Ann Thorac Surg 2005;79:682-685
© 2005 The Society of Thoracic Surgeons
a Department of Thoracic and Cardiovascular Surgery, Johann Wolfgang Goethe University, Frankfurt/Main, Germany
Accepted for publication October 28, 2003.
* Address reprint requests to Dr Doss, Department of Thoracic and Cardiovascular Surgery, Johann Wolfgang Goethe-University Frankfurt/Main, Theodor-Stern-Kai 7, Frankfurt/Main 60590, Germany
mirkodoss{at}aol.com
PURPOSE: Clinical trials with the new 3F stentless aortic bioprosthesis began October 2001, and as one of the first centers to implant this prosthesis in humans, we would like to present our experiences with this new device.
DESCRIPTION: The 3F aortic bioprosthesis is a stentless biological heart valve fabricated from three equal leaflets of equine pericardium, assembled in a tubular shape, and implanted in the native aortic root to replace the patient's diseased aortic leaflets. Between January 2002 and August 2002, 24 3F aortic bioprostheses were implanted at our institution. Effective orifice area, mean gradients, and ejection fraction were evaluated by echocardiography at discharge and at 12-month follow-ups after surgery.
EVALUATION: At 12-month follow-ups, the 3F bioprosthesis showed a good hemodynamic performance with a significant drop of mean gradients to 10.3 mm Hg, a mean effective orifice area of 1.7 cm2, and a mean ejection fraction of 61.5%.
CONCLUSIONS: The clinical performance of the new 3F aortic bioprosthesis is comparable with regular stentless aortic valves. However its unique design facilitates implantation.
This article has been cited by other articles:
![]() |
W. A. Goetz, T. E. Tan, K. H. Lim, S. L. H. Salgues, N. Grousson, F. Xiong, Y. L. Chua, and J. H. Yeo Truly stentless molded autologous pericardial aortic valve prosthesis with single point attached commissures in a sheep model Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 548 - 553. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stelzer Stentless Aortic Valve Replacement: Porcine and Pericardial Card. Surg. Adult, January 1, 2008; 3(2008): 915 - 934. [Full Text] |
||||
![]() |
D. Grandmougin and G. Fayad Implantation of a modified freestyle valve with a single inflow suture line: technical patterns and advantages. Ann. Thorac. Surg., September 1, 2006; 82(3): 1128 - 1130. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A. Goetz, K. H. Lim, R. Ibled, N. Grousson, S. L. H. Salgues, and J. H. Yeo Forces at single point attached commissures (SPAC) in pericardial aortic valve prosthesis Eur. J. Cardiothorac. Surg., February 1, 2006; 29(2): 150 - 155. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |