ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Etienne Tatou
Roger Brenot
Michel David
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tatou, E.
Right arrow Articles by David, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tatou, E.
Right arrow Articles by David, M.
Related Collections
Right arrow Valve disease

Ann Thorac Surg 2001;71:1366-1369
© 2001 The Society of Thoracic Surgeons


Case report

Fracture-embolization of Duromedics valve prosthesis and microscopic uncommon lesions

Etienne Tatou, MDa, Mountajab Saleh, MDa, Jean Christophe Eicher, MDb, Roger Brenot, MDa, Michel David, MDa

a Service de Chirurgie Cardio-vasculaire, Centre Hospitalier Universitaire de Dijon, Dijon, France
b Service de Cardiologie, Centre Hospitalier Universitaire de Dijon, Dijon, France

Accepted for publication May 2, 2000.

Address reprint requests to Dr Tatou, Service de Chirurgie Cardio-Vasculaire, 2 Bd maréchal de Lattre de Tassigny, BP 1542, 21034 Dijon, France
e-mail: etatou{at}ipac.fr

We report a sudden leaflet fracture of a Duromedics mitral valve 6 years after implantation. The patient had cardiogenic shock and complained of asthenia, orthopnea, and tachycardia. Transesophageal echocardiography showed the lack of one leaflet of the prosthesis and regurgitation. An emergency mitral replacement was successfully performed. Angiographic computed tomography scan localized the sequestrum that embolized the common iliac arteries. Examination of the deficient prosthesis showed multiple lesions and, in particular, a subsurface lesion that may be characteristic of carbon pyrolytic valves.




This article has been cited by other articles:


Home page
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in MedicineHome page
H. Mohammadi, R. J. Klassen, and W.-K. Wan
A finite element model on effects of impact load and cavitation on fatigue crack propagation in mechanical bileaflet aortic heart valve
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, October 1, 2008; 222(7): 1115 - 1125.
[Abstract] [PDF]


Home page
Asian Cardiovasc. Thorac. Ann.Home page
K. Yamazaki, K. Nishimura, A. Iwakura, K. Uehara, T. Okada, and T. Sugita
Fracture Embolization of an Edwards-Duromedics Valve with Asymmetrical Closure
Asian Cardiovasc Thorac Ann, June 1, 2008; 16(3): 242 - 245.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
L. Hansen, M. Danne, B. Hoffmann, and F.-C. Riess
Structural valve failure with every beat regurgitation observed using the Medtronic Advantage aortic valve.
J. Thorac. Cardiovasc. Surg., November 1, 2007; 134(5): 1344 - 1345.