ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Dominique Gossot
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gossot, D.
Right arrow Articles by Nana, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gossot, D.
Right arrow Articles by Nana, A.
Related Collections
Right arrow Lung - other

Ann Thorac Surg 2005;80:1898-1901
© 2005 The Society of Thoracic Surgeons


New technology

Computer-Controlled Stapling System for Lung Surgery

Dominique Gossot, MD * , Albert Nana, MD

Thoracic Department, Institut Mutualiste Montsouris, Paris, France

Accepted for publication February 3, 2005.

* Address correspondence to Dr Gossot, Institut Mutualiste Montsouris, Thoracic Department, 42 Blvd Jourdan, Paris, F-75014 France (Email: dominique.gossot{at}imm.fr).

PURPOSE: Current disposable hand-actuated staplers may pose reliability problems, especially with respect to the measurement of tissue thickness. We have evaluated a newly developed stapler with a computer-controlled placement of staples.

DESCRIPTION: The SurgAssist system (Power Medical Interventions, New Hope, PA) is comprised of a console that houses a computer, a remote control unit, a flexible shaft, and a cartridge. The remote control unit has two uses: (1) controlling the accurate placement of the cartridge by orientating the tip of the flexible shaft, and (2) controlling the closure of the stapler and the firing. Each cartridge contains a programmed electronic device that triggers the activation of the appropriate program in the main microprocessor. The compression level on the tissue is determined by the computer.

EVALUATION: The system was used in a consecutive series of 38 patients, 26 times during open lung surgery and 12 times during video-assisted thoracic surgery. The following open procedures were performed: three pneumonectomies, 15 lobectomies, three segmentectomies, and five wedge resections. The following video-assisted thoracic surgery procedures were performed: eight wedge resections and four bullectomies for pneumothorax. There was no stapling failure and no complication related to the use of the stapler. During video-assisted thoracic surgery, some ergonomic problems were encountered that will be overcome by redesign.

CONCLUSIONS: The computer-controlled stapling system may significantly improve tissue approximation during open and video-assisted thoracic surgery.




This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. Takata, G. Watanabe, H. Ohtake, T. Ushijima, S. Yamaguchi, Y. Kikuchi, and Y. Yamamoto
Automatic aortic anastomosis with an innovative computer-controlled circular stapler for surgical treatment of aortic aneurysm
J. Thorac. Cardiovasc. Surg., May 1, 2011; 141(5): 1265 - 1269.
[Abstract] [Full Text] [PDF]


Home page
MMCTSHome page
C. Komanapalli and M. S. Sukumar
Computer assisted surgical stapling
MMCTS, January 1, 2008; 2008(1110): mmcts.2008.003145 - mmcts.2008.003145.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
R. F. Heitmiller
Invited commentary
Ann. Thorac. Surg., November 1, 2005; 80(5): 1902 - 1902.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2005 by The Society of Thoracic Surgeons.