Ann Thorac Surg 1995;59:614-620
© 1995 The Society of Thoracic Surgeons
Department of Anesthesiology and the Section of Cardiothoracic Surgery, Department of Surgery, and Department of Neurosurgical Research, Mayo Clinic, Rochester, Minnesota
Accepted for publication November 4, 1994.
This study simultaneously compared the standard Kety-Schmidt and the modified xenon-133 (133Xe) clearance techniques for measuring cerebral blood flow (CBF) and cerebral metabolic rate for oxygen (CMRO2) during cardiac operations. The validity of the CBF method is important because our management of the patient during cardiopulmonary bypass (CPB) is based, in part, on our understanding of the cerebral hemodynamics during CPB. In 20 patients undergoing coronary artery bypass grafting, CBF and CMRO2 were determined by both methods. Measurements were made before onset of CPB and once during CPB. Ten patients underwent CPB with systemic normothermia (37°C) and 10 with systemic hypothermia (27°C). Anesthesia consisted of fentanyl and midazolam. CPB pump flows were kept at 2.2 to 2.4 L min-1 m-2 and
-stat pH management was used. Xenon-133 clearance significantly underestimated CBF and CMRO2 relative to the Kety-Schmidt technique before CPB and at both bypass temperatures. Values obtained by 133Xe clearance were approximately 50% of that measured by the Kety-Schmidt method. The modified 133Xe technique as typically used during cardiac operations does not appear to measure CBF accurately; this leads to corresponding errors in CMRO2 calculations. Determination of CMRO2 and cerebral autoregulatory function during cardiac operations appears to be more appropriate if based on the more direct Kety-Schmidt technique. Accordingly, our management of CPB with respect to cerebral perfusion as it has been determined by the modified 133Xe clearance method may require reassessment.
Related Article
Ann. Thorac. Surg. 1995 59: 558-561.
This article has been cited by other articles:
![]() |
I. MacVeigh, D. J. Cook, T. A. Orszulak, R. C. Daly, and D. E. Munnikhuysen Nitrous Oxide Method of Measuring Cerebral Blood Flow During Hypothermic Cardiopulmonary Bypass Ann. Thorac. Surg., March 1, 1997; 63(3): 736 - 740. [Abstract] [Full Text] |
||||
![]() |
U. H. Trivedi, R. L. Patel, M. R. J. Turtle, G. E. Venn, and D. J. Chambers Relative Changes in Cerebral Blood Flow During Cardiac Operations Using Xenon-133 Clearance Versus Transcranial Doppler Sonography Ann. Thorac. Surg., January 1, 1997; 63(1): 167 - 174. [Abstract] [Full Text] |
||||
![]() |
D. J. Cook, W. C. Oliver Jr, T. A. Orszulak, R. C. Daly, and R. D. Bryce Cardiopulmonary Bypass Temperature, Hematocrit, and Cerebral Oxygen Delivery in Humans Ann. Thorac. Surg., December 1, 1995; 60(6): 1671 - 1677. [Abstract] [Full Text] |
||||
![]() |
D. J. Cook, R. E. Anderson, J. D. Michenfelder, W. C. Oliver Jr, T. A. Orszulak, R. C. Daly, and R. D. Bryce Who's Afraid of Kety-Schmidt? Ann. Thorac. Surg., October 1, 1995; 60 (4): 1156 - 1157. [Full Text] |
||||
![]() |
W. L. Young, M. F. Newman, D. Amory, and J. G. Reves Cerebral Blood Flow Values During Cardiopulmonary Bypass: Relatively Absolute or Absolutely Relative? Ann. Thorac. Surg., March 1, 1995; 59(3): 558 - 561. [Full Text] |
||||