Ann Thorac Surg 1997;64:1177-1179
© 1997 The Society of Thoracic Surgeons
Medical Department III and Department of Cardiothoracic Surgery, Eberhard-Karls-University, Tübigen, Germany
Accepted for publication May 28, 1997.
Occasional patients have excessive defibrillation energy requirements despite appropriate transvenous defibrillation lead position and the use of biphasic shocks. A single-element subcutaneous array electrode was implanted in 2 patients with a high defibrillation threshold. The array electrode was implanted through the same infraclavicular incision that was used for implantation of the transvenous lead. The defibrillation threshold decreased from 30 J to 15 J and from 24 J to 9 J with the subcutaneous array electrode.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |