ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Igor Gregoric
Kamuran A. Kadipasaoglu
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gregoric, I.
Right arrow Articles by Frazier, O. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gregoric, I.
Right arrow Articles by Frazier, O. H.
Related Collections
Right arrow Valve disease

Ann Thorac Surg 2004;77:196-202
© 2004 The Society of Thoracic Surgeons


Original article: cardiovascular

Preclinical assessment of a trileaflet mechanical valve in the mitral position in a calf model

Igor Gregoric, MDa*, Jeff L. Conger, BSa, Helmut Reul, DScb, Daniel Tamez, BSa, Fred J. Clubb, Jr, DVM, PhDa, Raymond F. Stainback, MDa, Antonieta Hernandez, MDa, Roberto Cervera, MDa, Kazuhiro Eya, MDc, Denise Bylera, Kamuran A. Kadipasaoglu, PhDa, O. H. Frazier, MDa

a Cardiovascular Research Laboratories, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas, USA
b Helmholtz Institute, Aachen, Germany
c National Cardiovascular Institute, Osaka, Japan

Accepted for publication July 10, 2003.

* Address reprint requests to Dr Gregoric, Texas Heart Institute, PO Box 20345, MC3-147, Houston, TX 77225-0345, USA
e-mail: igregoric{at}hotmail.com

BACKGROUND: The bileaflet valve is currently the mechanical replacement valve of choice. Though durable, it does not closely mimic native valve hemodynamics and remains potentially thrombogenic.

METHODS: Prototype trileaflet valves (T1 and T2) were implanted in the mitral position in calves. Group I calves received either a T1 valve (n = 12) or a control bileaflet valve (n = 5); Group II, either a T2 valve (n = 7) or a control bileaflet valve (n = 5). Valve function, perivalvular leakage, and transvalvular pressure gradients were evaluated. Also, long-term prototype leaflet wear was evaluated in vivo in one Group I calf (502 days) and two Group II calves (385 and 366 days). Calves were euthanized and necropsied at study termination, and major organs weighed and examined.

RESULTS: Valve function was excellent and hematologic parameters remained normal in all calves that survived to study termination. Mean peak transvalvular pressure gradients were 10 ± 7 mm Hg for T1 valves, 6 ± 3 mm Hg for T2 valves, and 12 ± 4 mm Hg for bileaflet control valves. Clinically insignificant valvular regurgitation was observed in both prototypes. Explanted valves showed no thrombus-impaired leaflet motion, except in two T1-fitted calves and one T2-fitted calf. Major organs showed no evidence of clinically significant thromboembolic events. There were no other significant differences between the results of experimental and control groups.

CONCLUSIONS: Prototype trileaflet valves performed safely and effectively in the mitral position in calves, even without long-term anticoagulation. This warrants their evaluation as an equivalent alternative to bileaflet valves.




This article has been cited by other articles:


Home page
Ann. Thorac. Surg.Home page
D. M. Lapeyre
Mechanical Bioprosthesis
Ann. Thorac. Surg., August 1, 2005; 80(2): 788 - 788.