|
|
||||||||
Ann Thorac Surg 1999;68:1285-1289
© 1999 The Society of Thoracic Surgeons
a Cardiothoracic Surgical Unit, Green Lane Hospital, Auckland, New Zealand
Address reprint requests to Mr Willcox, Clinical Perfusion, Green Lane Hospital, Green Lane West, Auckland 3, New Zealand
e-mail: timw{at}ahsl.co.nz
| Abstract |
|---|
|
|
|---|
Methods. Using salvaged clinical cardiopulmonary bypass circuits, fixed volumes of air were introduced into the venous return line at unrestricted rates and at fixed rates using gravity venous drainage and vacuum-assisted venous drainage. Emboli counts were recorded distal to the arterial line filter using a 2-MHz pulsed-wave Doppler monitor. Emboli counts were similarly recorded after the introduction of carbon dioxide into the venous return line instead of air.
Results. The number of emboli rose with increasing volumes of entrained venous air (p < 0.001), and there was an almost tenfold increase with vacuum-assisted venous drainage (p < 0.0001) compared with gravity venous drainage. Venous air was entrained at a significantly faster rate under vacuum-assisted venous drainage (p < 0.0001). When the entrainment rate of venous air was fixed, the difference in emboli numbers recorded for gravity and assisted venous drainage was not significant. There was a significant reduction in arterial line emboli when carbon dioxide rather than air was entrained under both vacuum-assisted and gravity drainage (p < 0.001).
Conclusions. Entrained venous air during cardiopulmonary bypass is a potential hazard, particularly during vacuum-assisted venous drainage. Every effort should be made to avoid venous air entrainment.
| Introduction |
|---|
|
|
|---|
During a recent study involving open heart operations [4], increases in carotid artery Doppler signals typical of gaseous emboli were associated with air seen in the venous return line. Removal of venous line air is considered an advantage of hard-shell venous reservoir designs in CPB circuits [5]. We [6] have previously reported the variable ability of hard-shell venous reservoirs to remove venous line bubbles but the extent to which these still appear in the line downstream from the arterial filter has not been investigated.
Vacuum-assisted venous drainage (VAVD) is a relatively new clinical technique. It allows the use of smaller-diameter venous return cannulas for given blood flow rates, thus improving access in the surgical field, especially during minimally invasive surgical procedures. However, the method is based on a vacuum applied to the venous reservoir, and this may influence behavior of gaseous emboli in the circuit. The present in vitro study was undertaken to measure the extent to which air introduced into the venous return line affects the numbers of emboli detected in the arterial line downstream from a 40-µm filter. The effect of venous air entrainment under VAVD conditions and the impact of changing the gas composition of bubbles entering the venous line were also studied.
| Material and methods |
|---|
|
|
|---|
inch); the Maxima Forté hard-shell venous reservoir (Medtronic Inc, Anaheim, CA); a roller pump on a Stockert CAPS or S3 heart-lung machine (Stöckert Instrumente, Munich, Germany); the Maxima Forté hollow-fiber oxygenator (Medtronic Inc); and a Bentley AF-1040D 40-µm arterial line filter (Baxter Healthcare Corp, Irvine, CA). The outer arterial reservoir shell of a Bentley Ben 10 bubble oxygenator (Baxter Healthcare Corp) was positioned at a standard height above the venous reservoir and used to simulate the patient. The venous return line was connected to the "patient" reservoir outlet port. The arterial line cannula (24F straight Argyle; Baxter Healthcare Corp) was fixed to the inside of the "patient" reservoir at the 1,000-mL level obliquely opposing the direction of the outlet port. The circuit prime volume was restored with group O date-expired resuspended red cells and Plasmalyte 148 (Baxter Healthcare Corp) to a hematocrit of 23% to 26%. The prime was recirculated at 3.5 L/min and a line pressure of 100 mm Hg with the volume of the venous reservoir maintained at 1,000 mL and the "patient" reservoir at 1,500 mL. The arterial filter purge was open, and the oxygenator was ventilated with a sweep gas flow rate of 2.5 L/min at an inspired oxygen fraction of 0.50.
Doppler emboli detection
In each experiment, the circuit was monitored with a Flowlink 300 color-flow Doppler (Rimed, Tel Aviv, Israel), interfaced with a custom-built analog signal processor that was designed to count emboli signals as described previously [7]. The Doppler monitor was operated in the 2-MHz pulsed-wave mode. A purpose-built clamp was used to mount the Doppler probe on the circuit tubing so that the probe faced the direction of blood flow at 45 degrees. The space between the probe and the tubing was filled with ultrasonic gel. In all experiments, the probe was placed on the arterial line downstream from the filter. Beam depth was set at 27 mm, machine power at 40%, and gain at 6 for optimal audible signal and color-flow display. A sensitivity control on the signal processor was adjusted to obtain correlation between the audible and visible signals typical of emboli and the count from the signal processor. We cite arterial line emboli counts but recognize that there are factors that can confound emboli signal counters [8]. A baseline arterial-line emboli count was recorded over 180 seconds prior to each experimental intervention. The net experimental arterial-line emboli count was obtained in all cases by subtracting the baseline count from the count done after entrainment of venous gas.
Venous air entrainment
A venous air injection site was created by inserting a standard 12.5 mm (
inch) Luer-Lok connector into the venous line beyond the venous occluder and fitting an injection plug (B | Braun, Melsungen, Germany) into the Luer-Lok. A 37.5-mm (1
-inch) stainless steel 21-gauge hypodermic needle was introduced into the venous line through the injection plug, and a three-way stopcock was attached to the needle. Opening the stopcock enabled entrainment of "venous air". The volume entrained was limited by connecting a 1.0-L blood transfer pack (Baxter Healthcare Corp), already filled with the desired volume of gas, to the stopcock. In this mode, there was no restriction on the rate at which gas could be entrained. In experiments requiring control of both gas volume and entrainment rate, gas was introduced from a prefilled syringe at the desired rate.
Experiments
We conducted the following four experiments: increasing the volume of entrained venous air during gravity venous drainage (GVD) (experiment 1); increasing the volume of entrained venous air during VAVD (experiment 2); a fixed rate of venous air entrainment during VAVD and GVD (experiment 3); and entrainment of carbon dioxide (CO2) into the venous line (experiment 4). For all experiments, a p value of less than 0.05 was considered significant.
Experiment 1
We tested the hypothesis that entrained venous air is incompletely removed by the CPB circuit under gravity drainage and that greater volumes of entrained venous air result in larger numbers of arterial line emboli. Twenty-five milliliters of air was entrained at an unrestricted rate, and the entrainment time was recorded. The number of arterial line emboli counted was recorded over 180 seconds beginning at initiation of air entrainment. The procedure was repeated for 50, 75, and 100 mL of entrained air, and this sequence was repeated three times in each of three separate circuits (a total of nine studies for each volume of venous air). Although air was entrained at an unrestricted rate, the time for entrainment at each volume was virtually identical for each experimental run. One-way analysis of variance was used to assess the significance of any trend in the mean count of arterial line emboli with the increasing volumes of entrained venous air.
Experiment 2
We tested the hypothesis that venous air entrained during VAVD results in larger numbers of emboli in the arterial line than equal amounts of air entrained during GVD. The circuit was converted to VAVD by sealing all luers and inlet ports of the venous reservoir and applying a 60 mm Hg vacuum to the reservoir vent port. The vacuum was controlled using a thoracic low-vacuum regulator (Clements Medical Equipment PTY Ltd, Sydney, Australia), and the pressure within the venous reservoir was continuously monitored using a pressure transducer module on the heart-lung machine. The experimental protocol was identical to that in experiment 1, except that only two volumes of air (25 and 50 mL) were tested. An unpaired t test was used to compare the mean arterial-line emboli count recorded after entrainment of equal volumes of air during VAVD and GVD. The time for these volumes of air to be entrained during both VAVD and GVD were similarly compared.
Experiment 3
We tested the hypothesis that the rate of venous air entrainment accounts for any difference in arterial line emboli count recorded under conditions of vacuum-assisted and gravity drainage. A disposable syringe was filled with 50 mL of air and attached to the three-way stopcock on the venous air injection site. Injection of the air at 2 mL/s and a 180-second arterial-line emboli count recording were then begun simultaneously. This was repeated six times in each condition in two circuits. An unpaired t test was used to comparethe mean arterial-line emboli count recorded during vacuum-assisted drainage and gravity drainage at a fixed rate of entrainment.
Experiment 4
We tested the hypothesis that entrainment of CO2 instead of air reduces the resultant arterial-line emboli count under both VAVD and GVD. Arterial emboli counts were recorded over 180 seconds from the onset of gas entrainment (unrestricted rate) nine times under the following conditions: first using GVDentrainment of 75 mL of air and 100 mL of air and then 75 mL of CO2 and 100 mL of CO2; and second using VAVDentrainment of 25 mL of air and 50 mL of air and then 25 mL of CO2 and 50 mL of CO2. An unpaired t test was used to compare arterial line emboli count after entrainment of air and CO2 at both gas volumes under each drainage condition.
| Results |
|---|
|
|
|---|
|
|
|
Experiment 4
The mean arterial-line emboli count after the entrainment of equal volumes of air and CO2 under both gravity drainage and vacuum-assisted drainage is shown in Figure 4. At all volumes, under both VAVD and GVD, there was a significant reduction in the arterial line emboli count after entrainment of CO2 into the venous line (p < 0.001).
|
| Comment |
|---|
|
|
|---|
The contribution of venous air to arterial line emboli has not been investigated. However, small amounts of gas injected into the venous return line result in microemboli downstream from a membrane oxygenator proximal to the arterial filter, as does rapid injection of well-degassed saline solution into the venous sampling port of a venous reservoir [12]. Venous air during CPB is common. The perceived ability of venous reservoirs to remove such air during CPB has led to an "acceptance" that venous air is consequently benign. Typically, when a hard-shell venous reservoir is used, venous cannulas are not deaired, and initiation of bypass is associated with a bolus of air returning to the CPB circuit. During CPB, air can enter the venous line around the pursestring suture of a venous or retroplegia cannula, especially during retraction of the atrial wall. A single atrial cannula can be used in procedures for some congenital conditions to vent blood, and coincidentally air, from the left heart across a patent foramen ovale or an excised atrial septum. New deairing techniques involve venting air from the left heart and aortic root directly to the venous return line [13].
This study demonstrates that small amounts of venous air result in emboli distal to a 40-µm filter in an arterial line when conventional gravity drainage is used. Further, the numbers of arterial line emboli increase as greater volumes of air are entrained (see Fig 1). Whereas we used fixed volumes of venous air, the volume of air entering the venous line clinically is unrestricted.
Vacuum-assisted venous drainage is a clinical technique used to enhance venous drainage to the perfusion circuit when narrower cannulas and tubing diameters are used. In a study [14] using an animal model, reductions in circuit priming volume were achieved during VAVD, with the result that the CPB circuit could be simplified and miniaturized. Potential applications for VAVD include minimally invasive surgical procedures, pediatric operations, and femoral-femoral bypass.
We have shown that VAVD profoundly increases the number of emboli likely to reach the patient when venous air is entrained into the CPB circuit. Vacuum-assisted venous drainage produced an almost tenfold increase in the arterial line emboli count compared with GVD after entrainment of venous air (see Fig 2), despite use of the Medtronic Maxima Forté reservoir, which we [6] have demonstrated to be a comparatively effective venous air filter.
Because of the influence of a vacuum, the time for equivalent volumes of air to enter the venous return line was significantly shorter under VAVD than GVD (see Fig 3). However, when the rate of air entry was fixed, the arterial line emboli count under VAVD was not significantly greater than that for GVD. It appears that the passage of venous air through a CPB circuit is delivery-rate dependent.
Leaving the venous line unprimed prior to initiation of CPB under VAVD to reduce circuit prime volume has also been advocated [15]. This practice will result in up to 200 mL of air being rapidly delivered to the venous reservoir, potentially increasing arterial line emboli at the onset of CPB. Consequently, deairing the venous cannula during connection to the CPB circuit is recommended.
Carbon dioxide flooding of the surgical field to reduce air embolism was first used during CPB in the 1960s [16], and there is renewed interest in this technique [17]. It is reasonable to expect that if CO2 rather than air were entrained, then given the transit time of these bubbles through the CPB circuit and the increased solubility of CO2, fewer emboli would be detected in the arterial line. The reduction in arterial line emboli count seen in experiment 4 after entrainment of CO2 rather than air supports this hypothesis under both VAVD and GVD (see Fig 4). However, it is notable that after 50 mL of CO2 was entrained under VAVD, the arterial line emboli count was significantly greater than that for 100 mL of air entrained under GVD. Thus, although CO2 is advantageous, it does not offset the potentially deleterious effect of VAVD on arterial line emboli.
These results illustrate the potential hazard of venous air, particularly during VAVD, and suggest that every effort should be made to eliminate this problem when it is detected. Excessive negative pressure caused by gravity can be reduced by partially clamping the venous return line [5], which in turn will reduce the amount of air entrained. However, control of venous air entrainment usually requires intervention by the surgeon. In light of our findings, the priority for such interventions should be emphasized.
| Acknowledgments |
|---|
This work was supported by grants from the English Freemasons of New Zealand and the Health Research Council of New Zealand.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. S. Murphy, E. A. Hessel II, and R. C. Groom Optimal Perfusion During Cardiopulmonary Bypass: An Evidence-Based Approach Anesth. Analg., May 1, 2009; 108(5): 1394 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
I Kutschka, J Skorpil, A El Essawi, T Hajek, and W Harringer Beneficial effects of modern perfusion concepts in aortic valve and aortic root surgery Perfusion, January 1, 2009; 24(1): 37 - 44. [Abstract] [PDF] |
||||
![]() |
P. Nielsen, J. Funder, M. Jensen, and H Nygaard Influence of venous reservoir level on microbubbles in cardiopulmonary bypass Perfusion, November 1, 2008; 23(6): 347 - 353. [Abstract] [PDF] |
||||
![]() |
R. W. Issitt, J. W. Mulholland, M. D. Oliver, G. J. Yarham, P. J. Borra, P. Morrison, I. Dimarakis, and J. R. Anderson Aortic Surgery Using Total Miniaturized Cardiopulmonary Bypass Ann. Thorac. Surg., August 1, 2008; 86(2): 627 - 631. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Lynch and J. Riley Microemboli detection on extracorporeal bypass circuitsa Perfusion, January 1, 2008; 23(1): 23 - 32. [Abstract] [PDF] |
||||
![]() |
J. W. Hammon Extracorporeal Circulation: Perfusion System Card. Surg. Adult, January 1, 2008; 3(2008): 350 - 370. [Full Text] |
||||
![]() |
B. Kiaii, D. Bainbridge, and P. Fernandes Surgical, Anesthetic, Perfusion-Related Advances in Minimal Access Surgery Seminars in Cardiothoracic and Vascular Anesthesia, December 1, 2007; 11(4): 282 - 287. [Abstract] [PDF] |
||||
![]() |
J.W. Mulholland, J.R. Anderson, G.J. Yarham, S. Tuladhur, I. Saed, and M.D. Oliver Miniature cardiopulmonary bypass the Hammersmith experience Perfusion, May 1, 2007; 22(3): 161 - 166. [Abstract] [PDF] |
||||
![]() |
W. Riley, D. FitzGerald, and L. Cohn Single, percutaneous, femoral venous cannulation for cardiopulmonary bypass Perfusion, May 1, 2007; 22(3): 211 - 215. [Abstract] [PDF] |
||||
![]() |
K. Yoshitani, F. de Lange, Q. Ma, H. P. Grocott, and G. B. Mackensen Reduction in Air Bubble Size Using Perfluorocarbons During Cardiopulmonary Bypass in the Rat Anesth. Analg., November 1, 2006; 103(5): 1089 - 1093. [Abstract] [Full Text] [PDF] |
||||
![]() |
J W Mulholland The Great Britain and Ireland perspective: current perfusion safety issues, preparing for the future Perfusion, July 1, 2005; 20(4): 217 - 225. [Abstract] [PDF] |
||||
![]() |
M. Kurusz and B. D Butler Bubbles and bypass: an update Perfusion, January 1, 2004; 19(1_suppl): S49 - S55. [Abstract] [PDF] |
||||
![]() |
M. Schoenburg, B. Kraus, A. Muehling, U. Taborski, H. Hofmann, G. Erhardt, S. Hein, M. Roth, P. R. Vogt, G. F. Karliczek, et al. The dynamic air bubble trap reduces cerebral microembolism during cardiopulmonary bypass J. Thorac. Cardiovasc. Surg., November 1, 2003; 126(5): 1455 - 1460. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A Sobieski II, M. S Slaughter, D. E Hart, P. S Pappas, and A. J Tatooles Peripheral cardiopulmonary bypass with modified assisted venous drainage and transthoracic aortic crossclamp: optimal management for robotic mitral valve repair Perfusion, September 1, 2003; 18(5): 307 - 311. [Abstract] [PDF] |
||||
![]() |
I. Shen, C. Giacomuzzi, and R. M. Ungerleider Current strategies for optimizing the use of cardiopulmonary bypass in neonates and infants Ann. Thorac. Surg., February 1, 2003; 75(2): S729 - 734. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. A. Hessel II and L. H. Edmunds Jr. Extracorporeal Circulation: Perfusion Systems Card. Surg. Adult, January 1, 2003; 2(2003): 317 - 338. [Full Text] |
||||
![]() |
F. Merkle, W. Bottcher, and R. Hetzer Prebypass filtration of cardiopulmonary bypass circuits: an outdated technique? Perfusion, January 1, 2003; 18(1_suppl): 81 - 88. [Abstract] [PDF] |
||||
![]() |
T. W. Willcox and R. van Uden Best Practice for Cardiopulmonary Bypass in the High-Risk Elderly Patient Seminars in Cardiothoracic and Vascular Anesthesia, December 1, 2002; 6(4): 293 - 300. [Abstract] [PDF] |
||||
![]() |
T. J. Jones, D. D. Deal, J. C. Vernon, N. Blackburn, and D. A. Stump Does vacuum-assisted venous drainage increase gaseous microemboli during cardiopulmonary bypass? Ann. Thorac. Surg., December 1, 2002; 74(6): 2132 - 2137. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J Forest, R. C Groom, R. Quinn, J. Donnelly, and C. Clark Repair of hypoplastic left heart syndrome of a 4.25-kg Jehovah's witness Perfusion, May 1, 2002; 17(3): 221 - 225. [Abstract] [PDF] |
||||
![]() |
L. Aklog, J. Sepic, F. Filsoufi, J. G. Byrne, and D. H. Adams Open inferior vena caval anastomosis during bicaval heart transplantation Ann. Thorac. Surg., February 1, 2002; 73(2): 671 - 672. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Bevilacqua, S. Matteucci, M. Ferrarini, M. Kacila, A. Ripoli, A. Baroni, D. Mercogliano, M. Glauber, and P. Ferrazzi Biochemical evaluation of vacuum-assisted venous drainage: a randomized, prospective study Perfusion, January 1, 2002; 17(1): 57 - 61. [Abstract] [PDF] |
||||
![]() |
M. Jahangiri, A. Rayner, B. Keogh, and C. Lincoln Cerebrovascular accident after vacuum-assisted venous drainage in a Fontan patient: a cautionary tale Ann. Thorac. Surg., November 1, 2001; 72(5): 1727 - 1728. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nakanishi, T. Shichijo, Y. Shinkawa, S. Takeuchi, M. Nakai, G. Kato, and O. Oba Usefulness of vacuum-assisted cardiopulmonary bypass circuit for pediatric open-heart surgery in reducing homologous blood transfusion Eur. J. Cardiothorac. Surg., August 1, 2001; 20(2): 233 - 238. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Borger, C. M. Peniston, R. D. Weisel, M. Vasiliou, R. E. A. Green, and C. M. Feindel Neuropsychologic impairment after coronary bypass surgery: Effect of gaseous microemboli during perfusionist interventions J. Thorac. Cardiovasc. Surg., April 1, 2001; 121(4): 743 - 749. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. LaPietra, E. A. Grossi, B. B. Pua, R. A. Esposito, A. C. Galloway, C. C. Derivaux, L. R. Glassman, A. T. Culliford, G. H. Ribakove, and S. B. Colvin Assisted venous drainage presents the risk of undetected air microembolism J. Thorac. Cardiovasc. Surg., November 1, 2000; 120(5): 856 - 863. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Mitchell, T. Willcox, F. Paget Milsom, and D. F. Gorman Physical and Pharmacological Neuroprotection in Cardiac Surgery Seminars in Cardiothoracic and Vascular Anesthesia, July 1, 2000; 4(2): 80 - 85. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |