Department of Thoracic and Cardiovascular Surgery, JW Goethe University Hospital, Frankfurt, Germany
Accepted for publication January 21, 2009.
* Address correspondence to Dr Martens, Department of Thoracic and Cardiovascular Surgery, JW Goethe University Hospital, Thoedor Stern Kai 7, Frankfurt, D-60590, Germany (Email: martens.herz{at}gmx.de).
Purpose: Important comorbid conditions in elderly patients referred for aortic valve replacement require alternative treatment options with possible reduction of the extracorporeal circuit time and reliable hemodynamic features. We report on clinical experiences with the sutureless, nitinol-stented 3f Enable (ATS Medical, Minneapolis, MN) aortic valve prosthesis in 32 patients.
Description: The procedure was performed using cardiopulmonary bypass with cardioplegic arrest. After resection of the stenotic aortic valve and debridement of the annulus, the valve was inserted and released. Mean age was 78 ± 3; mean logistic Euroscore was 13.7. Concomitant procedures were mitral valve and tricuspid valve repair (n = 1), coronary artery bypass graft (n = 9), and subvalvular myectomy (n = 3).
Evaluation: Implantation of the valve required 9 ± 5 minutes. Cardiopulmonary bypass and aortic cross-clamp time were 87 ± 16 and 55 ± 11 minutes for stand-alone procedures. Combined procedures required 126 ± 42 and 84 ± 28 minutes, respectively. Two patients were abandoned intraoperatively and converted to standard procedures due to misalignment of the valve. In the other 30 patients, no paravalvular leakage was detected. The transvalvular gradient at discharge was 9 ± 6 mm Hg (mean) and 18 ± 9 mm Hg (peak). Six months after surgery, gradients were 10 ± 4 mm Hg (mean) and 18 ± 6 mm Hg (peak).
Conclusions: Sutureless valve implantation is feasible and safe with the 3f Enable (ATS Medical) bioprosthesis. Reduction of cardiopulmonary bypass and aortic cross-clamp time seems to be possible with increasing experience. Hemodynamic data are promising with low gradients at discharge and after 6 months.
Related Article
Ann. Thorac. Surg. 2009 87: 1917-1918.
This article has been cited by other articles:
![]() |
M. Kalejs, R. Lacis, V. Kasyanov, I. Ozolanta, P. Abdel Sayed, P. Stradins, and L. K. von Segesser Comparison of radial deformability of stent posts of different aortic bioprostheses Interact CardioVasc Thorac Surg, February 1, 2013; 16(2): 129 - 133. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Concistre, A. Miceli, F. Chiaramonti, and M. Glauber Delayed dislocation of a sutureless aortic bioprosthesis: the first case Interact CardioVasc Thorac Surg, June 1, 2012; 14(6): 892 - 893. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Folliguet, F. Laborde, K. Zannis, G. Ghorayeb, A. Haverich, and M. Shrestha Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers Ann. Thorac. Surg., May 1, 2012; 93(5): 1483 - 1488. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. H. Sepehripour, L. Harling, and T. Athanasiou What are the current results of sutureless valves in high-risk aortic valve disease patients? Interact CardioVasc Thorac Surg, May 1, 2012; 14(5): 615 - 621. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Concistre, P. Farneti, A. Miceli, and M. Glauber Sutureless aortic bioprosthesis in severe aortic root calcification: an innovative approach Interact CardioVasc Thorac Surg, May 1, 2012; 14(5): 670 - 672. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Risteski, C. Adami, N. Papadopoulos, A. S. Sirat, A. Moritz, and M. Doss Leaflet Replacement for Aortic Stenosis Using the 3f Stentless Aortic Bioprosthesis: Midterm Results Ann. Thorac. Surg., April 1, 2012; 93(4): 1134 - 1140. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Flameng, M.-C. Herregods, H. Hermans, G. Van der Mieren, M. Vercalsteren, G. Poortmans, J. Van Hemelrijck, and B. Meuris Effect of sutureless implantation of the Perceval S aortic valve bioprosthesis on intraoperative and early postoperative outcomes J. Thorac. Cardiovasc. Surg., December 1, 2011; 142(6): 1453 - 1457. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Berreklouw, B. Koene, F. De Somer, S. Bouchez, K. Chiers, Y. Taeymans, and G. J. Van Nooten Sutureless replacement of aortic valves with St Jude Medical mechanical valve prostheses and Nitinol attachment rings: Feasibility in long-term (90-day) pig experiments J. Thorac. Cardiovasc. Surg., May 1, 2011; 141(5): 1231 - 1237.e1. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Garatti, F. Mori, F. Innocente, A. Canziani, P. Gagliardotto, E. Mossuto, T. Santoro, V. Montericcio, A. Frigiola, and L. Menicanti Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient-Prosthesis Mismatch a Relevant Phenomenon? Ann. Thorac. Surg., January 1, 2011; 91(1): 71 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Breitenbach, G. Wimmer-Greinecker, L. A. Bockeria, J. Sadowski, C. Schmitz, B. Kapelak, K. Bartus, R. Muratov, and W. Harringer Sutureless aortic valve replacement with the Trilogy Aortic Valve System: Multicenter experience J. Thorac. Cardiovasc. Surg., October 1, 2010; 140(4): 878 - 884.e1. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Aymard, A. Kadner, N. Walpoth, V. Gober, L. Englberger, M. Stalder, F. Eckstein, C. Zobrist, and T. Carrel Clinical experience with the second-generation 3f Enable sutureless aortic valve prosthesis J. Thorac. Cardiovasc. Surg., August 1, 2010; 140(2): 313 - 316. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Magovern Jr Invited Commentary Ann. Thorac. Surg., June 1, 2009; 87(6): 1917 - 1918. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |