Ann Thorac Surg 2004;77:563-568
© 2004 The Society of Thoracic Surgeons
a Division of Cardiothoracic Surgery, University of Miami/Jackson Memorial Hospital, Miami, Florida, USA
* Address reprint requests to Dr Katariya, Cardiothoracic Surgery, University of Miami, 1611 NW 12th Ave, ET 3072, Miami, FL 33133, USA.
e-mail: kkatariya{at}med.miami.edu
Presented at the Thirty-ninth Annual Meeting of The Society of Thoracic Surgeons, San Diego, CA, Jan 31Feb 2, 2003.
BACKGROUND: We report our early experience with the Symmetry Aortic Connector (St. Jude Medical, St. Paul, MN) used for sutureless proximal aortosaphenous vein graft anastomoses without any cross clamp during coronary bypass procedures.
METHODS: Between November 2001 and August 2002, 206 saphenous vein to aorta proximal anastomoses were created in 132 patients using the Symmetry device. All procedures were performed as part of off-pump coronary artery bypass surgery without any aortic clamping. Intraoperative variables and postoperative data were collected and analyzed retrospectively.
RESULTS: All 206 anastomoses (100%) were successfully completed with the connector. Severe atherosclerotic disease of the aorta was documented in 16 patients (12%). Four anastomoses (2%) required additional suture placement. Predeployment problems occurred with 3 grafts (2.5%) during loading of the connector. Average number of distal bypasses was 3.2 per patient. One patient (0.7%) required reoperation for bleeding from a proximal anastomosis. Six patients (4.5%) had perioperative myocardial infarction documented by electrocardiographic changes. Thirty-day operative mortality was 3% (4 patients). Intraoperative transit time flow measurement was performed in all cases (100%). Postoperative angiography in 43 patients at a median 3 months postoperatively revealed occlusion of 9 of the 81 saphenous vein grafts (11%).
CONCLUSIONS: The initial experience with a proximal saphenous vein graft to aorta anastomosis using the Symmetry connector demonstrates safety and ease of use. There is however some concern with early graft closure. A prospective randomized study is needed to clarify these concerns.
This article has been cited by other articles:
![]() |
M. Ricci and T. A. Salerno eComment: Aortic connectors in coronary artery surgery Interactive CardioVascular and Thoracic Surgery, August 1, 2008; 7(4): 672 - 672. [Full Text] [PDF] |
||||
![]() |
T. Puehler, S. Fraund-Cremer, J. Cremer, and A. Boening Successful six-year follow-up of a sutureless device for proximal anastomoses in a severely calcified ascending aorta Interactive CardioVascular and Thoracic Surgery, August 1, 2008; 7(4): 670 - 672. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J.L. Suyker and C. Borst Coronary Connector Devices: Analysis of 1,469 Anastomoses in 1,216 Patients Ann. Thorac. Surg., May 1, 2008; 85(5): 1828 - 1836. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Bergmann, K. Meszaros, S. Huber, P. Oberwalder, H. Machler, G. Schaffler, R. Rienmueller, and B. Rigler Forty-one-month follow-up of the Symmetry aortic connector system for proximal venous anastomosis J. Thorac. Cardiovasc. Surg., July 1, 2007; 134(1): 23 - 28. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. G. Wolf, Y. Abu-Omar, B. P. Choudhary, D. Pigott, and D. P. Taggart Gaseous and solid cerebral microembolization during proximal aortic anastomoses in off-pump coronary surgery: The effect of an aortic side-biting clamp and two clampless devices J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 485 - 493. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Caputo, P. Narayan, and G. D. Angelini Conventional surgery with aortic cross-clamping MMCTS, March 15, 2006; 2006(0315): 828. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Skjelland, J. Bergsland, R. Lundblad, P. S. Lingaas, K. A. Rein, S. Halvorsen, J. L. Svennevig, E. Fosse, R. Brucher, and D. Russell Cerebral microembolization during off-pump coronary artery bypass surgery with the Symmetry aortic connector device J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1581 - 1585. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kitamura, H. Okabayashi, M. Hanyu, Y. Soga, T. Nomoto, H. Johno, J. Nakano, T. Matsuo, M. Kai, and E. Umehara Early and midterm patency of the proximal anastomoses of saphenous vein grafts made with a Symmetry Aortic Connector System J. Thorac. Cardiovasc. Surg., October 1, 2005; 130(4): 1028 - 1031. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. F. Aranki, P. S. Shekar, A. Ehsan, M. Byrne-Taft, and G. S. Couper Evaluation of the Enclose Proximal Anastomosis Device in Coronary Artery Bypass Grafting Ann. Thorac. Surg., September 1, 2005; 80(3): 1091 - 1095. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Izutani, M. Yoshitatsu, J. Kawamoto, and K. Katayama A case of ostial stenosis with the PAS-Port proximal anastomosis system in off-pump coronary artery bypass grafting Interactive CardioVascular and Thoracic Surgery, August 1, 2005; 4(4): 341 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. D.L. Keenan, Y. Abu-Omar, and D. P. Taggart Bypassing the Pump: Changing Practices in Coronary Artery Surgery Chest, July 1, 2005; 128(1): 363 - 369. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lev-Ran, R. Braunstein, R. Sharony, A. Kramer, Y. Paz, R. Mohr, and G. Uretzky No-touch aorta off-pump coronary surgery: The effect on stroke J. Thorac. Cardiovasc. Surg., February 1, 2005; 129(2): 307 - 313. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Cavendish, W. F. Penny, M. M. Madani, S. Keramati, O. Ben-Yehuda, D. G. Blanchard, E. Mahmud, A. Perricone, and S. Tsimikas Reply J. Am. Coll. Cardiol., September 1, 2004; 44(5): 1141 - 1142. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |