Ann Thorac Surg 2006;82:1322-1326
© 2006 The Society of Thoracic Surgeons
a Departments of Cardiac Surgery and Cardiology, Children's Hospital Boston and Harvard Medical School, Boston, Massachusetts
b Department of Pediatric Cardiology, Kardiozentrum, La Paz, Bolivia
Accepted for publication May 4, 2006.
* Address correspondence to Dr del Nido, Department of Cardiac Surgery, Children's Hospital Boston, Harvard Medical School, 300 Longwood Ave, Boston, MA 02115 (Email: pedro.delnido{at}cardio.chboston.org).
Presented at the Forty-second Annual Meeting of The Society of Thoracic Surgeons, Chicago, IL, Jan 30Feb 1, 2006.
BACKGROUND: Current real-time three-dimensional echocardiography systems (RT3DE) can provide sufficient visualization to permit atrial septal defect (ASD) closure. However, detailed visualization of small objects inside the heart is still suboptimal because of limited resolution, which is a limitation for clinical application. We evaluate the complementary use of videocardioscopy in image-guided ASD closure.
METHODS: In a pig model (n = 5), a 4-mm to 8-mm ASD was created with RT3DE guidance. Defect closure was accomplished with a catheter-based patch-delivery system fixed around the defect with mini-anchors under combined RT3DE and videocardioscopy guidance. The endoscope was inserted into the heart through a custom built port designed to allow visualization in the presence of blood.
RESULTS: All ASDs were successfully closed. The combination of RT3DE and videocardioscopy allowed detailed visualization of intracardiac structures, instruments, patch, and mini-anchors.
CONCLUSIONS: Beating-heart ASD closure can be achieved with combined RT3DE and videocardioscopy imaging. Use of videocardioscopy provides high-resolution imaging and likely improves safety of the image-guided procedure.
This article has been cited by other articles:
![]() |
S. G. Yuen, D. T. Kettler, P. M. Novotny, R. D. Plowes, and R. D. Howe Robotic Motion Compensation for Beating Heart Intracardiac Surgery The International Journal of Robotics Research, October 1, 2009; 28(10): 1355 - 1372. [Abstract] [PDF] |
||||
![]() |
R. Kozlik-Feldmann, N. Lang, A. Lehner, M. Sigler, C. Schmitz, R. Sodian, F. Freudenthal, R. Dalla-Pozza, N. V. Vasilyev, P. J. del Nido, et al. Evaluation of a new hybrid technique for closure of muscular ventricular septal defects in a long-term setting. J. Thorac. Cardiovasc. Surg., August 1, 2009; 138(2): 365 - 373. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. V. Vasilyev, P. M. Novotny, J. F. Martinez, H. Loyola, I. S. Salgo, R. D. Howe, and P. J. del Nido Stereoscopic vision display technology in real-time three-dimensional echocardiography-guided intracardiac beating-heart surgery. J. Thorac. Cardiovasc. Surg., June 1, 2008; 135(6): 1334 - 1341. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Kozlik-Feldmann, N. Lang, R. Aumann, A. Lehner, D. Rassoulian, R. Sodian, C. Schmitz, M. Hinterseer, R. Hinkel, E. Thein, et al. Patch closure of muscular ventricular septal defects with a new hybrid therapy in a pig model. J. Am. Coll. Cardiol., April 22, 2008; 51(16): 1597 - 1603. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Huang, J. K. Triedman, N. V. Vasilyev, Y. Suematsu, R. O. Cleveland, and P. E. Dupont Imaging Artifacts of Medical Instruments in Ultrasound-Guided Interventions J. Ultrasound Med., October 1, 2007; 26(10): 1303 - 1322. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |