|
|
||||||||
Ann Thorac Surg 1998;65:1631-1633
© 1998 The Society of Thoracic Surgeons
a Department of Cardiac Surgery, Medical University of Lübeck, Lübeck, Germany
Accepted for publication January 23, 1998.
Address reprint requests to Dr Bartels, Klinik für Herzchirurgie, Medical University of Lübeck, Ratzeburger Allee 160, 23538 Lübeck, Germany
| Abstract |
|---|
|
|
|---|
Methods. Sizer and valve dimensions of 22 different mechanical aortic prostheses (19 to 23 mm) were evaluated by caliper micrometer measurements.
Results. Nearly all sizers exceeded their marked dimensions by up to 1.0 mm. Measured tissue annulus diameters for 19-mm-labeled valves varied between 18.3 and 19.6 mm, for 21-mm valves from 20.5 to 21.6 mm, and for 23-mm valves from 22.4 to 23.5 mm, respectively. The orifice areas ranged from 1.5 to 2.06 cm2 for 19-mm valves, from 2.0 to 2.55 cm2 for 21-mm valves, and from 2.4 to 3.09 cm2 for 23-mm valves, respectively.
Conclusions. Actual sizer dimensions and tissue annulus diameters of various small mechanical aortic prostheses varied considerably from their marked diameters. These differences should be considered to ensure the optimal prosthesis selection for each patient.
| Introduction |
|---|
|
|
|---|
However, discrepancies between internal and tissue annulus diameters of different valves as well as diameter differences in sizing equipment make small mechanical aortic prostheses hardly comparable. Therefore, these frequently overlooked discrepancies might result in the selection of suboptimal prostheses or unnecessary aortic root enlargements.
The purpose of our study was to compare differences in sizer dimensions with tissue annulus diameters and orifice areas of various small mechanical aortic prostheses.
| Material and methods |
|---|
|
|
|---|
|
|
| Results |
|---|
|
|
|---|
|
|
|
21-mm- to 23-mm-labeled valves
Results for 21-mm and 23-mm prostheses were similar to those obtained for 19-mm valves. Smaller tissue annulus diameters were found for the St. Jude Medical Standard valves. Identical tissue annulus diameters were assessed for all CarboMedics prostheses and St. Jude Medical HP 21-mm valve. The measured tissue annulus diameters for 21-mm and 23-mm Baxter Edwards Tekna, 23-mm St. Jude Medical HP and Medtronic Hall valves were larger than indicated (see Tables 3 and 4 for details).
Regarding the comparison of measured sizer diameters versus measured tissue annulus diameters, our data indicate that in most cases the measured tissue annulus diameter did not correspond to the appropriate sizers as stated by the manufacturer. Actual sizer dimensions were both larger and smaller than the tissue annulus diameters of the corresponding valves due to incorrect labeling.
Orifice area
Orifice areas also varied considerably among the different studied artificial aortic valves indicated to have the same size. Modification of the external valve design (eg, reduced sewing ring) resulted in reduced tissue annulus diameters by constant orifice areas (Carbomedics Standard versus Carbomedics Reduced; Medtronic Hall 20-mm and 22-mm versus Medtronic Hall 21-mm and 23-mm). The largest orifice area stated for each valve size could be assessed in the St. Jude Medical HP series.
For 19-mm prostheses the orifice areas varied between 1.5 and 2.06 cm2. Orifice areas for 21-mm valves varied from 2.0 to 2.55 cm2. For 23-mm mechanical valves, the orifice areas varied from 2.4 to 3.09 cm2 (see Tables 2 to 4).
| Comment |
|---|
|
|
|---|
Our results are in agreement with Cochran and Kunzelman [3], who reported on discrepancies between labeled and actual dimensions of various aortic and mitral prosthetic valves and sizers. A larger number of different prostheses than in the study by Cochran and Kunzelman, including two supraannular prostheses, were evaluated in the present study. The present investigation also indicates different orifice areas of the studied prostheses. However, the significance of different calculated orifice areas of various prostheses (bileaflet valves, tilting disk valves) on hydrodynamic performance has to be evaluated in vivo.
In the small aortic root, accurate measurement of annulus size is essential for optimal mechanical valve selection and to decide whether the aortic root has to be enlarged. More than 10 years ago cardiac surgeons indicated that sizers were labeled incorrectly [4]. However, to date manufacturers have not standardized sizing equipment dimensions.
We found that in some cases sizers exceed their indicated dimension (eg, a labeled 19-mm CarboMedics Standard sizer was measured to be 19.8 mm). This under-labeling seems to be necessary for CarboMedics Standard and Medtronic Hall 21-, 22-, and 23-mm prostheses to allow implantation, as the tissue annulus diameters of these valves are larger than the labeled prosthesis size. The sizer equipment for Baxter Edwards Tekna valves is correctly labeled; however, the tissue annulus diameter of these valves are larger (0.5 mm) than the sizer dimensions and the stated valve dimensions. The variation between the sizer and tissue annulus diameter with the Baxter Edwards Tekna prostheses should be considered when implanting Baxter Edwards Tekna aortic valves. Moreover, for other valves (St. Jude Medical Standard, Medtronic Hall 20 mm), sizers are larger than the tissue annulus diameters, thus preventing optimal prosthesis/aortic root relation. Our data confirm the results of Walther and colleagues [5] who studied sizer dimensions of conventional and stentless aortic prostheses and observed significant differences between marked and actual sizer dimensions.
The use of small mechanical aortic prostheses raises concern about their potentially harmful effect on residual left ventricular outflow obstruction. The pressure gradient at high flow rates across a small aortic prosthesis is mainly determined by its internal diameter and its orifice area. Recently, González-Juanatey and colleagues [2] studied the influence of the size of aortic prostheses on hemodynamics and change in left ventricular mass using echocardiography. They concluded that 19-mm aortic prostheses create significant left ventricular outflow obstruction. Butterfield and coworkers [6] demonstrated that the pressure drop at high flow across 19-mm standard mechanical and bioprosthetic aortic prostheses are twice as high as across 21-mm valves. Similar results have been obtained from in vitro studies examining the hydrodynamic function of small mechanical aortic prostheses [7]. In their study, lower forward flow pressure drops were observed for the high-performance St. Jude Medical prostheses as compared to standard mechanical valves [7]. This favorable hydrodynamic performance was attributable to the larger orifice areas of supraannular valves. However, even with advances in valve design (eg, supraannular positioning, reduced sewing rings) resulting in increased orifice areas the optimal mechanical valve cannot be selected unless the true valve dimensions are considered.
The following conclusions can be drawn from our findings: (1) actual sizer dimensions and tissue annulus diameters of various small mechanical aortic prostheses vary considerably from their marked diameters. (2) these differences should be considered to ensure the optimal prosthesis selection. (3) standardized sizer equipment and standardized tissue annulus diameters are hardly needed to optimize prosthesis/patient aortic ring relation.
| Footnotes |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. A. Youdelman, H. Hirose, H. Jain, J. Y. Kresh, J. W.C. Entwistle III, and A. S. Wechsler Comparison of eight prosthetic aortic valves in a cadaver model. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1526 - 1532. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Taggart Prosthesis patient mismatch in aortic valve replacement: possible but pertinent? Eur. Heart J., March 2, 2006; 27(6): 644 - 646. [Full Text] [PDF] |
||||
![]() |
M. Turina Supra-annular aortic valve replacement with a mechanical prosthesis MMCTS, November 29, 2005; 2005(1129): 646. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-H. Sievers Prosthetic aortic valve replacement J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 961 - 965. [Full Text] [PDF] |
||||
![]() |
T. Walther, S. Lehmann, V. Falk, S. Metz, N. Doll, A. Rastan, M. Viehweg, M. Richter, J. Gummert, and F. W. Mohr Prospectively Randomized Evaluation of Stented Xenograft Hemodynamic Function in the Aortic Position Circulation, September 14, 2004; 110(11_suppl_1): II-74 - II-78. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. D. Zuckerman and W. Sapirstein Mismatching of annular sizers and cardiac prosthetic valve sewing cuff sizes Ann. Thorac. Surg., July 1, 2003; 76(1): 337 - 337. [Full Text] [PDF] |
||||
![]() |
B. D. Zuckerman and W. Sapirstein Prosthetic cardiac valve replacement J. Thorac. Cardiovasc. Surg., June 1, 2003; 125(6): 1563 - 1563. [Full Text] [PDF] |
||||
![]() |
M. B. Izzat, I. Kadir, B. Reeves, P. Wilde, A. J. Bryan, and G. D. Angelini Patient-prosthesis mismatch is negligible with modern small-size aortic valve prostheses Ann. Thorac. Surg., November 1, 1999; 68(5): 1657 - 1660. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Bartels and H.-H. Sievers Reply Ann. Thorac. Surg., September 1, 1998; 66(3): 985 - 985. [Full Text] [PDF] |
||||
![]() |
L. I. Bonchek Why aren't valve sizers and prostheses labeled accurately? Ann. Thorac. Surg., September 1, 1998; 66(3): 984 - 985. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |