ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rose, A. G.
Right arrow Articles by Miller, L. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rose, A. G.
Right arrow Articles by Miller, L. W.

Ann Thorac Surg 2000;70:1270-1274
© 2000 The Society of Thoracic Surgeons


Original articles: cardiovascular

Partial aortic valve fusion induced by left ventricular assist device

Alan G. Rose, MDa, Soon J. Park, MDb, Alan J. Bank, MDc, Leslie W. Miller, MDc

a Department of Laboratory Medicine and Pathology, University of Minnesota and Fairview-University Medical Center, Minneapolis, Minnesota, USA
b Division of Cardiovascular and Thoracic Surgery, University of Minnesota and Fairview-University Medical Center, Minneapolis, Minnesota, USA
c Division of Cardiology, University of Minnesota and Fairview-University Medical Center, Minneapolis, Minnesota, USA

Address reprint requests to Dr Rose, Department of Laboratory Medicine and Pathology, University of Minnesota Medical School, MMC 76, Mayo Building, 420 Delaware St SE, Minneapolis, MN 55455
e-mail: rosex031{at}tc.umn.edu

Background. Left ventricular assist devices (LVADs) may be used (1) as a bridging device to cardiac transplantation, (2) for permanent replacement of left ventricular function, and (3) as a bridge to recovery of ventricular function, for example, in recoverable myocardial disease. In this third group of patients, it is important that the LVAD does not produce changes in the heart that will have a deleterious effect on cardiac function once the device is removed. Furthermore, if the LVAD fails, survival depends on optimal function of the diseased heart.

Methods. All hearts with LVADs encountered as surgical specimens following heart transplantation or at autopsy at the Fairview-University of Minnesota Medical Center during the 5-month period August 1998 to January 1999 were examined for native valvular heart disease. The nature and extent of commissural fusion was noted and measured. Light microscopy was performed on any valve lesions.

Results. Four of 6 patients with HeartMate (Thermo Cardiosystems, Inc, Woburn, MA) LVADs showed evidence of commissural fusion (acquired aortic stenosis). In 1 patient, this condition was caused by an organizing thrombus uniting a 14-mm length of the commissural region of the right coronary and noncoronary cusps of the aortic valve. Fibrous commissural fusion due to totally organized thrombus in the other 3 patients affected one aortic commissure (2 patients, 2 mm and 4 mm, respectively) and two commissures (1 patient, 2 mm and 5 mm). Partial cuspal fusion in each case was due to permanent closure of the native aortic valve induced by the LVAD’s operating in its automatic setting. Mean length of commissural fusion was 5.4 mm (range, 2 to 14 mm; standard deviation [SD] = ±5.0 mm). Mean duration of implantation of the six LVADs was 180.3 days (range, 26 to 689 days; SD = ±253.8 days). The LVADs of the 3 patients with fibrous fusion of the commissures had been implanted for an average of 252.3 days (range, 26 to 689 days; SD = ±378.2 days).

Conclusions. Normal function of the LVAD produces permanent closure of the native aortic valve. Stasis on the ventricular aspect of the aortic valve, combined with a low level of anticoagulation, favors thrombosis at this site. Thrombus organization leads to aortic stenosis of variable severity. This previously unsuspected complication was not detected clinically in any of our patients. Aortic stenosis may hold serious implications for patients in whom the LVAD acts as a bridge to recovery or in those in whom the LVAD fails. Prevention may be achieved by intermittently reducing LVAD pumping action. A built-in venting cycle would be of value in long-term implants. Thrombi on the aortic valve may also predispose patients to infective endocarditis, because bloodstream infection is common in patients with LVADs.




This article has been cited by other articles:


Home page
Anesth. Analg.Home page
S. Chumnanvej, M. J. Wood, T. E. MacGillivray, and M. F. V. Melo
Perioperative Echocardiographic Examination for Ventricular Assist Device Implantation
Anesth. Analg., September 1, 2007; 105(3): 583 - 601.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
S. C. Horton, R. Khodaverdian, P. Chatelain, M. L. McIntosh, B. D. Horne, J. B. Muhlestein, and J. W. Long
Left Ventricular Assist Device Malfunction: An Approach to Diagnosis by Echocardiography
J. Am. Coll. Cardiol., May 3, 2005; 45(9): 1435 - 1440.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
S. C. Horton, R. Khodaverdian, A. Powers, J. Revenaugh, D. G. Renlund, S. A. Moore, B. Rasmusson, K. E. Nelson, and J. W. Long
Left ventricular assist device malfunction: a systematic approach to diagnosis
J. Am. Coll. Cardiol., May 5, 2004; 43(9): 1574 - 1583.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
R. Hetzer, E. V. Potapov, Y. Weng, H. Sinawski, F. Knollmann, T. Komoda, E. Hennig, and M. Pasic
Implantation of MicroMed DeBakey VAD through left thoracotomy after previous median sternotomy operations
Ann. Thorac. Surg., January 1, 2004; 77(1): 347 - 350.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
K. D. Aaronson, H. Patel, and F. D. Pagani
Patient selection for left ventricular assist device therapy
Ann. Thorac. Surg., June 1, 2003; 75(90060): S29 - 35.
[Abstract] [Full Text] [PDF]


Home page
SEMIN CARDIOTHORAC VASC ANESTHHome page
B. A. J. M. de Mol
Mechanical Support in Acute Perioperative Heart Failure: Are Assist Devices Smart Enough to Heal the Heart?
Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2003; 7(1): 105 - 109.
[PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. Nishimura, S. Ohtake, Y. Sawa, N. Fukushima, G. Matsumiya, and H. Matsuda
Severe aortic valve fusion after nearly three years of support with the Novacor left ventricular assist system
J. Thorac. Cardiovasc. Surg., July 1, 2002; 124(1): 179 - 180.
[Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
L. E. Samuels, M. P. Thomas, E. C. Holmes, J. Narula, J. Fitzpatrick, D. Wood, B. Fyfe, and A. S. Wechsler
Insufficiency of the native aortic valve and left ventricular assist system inflow valve after support with an implantable left ventricular assist system: Signs, symptoms, and concerns
J. Thorac. Cardiovasc. Surg., August 1, 2001; 122(2): 380 - 381.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2000 by The Society of Thoracic Surgeons.