Hongcan Shi
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shi, H.
Right arrow Articles by Lu, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shi, H.
Right arrow Articles by Lu, D.
Related Collections
Right arrow Trachea and bronchi

Ann Thorac Surg 2005;79:672-676
© 2005 The Society of Thoracic Surgeons


New technology

Experimental Study of Replacing Circumferential Tracheal Defects With New Prosthesis

Hongcan Shi, MD, PhDa,*, Zhifei Xu, MDb, Xiong Qin, MDb, Xuewei Zhao, MDb, Dan Lu, MDa

a Department of Thoracic and Cardiovascular Surgery, Medical College of Yangzhou University, Yangzhou, Jiangsu Province, People's Republic of China
b Department of Thoracic and Cardiovascular Surgery, Second Military Medical University, Shanghai, People's Republic of China

Accepted for publication January 9, 2004.

* Address reprint requests to Dr Shi, Department of Thoracic and Cardiovascular Surgery, Medical College of Yangzhou University, 6 Huaihai Rd, Yangzhou, Jiangsu Province, PR China, P R China
shcan111{at}yahoo.com.cn

PURPOSE: To investigate the feasibility of using a new tracheal prosthesis made of biomaterials to replace extensive circumferential tracheal defects.

DESCRIPTION: Three types of tracheal prostheses were developed and studied. Type I prosthesis was used in 8 mongrel dogs (group A), type II in 4 dogs (group B), and type III in 4 dogs (group C).

EVALUATION: In group A, one died from prosthetic dehiscence, another from anastomotic leakage, and the others had uneventful postoperative courses. The implanted prosthesis was completely incorporated with the recipient trachea, where different lengths of reepithelialization occurred on the luminal surface of the reconstructed trachea. Macroscopic examination showed scattered and different sizes of neo-ossification surrounding the implanted prosthesis. The prosthesis was radioopaque when exposed to routine x rays. In contrast, a relatively high number of complications occurred postoperatively in groups B and C.

CONCLUSIONS: Type I tracheal prosthesis, with further improvements, may be the optimal tracheal graft to replace circumferential tracheal defects, and appears very promising for clinical application.




This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
H. Shi, H. Xu, D. Lu, and J. Wu
Animal models of tracheal allotransplantation using vitrified cryopreservation
J. Thorac. Cardiovasc. Surg., November 1, 2009; 138(5): 1222 - 1226.