ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Isaac Kadir
Alan J. Bryan
Gianni D. Angelini
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kadir, I.
Right arrow Articles by Angelini, G. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kadir, I.
Right arrow Articles by Angelini, G. D.
Related Collections
Right arrow Valve disease

Ann Thorac Surg 2001;72:49-53
© 2001 The Society of Thoracic Surgeons


Original article: cardiovascular

Hemodynamic performance of the 21-mm Sorin Bicarbon mechanical aortic prostheses using dobutamine Doppler echocardiography

Isaac Kadir, FRCSa, Innes Y.P. Wan, FRCSa, Catherine Walsh, Dip-Rada,b, Peter Wilde, FRCRb, Alan J. Bryan, FRCSa, Gianni D. Angelini, MDa a Bristol Heart Institute, Bristol, United Kingdom
b Department of Clinical Radiology, University of Bristol, Bristol, United Kingdom

Accepted for publication March 23, 2001.

Address reprint requests to Dr Angelini, Bristol Heart Institute, Department of Cardiac Surgery, Bristol Royal Infirmary, Bristol BS2 8HW, United Kingdom
e-mail: g.d.angelini{at}bristol.ac.uk

Background. Small-sized mechanical aortic prostheses are commonly associated with generation of high transvalvular gradients, particularly in patients with large body surface area, and can result in patient–prosthesis mismatch. This study evaluates the hemodynamic performance of 21-mm Sorin Bicarbon bileaflet mechanical prostheses using dobutamine stress echocardiography.

Methods. Fourteen patients (7 women; mean age, 63 ± 8 years) who had undergone aortic valve replacement with a 21-mm Sorin Bicarbon bileaflet mechanical prosthesis 32.4 ± 5.1 months previously were studied. After a resting Doppler echocardiogram, a dobutamine infusion was started at a rate of 5 µg · kg-1 · min-1 and increased to 30 µg · kg-1 · min-1 at 15-minute intervals. Pulsed- and continuous-wave Doppler echocardiographic studies were performed at rest and at the end of each increment of dobutamine. Both peak and mean velocity and pressure gradient across the prostheses were measured, and effective orifice area, discharge coefficient, and performance index were calculated.

Results. Dobutamine stress increased heart rate and cardiac output by 83% and 81%, respectively (both p < 0.0001), and mean transvalvular gradient increased from 15.6 ± 5.5 mm Hg at rest to 35.4 ± 11.9 mm Hg at maximum stress (p < 0.0001). Although the indexed effective orifice area was significantly lower in patients with a larger body surface area, this was not associated with any significant pressure gradient. The performance index of this valve was unchanged throughout the study. Regression analyses demonstrated that the mean transvalvular gradient at maximum stress was independent of all variables except resting gradient (p = 0.05). Body surface area had no association with the changes in cardiac output, transvalvular gradient at maximum stress, and effective orifice area.

Conclusions. These data show that the 21-mm Sorin Bicarbon bileaflet mechanical prosthesis offers an excellent hemodynamic performance with full utilization of its available orifice when implanted in the aortic position. The lack of significant transvalvular gradient in patients with a larger body surface area suggests that patient–prosthesis mismatch is highly unlikely when this prosthesis is used.




This article has been cited by other articles:


Home page
Ann. Thorac. Surg.Home page
N. Vitale, G. Cappabianca, G. Visicchio, C. Fondacone, V. Paradiso, G. Mannatrizio, and L. d. L. T. Schinosa
Midterm evaluation of the Sorin Bicarbon heart valve prosthesis: single-center experience
Ann. Thorac. Surg., February 1, 2004; 77(2): 527 - 531.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. De Carlo, A. D. Milano, C. Nardi, G. Mecozzi, and U. Bortolotti
Serial Doppler echocardiographic evaluation of small-sized sorin bicarbon prostheses
J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 337 - 343.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2001 by The Society of Thoracic Surgeons.