|
|
||||||||
Ann Thorac Surg 1998;65:643-646
© 1998 The Society of Thoracic Surgeons
Infinity Heart Institute, St Lukes Medical Center, Milwaukee, Wisconsin, USA,
Thermo Cardiosystems, Inc, Woburn, Massachusetts, USA
Accepted for publication August 27, 1997.
Dr Tector, 2901 W Kinnickinnic River Parkway, Suite 511, Milwaukee, WI 53215.
| Abstract |
|---|
|
|
|---|
Methods. From January 1994 through July 1996, we used an extracorporeal membrane oxygenation right atrial-to-left ventricular shunt during 17 HeartMate implantations in 16 patients. The shunt consists of the existing right atrial two-stage cannula, the bypass circuit, and a separate aortic line that fills the left ventricle using a 21F cannula in the lateral ventricular wall. Air is monitored in the heart and aorta using transesophageal echocardiography.
Results. Ten of the 16 patients are living and 8 have undergone transplantation. Two patients are still using the device and are awaiting transplantation. None of the patients have experienced postoperative neurologic events suggestive of air emboli.
Conclusions. The extracorporeal membrane oxygenation right atrial-to-left ventricular shunt is simple and inexpensive to construct. It provides for a smoother and safer transition from cardiopulmonary bypass to the HeartMate left ventricular assist device.
| Introduction |
|---|
|
|
|---|
The process of deairing is tedious and can take more than 15 minutes. If the volume of blood in the left ventricular reservoir becomes too low during or after weaning of the patient from CPB, air can be aspirated into the system through the connectors, the needle holes, and the interstices in the Dacron grafts. When this occurs, the entire deairing procedure has to be repeated. The volume of blood in the left ventricular reservoir is totally dependent on the output of the right ventricle. Although assisting the patient with partial CPB during the transition from CPB to the HeartMate maintains satisfactory systemic perfusion, it offers minimal support to the right ventricle. When the flow to the left ventricle is less than 3 L/min, air can be aspirated into the system even while it is being supported partially with CPB.
We have developed an extracorporeal membrane oxygenation (ECMO) right atrial-to-left ventricular shunt that quickly purges all air from the system and prevents its reentry by sustaining a safe level of blood in the left ventricular reservoir. The shunt also has the ability to assist the right ventricle in supplying enough blood to the HeartMate so that it immediately can establish and maintain a normal cardiac output with well-oxygenated blood while the right ventricle is recovering. The circuit uses a two-stage right atrial cannula (the same cannula that is used for venous return in the CPB circuit) that returns venous blood from the body to the oxygenator. The oxygenated blood is pumped directly into the left ventricle through a cannula placed into the chamber through the wall of the ventricle.
This report describes the design and application of the ECMO right atrial-to-left ventricular shunt in patients undergoing implantation of the HeartMate pump.
| Material and Methods |
|---|
|
|
|---|
| How Can Air Be Aspirated Into the System? |
|---|
|
|
|---|
How the Shunt Works
The ECMO right atrial-to-left ventricular shunt removes venous blood from the right atrium through the two-stage venous cannula that was used for CPB. The venous blood passes through the oxygenator and is oxygenated. Then it is pumped into the arterial line through the arterial filter. Just past the filter, there is a Y-connector that creates two arterial lines. A flow probe is inserted into each arterial line, enabling the perfusionist to adjust accurately the flow in each line. One of the arterial lines is connected to the aortic cannula for CPB. The other arterial line is connected to the cannula that is inserted into the left ventricle, completing the ECMO shunt circuit (Fig 1). The ECMO shunt then can pump blood into the left ventricle through the HeartMate pump and exit the outflow port of the pump to purge the system of air rapidly and effectively. In addition, the shunt supplies the left ventricular reservoir (the collecting chamber for the HeartMate pump) with the amount of oxygenated blood necessary for the LVAD to maintain normal cardiac output without CPB. This eliminates the possibility of low volume in the left ventricular reservoir that can occur during weaning of the patient from CPB and create the potential for aspiration of air into the system.
|
After CPB has been initiated, a circular piece of muscle is removed from the apex of the left ventricle using a coring knife. The apical sewing ring of the inflow connector is sutured into the orifice created in the left ventricular apex. A descending arch 21F cannula (Medtronic, Anaheim, CA) is inserted into the left ventricle between the diagonal and left anterior descending coronary arteries proximal to the apical inflow cannulation site. The shunt cannula is placed similar to the way one would place a left ventricular vent and its position in the left ventricle is checked by palpating the tip of the cannula through the inflow connector. It is secured with two pursestring sutures. The driveline in the pneumatic pump or the electrical and vent lines in the vented electric pump are exited at the most optimal positions in the abdominal wall. The inflow conduit of the pump is inserted into the left ventricle through the inflow connector and is secured with multiple heavy, braided ligatures.
Before the outflow graft of the HeartMate blood pump is connected, air can be purged rapidly from the left ventricle and the HeartMate pump in 1 to 2 minutes. The shunt is started at a flow rate of 1 L/min, the outflow port of the pump is elevated, and the lungs are ventilated, allowing blood and air trapped in the system to flow through the pump and be discharged out of the outflow port into the patients mediastinum, where it can be suctioned and recirculated into the bypass circuit. The pump housing is tapped with an instrument to dislodge any air that is adherent to its walls. The hand crank of the pneumatic pump is turned several times or the hand pump of the vented electric device is pumped several times to move the pusher plate and dislodge any air that is attached to the diaphragm. All air must be removed from the patients heart, the HeartMate pump, the outflow graft, and the ascending aorta before the HeartMate LVAD can be started safely. Transesophageal echocardiography is used to demonstrate that all air has been removed from the cardiac chambers and the ascending aorta. Of equal importance, transesophageal echocardiography immediately detects the entry of air into the heart or ascending aorta during the transition from CPB to the HeartMate LVAD [3]. If air is seen, the surgeon can turn off the HeartMate immediately and restart CPB. After the air is removed, the transition from CPB to the HeartMate LVAD is restarted.
The outflow graft is flushed by releasing the clamp and the graft is connected to the HeartMate. The surgeon should make certain that the aortic vent is working properly. A needle is placed into the outflow graft at the highest point and the hand crank or the hand pump is pumped to remove any residual air at the outflow graft pump connection site. If air is not seen, the shunt flow is increased to 4 L/min, the HeartMate is turned on, and CPB is discontinued. Extracorporeal membrane oxygenation shunt flow is set to maintain at least a 5-L/min flow rate by the HeartMate pump. As the right ventricle recovers and increases its flow to the left ventricle, the shunt flow is decreased appropriately, maintaining a HeartMate output of 4 to 5 L/min, depending on the size of the patient.
Gradually reducing the ECMO shunt flow as the right ventricle recovers removes both the risk of having too low a reservoir of blood in the left ventricle and the potential for aspirating air. In addition, the need for large doses of inotropic medications to maintain an adequate reservoir in the left ventricle in patients who have a slowly recovering right ventricle is reduced significantly when the ECMO shunt is used. Adjusting the shunt flow of blood to the left ventricular reservoir to the minimum level necessary to maintain the output of the HeartMate pump at 5 L/min prevents distention of the left ventricle. Further, the aortic valve acts as a relief when the left ventricle is selected as the shunt site. The shunt flow is decreased gradually as the right ventricle increases its output. When the right ventricle can supply the left ventricle completely, the ECMO shunt is discontinued. If the patient needs circulating volume, it can be infused from the CPB into the ascending aorta.
We recently have begun to use inhaled nitric oxide, beginning with the highest concentration tolerated (usually 80 ppm), to reduce pulmonary vascular resistance to its lowest level in an attempt to prevent right ventricular failure [4]. Nitric oxide has been shown to diminish pulmonary vascular resistance significantly without affecting systemic vascular resistance [4][5].
| Results |
|---|
|
|
|---|
| Comment |
|---|
|
|
|---|
During the transition from CPB to the HeartMate, when the ECMO right atrial-to-left ventricular shunt flow is set at 1 to 2 L/min, all the existing air is purged quickly from the left ventricle and the HeartMate pump. When the HeartMate is started without ECMO, the right ventricle often is unable to supply enough blood to the HeartMate, resulting in a low cardiac output. When this situation becomes more extreme and flow to the left ventricular reservoir is less than 3 L/min, air can be aspirated into the system through the connectors, needle holes, and interstices of the graft. As a consequence, air emboli can produce strokes or obstruct the right coronary circulation and further diminish the output of the right ventricle.
The ECMO shunt assists the depressed right ventricle, allowing immediate conversion from CPB to the HeartMate with a 5-L/min cardiac output and perfusion of the body with blood that is well oxygenated. The fear of aspirating air into the circulation is eliminated. A normal cardiac output encourages diuresis and the removal of excess lung water, promoting a reduction in pulmonary vascular resistance, the major cause of right ventricular failure. The need to place a cannula in the left ventricle and to have an extra arterial line and flow probe are the main drawbacks of the ECMO shunt.
We have selected the left ventricle, the reservoir of the HeartMate pump, as the site for perfusion of oxygenated blood from the ECMO system. It is the safest chamber for infusion of the blood; if the left ventricle ever became overdistended, the aortic valve would act as a relief valve. Other possible cannulation sites might be the left atrium or the ascending aorta and placement of the cannula through the aortic valve into the left ventricle. We caution against infusing shunt flow into the left atrium because there is a risk of left atrial overload, and when the left atrial pressure is raised, it is transmitted directly to the pulmonary veins, alveolar capillaries, and pulmonary arteries [6]. If the aortic valve were rendered incompetent by the shunt cannula, the efficiency of the HeartMate could be decreased. Besides being the safest site for the shunt, the left ventricle is accessible and shunt placement and removal are easy and similar to the placement of a left ventricular vent. There have been no adverse events such as bleeding, arrhythmias, or aspiration of air associated with the shunt.
All patients who require LVAD implantation have increased pulmonary vascular resistance and right ventricular insufficiency. However, the degree is difficult to predict before operation. After CPB for LVAD implantation, lung edema [7] and the release of a constrictive prostanoid substance [8] can increase pulmonary vascular resistance. Often, the output of the LVAD is low because the amount of blood supplied to the left ventricular reservoir by the right ventricle is insufficient after weaning from CPB. This reduces systemic perfusion and can inhibit diuresis, preventing the removal of lung edema. The stress on the right ventricle multiplies and, sometimes, in spite of large doses of inotropic drugs, the right ventricle fails and a right ventricular assist device has to be implanted. Although CPB can assist the peripheral circulation adequately, it offers little support to the right ventricle during its recovery. The ECMO shunt supplies ample blood to the HeartMate by maintaining an adequate volume of blood in the left ventricular reservoir. A 5-L/min cardiac output is established immediately by the HeartMate pump for optimal perfusion of all the organs of the body, allowing the right ventricle to recover under these ideal conditions.
In conclusion, the ECMO right atrial-to-left ventricular shunt is simple and inexpensive to construct. It provides for a smoother and safer transition from CPB to the HeartMate pump by significantly reducing the chance of air embolism and providing optimal systemic perfusion while the right ventricle recovers.
| Footnotes |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. G. Licina, R. M. Savage, C. Hearn, and E. J. Kraenzler The Role of Transesophageal Echocardiography in Perfusion Management Seminars in Cardiothoracic and Vascular Anesthesia, November 1, 2001; 5(4): 321 - 334. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |