Background: Bioprostheses for supraannular placement have been developed to optimize the hemodynamic performance after aortic valve replacement. To evaluate the potential benefit of this design, we analyzed the influence of prosthesis–patient mismatch on diastolic function and left ventricular mass regression and evaluated the clinical performance of the Sorin Soprano and Medtronic Mosaic in the aortic position.
Methods: A total of 372 patients underwent aortic valve replacement between July 2004 and February 2007, receiving either a Sorin Soprano (n = 235) or a Medtronic Mosaic (n = 137) prosthetic valve. Echocardiographic and clinical data were collected prospectively, and follow-up was performed in April 2007. Multivariate analyses were used to identify differences in hemodynamic performance, diastolic function, left ventricular mass regression, and predictors of impaired survival. Kaplan–Meier survival curves and log-rank tests were used to compare postoperative outcomes.
Results: The 30-day mortality was 1.7% (4 of 235 patients) in the Sorin Soprano group and 2.9% (4 of 137 patients) in the Medtronic Mosaic group (p = 0.473). Neither prosthesis–patient mismatch nor type of prosthesis was a significant predictor of early or late mortality. Diastolic heart failure was a predictor of poor survival (p = 0.004); however, the recovery of diastolic function was not significantly influenced by prosthesis–patient mismatch. Neither moderate (indexed effective orifice area < 0.85 cm2/m2) nor severe (indexed effective orifice area < 0.65 cm2/m2) prosthesis–patient mismatch resulted in a significantly impaired left ventricular mass regression.
Conclusions: Prosthesis-patient mismatch was not an independent predictor of poor survival, impaired left ventricular mass regression, or recovery of diastolic function. The Sorin Soprano and the Medtronic Mosaic bioprostheses demonstrated comparable hemodynamic performance and excellent clinical outcome without signs of structural valve deterioration during follow-up.
Related Article
Ann. Thorac. Surg. 2008 85: 1317-1318.
This article has been cited by other articles:
![]() |
H. Jilaihawi, D. Chin, T. Spyt, M. Jeilan, M. Vasa-Nicotera, J. Bence, E. Logtens, and J. Kovac Prosthesis-patient mismatch after transcatheter aortic valve implantation with the Medtronic-Corevalve bioprosthesis Eur. Heart J., April 1, 2010; 31(7): 857 - 864. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Vohra, R. N. Whistance, M. Bolgeri, T. Velissaris, G. M. K. Tsang, C. W. Barlow, and S. K. Ohri Mid-term evaluation of Sorin Soprano bioprostheses in patients with a small aortic annulus <=20 mm Interact CardioVasc Thorac Surg, March 1, 2010; 10(3): 399 - 402. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. J. Howell, B. E. Keogh, D. Ray, R. S. Bonser, T. R. Graham, J. Mascaro, S. J. Rooney, I. C. Wilson, and D. Pagano Patient-Prosthesis Mismatch in Patients With Aortic Stenosis Undergoing Isolated Aortic Valve Replacement Does Not Affect Survival Ann. Thorac. Surg., January 1, 2010; 89(1): 60 - 64. [Abstract] [Full Text] [PDF] |
||||
![]() |
F.-C. Riess, E. Cramer, L. Hansen, S. Schiffelers, G. Wahl, J. Wallrath, S. Winkel, and P. Kremer Clinical results of the Medtronic Mosaic porcine bioprosthesis up to 13 years Eur J Cardiothorac Surg, January 1, 2010; 37(1): 145 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Ruzicka, I. Hettich, A. Hutter, S. Bleiziffer, C. C. Badiu, R. Bauernschmitt, R. Lange, and W. B. Eichinger The Complete Supraannular Concept: In Vivo Hemodynamics of Bovine and Porcine Aortic Bioprostheses Circulation, September 15, 2009; 120(11_suppl_1): S139 - S145. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. Bockeria, I. I. Skopin, I. M. Tsiskaridze, and D. V. Murysova eComment: Independent risk factors of in-hospital mortality in patients undergoing aortic valve replacement Interact CardioVasc Thorac Surg, September 1, 2009; 9(3): 518 - 519. [Full Text] [PDF] |
||||
![]() |
V. Rao Invited Commentary Ann. Thorac. Surg., April 1, 2008; 85(4): 1317 - 1318. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |