ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
George V. Letsou
Ernst Wolner
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nosé, Y.
Right arrow Articles by Schima, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nosé, Y.
Right arrow Articles by Schima, H.

Ann Thorac Surg 1999;68:775-779
© 1999 The Society of Thoracic Surgeons


Innovative Circulatory Support Systems

Development of a totally implantable biventricular bypass centrifugal blood pump system

Yukihiko Nosé, MD, PhDa, Kin-ichi Nakata, MD, PhDa, Masaharu Yoshikawa, MD, PhDa, George V. Letsou, MDa, Akira Fujisawaa, Ernst Wolner, MDa, Heinrich Schima, PhDa

a Department of Surgery, Baylor College of Medicine, Houston, Texas, USA

Address reprint requests to Dr Nosé, Department of Surgery, Baylor College of Medicine, One Baylor Plaza, Houston, TX 77030;
e-mail: ynose{at}bcm.tmc.edu

Presented at the Fourth International Conference on Circulatory Support Devices for Severe Cardiac Failure, Houston, TX, Oct 3–5, 1997.

Abstract

Background. During the past 2 years, the development of a totally implantable biventricular bypass rotary blood pump system has been made.

Methods. An extracorporeal gyro centrifugal pump, the CIE3, was miniaturized and developed into the PI601, a totally implantable plastic pump. Two-day anatomic and hemodynamic feasibility studies demonstrated that these two pump systems were easily implantable inside a calf’s abdominal wall, directly under the diaphragm. The priming volume of the pump was 20 mL, with sufficient cardiac outputs at approximately 2,000 rpm and requiring less than 10 W of power. Two-week antithrombogenic screening tests also revealed these pump systems to be quite antithrombogenic. In addition, 1-month system reliability studies demonstrated fail-safe reliable performances.

Results and Conclusions. Encouraged by these preliminary studies, the PI601 model was converted to the permanently implantable titanium gyro pump PI702 model. The long-term implantations were initiated approximately 3 months ago, and two such long-term LVAD studies are currently underway with no sign of difficulty (October 10, 1997). They were followed 283 days and 72 days, respectively. Both terminated due to functional inflow obstruction. There were no blood clots or emboli at autopsy.




This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
G. J. Endo, K. Kojima, K. Nakamura, Y. Matsuzaki, and T. Onitsuka
Nitric oxide inhalation prompts weaning from the right ventricular assist device: Evaluation under continuous-flow biventricular assistance
J. Thorac. Cardiovasc. Surg., October 1, 2002; 124(4): 739 - 749.
[Abstract] [Full Text] [PDF]


Home page
PerfusionHome page
H. M Reul and M. Akdis
Blood pumps for circulatory support
Perfusion, July 1, 2000; 15(4): 295 - 311.
[PDF]


Home page
PerfusionHome page
M. Loebe, T. Drews, E. Potapov, D.-V. Ngo, R. zu Dohna, and R. Hetzer
Device selection in mechanical circulatory support
Perfusion, July 1, 2000; 15(4): 313 - 318.
[PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 1999 by The Society of Thoracic Surgeons.