Ann Thorac Surg 2007;83:586-591
© 2007 The Society of Thoracic Surgeons
a Department of Cardiac Surgery, San Raffaele Hospital, Milan, Italy
b Department of Cardiac Anesthesiology, San Raffaele Hospital, Milan, Italy
Accepted for publication September 11, 2006.
* Address correspondence to Dr Verzini, Department of Cardiac Surgery, San Raffaele Hospital, Via Olgettina, 60 20100 Milan, Italy (Email: verzini.alessandro{at}hsr.it).
BACKGROUND: The purpose of this study was to evaluate the safety and clinical results of aortic valve replacement performed with minimally invasive closed circuit extracorporeal circulation technique (MECC system) versus standard cardiopulmonary bypass.
METHODS: Forty consecutive patients undergoing isolated aortic valve replacement at a single institution were randomly assigned to either miniaturized closed circuit cardiopulmonary bypass with the Maquet-Cardiopulmonary (Rastatt, Germany) minimal extracorporeal circulation (MECC) system (study group B, n = 17) or standard cardiopulmonary bypass (control group A, n = 23). The MECC system is a low priming circuit without blood-air interface. Technical feasibility, in particular the potential entry of air in the circuit, and clinical results were prospectively evaluated.
RESULTS: Demographic characteristics and surgical data were similar in both groups. Patients in the study group showed reduced chest tube drainage (217 ± 62 mL vs 420 ± 219 mL, p < 0.05) and blood transfusion requirements (5.1% vs 43.4%, p < 0.02) compared with patients in the control group. Moreover, the study group showed significantly higher time course of hematocrit at all time points during the operation and longer hospital stay (p < 0.02) than the control group; similarly, in the study group patients platelet count in intensive care unit admission was significantly higher than the control group (140 ± 29 x 109/L vs 119 ± 37 x 109/L, p < 0.05). Peak postoperative troponin C release was significantly lower in the study group (4.74 ± 2.82 vs 8.43 ± 6.25 ng/dL, p < 0.033). One patient undergoing the MECC system operation showed a major neurologic event on postoperative day four, which was probably secondary to severe aortic calcification.
CONCLUSIONS: The MECC system is suitable for aortic valve replacement and provides better clinical results than standard cardiopulmonary bypass as regards blood transfusion requirements, platelets consumption, and myocardial damage.
This article has been cited by other articles:
![]() |
G. F.V. Panday, S. Fischer, A. Bauer, D. Metz, J. Schubel, N. E. Shouki, T. Eberle, and H. Hausmann Minimal extracorporeal circulation and off-pump compared to conventional cardiopulmonary bypass in coronary surgery Interactive CardioVascular and Thoracic Surgery, November 1, 2009; 9(5): 832 - 836. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Colli, C. Fernandez, L. Delgado, B. Romero, M. L. Camara, and X. Ruyra Aortic valve replacement with minimal extracorporeal circulation versus standard cardiopulmonary bypass Interactive CardioVascular and Thoracic Surgery, October 1, 2009; 9(4): 583 - 587. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. P. Simons, K. D. Reesink, M. D. Lance, F. H. van der Veen, D. S. de Jong, P. W. Weerwind, and J. G. Maessen Dynamic filling index: a novel parameter to monitor circulatory filling during minimized extracorporeal bypass Eur. J. Cardiothorac. Surg., August 1, 2009; 36(2): 330 - 334. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Biancari and R Rimpilainen Meta-analysis of randomised trials comparing the effectiveness of miniaturised versus conventional cardiopulmonary bypass in adult cardiac surgery Heart, June 15, 2009; 95(12): 964 - 969. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Formica, F. Broccolo, A. Martino, J. Sciucchetti, V. Giordano, L. Avalli, G. Radaelli, O. Ferro, F. Corti, C. Cocuzza, et al. Myocardial revascularization with miniaturized extracorporeal circulation versus off pump: Evaluation of systemic and myocardial inflammatory response in a prospective randomized study. J. Thorac. Cardiovasc. Surg., May 1, 2009; 137(5): 1206 - 1212. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Fernandes, J. MacDonald, A. Cleland, R. Mayer, S. Fox, and B. Kiaii The use of a mini bypass circuit for minimally invasive mitral valve surgery Perfusion, May 1, 2009; 24(3): 163 - 168. [Abstract] [PDF] |
||||
![]() |
T. Puehler, A. Haneya, A. Philipp, K. Wiebe, A. Keyser, L. Rupprecht, S. Hirt, R. Kobuch, C. Diez, M. Hilker, et al. Minimal extracorporeal circulation: an alternative for on-pump and off-pump coronary revascularization. Ann. Thorac. Surg., March 1, 2009; 87(3): 766 - 772. [Abstract] [Full Text] [PDF] |
||||
![]() |
I Kutschka, J Skorpil, A El Essawi, T Hajek, and W Harringer Beneficial effects of modern perfusion concepts in aortic valve and aortic root surgery Perfusion, January 1, 2009; 24(1): 37 - 44. [Abstract] [PDF] |
||||
![]() |
I Dimarakis, D Stefanou, G Yarham, J Mulholland, and J Anderson Total miniaturized cardiopulmonary bypass: the next step in minimally invasive aortic valve replacement Perfusion, September 1, 2008; 23(5): 275 - 278. [Abstract] [PDF] |
||||
![]() |
N Colangelo, S Moriggia, T Kieser, F De Simone, A Vescovo, L Palumbo, P Denti, A Guidotti, A Blasio, F Maisano, et al. A cardioplegia circuit with versatility: the 'ReVerse' system. How to do it Perfusion, July 1, 2008; 23(4): 205 - 207. [Abstract] [PDF] |
||||
![]() |
R. H. Habib, A. Zacharias, and T. A. Schwann Minimally Invasive Closed Circuit Versus Standard Cardiopulmonary Bypass: Is It Renoprotective? Ann. Thorac. Surg., October 1, 2007; 84(4): 1426 - 1427. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |