Ann Thorac Surg 2006;81:1879-1880
© 2006 The Society of Thoracic Surgeons
Department of Cardiovascular Surgery, Heart Center Brandenburg, Bernau, Berlin, Germany
Accepted for publication May 19, 2005.
* Address correspondence to Dr Hartrumpf, Department of Cardiovascular Surgery, Heart Center Brandenburg, Ladeburger Strasse 17, Bernau bei Berlin, 16321 Germany (Email: m.hartrumpf{at}immanuel.de).
The first case of a fracture of a semi-rigid annuloplasty device is reported. A 56-year-old man presented with recurrence of severe mitral and tricuspid regurgitation 10 months after proper implantation of a Colvin-Galloway Future band (Medtronic Inc, Minneapolis, MN). During redo surgery we made the unexpected finding that its metal skeleton was fractured near the posterior segment. A tear in the anterior mitral leaflet was also found. It is speculated that the device fracture may have put additional strain on the native annulus.
This article has been cited by other articles:
![]() |
R. Lange, T. Guenther, B. Kiefer, C. Noebauer, W. Goetz, R. Busch, P. Tassani-Prell, B. Voss, and R. Bauernschmitt Mitral valve repair with the new semirigid partial Colvin-Galloway Future annuloplasty band. J. Thorac. Cardiovasc. Surg., May 1, 2008; 135(5): 1087 - 1093.e4. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |