ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Joseph Forbess
Ajit P. Yoganathan
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jimenez, J. H.
Right arrow Articles by Yoganathan, A. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jimenez, J. H.
Right arrow Articles by Yoganathan, A. P.
Related Collections
Right arrow Valve disease

Ann Thorac Surg 2006;82:1362-1368
© 2006 The Society of Thoracic Surgeons


Original Articles: Cardiovascular

Effects of Annular Size, Transmitral Pressure, and Mitral Flow Rate on the Edge-To-Edge Repair: An In Vitro Study

Jorge H. Jimenez, MSa, Joseph Forbess, MDb, Laura R. Crofta, Lisa Smalla, Zhaoming He, PhDc, Ajit P. Yoganathan, PhDa,*

a Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, Georgia
b University of Texas Southwestern Medical Center, Lubbock, Texas
c Texas Tech University, Lubbock, Texas

Accepted for publication May 4, 2006.

* Address correspondence to Dr Yoganathan, Wallace H. Coulter School of Biomedical Engineering, Georgia Institute of Technology and Emory University, Room 2119, U.A. Whitaker Bldg, 313 Ferst Dr, Atlanta, GA 30332-0535 (Email: ajit.yoganathan{at}bme.gatech.edu).

BACKGROUND: Although edge-to-edge repair is an established adjunctive procedure, there is still debate on its long-term durability and efficacy.

METHODS: Fifteen porcine mitral valves were studied in a physiologic left heart simulator with a variable size annulus (dilated = 8.22 cm2, normal = 6.86 cm2, contracted = 5.5 cm2). Mitral valves were tested under steady and physiologic pulsatile flow conditions (cardiac outputs: 4 to 6 L/min), at peak transmitral pressures between 100 mm Hg and 140 mm Hg. A miniature force transducer was used to measure the Alfieri stitch force (FA). Mitral flow rate (MFR), transmitral pressure, effective orifice area, mitral regurgitation, and FA were monitored.

RESULTS: The edge-to-edge repair led to a decrease in effective orifice area of 16.55% ± 8.22%; further reduction in effective orifice area was attained with annular contraction. Mitral regurgitation after the edge-to-edge repair was significantly higher (p <0.05) with annular dilation. In the pulsatile experiments, two peaks in FA were observed: one during systole (FA = 0.059 ± 0.024 N) and a second during diastole (FA = 0.072 ± 0.021 N). Multivariate analysis of variance analysis showed that during systole, transmitral pressure and mitral annular area (MAA) had significant effects on FA [FA = (4.40 x 10–4) transmitral pressure (mm Hg) + (5.0 x 10–3) MAA (cm2) – 0.05 (R2= 0.80)], whereas during diastole MFR and MAA had significant effects on FA [FA = (1.03 x 10–4) MFR2 (L/min) – (1.60 x 10–3) MAA (cm2) + 0.02 (R2 = 0.90)].

CONCLUSIONS: With annular dilation, mitral regurgitation persisted even after the edge-to-edge repair. The edge-to-edge repair does not cause clinically relevant mitral valve stenosis in a normal size mitral valve. Mitral flow rate and transmitral pressure are the main determinants of FA during the cardiac cycle. Increasing annular area increases FA during systole but decreases FA during diastole. Systolic FA may become dominant with increases in MAA or peak transmitral pressure, or both.




This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
H. Hasegawa, Y. Araki, A. Usui, J. Yokote, S. Saito, H. Oshima, and Y. Ueda
Mitral valve motion after performing an edge-to-edge repair in an isolated swine heart
J. Thorac. Cardiovasc. Surg., September 1, 2008; 136(3): 590 - 596.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
L. P. Dasi, P. Sucosky, and A. P. Yoganathan
Letter by Dasi et al Regarding Article, "Effect of Chronotropy and Inotropy on Stitch Tension in the Edge-to-Edge Mitral Repair"
Circulation, July 22, 2008; 118(4): e78 - e78.
[Full Text] [PDF]


Home page
CirculationHome page
T. A. Timek, S. L. Nielsen, D. T. Lai, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller
Effect of Chronotropy and Inotropy on Stitch Tension in the Edge-to-Edge Mitral Repair
Circulation, September 11, 2007; 116(11_suppl): I-276 - I-281.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2006 by The Society of Thoracic Surgeons.