|
|
||||||||
Ann Thorac Surg 2000;69:536-540
© 2000 The Society of Thoracic Surgeons
a Heart Institute of University of São Paulo Medical School, São Paulo, Brazil
Address reprint requests to Dr Pêgo-Fernandes, Instituto do Coração do Hospital das Clínicas da Universidade de São Paulo, Divisão Cirúrgica, 2nd andar Av Dr Eneas de Carvalho Aguiar, 44, São Paulo-SP, Brazil, CEP 05403-000
e-mail: paulopego{at}incor.usp.br
| Abstract |
|---|
|
|
|---|
Methods. Fifteen dogs were studied. The hemodynamic evaluation included serial measurements of the classic parameters. Carotid and coronary blood flows were obtained by electromagnetic transducers.
Results. The hemodynamic evaluation did not show significant statistical changes. The use of circulatory-isolated assistance with biopump shows reduction (24.6% ± 6.1%) in coronary flow, in relation to the control situation and the concomitant use of biopump and intraaortic balloon pump showed similar coronary flow. Regarding carotid flow, a similar trend was observed in relation to the positive influence of the pulsatile flow with an intraaortic balloon pump without statistical significance (p = 0.0582).
Conclusions. The biopump reduces the coronary flow in dogs. The use of intraaortic balloon pump with the biopump increases the coronary flow significantly, reaching similar values to those observed without the circulatory assistance.
| Introduction |
|---|
|
|
|---|
The concomitant use of an intraaortic balloon pump and a biopump has been studied, as the combination increases the coronary blood flow produced by the intraaortic balloon and increases the arterial blood flow produced by the biopump. It would be even more convenient to combine two easy to handle and inexpensive devices. This would be essential to increase myocardial perfusion to relieve ventricular load and maintain a mean blood pressure and blood flow, allowing the recovery of organ failure, and for the undesirable effects of the continuous biopump flow to be minimized by the pulsatility of the intraaortic balloon pump.
The objective of this study was to evaluate the effects of the centrifugal pump on the coronary and carotid flows in left ventricular circulatory assistance in dogs with normal hearts. In addition, it was aimed at evaluating whether the concomitant use of the intraaortic balloon is sufficient to counterbalance eventual undesirable effects of circulatory assistance with continuous flow.
| Material and methods |
|---|
|
|
|---|
The anesthesia technique used hypnotic and halogenated drugs. After tracheal intubation, controlled mechanical ventilation was started. Standardized monitoring included electrocardioscopy, invasive blood pressure (systolic, diastolic, and mean blood pressure), capnography, central venous pressure, central temperature, vesical catheter, and Swan-Ganz catheter.
The experiment included the direct evaluation of coronary artery flow: left coronary artery circumflex branch and the carotid arteries; right carotid artery, with circulatory assistance with a centrifugal pump to the left ventricle, with or without intraaortic balloon assistance.
To measure carotid flow to the right carotid artery, a cervicotomy was performed from the anterior edge of the right sternocleidomastoid muscle, and an extension of 3 cm was released. To measure coronary flow, the heart was approached by a longitudinal thoracotomy with a median sternotomy, while the animal was kept in horizontal dorsal decubitus. The initial portion of the coronary artery circumflex branch was exposed and dissected. Flow rates were recorded with a Hewlett-Packard electromagnetic flowmeter and a Hewlett-Packard polygraph (Hewlett Packard, Waltham, MD).
The intraaortic balloon pump was placed at the thoracic descending aorta, and inserted through the previously dissected left femoral artery. The transsternal access was used to install the biopump, after dissection of the left coronary artery circumflex branch. Two concentric sutures were performed at the beginning of the ascending aorta where a no. 18 infusion cannula was inserted in the middle of the ascending aorta. After a suture was made at the left auricle to insert a
-inch cannula. The cannulas are routinely used in cardiac operations and in cases of circulatory assistance with the biopump.
A Biomedicus biopump (Medtronic, Minneapolis, MN) was used. In addition to the measurements of mean coronary and carotid flow, mean right atrial pressure, mean pulmonary capillary pressure, mean pulmonary artery pressure, mean blood pressure, and heart rate were made. Pressure curves were recorded in a Hewlett-Packard polygraph.
To measure the cardiac output a thermodilution technique was used. Cardiac output value, calculated as liters per minute, was obtained by the average of three measurements, as long as the differences among them were less than 10%.
The following hemodynamic parameters were calculated using classic formulas: cardiac index, systolic index, systemic vascular resistance, and pulmonary vascular resistance. All measurements were taken at four time periods: (1) control (C1), the dog has the intraaortic balloon pump and the biopump installed, but they are not in use; (2) biopump test (C2), the intraaortic balloon pump and biopump are installed, but only the biopump is working. This evaluates the isolated effect of the biopump on left ventricular circulatory assistance; (3) biopump and intraaortic balloon pump test (C3), the intraaortic balloon pump and the biopump are working, and the balloon in inflated once at each diastole of the animals heart; (4) final condition (C4), the intraaortic balloon pump and the biopump are installed, but only the biopump is working. This condition repeats condition 2 (a second proof condition).
In C2, C3, and C4, the blood flow maintained by the biopump was such that the relation between animals blood flow and cardiac output was on average 75%.
The studied variables were presented as mean and standard deviation. Differences between the conditions were evaluated by one-way analysis of variance. The significance level was 0.05.
| Results |
|---|
|
|
|---|
Table 1 shows the mean values and standard deviation for the heart rate, mean right atrial pressure, mean pulmonary artery pressure, mean pulmonary capillary pressure, mean blood pressure, cardiac output, cardiac index, systemic vascular resistance, pulmonary vascular resistance. Heart rate was significantly greater in the control animals than in the other three conditions. Statistically significant differences were not found when the three conditions with circulatory assistance were compared to each other.
|
Pulmonary artery mean pressure was lower in the control animals when compared with C2 (biopump). There was no statistically significant change in the remaining comparisons for this parameter.
Mean pulmonary capillary pressure did not show statistically significant change in the different conditions, as well as the mean blood pressure, cardiac output, cardiac index, systemic vascular resistance, and pulmonary vascular resistance.
Table 2 shows mean values and standard deviation in biopump flows, cardiac output, and biopump flow to total cardiac output ratio (the output produced by the heart in addition to the output produced by the biopump). No statistically significant change for these parameters was observed in the different conditions. The ratio of the biopump mean flow and the total cardiac output was on average 75%.
|
|
For the carotid flow, the profile analysis strongly suggests that there is a difference, although not significant (p = 0.0582), among the different steps. In Figure 2 the carotid blood flow may be better visualized in the four conditions. A trend to carotid flow may be observed when C1 is compared to C2 where assisted circulation with the biopump is started (p = 0.0540), and to C4 where the intraaortic balloon is disconnected (p = 0.0602). C1 and C3 have the same results in this parameter (p = 0.8327). A statistically significant difference of the carotid flow is observed when C2 is compared with C3 (p = 0.0298). This reversal of the carotid flow decrease is not confirmed in the final condition (second proof) (p = 0.1186) between C3 and C4, showing this parameter only had a significant trend to decrease with the biopump.
|
| Comment |
|---|
|
|
|---|
In the English literature, the use of the biopump, a continuous flow device, whether or not used in combination with an intraaortic balloon pump, a pulsatile device, is discussed in a broad manner, pulsatile flow devices versus nonpulsatile flow devices [3, 4]. There are several clinical studies on this issue. Even in acute conditions of mechanical circulatory assistance, as with extracorporeal circulation, there are evaluations comparing the use of pulsatile extracorporeal circulation with continuous flow circulation, which is usually used [5, 6]. Several researchers support the use of pulsatile flow during extracorporeal circulation [57], although others consider it absolutely unnecessary [3]. The present study might provide physiopathologic data to explain these differences [7], thanks to better coronary flow resulting from the pulsatility of the intraaortic balloon.
Mickleborough and colleagues [8] subjected two groups of seven dogs to normothermic cardiac arrest to produce myocardial injury and left ventricular insufficiency. Intraaortic balloon pump and inotropic drugs were used in one group of dogs and in the other group, the biopump was used. Both methods achieved good hemodynamic rates at the beginning of left ventricular assistance. However, after 3 hours of assistance, the dogs treated with intraaortic balloon pump and inotropic drugs had lower left ventricular compliance and worse left ventricular systolic function, and a greater degree of necrosis than the group of dogs treated with the biopump. These researchers suggest that even patients who are hemodynamically stable with the use of intraaortic balloon pump and drugs could benefit from early circulatory assistance with the biopump. In our clinical experience at the Heart Institute [9, 10], we observed that late introduction of circulatory assistance devices is an important factor for high morbidity and mortality. However, we believe both methods may be used concurrently in patients where inotropic drugs and intraaortic balloon cannot adequately resolve low cardiac output. This study supports the suggestion made by Penningtons group: the earlier use of the biopump allows a decrease of ventricular load and produces a better recovery of the insufficient ventricle [11].
Other investigators [12, 13] prefer the concurrent use of mechanical circulatory assistance devices to potentiate the advantages of each method. Ide and colleagues [12] evaluated the use of a catheter combining three ventricular mechanical assistance devices in dogs: the intraaortic balloon pump, the biopump, and the "hemopump." The intraaortic balloon pump was aimed at providing pulsatility to blood flow, and the other devices were continuous flow devices. They studied hemodynamic and myocardial blood flow conditions in 10 dogs with heart failure, comparing these parameters in pulsatile (with intraaortic balloon pump) and nonpulsatile conditions. They concluded that there was improvement in cardiac output, mean aorta pressure, and myocardial blood flow as measured by a Doppler laser flowmeter. This study also highly supports the combination of intraaortic balloon pump and biopump as circulatory assistance devices.
Nagasaka and associates [13] developed a device combining the biopump and a synchronized intraaortic balloon pump. They studied the influence of this device on endocardial viability and renal flow rate. The system was used in 12 pigs and evaluated the devices alone and in combination. The researchers concluded that cardiac assistance was more effective when the two mechanisms were combined, whether or not they were synchronized. Although these studies used different indexes to evaluate the effects on the heart, they corroborate our findings.
Using carotid flow measurements and intermittent hypercapnia, a study with a biopump in 17 calves showed that there was no damage to the cerebral flow autoregulation with the use of continuous flow (fibrillating heart and biventricular assistance with biopump). Onoe and colleagues [14] studied the influence of pulsatile extracorporeal circulation on cerebral blood flow in dogs undergoing periods of circulatory arrest with profound hypothermia, and observed that after circulation resumption, the dogs with pulsatile flow tended to have higher cerebral flow compared to those dogs receiving continuous flow, especially with a longer circulatory arrest time. Sadahiro and associates [15] studied pulsatile and nonpulsatile extracorporeal circulation with normothermia and moderate hypothermia (25°C) in dogs. They observed that the autoregulation of the cerebral flow was kept intact at pressures higher than 50 mm Hg, but below this value the dogs undergoing perfusion with pulsatile flow had a higher cerebral blood flow than the dogs with continuous flow.
In our study, significant carotid blood flow changes were observed when C2 was compared to C3. The significant carotid flow improvement observed with the use of the combination of both methods was not confirmed in the second proof (C4) where the intraaortic balloon pump is turned off. Although there was a trend with p values close to 0.05, the significance level was not achieved probably because the hemodynamic conditions were satisfactory and therefore, did not affect the autoregulation, as shown in other studies.
In summary, (1) left ventricular circulatory assistance with the biopump decreases coronary flow in dogs with normal heart; (2) left ventricular circulatory assistance with the biopump decreased the carotid flow in dogs with normal hearts; (3) the intraaortic balloon pump combined with left ventricular circulatory assistance using a biopump significantly increased the dogs coronary flow, achieving similar values to those observed without circulatory assistance; and (4) the intraaortic balloon pump combined with left ventricular circulatory assistance using a biopump tended to normalize the carotid flow in dogs with normal hearts. The intraaortic balloon must be combined with a biopump whenever possible, when it is used for left ventricular circulatory assistance, as it improves coronary flow and consequently myocardial perfusion.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Y. Ootaki, K. Kamohara, M. Akiyama, F. Zahr, M. W. Kopcak Jr., R. Dessoffy, and K. Fukamachi Phasic coronary blood flow pattern during a continuous flow left ventricular assist support Eur. J. Cardiothorac. Surg., November 1, 2005; 28(5): 711 - 716. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |