Ann Thorac Surg 2005;80:1898-1901
© 2005 The Society of Thoracic Surgeons
Thoracic Department, Institut Mutualiste Montsouris, Paris, France
Accepted for publication February 3, 2005.
* Address correspondence to Dr Gossot, Institut Mutualiste Montsouris, Thoracic Department, 42 Blvd Jourdan, Paris, F-75014 France (Email: dominique.gossot{at}imm.fr).
PURPOSE: Current disposable hand-actuated staplers may pose reliability problems, especially with respect to the measurement of tissue thickness. We have evaluated a newly developed stapler with a computer-controlled placement of staples.
DESCRIPTION: The SurgAssist system (Power Medical Interventions, New Hope, PA) is comprised of a console that houses a computer, a remote control unit, a flexible shaft, and a cartridge. The remote control unit has two uses: (1) controlling the accurate placement of the cartridge by orientating the tip of the flexible shaft, and (2) controlling the closure of the stapler and the firing. Each cartridge contains a programmed electronic device that triggers the activation of the appropriate program in the main microprocessor. The compression level on the tissue is determined by the computer.
EVALUATION: The system was used in a consecutive series of 38 patients, 26 times during open lung surgery and 12 times during video-assisted thoracic surgery. The following open procedures were performed: three pneumonectomies, 15 lobectomies, three segmentectomies, and five wedge resections. The following video-assisted thoracic surgery procedures were performed: eight wedge resections and four bullectomies for pneumothorax. There was no stapling failure and no complication related to the use of the stapler. During video-assisted thoracic surgery, some ergonomic problems were encountered that will be overcome by redesign.
CONCLUSIONS: The computer-controlled stapling system may significantly improve tissue approximation during open and video-assisted thoracic surgery.
This article has been cited by other articles:
![]() |
M. Takata, G. Watanabe, H. Ohtake, T. Ushijima, S. Yamaguchi, Y. Kikuchi, and Y. Yamamoto Automatic aortic anastomosis with an innovative computer-controlled circular stapler for surgical treatment of aortic aneurysm J. Thorac. Cardiovasc. Surg., May 1, 2011; 141(5): 1265 - 1269. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Komanapalli and M. S. Sukumar Computer assisted surgical stapling MMCTS, January 1, 2008; 2008(1110): mmcts.2008.003145 - mmcts.2008.003145. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. F. Heitmiller Invited commentary Ann. Thorac. Surg., November 1, 2005; 80(5): 1902 - 1902. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |