Ann Thorac Surg 2004;77:196-202
© 2004 The Society of Thoracic Surgeons
a Cardiovascular Research Laboratories, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas, USA
b Helmholtz Institute, Aachen, Germany
c National Cardiovascular Institute, Osaka, Japan
Accepted for publication July 10, 2003.
* Address reprint requests to Dr Gregoric, Texas Heart Institute, PO Box 20345, MC3-147, Houston, TX 77225-0345, USA
e-mail: igregoric{at}hotmail.com
BACKGROUND: The bileaflet valve is currently the mechanical replacement valve of choice. Though durable, it does not closely mimic native valve hemodynamics and remains potentially thrombogenic.
METHODS: Prototype trileaflet valves (T1 and T2) were implanted in the mitral position in calves. Group I calves received either a T1 valve (n = 12) or a control bileaflet valve (n = 5); Group II, either a T2 valve (n = 7) or a control bileaflet valve (n = 5). Valve function, perivalvular leakage, and transvalvular pressure gradients were evaluated. Also, long-term prototype leaflet wear was evaluated in vivo in one Group I calf (502 days) and two Group II calves (385 and 366 days). Calves were euthanized and necropsied at study termination, and major organs weighed and examined.
RESULTS: Valve function was excellent and hematologic parameters remained normal in all calves that survived to study termination. Mean peak transvalvular pressure gradients were 10 ± 7 mm Hg for T1 valves, 6 ± 3 mm Hg for T2 valves, and 12 ± 4 mm Hg for bileaflet control valves. Clinically insignificant valvular regurgitation was observed in both prototypes. Explanted valves showed no thrombus-impaired leaflet motion, except in two T1-fitted calves and one T2-fitted calf. Major organs showed no evidence of clinically significant thromboembolic events. There were no other significant differences between the results of experimental and control groups.
CONCLUSIONS: Prototype trileaflet valves performed safely and effectively in the mitral position in calves, even without long-term anticoagulation. This warrants their evaluation as an equivalent alternative to bileaflet valves.
This article has been cited by other articles:
![]() |
D. M. Lapeyre Mechanical Bioprosthesis Ann. Thorac. Surg., August 1, 2005; 80(2): 788 - 788. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |