ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Abdusalam El-Gatit
Zsolt L. Nagy
Kevin G. Watterson
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jennings, L. M.
Right arrow Articles by Watterson, K. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jennings, L. M.
Right arrow Articles by Watterson, K. G.
Related Collections
Right arrow Valve disease

Ann Thorac Surg 2002;74:63-68
© 2002 The Society of Thoracic Surgeons


Original article: cardiovascular

Hydrodynamic function of the second-generation mitroflow pericardial bioprosthesis

Louise M. Jennings, MEngb, Abdusalam El-Gatit, MDa, Zsolt L. Nagy, PhDa, John Fisher, DEngb, Peter G. Walker, PhDb, Kevin G. Watterson, FRACS*a

a Yorkshire Heart Centre, University of Leeds, Leeds, United Kingdom
b School of Mechanical Engineering, University of Leeds, Leeds, United Kingdom

Accepted for publication March 19, 2002.

* Address reprint requests to Mr. Watterson, Yorkshire Heart Centre, Leeds Teaching Hospital, Calverley St, Leeds LS1 3EX, UK
e-mail: kevin.watterson{at}leedsth.nhs.uk

Background. The hydrodynamic function of the smaller size Mitroflow Synergy stented pericardial bioprostheses has been studied in an in vitro fresh tissue aortic root model and compared with previous studies of free-sewn bioprostheses.

Methods. Three valves of each of the sizes 19, 21, and 23 mm were sutured into fresh tissue aortic roots and tested in a pulsatile flow simulator using two different ventricular input impedance conditions. A high-speed camera was used to study the leaflet opening and closing configurations. Mean pressure difference as a function of root mean square forward flow, effective orifice area, regurgitant volumes, and total energy loss across the valves was measured.

Results. Mean pressure difference with respect to root mean square forward flow decreased as the valve size increased. Thus effective orifice area increased as the valve size increased. The open leaflet configuration images showed that all three sizes of Mitroflow valves had a large circular orifice with minimal open leaflet deformation. All valves closed competently with no visible leakage and no closed regurgitant volume. The Mitroflow valves showed better effective orifice areas compared with previously tested frame-mounted porcine bioprostheses but lower effective orifice areas compared with porcine stentless bioprostheses; however, the open leaflet bending deformation was better than for any of the previously tested bioprosthetic valves.

Conclusions. The hydrodynamic function of the Mitroflow Synergy stented pericardial bioprosthesis shows potential for good in vivo hemodynamic performance. The good hemodynamic performance combined with relative ease of implantation technique makes the pericardial valve a good valve in the aortic position, particularly in older patients with small annuli.




This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
F. Bakhtiary, O. Dzemali, U. Steinseiffer, C. Schmitz, B. Glasmacher, A. Moritz, and P. Kleine
Opening and closing kinematics of fresh and calcified aortic valve prostheses: An in vitro study
J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 657 - 662.
[Abstract] [Full Text] [PDF]


Home page
Card Surg AdultHome page
N. D. Desai and G. T. Christakis
Stented Mechanical/Bioprosthetic Aortic Valve Replacement
Card. Surg. Adult, January 1, 2003; 2(2003): 825 - 856.