Ann Thorac Surg 2005;80:1091-1095
© 2005 The Society of Thoracic Surgeons
Division of Cardiac Surgery, Brigham and Womens Hospital, Boston, Massachussetts
Accepted for publication October 22, 2004.
* Address reprint requests to Dr Aranki, Brigham and Womens Hospital, 75 Francis St, Boston, MA 02115. (Email: saranki{at}partners.org).
PURPOSE: Minimal manipulation of the ascending aorta during coronary artery bypass grafting may reduce the incidence of adverse perioperative events related to atheroembolic events. The objective of this study was to evaluate a new proximal anastomotic device (Novare Enclose device [Novare Surgical Systems, Cupertino, CA]).
DESCRIPTION: Over a 23-month period, a total of 76 proximal anastomoses were performed in 50 patients (40 males and 10 females; mean age, 74.36 years). In 45 patients, the proximal anastomosis was constructed before the distal (25 off-pump and 20 on-pump), and in 5 patients the proximal anastomosis was constructed on cardiopulmonary bypass because of diseased ascending aortas.
EVALUATION: The average time per anastomosis was <8 minutes. There were no device related complications. There were no atheroembolic complications in this cohort of patients. One patient underwent cardiac catheterization postoperatively, and the proximal vein anastomosis in this patient was widely patent.
CONCLUSIONS: The Novare Enclose proximal anastomotic device (Novare Surgical Systems) appears to be safe and effective as a tool for proximal anastomosis within a reasonable time period and is comparable with accepted techniques.
This article has been cited by other articles:
![]() |
T. P. Martens, M. Argenziano, and M. C. Oz New Technology for Surgical Coronary Revascularization Circulation, August 8, 2006; 114(6): 606 - 614. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |