Ann Thorac Surg 2006;82:2133-2138
© 2006 The Society of Thoracic Surgeons
Department of Cardiopulmonary Sciences, A.O. S Maria della Misericordia, Udine, Italy
Accepted for publication July 6, 2006.
* Address correspondence to Dr Pavoni, Cardiology Unit, Echo-lab, Department of Cardiopulmonary Sciences, A.O. S Maria della Misericordia, P. le S Maria della Misericordia 15, Udine 33100, Italy (Email: daisy_pav{at}libero.it).
BACKGROUND: The stented Soprano valve (Sorin Biomedica S.pA., Saluggia, Italy) is a pericardial tissue valve designed for a totally supra-annular implant with the aim of improving hemodynamic performance, particularly in patients with a reduced aortic annulus. However, its hemodynamics and early clinical outcome are poorly understood.
METHODS: Seventy-seven patients (mean age 76 ± 5 years, 56% males) underwent aortic valve replacement with the Soprano valve. All patients were monitored with clinical examination and serial echocardiography at 1, 6, and 12 months and yearly afterward.
RESULTS: At preoperative echocardiography, average left ventricular outflow tract diameter was 2.1 ± 0.2 cm. At operation, 35% of patients received a 20-mm valve, 54% a 22-mm valve, and 11% a 24-mm valve. At 6-month follow-up, peak and mean transprosthetic gradients were 18 ± 8 and 9 ± 4 mm Hg, respectively; effective orifice area (EOA) and EOA index were 1.84 ± 0.6 cm2 and 0.9 ± 0.2 cm2/m2, respectively. Incidence of patient-prosthesis mismatch (ie, EOA index < 0.85 cm2/m2) was 23%, with no case of severe mismatch (ie, EOA index < 0.6 cm2/m2). In addition, left ventricular hypertrophy showed a significant regression (mass index from 214 ± 98 g to 129 ± 41 g; p = 0.001), and ejection fraction increased (from 58% ± 17% to 67% ± 8%; p = 0.001). Cumulative follow-up was 7.9 months per patient. Thirty-day mortality rate was 2.6% (2 of 77). Cumulative survival at 6, 12, and 24 months was 92% ± 3.7%, 85% ± 5.7%, and 85% ± 5.7%, respectively. Two patients experienced embolic episodes. One case of anticoagulant-related bleeding occurred.
CONCLUSIONS: In our series, the Soprano bioprosthesis showed a good hemodynamic performance, low incidence of patient-prosthesis mismatch, and favorable early clinical results.
This article has been cited by other articles:
![]() |
S. Nozohoor, J. Nilsson, C. Luhrs, A. Roijer, and J. Sjogren Influence of Prosthesis-Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Ann. Thorac. Surg., April 1, 2008; 85(4): 1310 - 1317. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bottio and G. Gerosa Conflicting Echocardiographic Evaluations of Supra-Annular Tissue Valves: A Mystery Not Yet Resolved Ann. Thorac. Surg., August 1, 2007; 84(2): 713 - 714. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |