - 338. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |