The Annals of Thoracic Surgery, Vol 57, 1335-1337, Copyright © 1994 by The Society of Thoracic Surgeons
Fracture of the internal lumen of a Datascope Percor Stat-DL Balloon, resulting in stroke
GJ Myers, RW Landymore, RB Leadon and C Squires
Department of Surgery, Dalhousie University, Halifax, Nova Scotia, Canada.
Intraaortic balloon counterpulsation was used to wean a 71-year-old man
from cardiopulmonary bypass. Thirty-six hours after insertion, the internal
lumen of the Datascope balloon fractured, releasing an unknown quantity of
helium into the patient's circulation. Left-sided hemiparesis developed,
secondary to a right hemispheric infarction.