Ann Thorac Surg 2002;74:S1408-S1412
© 2002 The Society of Thoracic Surgeons
a The Department of Cardiovascular and Thoracic Surgery, University of Texas Southwestern Medical Center, Dallas, Texas, USA
* Address reprint requests to Dr Jessen, Department of Cardiovascular and Thoracic Surgery, University of Texas Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas TX 75390-8879, USA.
e-mail: michael.jessen{at}utsouthwestern.edu
Presented at the Eighth Annual Cardiothoracic Techniques and Technologies Meeting 2002, Miami Beach, FL, Jan 2326, 2002.
BACKGROUND: Cardiac output is often monitored after cardiac operations with a pulmonary artery catheter. A new method has been introduced that measures cardiac output by lithium dilution (LiDCO) and uses these data to calibrate a system (PulseCO) that calculates cardiac output continuously from the energy of the arterial pressure waveform. It is unknown whether PulseCO measurements are valid early after cardiac surgery when changes in temperature and vascular tone or intermittent use of the arterial line for blood sampling may occur. This study assessed the reliability of cardiac output determinations by PulseCO in the first 8 hours after cardiac surgery.
METHODS: After a one-time PulseCO calibration, cardiac output was measured in 20 patients who had undergone coronary artery bypass grafting at 0, 2, 4, 6, and 8 hours after arrival in the intensive care unit using (1) thermodilution through a pulmonary artery catheter (Thermo); (2) lithium dilution (LiDCO); and (3) PulseCO. Concordance correlations were calculated between methods, and differences were compared by Wilcoxon paired rank test and Bland-Altman analysis.
RESULTS: Cardiac output ranged from 3.4 to 8.5 L/min. No significant differences were noted between measurements obtained by each technique at any time point. Concordance correlations and Bland-Altman analysis confirmed good agreement between PulseCO and Thermo determinations of cardiac output during the study interval.
CONCLUSIONS: PulseCO measurements remain reliable without recalibration for at least 8 hours after cardiac surgery and may offer a less-invasive approach for early postoperative cardiac output monitoring.
This article has been cited by other articles:
![]() |
J. Vuyk, B. J. Lichtenbelt, E. Olofsen, J. W. van Kleef, and A. Dahan Mixed-Effects Modeling of the Influence of Midazolam on Propofol Pharmacokinetics Anesth. Analg., May 1, 2009; 108(5): 1522 - 1530. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Compton and J.-H. Schafer Noninvasive Cardiac Output Determination: Broadening the Applicability of Hemodynamic Monitoring Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2009; 13(1): 44 - 55. [Abstract] [PDF] |
||||
![]() |
H. M. C. Kemps, E. J. M. Thijssen, G. Schep, B. T. H. M. Sleutjes, W. R. De Vries, A. R. Hoogeveen, P. F. F. Wijn, and P. A. F. M. Doevendans Evaluation of two methods for continuous cardiac output assessment during exercise in chronic heart failure patients J Appl Physiol, December 1, 2008; 105(6): 1822 - 1829. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Matthieu, N.-G. Karine, C. Vincent, V. Alain, C. J. Francois, R. Philippe, and S. Francois Cardiac Output Measurement in Patients Undergoing Liver Transplantation: Pulmonary Artery Catheter Versus Uncalibrated Arterial Pressure Waveform Analysis Anesth. Analg., May 1, 2008; 106(5): 1480 - 1486. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Giomarelli, B. Biagioli, and S. Scolletta Cardiac output monitoring by pressure recording analytical method in cardiac surgery Eur. J. Cardiothorac. Surg., September 1, 2004; 26(3): 515 - 520. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |