Ann Thorac Surg 2000;70:212-217
© 2000 The Society of Thoracic Surgeons
a Department of Cardiovascular Surgery, University Hospital Insel, Bern, Switzerland
b Division of Cardiology, University Hospital Insel, Bern, Switzerland
Address reprint requests to Dr Walpoth, Department of Cardiovascular Surgery, University Hospital Insel, CH-3010, Bern, Switzerland
e-mail: beat.walpoth{at}insel.ch
Background. Bloodflow measurements are of major clinical importance for quality control in vascular surgery. They allow detection of low-flow situations which may influence outcome adversely. The purpose of the present study was to validate three different flow systems for measuring absolute blood flow.
Methods. Measurements were performed in an experimental flow model using arteries and veins and blood or saline at two different temperatures. As a reference method true flow was measured by volume sampling.
Results. Correlation coefficients between transit time flow and true flow measurements ranged between 0.71 and 0.92. Systematic overestimation and underestimation of transit time flow were observed, but after second-order correction all correlations were excellent, ranging from 0.93 to 0.95 irrespective of flow medium and temperature.
Conclusions. Transit time flow measurements are exact and reproducible. Second-order correction yields good accuracy and high precision, with minimal differences among the three systems evaluated.
This article has been cited by other articles:
![]() |
G. D'Ancona, J. M. Hartman, F. Bartolozzi, E. van Deel, D. J.G.M. Duncker, A. J.J.C. Bogers, M. Pilato, and A. P. Kappetein Epicardial coronary artery Doppler: validation in the animal model Interactive CardioVascular and Thoracic Surgery, August 1, 2008; 7(4): 634 - 637. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-P. Verhoye, I. Abouliatim, A. Drochon, B. de Latour, C. Leclercq, A. Leguerrier, and H. Corbineau Collateral blood flow between left coronary artery bypass grafts and chronically occluded right coronary circulation in patients with triple vessel disease.: Observations during complete revascularisation of beating hearts Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 49 - 54. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Winkler, K. D. Calligaro, S. Kolakowski, K. J. Doerr, S. McAffee-Bennett, K. Muller, and M. J. Dougherty Comparison of Intraoperative Completion Flowmeter Versus Duplex Ultrasonography and Contrast Arteriography for Carotid Endarterectomy Vascular and Endovascular Surgery, January 1, 2007; 40(6): 482 - 486. [Abstract] [PDF] |
||||
![]() |
G. Di Giammarco, M. Pano, S. Cirmeni, P. Pelini, G. Vitolla, and M. Di Mauro Predictive value of intraoperative transit-time flow measurement for short-term graft patency in coronary surgery. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 468 - 474. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.-B. Kim, C. H. Kang, and C. Lim Prediction of Graft Flow Impairment by Intraoperative Transit Time Flow Measurement in Off-Pump Coronary Artery Bypass Using Arterial Grafts Ann. Thorac. Surg., August 1, 2005; 80(2): 594 - 598. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. H. Walpoth INVITED COMMENTARY Ann. Thorac. Surg., March 1, 2005; 79(3): 857 - 858. [Full Text] [PDF] |
||||
![]() |
H. Kamiya, G. Watanabe, T. Saito, T. Doi, S. Tomita, H. Ohtake, and T. Kanamori Real-time and continuous monitoring of myocardial blood flow using a thermal diffusion method Eur. J. Cardiothorac. Surg., April 1, 2002; 21(4): 748 - 752. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Wolfe The coronary artery bypass conduit: II. Assessment of the quality of the distal anastomosis Ann. Thorac. Surg., December 1, 2001; 72(6): S2253 - 2258. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |