Ann Thorac Surg 1996;61:1418-1422
© 1996 The Society of Thoracic Surgeons
Clinic for Cardiovascular Surgery and Department of Cardiology, University Hospital, Zürich, and Clinic for Thoracic and Cardiovascular Surgery and Department of Cardiology, University Hospital, Berne, Switzerland
Accepted for publication February 2, 1996.
Background. Small aortic orifice primarily resulted in heart prosthesis mismatch in a significant number of patients. The Hemodynamic Plus (HP) series of St. Jude Medical heart valves represents an interesting innovation, allowing a larger valve orifice area with an equivalent tissue annulus diameter.
Methods. Hemodynamic characteristics of the 21-mm HP St. Jude Medical valve were prospectively compared with those of the standard 21-mm and 23-mm St. Jude Medical valves in three groups of 22 patients. Patients were selected from a database to be rigorously matched for age, sex, body surface area, functional class, underlying lesion, native valve opening area, left ventricular function, and preoperative peak and mean valve gradients. Postoperative evaluation (follow-up ranging from 3 to 24 months; mean, 11.5 months) included clinical examination and echocardiographic studies.
Results. There was no operative mortality or significant perioperative complications. Short-term clinical follow-up was marked by a complete absence of valve-related complications. Presently, all but 1 patient in the 21-mm HP group and 2 in the 21-mm standard group are in New York Heart Association functional class I. Doppler echocardiography-derived mean and maximal pressure gradients were significantly lower in the 21-mm HP group (8.1 ± 1.9 and 16.4 ± 3.4 mm Hg) than in the 21-mm standard group (13.4 ± 3.9 and 21.2 ± 4.3 mm Hg; p = 0.002 and p = 0.0004, respectively), confirming the better hemodynamic performance already described in in vitro studies. Pressure gradients did not differ significantly between the 21-mm HP and the 23-mm standard groups. The 21-mm HP valve demonstrated the highest performance index: 0.66 ± 0.08, compared with 0.49 ± 0.09 for the 21-mm standard valve (p < 0.001) and 0.59 ± 0.07 for the 23-mm standard valve (p < 0.001).
Conclusions. In vivo hemodynamic performance of the 21-mm HP valve corresponds closely to that of the 23-mm standard valve and is substantially better than that of the 21-mm standard valve. The 21-mm HP St. Jude Medical valve demonstrates excellent hemodynamic characteristics and can be recommended in normal-sized adult patients with narrow aortic root. This valve will minimize the need for aortic annulus enlargement.
This article has been cited by other articles:
![]() |
M. R. Moon, M. K. Pasque, N. A. Munfakh, S. J. Melby, J. S. Lawton, N. Moazami, J. E. Codd, T. D. Crabtree, H. B. Barner, and R. J. Damiano Jr Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival Ann. Thorac. Surg., February 1, 2006; 81(2): 481 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Niinami, S. Aomi, H. Tomioka, K. Nakano, and H. Koyanagi A comparison of the in vivo performance of the 19-mm St. Jude medical hemodynamic plus and 21-mm standard valve Ann. Thorac. Surg., October 1, 2002; 74(4): 1120 - 1124. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Harada, T. Ito, T. Mawatari, and T. Abe Replacement of a thrombosed valve after the Bentall procedure Ann. Thorac. Surg., March 1, 2002; 73(3): 965 - 967. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Vitale, I. Caldarera, C. Muneretto, R. Sinatra, A. Scafuri, E. Di Rosa, A. Contini, N. Tedesco, A. Pierangeli, M. Abbate, et al. Clinical evaluation of St Jude Medical Hemodynamic Plus versus standard aortic valve prostheses: The Italian multicenter, prospective, randomized study J. Thorac. Cardiovasc. Surg., October 1, 2001; 122(4): 691 - 698. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pibarot and J. G. Dumesnil Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Aydogan, H. Akay, G. Orhan, U. Filizcan, N. Kaynarca, O. Yucel, Y. Comertoglu, S. Aykut-Aka, and E. E Eren Early and Midterm Results of Aortic Annular Enlargement Asian Cardiovasc Thorac Ann, September 1, 1999; 7(3): 182 - 185. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Vijayanagar, G. Chan, R. Musunuru, N. Sastry, I. Siegman, N. Rattehalli, and M. Cortelli Aortic Valve Replacement Without Annular Enlargement in Patients with Small Aortic Roots Asian Cardiovasc Thorac Ann, December 1, 1998; 6(4): 279 - 284. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Demirkilic, E. Kuralay, E. Ozal, H. Bingol, C. Sag, A. T Yilmaz, and H. Tatar Medium-Term Outcome With Small Size ATS Medical Valves in Aortic Position Asian Cardiovasc Thorac Ann, September 1, 1998; 6(3): 183 - 187. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |