ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Robert L. Kormos
Bartley P. Griffith
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Litwak, P.
Right arrow Articles by Borovetz, H. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Litwak, P.
Right arrow Articles by Borovetz, H. S.

Ann Thorac Surg 1996;61:448-451
© 1996 The Society of Thoracic Surgeons


Pumps in Progress

Development and Initial Testing of a Pediatric Centrifugal Blood Pump

Philip Litwak, DVM, PhD, Kenneth C. Butler, MS, Douglas C. Thomas, MS, Lynn P. Taylor, Mahendar Macha, MD, Kenji Yamazaki, MD, Hiroaki Konishi, MD, Robert L. Kormos, MD, Bartley P. Griffith, MD, Harvey S. Borovetz, PhD

McGowan Center, Artificial Heart and Lung Program, Department of Surgery, Pittsburgh, Pennsylvania, and Nimbus, Inc, Rancho Cordova, California

Abstract

Background. We are developing a miniaturized centrifugal blood pump for use as a temporary cardiac assist device in neonatal and pediatric sized patients. This pump has a very low priming volume of 13 mL. A small motor stator has also been designed, which resulted in a device that can be placed very close to the patient, thereby minimizing overall circuit volume.

Methods. Testing to date has included in vitro hemodynamic performance, in vitro hemolysis generation, and in vivo evaluation in 5 lambs weighing 5.5 to 21 kg. Two lambs underwent peripheral cannulation from external jugular vein to carotid artery, whereas 3 others were cannulated from left atrium to carotid artery.

Results. In vitro data demonstrated pump capacity spanning 0.3 to 3.0 L/min and very low hemolysis generation at these conditions. In vivo, the pump functioned satisfactorily for periods up to 148 hours, and the bypass appeared to be well tolerated by the animals. Plasma free hemoglobin levels remained less than 25 mg/dL during all animal experiments. All devices were thrombus-free at explantation.

Conclusions. We conclude that this device has merit as an alternative to current oversized systems used for neonatal and pediatric cardiac assistance. In addition, a chronic neonatal lamb model in which to evaluate pediatric circulatory assist devices has been developed successfully.




This article has been cited by other articles:


Home page
PerfusionHome page
A K Mahmood, J M Courtney, S Westaby, M Akdis, and H Reul
Critical review of current left ventricular assist devices
Perfusion, September 1, 2000; 15(5): 399 - 420.
[PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 1996 by The Society of Thoracic Surgeons.