Ann Thorac Surg 1999;68:1380-1383
© 1999 The Society of Thoracic Surgeons
a Division of Pediatric Cardiology, Childrens University Hospital, Zurich, Switzerland
b II Medical Clinic, Johannes Gutenberg University, Mainz, Germany
c Department of Cardiothoracic Surgery, University Hospital, Lund, Sweden
Address reprint requests to Dr Bauersfeld, Division of Pediatric Cardiology, Childrens University Hospital, Steinwiesstr 75, 8032 Zurich Switzerland
e-mail: bauersfe{at}kispi.unizh.ch
Background. Permanent cardiac pacing in children results commonly in augmented energy consumption because of the high pacing rates and the ample stimulation safety margin applied in children. Cardiovascular anatomy and limited venous access sometimes preclude the otherwise preferred endocardial approach. In this multicenter patient series, we studied the feasibility, safety, and energy saving obtained by a combination of steroid-eluting epicardial leads with autocapture devices capable of ongoing adjustment of the stimulation output to the prevailing threshold.
Methods. Autocapture devices (Pacesetter Microny SR+ and Regency SR+; Pacesetter, Solna, Sweden) and steroid-eluting epicardial pacing leads (Medtronic CapSure Epi 10366; Medtronic, Inc, Minneapolis, MN) were implanted in 14 children. Thresholds, telemetry data, evoked response, and polarization signals were obtained at discharge and follow-up, and battery service life was calculated.
Results. During a median follow-up of 6.5 months, autocapture pacing was applied in 12 of 14 children. The automatically adjusted pulse amplitude of autocapture devices demonstrated low-energy pacing with no significant changes between discharge and 6 months follow-up (1.1 ± 0.3 versus 0.9 ± 0.3 V). Autocapture-programmed pacemakers had calculated life spans of 7.8 ± 1.4 years (Microny) and 21.0 ± 1.6 years (Regency). No adverse effects were noted.
Conclusions. Autocapture-controlled pacing with bipolar epicardial pacing leads is feasible and safe in children. Autocapture programming results in substantial energy savings and extends battery life markedly.
This article has been cited by other articles:
![]() |
M. Tomaske, B. Gerritse, L. Kretzers, R. Pretre, A. Dodge-Khatami, M. Rahn, and U. Bauersfeld A 12-Year Experience of Bipolar Steroid-Eluting Epicardial Pacing Leads in Children Ann. Thorac. Surg., May 1, 2008; 85(5): 1704 - 1711. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Odim, B. Suckow, B. Saedi, H. Laks, and K. Shannon Equivalent Performance of Epicardial Versus Endocardial Permanent Pacing in Children: A Single Institution and Manufacturer Experience Ann. Thorac. Surg., April 1, 2008; 85(4): 1412 - 1416. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Silvetti, A. De Santis, S. Marcora, T. De Santo, N. Grovale, and F. Drago Circadian pattern of atrial pacing threshold in the young Europace, February 1, 2008; 10(2): 147 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tomaske, P. Harpes, R. Pretre, A. Dodge-Khatami, and U. Bauersfeld Long-term experience with AutoCapture(R)-controlled epicardial pacing in children Europace, August 1, 2007; 9(8): 645 - 650. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. C. Aellig, C. Balmer, A. Dodge-Khatami, M. Rahn, R. Pretre, and U. Bauersfeld Long-Term Follow-Up After Pacemaker Implantation in Neonates and Infants Ann. Thorac. Surg., April 1, 2007; 83(4): 1420 - 1423. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Silvetti, F. Drago, G. Grutter, A. De Santis, V. Di Ciommo, and L. Rava Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients Europace, July 1, 2006; 8(7): 530 - 536. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Çeliker, N. Ceviz, and O. Küçükosmanoglu Long-term results of endocardial pacing with AutocaptureTM threshold tracking pacemakers in children Europace, January 1, 2005; 7(6): 569 - 575. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. I. Cohen, K. Buck, R. E. Tanel, V. L. Vetter, L. A. Rhodes, J. Cox, T. Sheldon, and L. Ruetz Capture management efficacy in children and young adults with endocardial and unipolar epicardial systems, Europace, January 1, 2004; 6(3): 248 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-H. Nurnberg, H. Abdul-Khaliq, P. Ewert, and P. E. Lange Antibradycardia pacing in patients with congenital heart disease: experience with automatic threshold determination and output regulation (AutocaptureTM) Europace, January 1, 2003; 5(2): 199 - 205. [Abstract] [PDF] |
||||
![]() |
C. Balmer, M. Fasnacht, M. Rahn, L. Molinari, and U. Bauersfeld Long-term follow up of children with congenital complete atrioventricular block and the impact of pacemaker therapy Europace, January 1, 2002; 4(4): 345 - 349. [Abstract] [PDF] |
||||
![]() |
M. I. Cohen, D. M. Bush, V. L. Vetter, R. E. Tanel, T. S. Wieand, J. W. Gaynor, and L. A. Rhodes Permanent Epicardial Pacing in Pediatric Patients : Seventeen Years of Experience and 1200 Outpatient Visits Circulation, May 29, 2001; 103(21): 2585 - 2590. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Valsangiacomo, L. Molinari, M. Rahn-Schonbeck, and U. Bauersfeld DDD pacing mode survival in children with a dual-chamber pacemaker Ann. Thorac. Surg., December 1, 2000; 70(6): 1931 - 1934. [Abstract] [Full Text] [PDF] |
||||
![]() |